1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
14 Page 14
15 Page 15
16 Page 16
17 Page 17
18 Page 18
19 Page 19
20 Page 20
21 Page 21
22 Page 22
23 Page 23
Page: / 23

Table of Contents:

  • Page 2 – SECTION II; SECTION I; This bulletin provides instructions for the; b. Time and Technique; . The time required for this calibration is approximately 4
  • Page 4 – SECTION II; Table 2. identifies the specific equipment to be used in this
  • Page 15 – ENTRY AMPTD
  • Page 18 – Test instrument; DC OFFSET; Multimeter indications
  • Page 22 – Figure 5. Power supply assembly A2 – test instrument top view.
  • Page 23 – By Order of the Secretary of the Army:; General, United States Army
Loading the manual

__________

*This bulletin supersedes TB 9-6625-2049-35, 15 January 1996.

*TB 9-6625-2049-35

SUPERSEDED COPY DATED 15 JANUARY 1996

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

CALIBRATION PROCEDURE FOR

SYNTHESIZER/FUNCTION GENERATOR

HEWLETT-PACKARD

MODELS 3325A( ) AND 3325B( )

Headquarters, Department of the Army, Washington, DC

2 January 2001

Approved for public release; distribution is unlimited

Paragraph

Page

SECTION

I.

IDENTIFICATION AND DESCRIPTION

Test instrument identification ............................

1

2

Forms, records, and reports ................................

2

2

Calibration description .......................................

3

2

II.

EQUIPMENT REQUIREMENTS

Equipment required ............................................

4

4

Accessories required ...........................................

5

5

III

CALIBRATION PROCESS

Preliminary instructions.....................................

6

6

Equipment setup .................................................

7

6

Harmonic distortion ............................................

8

7

Amplitude modulation distortion........................

9

9

Square wave characteristics ...............................

10

9

Ramp retrace.......................................................

11

10

Frequency accuracy ............................................

12

11

Phase increment .................................................

13

13

Sine wave amplitude and flatness ......................

14

14

Square wave amplitude and flatness..................

15

16

REPORTING OF ERRORS AND RECOMMENDED IMPROVEMENTS

You can help improve this publication. If you find any mistakes or if you
know of a way to improve the procedure, please let us know. Mail your letter
or DA Form 2028 to: Commander, U. S. Army Aviation and Missile
Command, ATTN: AMSAM-MMC-LS-LP, Redstone Arsenal, AL 35898-5230.
A reply will be furnished to you. You may also send in your comments
electronically to our e-mail address:

[email protected]

or by FAX (256)

842-6546/DSN 788-6546

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - SECTION II; SECTION I; This bulletin provides instructions for the; b. Time and Technique; . The time required for this calibration is approximately 4

TB 9-6625-2049-35 2 Paragraph Page SECTION II I. CALIBRATION PROCESS - Continued Triangle and ramp amplitude ............................ 16 17 Dc offset ............................................................... 17 18 Dc offset with ac .................................................. 18 19 H...

Page 4 - SECTION II; Table 2. identifies the specific equipment to be used in this

TB 9-6625-2049-35 4 Table 1. Calibration Description - Continued Test instrument parameters Performance specifications Amplitude accuracy with dc offset and no attenuation (range 1) into 50 Ω load Tolerance relative to programmed amplitude Sine wave: .001 Hz to 100 kHz ± 0.3 dB Square: .001 Hz to 10...

Page 15 - ENTRY AMPTD

TB 9-6625-2049-35 15 Table 3. Amplitude Test instrument settings Multimeter indications ENTRY AMPTD ENTRY FREQ (V ac) 3.536 V RMS 1 kHz 3.495 3.577 3.536 V RMS 100 kHz 3.495 3.577 1.061 V RMS 100 kHz 1.048 1.073 1.061 V RMS 1 kHz 1.048 1.073 1.061 V RMS 100 Hz 1.048 1.073 .3536 V RMS 1 100 Hz 0.3416...