HP 3325B - Manual
HP 3325B – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 2 – SECTION II; SECTION I; This bulletin provides instructions for the; b. Time and Technique; . The time required for this calibration is approximately 4
- Page 4 – SECTION II; Table 2. identifies the specific equipment to be used in this
- Page 15 – ENTRY AMPTD
- Page 18 – Test instrument; DC OFFSET; Multimeter indications
- Page 22 – Figure 5. Power supply assembly A2 – test instrument top view.
- Page 23 – By Order of the Secretary of the Army:; General, United States Army
__________
*This bulletin supersedes TB 9-6625-2049-35, 15 January 1996.
SUPERSEDED COPY DATED 15 JANUARY 1996
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
SYNTHESIZER/FUNCTION GENERATOR
HEWLETT-PACKARD
MODELS 3325A( ) AND 3325B( )
Headquarters, Department of the Army, Washington, DC
2 January 2001
Approved for public release; distribution is unlimited
Paragraph
Page
SECTION
I.
IDENTIFICATION AND DESCRIPTION
Test instrument identification ............................
Forms, records, and reports ................................
Calibration description .......................................
II.
EQUIPMENT REQUIREMENTS
Equipment required ............................................
Accessories required ...........................................
III
CALIBRATION PROCESS
Preliminary instructions.....................................
Equipment setup .................................................
Harmonic distortion ............................................
Amplitude modulation distortion........................
Square wave characteristics ...............................
Ramp retrace.......................................................
Frequency accuracy ............................................
Phase increment .................................................
Sine wave amplitude and flatness ......................
Square wave amplitude and flatness..................
REPORTING OF ERRORS AND RECOMMENDED IMPROVEMENTS
You can help improve this publication. If you find any mistakes or if you
know of a way to improve the procedure, please let us know. Mail your letter
or DA Form 2028 to: Commander, U. S. Army Aviation and Missile
Command, ATTN: AMSAM-MMC-LS-LP, Redstone Arsenal, AL 35898-5230.
A reply will be furnished to you. You may also send in your comments
electronically to our e-mail address:
or by FAX (256)
842-6546/DSN 788-6546
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
TB 9-6625-2049-35 2 Paragraph Page SECTION II I. CALIBRATION PROCESS - Continued Triangle and ramp amplitude ............................ 16 17 Dc offset ............................................................... 17 18 Dc offset with ac .................................................. 18 19 H...
TB 9-6625-2049-35 4 Table 1. Calibration Description - Continued Test instrument parameters Performance specifications Amplitude accuracy with dc offset and no attenuation (range 1) into 50 Ω load Tolerance relative to programmed amplitude Sine wave: .001 Hz to 100 kHz ± 0.3 dB Square: .001 Hz to 10...
TB 9-6625-2049-35 15 Table 3. Amplitude Test instrument settings Multimeter indications ENTRY AMPTD ENTRY FREQ (V ac) 3.536 V RMS 1 kHz 3.495 3.577 3.536 V RMS 100 kHz 3.495 3.577 1.061 V RMS 100 kHz 1.048 1.073 1.061 V RMS 1 kHz 1.048 1.073 1.061 V RMS 100 Hz 1.048 1.073 .3536 V RMS 1 100 Hz 0.3416...