Page 2 - Table of contents
2 Table of contents Definitions and test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3System performance summaries . . . . . . . . . . . . . . . . . . . . . . . . . . . .4Specifications and characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . .10Measurement thr...
Page 3 - Definitions and test conditions; Dynamic range
3 Definitions and test conditions Specifications describe the instrument's warrantedperformance after a half-hour warm-up and over thetemperature range of 25 ˚C ± 5 ˚C, unless otherwisestated. Specifications for frequencies above 3 GHz donot apply to instruments with Option 075 (75-ohmimpedance). Su...
Page 4 - System performance summaries; Configuration; Network analyzer; System dynamic range; GHz to 6 GHz; Reflection uncertainty; Transmission uncertainty; Measurement port characteristics
4 System performance summaries HP 8753ES (50-ohm systems) 7-mm test ports Configuration Network analyzer HP 8753ES Standard and Option 006 Calibration kit HP 85031B Test-port cables HP 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3...
Page 10 - MHz at nominal frequency
10 HP 8753ES specifications Test-port output characteristics 7 Frequency characteristics Range 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability 9 ±7.5 ppm (0˚ to 55 ˚C)±3 ppm/year Accuracy ±10 ppm at 25 ˚C ± 5 ˚C Option 1D5 ±0.05 ppm (0˚ to 55 ˚C),±0.5 ppm/year Power range 2 -85 ...
Page 11 - Test port input characteristics (continued); Maximum; Phase characteristics
11 Test port input characteristics (continued) External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical) Power level 0 to -25 dBm Spectral purity Maximum <-30 dBc spurious input Residual FM <20 kHz Typical settling 500 ms (automatic) ti...
Page 12 - Test port output characteristics; nominal; Test port input characteristics Option 011; 0 ohms nominal
12 HP 8753ES Option 011 specifications Test port output characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.5 ppm 0˚ to 55 ˚C typically ±3 ppm/year Option 1D5 typically ±0.05 ppm 0˚ to 55 ˚C typically ±0.5 ppm/year Accura...
Page 13 - Maximum spurious
13 HP 8753ES Option 011 specifications Test port input characteristics (continued) Frequency offset mode 3 Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 0 to -35 dBm 3 GHz to 6 GHz 0 to -30 dBm LO spectral purity Maximum spurio...
Page 15 - Magnitude characteristics
15 HP 8753ET specifications Test-port input characteristics (continued) External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements 2 Power level 0 to -25 dBm Spectral purity Maximum spurious <-30 dBc inputResidual FM <20 kHz Typical settling time ...
Page 16 - Measurement; Four display channels available.; Group delay characteristics; selectable; Source control; Set sweep time in seconds, minutes or hours.
16 HP 8753ET/ES supplementalcharacteristics Measurement Number of display channels Four display channels available. Number of measurement channels Two primary and two auxiliary measurement channelsavailable. Measurement parameters HP 8753ET: Reflection, transmission, A, B, R, A/R, B/R,A/B. Conversio...
Page 17 - The IF bandwidth is selectable from 6 kHz to 10 Hz; Display control
17 HP 8753ET/ES supplementalcharacteristics (continued) Sweep trigger Set to either continuous, single, group sweep, or exter-nal trigger. Set external trigger to take a completesweep or to measure individual points in a frequency,power or list sweep. Power Set source power from -20 to +5 dBm for th...
Page 18 - Automatic
18 HP 8753ET/ES supplementalcharacteristics (continued) Trace functions Display data Display current measurement data, memory data, orcurrent measurement with measurement and memorydata simultaneously. Trace math Vector division or subtraction of current linear mea-surement values and memory data. D...
Page 19 - Test sequences; This stimulus is also used to measure low-pass devices.; Remote programming
19 HP 8753ET/ES supplementalcharacteristics (continued) Frequency offset on/off Sets the RF source to be swept at a fixed offset fre-quency above the receiver as required in a sweptRF/IF, fixed LO, mixer test. Service menu Select the desired service test, service diagnostic, ser-vice or verification...
Page 20 - General characteristics; dBm minimum
20 General characteristics Front panel connectors HP 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) HP 8753ES Option 011 test ports Connector type Type-N female Impedance 50 ohms (nominal) HP 8753ES Option 075 test ports Connector type Type-N female...
Page 21 - Environmental characteristics; General conditions; Measurement throughput summary; Number of Points; Conversion; Options
21 Environmental characteristics General conditions RFI and EMI susceptibility: defined by VDE 0730,CISPR Publication 11, and FCC Class B Standards. ESD (electrostatic discharge): must be eliminated byuse of static-safe work procedures and an anti-staticbench mat. The flexible rubber keypad protects...
Page 22 - Test set specifications for Option 011
22 Test set specifications for Option 011 HP 85046A/B S-parameter test sets The HP 85046A/B S-parameter test sets provide thecapability to measure reflection and transmission char-acteristics (including S-parameters) of two-portdevices in either direction with a single connection.The test sets are c...
Page 23 - Specifications
23 HP 85047A S-parameter test set The HP 85047A S-parameter test set provides thecapability to simultaneously measure the reflectionand transmission characteristics of two-port devices ineither direction with a single connection. This test setincludes a frequency doubler that can be switched inby an...
Page 24 - Accessories; Calibration kits
24 Accessories Calibration kits Vector accuracy enhancement procedures require thatthe systematic errors of the measurement system becharacterized by measuring known devices (standards)on the system over the frequency range of interest.Return loss specifications or typical values are provid-ed where...
Page 25 - Description
25 HP 85038 family of calibration kits The HP 85038A calibration kit contains open circuits,short circuits, and terminations with both male andfemale 7-16 connectors. More economical versions ofthis kit are available: the HP 85038M contains onlymale devices, and the HP 85038F provides only femaledev...
Page 28 - 5 dB minimum; Dimensions
28 HP 8347A RF amplifier The HP 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The HP 8347A provides leveled output power without using anexternal coupler and detector, since these parts are built-in. Theexternal ALC can be directly connected to the external A...