Teledyne T801 - Manuals
Teledyne T801 – Manual in PDF format online.
Manuals:
Manual Teledyne T801
Summary
i A BOUT T ELEDYNE A DVANCED P OLLUTION I NSTRUMENTATION (TAPI) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., is a worldwide market leader in the design and manufacture of precision analytical instrumentation used for air quality monitoring, conti...
v W ARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides that: Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment failure occur, TAPI assures its customers that prompt service ...
vii A BOUT THIS M ANUAL This operation manual, PN 07274, is comprised of multiple documents in PDF format, as listed below. Part No. Rev Name/Description 07274 B T801 Operation manual 06843 B Menu Trees and Software Documentation (inserted as Appendix A in this manual) 07269 A Spare Parts List (loca...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual viii CONVENTIONS USED In addition to the safety symbols as presented in the Important Safety Information page, this manual provides special notices related to the safety and effective use of the analyzer and other pertinent information. Special No...
ix T ABLE OF C ONTENTS A BOUT T ELEDYNE A DVANCED P OLLUTION I NSTRUMENTATION (TAPI) .......................................................................... i S AFETY M ESSAGES ...........................................................................................................................
Table of Contents Teledyne API T801 NDIR CO2 Analyzer Operation Manual xii 12.2.1. Electronic Operation of the CO 2 Sensor............................................................................................ 204 12.3. Pneumatic Operation ..........................................................
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Table of Contents xiii Figure 5-4: Alternative Setup Using 250 Ω Resistor for Checking Current Output Signal Levels ..... 95 Figure 7-1: Default DAS Channel Setup .......................................................................................
Table of Contents Teledyne API T801 NDIR CO2 Analyzer Operation Manual xiv Table 5-9: Concentration Alarm Default Settings .......................................................................... 102 Table 6-1: COM Port Communication Modes..............................................................
16 This page intentionally left blank. 07274B DCN6418
17 1. INTRODUCTION, FEATURES, AND OPTIONS 1.1. T801 OVERVIEW The Model T801 NDIR CO 2 Analyzer is a microprocessor-controlled analyzer that employs Non-Dispersive Infrared (NDIR) spectroscopy to determine the concentration of molecular carbon dioxide (CO 2 ) in the sample gas drawn through the instr...
Introduction, Features, and Options Teledyne API T801 NDIR CO2 Analyzer Operation Manual 18 1.3. OPTIONS Table 1-1 presents the options available with the T801 analyzer. For assistance with ordering, please contact the Sales department of Teledyne API at: PHONE (toll free, North America) 800-324-519...
21 2. SPECIFICATIONS, APPROVALS & COMPLIANCE This section presents specifications, agency approvals, and safety compliance details. 2.1. SPECIFICATIONS Table 2-1: T801 CO 2 Specifications Parameter Description Ranges Min: 0-1% Full scale Max: 0-20% Full scale (selectable, dual ranges and auto-ra...
Specifications, Approvals & Compliance Teledyne API T801 NDIR CO2 Analyzer Operation Manual 22 2.2. APPROVALS AND CERTIFICATIONS The Teledyne API Model T801 NDIR CO 2 Analyzer was tested and certified for Safety and Electromagnetic Compatibility (EMC). This section presents the compliance statem...
23 3. GETTING STARTED This section addresses the procedures for unpacking the instrument and inspecting for damage, presents clearance specifications for proper ventilation, introduces the instrument layout, then presents the procedures for getting started: making electrical and pneumatic connection...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 24 Verify that there is no apparent external shipping damage. If damage has occurred, please advise the shipper first, then Teledyne API. Included with your analyzer is a printed record of the final performance characterization per...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 25 3.2. INSTRUMENT LAYOUT Instrument layout shows front panel and display, rear panel connectors, and internal chassis layout. 3.2.1. FRONT PANEL Figure 3-1 shows the analyzer’s front panel layout, followed by a close-up of the dis...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 26 Figure 3-2: Display Screen and Touch Control The front panel liquid crystal display screen includes touch control. Upon analyzer start-up, the screen shows a splash screen and other initialization indicators before the main disp...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 27 Table 3-2: Display Screen and Touch Control Description Field Description/Function LEDs indicating the states of Sample, Calibration and Fault, as follows: Name Color State Definition SAMPLE Green Off On Blinking Unit is not ope...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 28 Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts Note The menu charts in this manual contain condensed representations of the analyzer’s display during the various operations being described. These menu charts are ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 29 3.2.2. REAR PANEL Figure 3-4: Rear Panel Layout 07274B DCN6418
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 31 3.2.3. INTERNAL CHASSIS LAYOUT Figure 3-5: Internal Layout 07274B DCN6418
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 32 3.3. CONNECTIONS AND SETUP This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2) connections for setup and preparing for instrument operation. 3.3.1. ELECTRICAL CONNECTIONS Note To maintain complianc...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 33 3.3.1.2. CONNECTING ANALOG INPUTS (OPTION 64) The Analog In connector is used for connecting external voltage signals from other instrumentation (such as meteorological instruments) and for logging these signals in the analyzer’...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 34 Channel A4 is special. It can be set by the user (see Section 5.9.3.10) to output any one of the parameters accessible through the <TST TST> buttons of the unit’s front panel menu. To access these signals attach a strip ch...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 35 Figure 3-8: Current Loop Option Installed C ONVERTING C URRENT L OOP A NALOG O UTPUTS TO S TANDARD V OLTAGE O UTPUTS CAUTION – Avoid Invalidating Warranty Servicing or handling of circuit components requires electrostatic discha...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 37 The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled STATUS (see Figure 3-4). Pin-outs for this connector are: STATUS 1 2 3 4 5 6 7 8 D + SY STEM O K CO NC V AL ID C A L M OD E / ME AS MO D...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 38 3.3.1.6. CONNECTING THE CONTROL INPUTS If you wish to use the analyzer to remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the an...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 39 3.3.1.7. CONCENTRATION ALARM RELAY (OPTION 61) The Teledyne API T-Series analyzers have an option for four (4) “dry contact” relays on the rear panel of the instrument. This relay option is different from and in addition to the ...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 40 The software for this instrument is flexible enough to allow you to configure the alarms so that you can have 2 alarm levels. CO 2 Alarm 1 = 5% CO 2 Alarm 2 = 10 % Another likely configuration for this would be to disable “Alarm...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 41 device. Adjust the DCE-DTE switch (Figure 3-4) to select DTE or DCE as appropriate. Configuration: Section 6.1 and Section 6.5.2 (for Hessen protocol). I MPORTANT I MPACT ON R EADINGS OR D ATA Cables that appear to be compatible...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 43 RS-232 M ULTIDROP O PTION C ONNECTION When the RS-232 Multidrop option is installed, connection adjustments and configuration through the menu system are required. This section provides instructions for the internal connection a...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 44 Figure 3-14: Jumper and Cables for Multidrop Mode (Note: If you are adding an instrument to the end of a previously configured chain, remove the shunt between Pins 21 22 of JP2 on the Multidrop/LVDS PCA in the instrument that ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 45 Analyzer Analyzer Analyzer Last Analyzer Female DB9 Male DB9 RS-232 COM2 RS-232 COM2 RS-232 COM2 RS-232 COM2 Host RS-232 port Ensure jumper is installed between JP2 pins 21 22 in last instrument of multidrop chain. Figure 3-15...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 46 RS-485 C ONNECTION As delivered from the factory, COM2 is configured for RS-232 communications. This port can be reconfigured for operation as a non-isolated, half-duplex RS-485 port. Using COM2 for RS-485 communication will dis...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 48 Calib rated N 2 100 % C once ntratio n Calib rated g a s ( CO 2 at 16% Span C oncentration) VENT Figure 3-16: Pneumatic Connections, Using Bottled Span Gas 3.3.2.3. SAMPLE GAS SOURCE Attach a sample inlet line to the SAMPLE inle...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 49 3.3.2.6. EXHAUST OUTLET Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line should be: PTEF tubing; minimum O.D ¼” A maximum of 10 meters long. Vented outside the T801 analyzer’s enclosure, pr...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 50 3.4. STARTUP, FUNCTIONAL CHECKS, AND INITIAL CALIBRATION I MPORTANT I MPACT ON R EADINGS OR D ATA The analyzer’s cover must be installed to ensure that the temperatures of the internal components are properly controlled. If you ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Getting Started 51 after the 60-minute warm up period, investigate their cause using the troubleshooting guidelines in Section 11.1.1. To view and clear warning messages, press: Figure 3-18: Viewing and Clearing WARNING Messages Table 3-9 lists br...
Getting Started Teledyne API T801 NDIR CO2 Analyzer Operation Manual 52 MESSAGE DEFINITION SAMPLE PRESS WARN Sample gas pressure outside of operational parameters. SYSTEM RESET 1 The analyzer was rebooted or the CPU was reset. 1 Does not clear after power up. 2 Clears the next time successful span c...
57 4. BASIC OPERATION OF THE T801 ANALYZER The T801 analyzer is a computer-controlled analyzer with a dynamic menu interface that allows all major operations to be controlled from the front panel touchscreen through user-friendly menus (A complete set of menu trees is located in Appendix A of this m...
Basic Operation of the T801 Analyzer Teledyne API T801 NDIR CO2 Analyzer Operation Manual 58 Figure 4-1: Front Panel Touchscreen and Display The mode field of the front panel display indicates to the user which operating mode the unit is currently running. In addition to SAMPLE and SETUP , other mod...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Basic Operation of the T801 Analyzer 59 4.2.1. TEST FUNCTIONS A series of TEST functions are available for viewing at the front panel whenever the analyzer is SAMPLE MODE. These functions provide information about the various functional parameters...
Basic Operation of the T801 Analyzer Teledyne API T801 NDIR CO2 Analyzer Operation Manual 60 Table 4-2: Test Functions Defined PARAMETER DISPLAY TITLE UNITS DEFINITION Range RNG RN1 RN2 % The full scale limit at which the reporting range of the analyzer is currently set. THIS IS NOT the Physical Ran...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Basic Operation of the T801 Analyzer 61 Setup Mode can be protected by password security through the SETUP>PASS menu (Section 5.5) to prevent unauthorized or inadvertent configuration adjustments. 4.4.1. PRIMARY SETUP MENU For a visual represen...
63 5. SETUP MENU The SETUP menu is used to set instrument parameters for performing configuration, calibration, reporting and diagnostics operations. 5.1. SETUP CFG: CONFIGURATION INFORMATION Pressing the CFG button displays the instrument’s configuration information. This display lists the analyz...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 64 5.3. SETUP DAS: INTERNAL DATA ACQUISITION SYSTEM Use the SETUP>DAS menu to capture and record data. Refer to Section 7 for configuration and operation details. 5.4. SETUP RNGE: ANALOG OUTPUT REPORTING RANGE CONFIGURATION Use t...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 65 5.4.2. ANALOG OUTPUT RANGES FOR CO 2 CONCENTRATION The analyzer has several active analog output signals accessible through a connector on the rear panel (see Figure 3-4). Not Used ANALOG OUT A1 A2 A3 A4 + - + - + - + - CO 2 concentr...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 66 5.4.3. REPORTING RANGE MODES The T801 provides three analog output range modes to choose from. Single range ( SNGL ) mode sets a single maximum range for the analog output. If single range is selected, all three outputs are slaved ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 67 5.4.3.1. RNGE MODE SNGL: CONFIGURING THE T801 ANALYZER FOR SINGLE RANGE MODE Single Range Mode (SNGL) is the default reporting range mode for the analyzer. When the single range mode is selected ( SNGL ), all concentration output...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 69 To set the upper range limit for each independent reporting range, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 5.4.3.3. RNGE MODE AUTO: CONFIGURING THE T801 ANALYZER FOR AUTO R...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 70 To set individual ranges press the following menu sequence: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 5.4.4. SETUP RNGE DIL: USING THE OPTIONAL DILUTION RATIO FEATURE This feature i...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 71 2. Select the reporting range mode and set the reporting range upper limit (see Section 5.4.3). 3. Ensure that the upper span limit entered for the reporting range is the maximum expected concentration of the non-diluted gas. 4. Set ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 73 Example: If all passwords are enabled, the following menu sequence would be required to enter the SETUP menu: SYSTEM ENTER PASSWORD:101 1 0 1 ENTR EXIT SYSTEM ENTER PASSWORD:0 0 0 0 ENTR EXIT EXAMPLE: This password enables the SETUP ...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 74 5.6. SETUP CLK: SETTING THE ANALYZER’S INTERNAL CLOCK The analyzer has an internal clock for setting the time and day; it’s speed can be adjusted to compensate for faster or slower CPU clocks. Press SETUP>CLK to access the clock...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 75 The CLOCK_ADJUST variable is accessed via the VARS submenu: To change the value of this variable, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07274B DCN6418
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 76 5.7. SETUP MORE COMM: COMMUNICATION PORTS This section introduces the communications setup menu; Section 6 provides the setup instructions and operation information. To arrive at the communications menu, press SETUP>ENTR>MO...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 77 To edit the instrument’s ID code, press: The ID can also be used for to identify any one of several analyzers attached to the same network, e.g., by location number, company asset number, etc.). (The ACAL submenu in the Primary Setup...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 80 To access and navigate the VARS menu, use the following key sequence. SETUP X.X 6) CLOCK_ADJUST=0 Sec/Day PREV NEXT JUMP EDIT ENTR EXIT SETUP X.X CLOCK_ADJUST=0 Sec/Day + 0 0 ENTR EXIT Enter sign and number of seconds per day the clo...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 81 5.9. SETUP MORE DIAG: DIAGNOSTICS FUNCTIONS A series of diagnostic tools is grouped together under the SETUP MORE DIAG menu. These tools can be used in a variety of troubleshooting and diagnostic procedures and are referred t...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 82 To access the DIAG functions press the following menu sequence: <TST TST> CAL MSG SETUP X.X CFG DAS ACAL RNGE PASS CLK EXIT SETUP X.X COMM VARS EXIT SETUP X.X EXIT PREV returns to the Activates the selected submenu PREV PREV PR...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 83 5.9.1. SIGNAL I/O The signal I/O diagnostic mode allows a user to review and change the digital and analog input/output functions of the analyzer. It can also be used for troubleshooting purposes (Section 11.1.3). Refer to Appendix A...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 84 5.9.2. ANALOG OUTPUT The T801 analyzer comes equipped with four analog outputs. The first two analog output ( A1 & A2) signals represent the currently measured CO 2 concentration (see Section 5.4.2). The third analog output (...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 86 5.9.3.1. ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION In its standard configuration the analog outputs are set to output a 0 – 5 VDC signals. Several other output ranges are available. Each range is usable from -5% to + 5% of the ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 87 To change the output type and range, select the ANALOG I/O CONFIGURATION submenu (see Section 5.9.3) then press, 5.9.3.2. CALIBRATION OF THE ANALOG OUTPUTS Analog output calibration should to be carried out on first startup of the an...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 88 5.9.3.3. ENABLING OR DISABLING THE AUTOCAL FOR AN INDIVIDUAL ANALOG OUTPUT To enable or disable the AutoCal feature for an individual analog output, elect the ANALOG I/O CONFIGURATION submenu (see Section 5.9.3) then press: 07274B DC...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 89 5.9.3.4. AUTOMATIC GROUP CALIBRATION OF THE ANALOG OUTPUTS I MPORTANT I MPACT ON R EADINGS OR D ATA Manual calibration should be used for any analog output set for a 0.1V output range or in cases where the outputs must be closely mat...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 90 5.9.3.5. AUTOMATIC INDIVIDUAL CALIBRATION OF THE ANALOG OUTPUTS To use the AUTO CAL feature to initiate an automatic calibration for an individual analog output, select the ANALOG I/O CONFIGURATION submenu (see Section 5.9.3) then pr...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 91 5.9.3.6. MANUAL CALIBRATION OF THE ANALOG OUTPUTS CONFIGURED FOR VOLTAGE RANGES For highest accuracy, the voltages of the analog outputs can be manually calibrated. Note The menu for manually adjusting the analog output signal level ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 93 5.9.3.7. MANUAL ADJUSTMENT OF CURRENT LOOP OPTION OUTPUT SPAN AND OFFSET A current loop option may be purchased for the A1 and A2 analog outputs of the analyzer. This option places circuitry in series with the output of the A-to-D co...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 95 An alternative method for measuring the output of the Current Loop converter is to connect a 250 ohm 1% resistor across the current loop output in lieu of the current meter (see Figure 3-7 for pin assignments and diagram of the ana...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 96 5.9.3.8. TURNING AN ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF In its default configuration, a ± 5% over-range is available on each of the T801’s analog outputs. This over-range can be disabled if your recording device is sensitive to e...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 97 5.9.3.9. ADDING A RECORDER OFFSET TO AN ANALOG OUTPUT Some analog signal recorders require that the zero signal be significantly different from the baseline of the recorder in order to record slightly negative readings from noise aro...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 98 5.9.3.10. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4 The test functions available to be reported are: Table 5-8: Test Channels Functions available on the T801’s Analog Output TEST CHANNEL DESCRIPTION ZERO FULL SCALE NONE Test Ch...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 99 To activate the TEST Channel and select a function (in this example SAMPLE PRESSURE ), press: SAMPLE CO2 RNG=20.00 % CO2=XXX.XX <TST TST> CAL SETUP DIAG TEST CHAN:NONE PREV NEXT ENTR EXIT DIAG SIGNAL I/O PREV NEXT ENTR EXIT DIA...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 100 5.9.3.11. AIN CALIBRATION This is the sub-menu to conduct a calibration of the T801 analyzer’s analog inputs. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Setup Menu 101 5.9.3.12. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGURATION To configure the analyzer’s optional analog inputs, define for each channel : gain (number of units represented by 1 volt) offset (volts) engineering units to be represe...
Setup Menu Teledyne API T801 NDIR CO2 Analyzer Operation Manual 102 5.10. SETUP MORE ALRM: USING THE GAS CONCENTRATION ALARMS (OPTION 61) The T801 concentration alarm option includes two alarms. Each alarm has a user-settable limit, and is associated with an opto-isolated TTL relay accessible vi...
105 6. COMMUNICATIONS SETUP AND OPERATION This instrument’s rear panel connections include an Ethernet port, a USB port (option) and two serial communication (COM) ports labeled RS232 , which is the COM1 port, and COM2 (see Figure 3-2). These ports operate similarly and give the user the ability to ...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 106 6.2.1. COMMUNICATION MODES Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in Table 6-1. As modes are selected, the analyzer sums the mode ID numbers an...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Communications Setup and Operation 107 Press the following menu buttons to select communication modes for one of the COM Ports, such as the following example where HESSEN is enabled: (The ACAL submenu in the Primary Setup Menu is a special configu...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 108 6.2.2. COM PORT BAUD RATE To select the baud rate of either COM Port, go to SETUP>MORE>COMM and select either COM1 or COM2 as follows: EXAMPLE SETUP X.X COM1 BAUD RATE:19200 PREV NEXT ENTR EXIT SETUP X....
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Communications Setup and Operation 109 6.2.3. COM PORT TESTING The serial ports can be tested for correct connection and output in the COMM menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, t...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 110 6.3. REMOTE ACCESS VIA THE ETHERNET For network or Internet communication, the analyzer can be connected to any standard 10BaseT or 100BaseT Ethernet network via low-cost network hubs, switches or routers. Th...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Communications Setup and Operation 113 6.3.2. MANUALLY CONFIGURING THE NETWORK IP ADDRESSES There are several circumstances when you may need to configure the interface settings of the analyzer’s Ethernet card manually: your LAN is not running a...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Communications Setup and Operation 115 6.3.2.1. CHANGING THE ANALYZER’S HOSTNAME The HOSTNAME is the name by which the analyzer appears on your network. The default name for all Teledyne API T801 analyzers iseither 0 or T801 . To change this nam...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 116 6.4. USB PORT FOR REMOTE ACCESS The analyzer can be operated through a personal computer by downloading the TAPI USB driver and directly connecting their respective USB ports. 1. Install the Teledyne T-Series...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 118 6.5. COMMUNICATIONS PROTOCOLS This section presents MODBUS and HESSEN information. 6.5.1. MODBUS The following set of instructions assumes that the user is familiar with MODBUS communications, and provides mi...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 120 6.5.2. HESSEN The Hessen protocol is a multidrop protocol, in which several remote instruments are connected via a common communications channel to a host computer. The remote instruments are regarded as slav...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 122 6.5.2.3. SELECTING A HESSEN PROTOCOL TYPE Currently there are two versions of Hessen Protocol in use. The original implementation, referred to as TYPE 1 , and a more recently released version, TYPE 2 that has...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Communications Setup and Operation 123 6.5.2.4. SETTING THE HESSEN PROTOCOL RESPONSE MODE The Teledyne API implementation of Hessen Protocol allows the user to choose one of several different modes of response for the analyzer. Table 6-6: Teledyne...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Communications Setup and Operation 125 6.5.2.6. EDITING OR ADDING HESSEN GAS LIST ENTRIES To add or edit an entry to the Hessen Gas List, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07274B DCN64...
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 126 6.5.2.7. DELETING HESSEN GAS LIST ENTRIES To delete an entry from the Hessen Gas list, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07274B DCN6418
Communications Setup and Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 128 To assign or reset the status flag bit assignments, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 6.5.2.9. INSTRUMENT ID Each instrument on a Hessen Protocol...
129 7. DATA ACQUISITION SYSTEM (DAS) & APICOM The T801 analyzer contains a flexible and powerful, internal data acquisition system (DAS) that enables the analyzer to store concentration and calibration data as well as a host of diagnostic parameters. The DAS of the T801 can store several months ...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 130 I MPORTANT I MPACT ON R EADINGS OR D ATA DAS operation is suspended whenever its configuration is edited using the analyzer’s front panel and therefore data may be lost. To prevent such data loss, it ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 131 Table 7-2: DAS Data Channel Properties PROPERTY DESCRIPTION DEFAULT SETTING SETTING RANGE NAME The name of the data channel. “NONE” Up to 6 letters or digits 1. TRIGGERING EVENT The event that trigger...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 132 DETAILED: Samples seven different parameters related to the operating status of the analyzer’s. For each parameter: A value is logged once every minute; An average of the last 60 readings is calcu...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 133 Triggering Events and Data Parameters/Functions for these default channels are: Figure 7-1: Default DAS Channel Setup 07274B DCN6418
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 135 7.1.4. SETUP DAS EDIT: ACCESSING THE DAS EDIT MODE DAS configuration is most conveniently done through the APICOM remote control program. The following sequence of menu items shows how to edit usi...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 136 7.1.4.1. EDITING DAS DATA CHANNEL NAMES To edit the name of a DAS data channel, follow the instruction shown in Section 7.1.4 then press: 7.1.4.2. EDITING DAS TRIGGERING EVENTS Triggering events defin...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 137 WARNINGS : Some data may be useful when stored if one of several warning messages appears such as CO2TMW (CO 2 sensor temperature warning). This is helpful for trouble-shooting by monitoring when a pa...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 138 7.1.4.3. EDITING DAS PARAMETERS Data parameters are types of data that may be measured and stored by the DAS. For each Teledyne API analyzer model, the list of available data parameters is different, ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 141 To define the REPORT PERIOD , follow the instruction shown in Section 7.1.4 then press: The SAMPLE PERIOD and REPORT PERIOD intervals are synchronized to the beginning and end of the appropriate inter...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 143 To set the NUMBER OF RECORDS , follow the instruction shown in Section 7.1.4 then press: 07274B DCN6418
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 144 7.1.4.7. RS-232 REPORT FUNCTION The DAS can automatically report data to the communications ports, where they can be captured with a terminal emulation program or simply viewed by the user using the A...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 146 7.1.4.9. THE COMPACT REPORT FEATURE When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each parameter in one channel on a separate line, up to five pa...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 148 7.2. REMOTE DAS CONFIGURATION The DAS can be configured and operated remotely via either the APICOM interface or a terminal emulation program. 7.2.1. DAS CONFIGURATION VIA APICOM Figure 7-2 shows an e...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Data Acquisition System (DAS) & APICOM 149 Figure 7-3: APICOM User Interface for Configuring the DAS Once a DAS configuration is created, it is conveniently saved to your computer and can be uploaded to any instrument; it can be stored on a co...
Data Acquisition System (DAS) & APICOM Teledyne API T801 NDIR CO2 Analyzer Operation Manual 150 7.2.2. DAS CONFIGURATION VIA TERMINAL EMULATION PROGRAMS Although Teledyne API recommends the use of APICOM, the DAS can also be accessed and configured through a terminal emulation program such as Hy...
151 8. REMOTE OPERATION This section provides information needed when using external digital and serial I/O for remote operation. It assumes that the electrical connections have been made as described in Section 3.3.1. Your analyzer can be remotely configured, calibrated or queried for stored data t...
Remote Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 152 8.2. INTERACTIVE MODE Interactive mode is used with a terminal emulation programs or a “dumb” computer terminal. 8.2.1. REMOTE CONTROL VIA A TERMINAL EMULATION PROGRAM Start a terminal emulation programs such as HyperTerminal....
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Remote Operation 153 may have additional arguments that define how the command is to be executed. Press ? <CR> or refer to Appendix A-6 for a list of available command designators <CR> is a carriage return. All commands must be termina...
Remote Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 154 Text strings : Used to represent data that cannot be easily represented by other data types, such as data channel names, which may contain letters and numbers. They consist of a quotation mark, followed by one or more printa...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Remote Operation 155 8.3. REMOTE ACCESS BY MODEM The T801 can be connected to a modem for remote access. This requires a cable between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable (available from Teledyne API with PN WR...
Remote Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 156 To initialize the modem press: SAMPLE CO2 RNG=20.00 % CO2=XXX.XX <TST TST> CAL SETUP If there is a problem initializing the modem the message, “MODEM NOT INITIALIZED” will appear. SETUP X.X COMMUNICATIONS MENU ID INET CO...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Remote Operation 157 8.4. COM PORT PASSWORD SECURITY In order to provide security for remote access of the T801, a LOGON feature can be enabled to require a password before the instrument will accept commands. This is done by turning on the SECURI...
159 9. CALIBRATION PROCEDURES This section contains a variety of information regarding the various methods for calibrating a T801 as well as other supporting information This section is organized as follows: Section 9.1 – Before Calibration - contains general information you should know before calib...
Calibration Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 160 9.1. BEFORE CALIBRATION The calibration procedures in this section assume that the range mode, analog range and units of measure have already been selected for the analyzer. If this has not been done, please do so before...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Calibration Procedures 161 Cylinders of calibrated CO 2 gas traceable to NIST-Standard Reference Material specifications (also referred to as SRMs or EPA protocol calibration gases) are commercially available. Table 9-1: NISTSRM's Available for Tr...
Calibration Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 162 9.2. MANUAL CALIBRATION CHECKS AND CALIBRATION I MPORTANT I MPACT ON R EADINGS OR D ATA ZERO/SPAN CALIBRATION CHECKS VS. ZERO/SPAN CALIBRATION Pressing the ENTR button during the following procedure resets the stored val...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Calibration Procedures 163 9.2.2. PERFORMING A MANUAL CALIBRATION CHECK Note If the ZERO or SPAN menu buttons are not displayed, the measurement made during this cal check is out of the allowable range allowed for a reliable calibration. See Secti...
Calibration Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 164 9.2.3. PERFORMING A MANUAL CALIBRATION The following section describes the basic method for manually calibrating the T801. If the analyzer’s reporting range is set for the DUAL or AUTO range modes, a step will appear for...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Calibration Procedures 165 9.2.3.2. ZERO/SPAN POINT CALIBRATION PROCEDURE To perform the zero/span calibration procedure: 07274B DCN6418
Calibration Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 166 9.3. ASSESSING CALIBRATION QUALITY After completing one of the calibration procedures described above, it is important to evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Calibration Procedures 167 To carry out this adjustment, the current ambient atmospheric pressure must be known. Before performing the following pressure calibration, ensure that the pressure being measured by the analyzer’s internal sensor is equ...
Calibration Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 168 9.4.2. FLOW CALIBRATION The flow calibration allows the user to adjust the values of the sample flow rates as they are displayed on the front panel and reported through COM ports to match the actual flow rate measured at...
170 This page intentionally left blank 07274B DCN6418
171 10. MAINTENANCE SCHEDULE & PROCEDURES The T801 Analyzer utilizes technologies that are non-depleting and require very little maintenance. However, there are a minimal number of simple procedures that when performed regularly will ensure that the analyzer continues to operate accurately and r...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Maintenance Schedule & Procedures 175 10.2. PREDICTIVE DIAGNOSTICS The Test Functions can be used to predict failures by looking at how their values change over time. Initially it may be useful to compare the state of these Test Functions to t...
Maintenance Schedule & Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 176 To change the filter: 1. Turn OFF the analyzer to prevent drawing debris into the instrument. 2. Open the T801’s hinged front panel and unscrew the knurled retaining ring on the filter assembly. Figure 10-...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Maintenance Schedule & Procedures 177 10.3.3. PERFORMING LEAK CHECKS Leaks are the most common cause of analyzer malfunction; Section 10.3.3.1 presents a simple leak check procedure. Section 10.3.3.2 details a more thorough procedure. 10.3.3.1...
Maintenance Schedule & Procedures Teledyne API T801 NDIR CO2 Analyzer Operation Manual 178 10.3.4. PERFORMING A SAMPLE FLOW CHECK CAUTION G ENERAL S AFETY H AZARD Always use a separate calibrated flow meter capable of measuring flows in the 0 – 1000 cm 3 /min range to measure the gas flow rate t...
179 11. TROUBLESHOOTING AND SERVICE This section describes a variety of methods for identifying the source of performance problems with the analyzer. Also included here are procedures that are used to repair the instrument. Q UALIFIED P ERSONNEL O NLY The operations outlined in this Section must be ...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 180 1. Note any WARNING MESSAGES and take corrective action as necessary. 2. Examine the values of all TEST functions and compare them to factory values. Note any major deviations from the factory values and take correc...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 181 As shown in the top screen in the illustration below, the analyzer will alert the user that a Warning message is active by flashing the FAULT LED, displaying the the Warning message in the Param field along with the...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 182 To view or clear the various warning messages press: (Please show CO2 RNG throughout document = 20%) Suppresses the warning messages Press CLR to clear the current message. If more than one warning is active, the ne...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 183 WARNING MESSAGE FAULT CONDITION POSSIBLE CAUSES RELAY BOARD WARN The CPU cannot communicate with the Relay Board. I 2 C bus failure Failed relay board Loose connectors/wiring SAMPLE FLOW WARN Sample flow rate is <...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 184 Table 11-2: Test Functions - Indicated Failures TEST FUNCTIONS ( As Displayed) INDICATED FAILURE(S) TIME Time of day clock is too fast or slow To adjust See Section 5.6. Battery in clock chip on CPU board may be dea...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 185 Figure 11-2: Example of Signal I/O Function (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). Note Any I/O signals changed while in the signal I/O menu will remain in effect O...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 186 11.2. USING THE INTERNAL ELECTRONIC STATUS LEDS Several LEDs are located inside the instrument to assist in determining if the analyzer’s CPU, I 2 C bus and relay board are functioning properly. 11.2.1. CPU STATUS I...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 187 11.2.2.2. RELAY PCA STATUS LED S D5 (Yellow) – CO2 Sensor Heater D1 (RED) Watchdog Indicator Figure 11-4: Relay PCA Status LEDS Used for Troubleshooting Table 11-4: Relay PCA Status LED Failure Indications LED Color...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 188 In general, flow problems can be divided into three categories: 1. Flow is too high 2. Flow is greater than zero, but is too low, and/or unstable 3. Flow is zero (no flow) When troubleshooting flow problems, it is c...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 190 upstream of the flow sensor. First, check the sample filter and make sure it is not plugged and then systematically check all the other components upstream of the orifice to ensure that they are not obstructed. If t...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 191 1. Check for leaks in the pneumatic systems as described in Section 10.3.3. Don’t forget to consider pneumatic components in the gas delivery system outside the T801 Such as: A change in zero air source such as am...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 192 11.5.1. TEMPERATURE PROBLEMS Individual control loops are used to maintain the set point of the temperatures to both sensors. If any of these temperatures are out of range or are poorly controlled, the T801 will per...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 193 Table 11-5: DC Power Test Point and Wiring Color Codes NAME TEST POINT# TP AND WIRE COLOR Dgnd 1 Black +5V 2 Red Agnd 3 Green +15V 4 Blue -15V 5 Yellow +12V Ret (ground) 6 Purple +12V 7 Orange A voltmeter should be ...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 194 11.6.4. TOUCH SCREEN INTERFACE Verify the functioning of the touchscreen by observing the display when pressing a touchscreen control button. Assuming that there are no wiring problems and that the DC power supplies...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 195 11.6.8. PRESSURE/FLOW SENSOR ASSEMBLY The pressure/flow sensor PCA can be checked with a Voltmeter using the following procedure which, assumes that the wiring is intact, and that the motherboard and the power suppl...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 196 11.6.9. MOTHERBOARD 11.6.9.1. A/D FUNCTIONS The simplest method to check the operation of the A-to-D converter on the motherboard is to use the Signal I/O function under the DIAG menu to check the two A/D reference ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 197 11.6.9.3. STATUS OUTPUTS The procedure below can be used to test the Status outputs: 1. Connect a jumper between the “D“ pin and the “ ” pin on the status output connector. 2. Connect a 1000 ohm resistor between t...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 198 11.6.11. RS-232 COMMUNICATIONS 11.6.11.1. GENERAL RS-232 TROUBLESHOOTING Teledyne API analyzers use the RS-232 communications protocol to allow the instrument to be connected to a variety of computer-based equipment...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Troubleshooting and Service 199 11.6.12. CO 2 SENSOR STATUS LED’S There are Two LEDs located on the CO 2 sensor PCA. Figure 11-6: Location of Diagnostic LEDs on CO 2 Sensor PCA Normal Operation: V8 is not lit – V9 is Blinking Error State: Both...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 200 Figure 11-7: Critical Flow Restrictor Assembly / Disassembly 11.7.2. DISK-ON-MODULE REPLACEMENT PROCEDURE ATTENTION COULD DAMAGE INSTRUMENT AND VOID WARRANTY Servicing of circuit components requires electrostatic di...
Troubleshooting and Service Teledyne API T801 NDIR CO2 Analyzer Operation Manual 202 QUESTION ANSWER How do I perform a leak check? See Section 10.3.3. How do I measure the sample flow? Sample flow is measured by attaching a calibrated rotameter, wet test meter, or other flow-measuring device to the...
203 12. PRINCIPLES OF OPERATION The T801 analyzer is a microprocessor-controlled analyzer that determines the percent concentration of carbon dioxde (CO 2 ) in a sample gas drawn through the instrument. The carbon dioxide measurement is achieved using infrared absorption technology. Calibration of t...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 204 Figure 12-1. CO 2 Sensor Theory of Operation The sensor computes the ratio between the reference signal and the measurement signal to determine the degree of light absorbed by CO 2 present in the sensor chamber. This du...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 205 Figure 12-2. CO 2 Sensor PCA Layout and Electronic Connections 12.3. PNEUMATIC OPERATION CAUTION G ENERAL S AFETY H AZARD It is important that the sample airflow system is both leak-tight and not pressurized over ambien...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 206 Figure 12-3: Internal Pneumatic Flow 12.4. FLOW RATE CONTROL To maintain a constant flow rate of the sample gas through the instrument, the T801 uses a special flow control assembly located in the exhaust gas line just ...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 207 the ratio of upstream pressure to downstream pressure is greater than 2:1, the velocity of the gas through the orifice reaches the speed of sound. As long as that ratio stays at least 2:1 the gas flow rate is unaffected...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 208 12.4.3. PNEUMATIC SENSORS 12.4.3.1. SAMPLE PRESSURE SENSOR An absolute value pressure transducer plumbed to the outlet of the sample chamber is used to measure sample pressure. The output of the sensor is used to compen...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 209 Finally, the CPU issues commands via a series of relays and switches (also over the I 2 C bus) located on a separate printed circuit assembly to control the function of key electromechanical devices such as heaters. Flo...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 210 Figure 12-6. CPU Card The CPU includes two types of non-volatile data storage: an embedded 2MB flash chip and a Disk on Module (DOM). 12.5.2.1. DISK-ON-MODULE (DOM) The DOM is a 44-pin IDE flash disk with a storage capa...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 211 Power Connection for DC Heaters Status LED’s (D2 through D16) DC Power Supply Test Points Watchdog Status LED (D1) (JP5) Thermocouple Configuration Jumpers Thermocouple Signal Output I 2 C Connector Valve Control Driver...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 212 AC Relay Retainer Plate Retainer Mounting Screws Figure 12-8: Relay PCA with AC Relay Retainer in Place 12.5.3.1. STATUS LEDS LEDs located on the Analyzer’s relay PCA, show the current status of various control function...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 213 D5 (Yellow) –CO2 Sensor Heater D1 (RED) Watchdog Indicator Figure 12-9: Status LED Locations – Relay PCA 12.5.3.2. WATCHDOG CIRCUITRY The most important of the status LEDs on the relay board is the red I 2 C Bus watch-d...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 214 the analog to digital converter (A/D). Under the control of the CPU, this functional block selects a particular signal input (e.g. BOX TEMP) and then converts the selected voltage into a digital word. The A/D consists o...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 215 connected back to the A/D converter through a Loop-back circuit. This permits the voltage outputs to be calibrated by the CPU without need for any additional tools or fixtures. 12.5.4.6. EXTERNAL DIGITAL I/O This Extern...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 216 CAUTION G ENERAL S AFETY H AZARD Should the AC power circuit breaker trip, investigate and correct the condition causing this situation before turning the analyzer back on. Figure 12-10: Power Distribution Block Diagram...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 217 Figure 12-11: Front Panel and Display Interface Block Diagram 12.5.6.1. LVDS TRANSMITTER BOARD The LVDS (low voltage differential signaling) transmitter board converts the parallel display bus to a serialized, low volta...
Principles of Operation Teledyne API T801 NDIR CO2 Analyzer Operation Manual 218 12.5.7. SOFTWARE OPERATION The T801 Analyzer is at its heart a high performance, VortexX86-based microcomputer running Windows CE. Inside Windows CE special software developed by Teledyne API interprets user commands vi...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual Principles of Operation 219 12.5.9. CALIBRATION - SLOPE AND OFFSET Calibration of the analyzer is performed exclusively in software. During instrument calibration the user enters expected values for span via the front panel keypad; values are not ...
221 13. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne API considers the prevention of damage caused by the discharge of static electricity to be extremely important part of making sure that your analyzer continues to provide reliable service for a long time. This section describes how static electri...
A Primer on Electro-Static Discharge Teledyne API T801 NDIR CO2 Analyzer Operation Manual 222 Table 13-1: Static Generation Voltages for Typical Activities MEANS OF GENERATION 65-90% RH 10-25% RH Walking across nylon carpet 1,500V 35,000V Walking across vinyl tile 250V 12,000V Worker at bench 100V 6...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual A Primer on Electro-Static Discharge 223 Any time a charged surface (including the human body) discharges to a device. Even simple contact of a finger to the leads of a sensitive device or assembly can allow enough discharge to cause damage. A s...
A Primer on Electro-Static Discharge Teledyne API T801 NDIR CO2 Analyzer Operation Manual 224 o A charge can be induced onto the conductive surface and/or discharge triggered in the presence of a charged field such as a large static charge clinging to the surface of a nylon jacket of someone walking...
A Primer on Electro-Static Discharge Teledyne API T801 NDIR CO2 Analyzer Operation Manual 228 13.4.2.5. PACKING COMPONENTS FOR RETURN TO TELEDYNE API TECHNICAL SUPPORT Always pack electronic components and assemblies to be sent to Teledyne API Technical Support in anti-ESD bins, tubes or bags. CAUTI...
229 GLOSSARY Note: Some terms in this glossary may not occur elsewhere in this manual. Term Description/Definition 10BaseT an Ethernet standard that uses twisted (“T”) pairs of copper wires to transmit at 10 megabits per second (Mbps) 100BaseT same as 10BaseT except ten times faster (100 Mbps) APICO...
233 INDEX A AC Power 60 Hz, 32 AIN, 99 ALRM, 61, 101 ANALOG CAL WARNING, 51 Analog Inputs, 99 Analog Outputs, 33, 35, 61, 64, 65, 80 Ain Calibration, 99 CONC1, 52 CONC2, 52 Configuration & Calibration, 61, 86, 87, 88, 89, 91, 93, 95, 96, 99 Automatic, 27, 61, 89 Manual-Current Loop, 92, 94 Manua...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual 234 CONC ALRM1 WARNING, 125 CONC ALRM2 WARNING, 125 CONC Key, 78 CONC VALID, 195 CONC_PRECISION, 78 CONC1, 52, 53 CONC2, 52, 53 Concentration Alarms, 101 Concentration Field, 27 CONFIG INITIALIZED, 51, 180 Contact, 225 Continuous Emission Monitori...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual A Primer on Electro-Static Discharge 235 Current Loop, 92 Voltage Ranges, 90 Analog Outputs pin assignments, 65 Control Inputs, 38 Ethernet, 108 Modem, 153 MULTIDROP, 46 Serial/COMM Ports, 41 Status Outputs, 36 Electro-Static Discharge, 23, 43, 22...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual 236 CO2 M-P CAL, 58 SPAN CAL [type], 58 ZERO CAL [type], 58 DIAG Mode, 58 Diagnostic Mode (DIAG), 80 SAMPLE A1, 58 Sample Mode, 27, 58, 78 Secondary Setup, 61 SETUP [X.X], 58 Optional Sensors CO2 Internal Pneumatics, 186 Outlet, 49 P Particulate F...
Teledyne API T801 NDIR CO2 Analyzer Operation Manual A Primer on Electro-Static Discharge 237 Status CAL MODE, 37 CO 2 Output, 37 CONC VALID, 37 RANGE2 CAL, 37 SPAN CAL, 37 SYSTEM OK, 37 Status LED’s I 2 C, 184 Relay PCA, 184, 210, 211 Watchdog, 184, 185, 210, 211 Status LED's CO 2 Sensor, 197 CPU, ...
Appendix A Models T801, 801E Software Menus (Reference: 06843C DCN6418) A-1 APPENDIX A – Version Specific Software Documentation NOTE: Appendix A-2 through Appendix A-7 include parameters for all models in the 800 Series; the footnotes in each table clarify differences among the models. APPENDIX A-1...
Appendix A Models T801, 801E Software Menus (Reference: 06843C DCN6418) A-3 APPENDIX A-1: Models T80X, 80XE Menu Trees, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) PRIMARY SETUP MENU SAMPLE MSG 1 CLR 1 SETUP <TST TST> CAL TEST 1 CO2 CONC ZERO SPAN DAS CFG ACAL CLK RNGE PASS MORE 1 Only a...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A A-4 SETUP PASS DAS RNGE CLK MORE ACAL CFG <TST TST> MODEL TYPE AND NUMBER PART NUMBER SERIAL NUMBER SOFTWARE REVISION LIBRARY REVISION iCHIP SOFTWARE REVISION HESSEN PROTOCOL REVISION 2 CPU TYPE & OS REVISION DATE FACT...
Appendix A Models T801, 801E Software Menus (Reference: 06843C DCN6418) A-7 SETUP PASS DAS RNGE CLK MORE ACAL CFG SAMPLE COMM VARS DIAG Go to DIAG Menu Tree HESN 2 INET 1 ID ENTER PASSWORD: 818 ENTER PASSWORD: 818 ENTER PASSWORD: 818 1 E-Series: only appears if Ethernet Option is installed. 2 Only a...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-9 APPENDIX A-2: Setup Variables For Serial I/O, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) Table A-1: Setup Variables Setup Variable Numeric U...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-17 APPENDIX A-3: Warnings and Test Measurements, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) Table A-2: Warning Messages Name 1 Message Text De...
Appendix A Models T801, 801E Software Menus (Reference: 06843C DCN6418) A-18 Error! Unknown document property name.Error! Unknown document property name.Error! Unknown documen Table A-3: Test Measurements Name 1 Message Text Description Test Measurements O2RANGE 5 CO2RANGE 4 O2 RNG=500.0 % CO2 RNG =...
Appendix A Models T801, 801E Software Menus (Reference: 06843C DCN6418) A-20 Error! Unknown document property name.Error! Unknown document property name. Error! Unknown document property name.Error! Unknown document property name. APPENDIX A-4: Signal I/O Definitions, Software Versions 1.0.3 (T-Seri...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-21 Signal Name Bit or Channel Number Description A status outputs, U24, J1017, pins 1 – 8 = bits 0 – 7, default I/O address 323 hex ST_SYSTEM_OK 0 0 ...
Appendix A Models T801, 801E Software Menus (Reference: 06843C DCN6418) A-22 Error! Unknown document property name.Error! Unknown document property name. Error! Unknown document property name.Error! Unknown document property name. Signal Name Bit or Channel Number Description CO2_SPAN_VALVE 2, 6 8 0...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-23 Signal Name Bit or Channel Number Description External analog input board, default I 2 C address 5C hex XIN1 7 0 External analog input 1 XIN2 7 1 ...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-27 APPENDIX A-6: Terminal Command Designators, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) Table A-7: Terminal Command Designators COMMAND ADDI...
Models T801, 801E Software Menus (Reference: 06843C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-29 APPENDIX A-7: MODBUS ® Register Map, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) MODBUS Register Address (dec., 0-based) Description Units M...
APPENDIX B - Spare Parts Note Use of replacement parts other than those supplied by Teledyne Advanced Pollution Instrumentation (TAPI) may result in non-compliance with European standard EN 61010-1. Note Due to the dynamic nature of part numbers, please refer to the TAPI Website at
T80X Spare Parts List (Ref: 072690000 A DCN6431, 2012 April 12) PARTNUMBER DESCRIPTION 000940700 CD, ORIFICE, .005 YELLOW 001763500 ASSY, FLOW CTL, 110CC, 1/4" ELBOW ‐ B 003290000 THERMISTOR, BASIC (VENDOR ASSY)(KB) 009690200 AKIT, TFE FLTR ...
Appendix C Warranty/Repair Questionnaire T80X, M80XE (06532C DCN 5798) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 C-1 CUSTOMER: _______________________________ PHONE: _____________________________________ ...
Appendix C Warranty/Repair Questionnaire T80X, M80XE (06532C DCN 5798) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 C-2 ____________________________________________________________________________________ If...
APPENDIX D – Wire List and Electronic Schematics 07274B DCN6418 D-1
Advanced Pollution Instrumentation A Teledyne Technologies Company TELEDYNE ' & % $ % 7KHLQIRUPDWLRQKHUHRQLVWKHSURSHUW\RI $3,DQGLVVXEPLWWHGLQVWULFWHVW FRQILGHQFHIRUUHIHUHQFHRQO\ 8QDXWKRUL]HGXVHE\DQ\RQHIRUDQ\ RWKHUSXUSRVHVLVSURKLWHGWKLV GRFXPHQWRUDQLQIRUPDWLRQFRQWDLQHG LQLWPD\QRWEHGXSOLFDWHGZLWKR...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P1.R3.schdoc Drawn By: GUI Interface 06698 D 1 4 2 3 4 5 6 7 8 1 51 52 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 ...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P2.R3.schdoc Drawn By: 06698 D 2 4 22uH L1 Vin 5 SHDN 4 SW 1 GN D 2 FB 3 U1 CAT4139TD-GT3 4.7uF/16V C9 1K R12 2.0 R14 5V-GND +5V 9.76 R13 3.9uH L2 487K R9 309K R8 80.6K R18 66.5K R1...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P3.R3.schdoc Drawn By: 06698 D 4 3 12345 To old TScreen 70553-004 NI J11 LLRLSDRTLT +5V CHASSIS 1 2 NI J12 BUS +5 12345 To new TScreen 70553-004 J10 LT 1 RT 2 SH L D 3 RL 4 LL 5 GN ...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P4.R3.schdoc Drawn By: 06698 D 4 4 TOUCH SCREEN INTERFACE CIRCUITRY ( TBD) GUI Interface R40 10K O p tion CLKOUT_P CLKOUT_N 3.3V C49 0.1 R41100 FB6 FB14 Vcc PIN 28 C47 0.01 22uF/6.3...
1 1 2 2 3 3 4 4 D D C C B B A A Title Number Revision Size A Date: 5/7/2010 Sheet of File: N:\PCBMGR\..\06882-P1-R0.SchDoc Drawn By: LVDS, Transmitter Board B 1 1 RT 06882 VAD6VAD7VAD8VAD9VAD10VAD11VBD10VBD11VAD0VAD1VAD2VAD3VBD2VBD3VBD4VBD5VBD6VBD7 VBDE Y0_NY0_PY1_NY1_PY2_NY2_P CLKOUT_NCLKOUT_P R2 2...
1 1 2 2 3 3 4 4 D D C C B B A A Title Number Revision Size A Date: 5/6/2011 Sheet of File: N:\PCBMGR\..\06731-1_ETHERNET.SchDoc Drawn By: Auxiliary I/O Board (PWR-ETHERNET) B 3 1 RT 06731 STRAIGHT THROUGH ETHERNET GND +5V SCL SDA +5V-ISO + C17 100uF GND DS3 GRN 12345678 P2 Header 8 1 2 3 4 5 6 U6 SP...
Teledyne Manuals
-
Teledyne 4220
Manual
-
Teledyne 311-D
Manual
-
Teledyne HPM 4/6
Manual
-
Teledyne T802
Manual
-
Teledyne 318R
Manual
-
Teledyne 9700
Manual
-
Teledyne BDS-3960
Manual
-
Teledyne C-2000
Manual
-
Teledyne GFC 7001EM
Manual
-
Teledyne 2103
Manual
-
Teledyne 6400E
Manual
-
Teledyne AX300-I
Manual
-
Teledyne 6600
Manual
-
Teledyne 9950
Manual
-
Teledyne PEM 9002
Manual
-
Teledyne HPM-2002-OBE
Manual
-
Teledyne Infrared Gas Analyzer
Manual
-
Teledyne 102E
Manual
-
Teledyne OT-3
Manual
-
Teledyne LLG
Manual