Summary

Page 2 - SECTION II; SECTION I; This bulletin provides instructions for the; b. Time and Technique; . The time required for this calibration is approximately 4

TB 9-6625-2049-35 2 Paragraph Page SECTION II I. CALIBRATION PROCESS - Continued Triangle and ramp amplitude ............................ 16 17 Dc offset ............................................................... 17 18 Dc offset with ac .................................................. 18 19 H...

Page 4 - SECTION II; Table 2. identifies the specific equipment to be used in this

TB 9-6625-2049-35 4 Table 1. Calibration Description - Continued Test instrument parameters Performance specifications Amplitude accuracy with dc offset and no attenuation (range 1) into 50 Ω load Tolerance relative to programmed amplitude Sine wave: .001 Hz to 100 kHz ± 0.3 dB Square: .001 Hz to 10...

Page 15 - ENTRY AMPTD

TB 9-6625-2049-35 15 Table 3. Amplitude Test instrument settings Multimeter indications ENTRY AMPTD ENTRY FREQ (V ac) 3.536 V RMS 1 kHz 3.495 3.577 3.536 V RMS 100 kHz 3.495 3.577 1.061 V RMS 100 kHz 1.048 1.073 1.061 V RMS 1 kHz 1.048 1.073 1.061 V RMS 100 Hz 1.048 1.073 .3536 V RMS 1 100 Hz 0.3416...