Page 3 - TABLE OF CONTENTS
M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) TABLE OF CONTENTS 1. PREFACE ...................................................................................................................................... 1 1.1. Reference Numbering convention..........................................
Page 4 - LIST OF APPENDICES; ii
M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 8. INSTRUMENT MAINTENANCE ..................................................................................................... 39 8.1. Additional and Updated Maintenance Procedures .............................................................
Page 5 - PREFACE; LIST OF FIGURES; iii
M102E/M501 TRS PREFACE (Addendum to M101E Manual - P/N 04740 Rev A) LIST OF FIGURES Figure 3-1: M102E Internal Layout ................................................................................... 7 Figure 3-2: M501-TRS Internal Layout...............................................................
Page 7 - NOTE
M102E/M501 TRS PREFACE (Addendum to M101E Manual - P/N 04740 Rev A) 1. PREFACE NOTE The information contained in this addendum is pertinent to M102E analyzers running software revision A.2. Some or all of the information may not be applicable to previous revision of that software. The software revis...
Page 8 - Reference Numbering convention
PREFACE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 1.1. Reference Numbering convention Unless otherwise specified, chapter, section, figure and table reference numbers referred to within this text are relative to this document. EXAMPLE: “Figure 2-1” refers to the figure, within this...
Page 9 - SPECIFICATIONS, APPROVALS AND WARRANTY; SPECIFICATIONS, APPROVALS AND; Specifications; M501-TRS Specifications; Table 2-1: Model 102E Basic Unit Specifications; EPA Equivalency Designation; when operated under the following conditions:
M102E/M501 TRS SPECIFICATIONS, APPROVALS AND WARRANTY (Addendum to M101E Manual - P/N 04740 Rev A) 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. Specifications There are no significant differences between the performance specifications for the M102E and the M102E as listed in Section 2.1 of the M10...
Page 10 - CE Mark Compliance
SPECIFICATIONS, APPROVALS AND WARRANTY M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) • Sample filter: Equipped with PTFE filter element in the internal filter assembly. • Sample flow of 650 +/- 65 cc/min. • Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 s...
Page 11 - GETTING STARTED; There are no shipping screws to be removed in the M102E.; Unpacking the M501-TRS; Remove the screws fastening the top cover to the unit (four per side).
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3. GETTING STARTED 3.1. Unpacking the M102E Unpack the M102E as per the directions ins Section 3.1 of the M101E Manual - P/N 04740 Rev A. 1. There are no shipping screws to be removed in the M102E. 3.2. Unpacking the M501-TR...
Page 13 - Internal Layouts
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.3. Internal Layouts Figures 3-1 & 3-3 supersede Figure 3-9 of the M101E Manual - P/N 04740 Rev A. Front Panel UV Source Lap Particulate Filter Reaction Cell Hydrocarbon Scrubber Hidden from view beneath Reaction Cell P...
Page 14 - Internal Pneumatic Flow of the M102E & the; Switching Valve Operating Modes
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.4. Internal Pneumatic Flow of the M102E & the M501-TRS Figure 3-3 shows the internal pneumatic flow of the M102E in its Standard configuration. For information on instruments in which one of the various zero/span valve...
Page 15 - Rear Panel Layout for the M102E & M501-TRS; Amp
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.5. Rear Panel Layout for the M102E & M501-TRS Figures 3-4 & 3-5 supersede Figure 3-2 of the M101E Manual - P/N 04740 Rev A. Figure 3-4: M102E Rear Panel Layout 5 Amp Slow-Blow Fuse Alarm Output Connections Figure 3...
Page 16 - Initial Setup; Electrical Connections:; M102E Analog Output Connections; Table 3–2: Analog output Pin Outs
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.6. Initial Setup 3.6.1. Electrical Connections: The electrical connections for the M102E are the same as those described in Section 3.1.1 of the M101E Manual - P/N 04740 Rev A except for the test channel analog output: 3.6...
Page 17 - M501-TRS Alarm Output Connections; Make sure the load does not exceed the rated capacity of the relay.; Pneumatic Connections:; CAUTION; M102E PNEMATIC CONNECTERS; M501-TRS PNEMATIC CONNECTERS
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) • TO change these settings, see Sections 6.9.4 and 6.7 of the M101E Manual - P/N 04740 Rev A respectively. 3.6.1.2. M501-TRS Alarm Output Connections The rear panel of the M501-TRS includes a terminal strip by which connecti...
Page 18 - MODEL
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) FROM ANALYZER Sample gas enters the M501-TRS from the M102E via this port. TO ANALYZER Sample gas leaves the M501-TRS to return to the M102E via this port. MODEL 102E Sample Zero Air Exhaust Span MODEL 701 Zero Air Generator...
Page 20 - Calibration Gases; ZERO AIR; , Dimethyl disulfide; Connections with Internal Valve Options Installed
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Table 3-4: NIST-SRM's Available for Traceability of H 2 S & SO 2 Calibration Gases NIST-SRM 4 TYPE NOMINAL CONCENTRATION 2730 2731 Hydrogen sulfide in N 2 Hydrogen sulfide in N 2 5000 ppb 20 ppm ZERO AIR • A gas that is ...
Page 22 - Option; Valve Option
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Source of SAMPLE Gas VENT if input is pressurized MODEL 700 Gas Dilution Calibrator (with O 3 generator option) VENT Calibrated H 2 S gas (At high concentration) MODEL 102E From Converter Sample Zero Air Exhaust M501-TRS To ...
Page 23 - Initial Operation; Startup / Warm Up of the M102E; Possible Warning Messages at Start-Up; Functional Check of the M102E; Test Functions
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.7. Initial Operation 3.7.1. Startup / Warm Up of the M102E Startup procedures and warm up behavior of the M102E are identical to those described in Sections 3.2.1 and 3.2.2 of the M101E Manual - P/N 04740 Rev A. Possible W...
Page 24 - the converter oven temperature rises above the lower alarm limit.; Display area
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.7.3. Startup / Warm Up of the M501-TRS After electrical and pneumatic connections are made, turn on the instrument and pump power. The exhaust fan should start. The M501-TRS’ temperature controller is preprogrammed at the ...
Page 25 - Initial Calibration; Initial calibration of the M102E should be performed with:
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.8. Initial Calibration Initial calibration of the M102E should be performed with: • Zero air supplied by a zero air generator such as the Teledyne Instruments’ M701; • Calibrated H 2 S span gas of the appropriate concentra...
Page 27 - OPTIONAL HARDWARE AND SOFTWARE; Rack mount brackets with 26 in. chassis slides.; Calibration Valves Options
M102E/M501 TRS OPTIONAL HARDWARE AND SOFTWARE (Addendum to M101E Manual - P/N 04740 Rev A) 4. OPTIONAL HARDWARE AND SOFTWARE This section includes descriptions of the hardware and software options available for the Model 102E analyzer and M501-TRS converter that are different from or not included in...
Page 29 - Table 4-2: IZS Valve Operating States
M102E/M501 TRS OPTIONAL HARDWARE AND SOFTWARE (Addendum to M101E Manual - P/N 04740 Rev A) ` FLOW SENSOR FLOW / PRESSURE SENSOR PCA SAMPLE PRESSURE SENSOR EXHAUST TO OUTER LAYER OF KICKER 3 2 1 SAMPLE/C AL VALVE Z ERO/SPAN VALVE 3 2 1 2 3 1 ZERO AIR INLET V ACU UM M ANI F O L D CRITICAL FLOW ORIFICE...
Page 30 - Additional Manuals; The CD version of the addendum has many advantages:
OPTIONAL HARDWARE AND SOFTWARE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 4.3. Additional Manuals 4.3.1. Printed Manuals (P/N 049880000) Additional printed copies of this addendum are available from Teledyne Instruments 4.3.2. Addendum on CD (Part number 049880200) This addendum is ...
Page 31 - M102E OPERATING INSTRUCTIONS; M102E Analog Output Signals; rather than
M102E/M501 TRS M102E OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 5. M102E OPERATING INSTRUCTIONS NOTE For the most part the operation instruction for the M102Eare the same as those described in Chapter 6 of the M101E Manual - P/N 04740 Rev A with the exception that the terms ...
Page 32 - Setting the M102E Gas Measurement Mode; Table 5-1 M102E gas Measurement Modes; SETUP – DIAG: Using the Diagnostics Functions; M102E Analog I/O Configuration; Table 5-2: Analog Output Pin Assignments
M102E OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) The output, labeled A3 is special. It can be set by the user (see Section 6.9.10 of the M101E Manual - P/N 04740 Rev A) to output many of the parameters accessible through the <TST TST> keys of the units S...
Page 33 - M102E Test Channel Output; Table 5-3: Test Parameters Available for Analog Output A4; output and; SETUP – COMM: Setting Up the M102E’s; The default ID code for all M102E analyzers is; M102E Ethernet Host Name
M102E/M501 TRS M102E OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 5.2.2. M102E Test Channel Output The following table supercedes Table 6-14 of the M101E Manual - P/N 04740 Rev A Table 5-3: Test Parameters Available for Analog Output A4 TEST CHANNEL TEST PARAMETER RANGE 1 NONE...
Page 34 - Remote Operation of the Analyzer; Control Inputs; Table 5-4: M102E Control Input Pin Assignments; CONTROL IN; Figure 5-2: Control Inputs with local 5 V power supply
M102E OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 5.4. Remote Operation of the Analyzer 5.4.1. Control Inputs The description of the control inputs in Section 6.12.1.2 of the M101E Manual - P/N 04740 Rev A in correctly shows an external low span cal input. This...
Page 35 - Figure 5-3: Control Inputs with external 5 V power supply; Using the M102E with a Hessen Protocol Network; M102E Hessen Protocol Gas ID List.; The default Hessen Gas Id’s for all M102E analyzers are:; Table 5-5: M102E Default Hessen Gas ID’s
M102E/M501 TRS M102E OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) CONTROL IN - + 5 VDC Power Supply LO W S P A N ZE R O SPAN A B C D E F U + Figure 5-3: Control Inputs with external 5 V power supply 5.4.2. Using the M102E with a Hessen Protocol Network 5.4.2.1. M102E Hessen Pr...
Page 36 - Setting Hessen Protocol Status Flags; Table 5-6: Default Hessen Status Bit Assignments
M102E OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 5.4.2.2. Setting Hessen Protocol Status Flags The following table supercedes Table 6-29 of the M101E Manual - P/N 04740 Rev A/ Table 5-6: Default Hessen Status Bit Assignments STATUS FLAG NAME DEFAULT BIT ASSIGN...
Page 37 - M501-TRS OPERATING INSTRUCTIONS; DO NOT OPERATE WITHOUT THE COVER OF THE M501TS CONVERTER; Basic M501-TRS Controls; Figure 6–1: M501-TRS Temperature Controls
M102E/M501 TRS M501-TRS OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 6. M501-TRS OPERATING INSTRUCTIONS CAUTION ! DO NOT OPERATE WITHOUT THE COVER OF THE M501TS CONVERTER INSTALLED. OVEN TEMPERATURE WILL NOT REGULATE PROPERLY WITHOUT THE COVER IN PLACE. NOTE: Changing the Conv...
Page 38 - Table 6-1: M501-TRS Temperature Controls and Definitions; To Display The Current Temperature:; If the
M501-TRS OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Table 6-1: M501-TRS Temperature Controls and Definitions NAME FUNCTION Main Display A 4-digit, 7 segment LED display on which the current value of the PV and the SV as well as error codes and programming para...
Page 39 - To Manually Adjust the Converter Oven; DO NOT SET THE TEMPERATURE HIGHER THAN 1050OC; Press the up-arrow under that digit once. The digit will flash.
M102E/M501 TRS M501-TRS OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 6.3. To Manually Adjust the Converter Oven Temperature: CAUTION ! DO NOT SET THE TEMPERATURE HIGHER THAN 1050OC 1. Set the main display to show the current value of the set variable by pressing the PV/SV mode...
Page 40 - Autotune the Temperature Controller:; Initiating the Autotune Process:; Note
M501-TRS OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 6.4. Autotune the Temperature Controller: The M501-TRS controller includes an auto tune feature which allows the controller to find and set optimum values for various process control parameters so that the co...
Page 41 - Aborting the Autotune Process:; M501TRS Alarm Relay Adjustment
M102E/M501 TRS M501-TRS OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 6.4.2. Aborting the Autotune Process: 1. Press the DATA key once. The display will show (1). 2. Press the 1’s UP key once. The will begin blinking 3. Press the DOWN key once. The display will show (zero). 4. ...
Page 43 - CALIBRATION PROCEDURES; If you are using the M102E for US-EPA controlled monitoring of SO; M501-TRS Calibration; The M501-TRS converter does not require field calibration.
M102E/M501 TRS CALIBRATION PROCEDURES (Addendum to M101E Manual - P/N 04740 Rev A) 05514 Rev A1 37 7. CALIBRATION PROCEDURES 7.1. M102E Calibration Calibration of the M102E should be performed according to the procedures described in Chapters 7 & 8 of the M101E Manual - P/N 04740 Rev A. NOTE It ...
Page 45 - INSTRUMENT MAINTENANCE
M102E/M501 TRS INSTRUMENT MAINTENANCE (Addendum to M101E Manual - P/N 04740 Rev A) 05514 Rev A1 39 8. INSTRUMENT MAINTENANCE The following table supercedes Table 9-1 of the M101E Manual - P/N 04740 Rev A Table 8-1: M102E Preventive Maintenance Schedule ITEM ACTION FREQUENCY CAL CHECK M101E MANUAL SE...
Page 46 - Additional and Updated Maintenance Procedures; Maintaining the SO; Predicting When the SO; MEASURE MODE
INSTRUMENT MAINTENANCE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 8.1. Additional and Updated Maintenance Procedures The following procedures need to be performed regularly as part of the standard maintenance of the Model 102E. 8.1.1. Maintaining the SO 2 Scrubber This section REPLA...
Page 47 - Checking the Function of the SO
M102E/M501 TRS INSTRUMENT MAINTENANCE (Addendum to M101E Manual - P/N 04740 Rev A) 8.1.1.2. Checking the Function of the SO 2 Scrubber To check to see if your SO 2 scrubber is operating properly perform the following test: 1. Set the analyzer for TRS gas measurement mode (see 6.8.1 of the M101E Manu...
Page 49 - THEORY OF OPERATION; Measurement Principle; TRS Conversion
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) 9. THEORY OF OPERATION The M102E is a modified M101E which, when used in conjunction with a M501-TRS determines the concentration of total reduced sulfur (TRS), in a sample gas drawn through the instrument In most ways t...
Page 50 - Figure 9-1: UV Absorption in the M102E Reaction Cell; Ia; nm
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) nm. The SO 2 molecules absorbs some of energy from the UV light causing one of the electrons of each of the affected molecules to move to a higher energy orbital state. * 2 SO hv SO Ia nm 214 2 ⎯ ⎯ → ⎯ + (Equation 9-2) T...
Page 51 - hv
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) The amount of detectable UV given off by the decay of the SO 2 * is affected by the rate at which this reaction occurs ( k ). ( ) * SO k F 2 = (Equation 9-5) Where: F = the amount of fluorescent light given off. k = The ...
Page 52 - The UV Light Path; UV Lamp Shutter & PMT Offset; DRK PMT; Pneumatic Operation; Sample gas Flow
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9.2. The UV Light Path The following information is in addition to that contained in Section 10.2 of the M101E Manual - P/N 04740 Rev A. 9.2.1. UV Lamp Shutter & PMT Offset Inherent in the operation of both the refer...
Page 53 - Electronic Operation; Sensor Module; PMT HOUSING
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) 9.4. Electronic Operation 9.4.1. Sensor Module This Section replaces Section 10.4.2 of the M101E Manual - P/N 04740 Rev A. Electronically, the M102E sensor module is a group of components that: create the UV light that i...
Page 54 - Sample Chamber
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9.4.1.1. Sample Chamber The main electronic components of the sample chamber are the reference detector(see Section 10.2.2 of the M101E Manual - P/N 04740 Rev A); the UV Lamp (see Section 10.2.1 of the M101E Manual - P/N...
Page 55 - M501-TRS electronics; HEATER; CONTROLLER; Thermal Switch
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) 9.4.2. M501-TRS electronics Electronically the M501-TRS is a simple device. The nucleus of the instrument is a programmable, P-I-D temperature controller which manages the temperature of the instruments’ AC powered conve...
Page 56 - Temperature Alarms and Alarm Output; fall below the level of the low alarm point.
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9.4.2.2. Temperature Alarms and Alarm Output The M501-TRS’ controller has two user settable alarm points: High Alarm and Low Alarm. A single SPST alarm output is triggered should the PV rise above the set level of the hi...
Page 57 - TROUBLESHOOTING & REPAIR; Fault Diagnosis with Warning Messages; M102E Warning Messages
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 10. TROUBLESHOOTING & REPAIR This section includes various troubleshooting and repair information that is either in addition to that included in Chapter 11 of the M101E Manual (P/N 04740 Rev A) CAUTION The o...
Page 58 - Fault Diagnosis with Test Functions
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 10.1.2. Fault Diagnosis with Test Functions The Following table supercedes Table 11.2 of the M101E Manual - P/N 04740 Rev A. Table 10-2: Test Functions - Possible Causes for Out-Of-Range Values TEST FUNCTION NOM...
Page 59 - M501-TRS Trouble shooting; TRS Converter Not Heating:; Other Performance Problems; Excessive noise
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 10.2. M501-TRS Trouble shooting 10.2.1. TRS Converter Not Heating: Problems with heating the If the TRS converter oven can have several causes. • The “UUUU” error code will is displayed on the M501-TRS temperatu...
Page 60 - Subsystem Checkout; Checking the Efficiency of the M501-TRS SO
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 10.4. Subsystem Checkout In addition to the information contained in Section 11.5 of the M101E Manual - P/N 04740 Rev A, the following diagnostic procedures are useful for troubleshooting and diagnosing problems...
Page 61 - Additional Repair Procedures; UV Lamp Adjustment and/or Replacement; Lamp Aging
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 10. If TRS Æ SO 2 converter efficiency is below 90% check for one of the following possible causes: • Pneumatic leak • Perform a leak check (with the M501-TRS connected) as described in Section 11.5.1 of the M10...
Page 62 - Lamp Positioning; Adjusting the UV Lamp; Peaking the Lamp; ALWAYS wear UV-Protective, Safety Glasses when working with the UV
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Lamp Positioning – The UV output level of the lamp is not even across the entire length of the lamp. Some portions of the lamp shine slightly more brightly than others. At the factory the position of the UV lamp...
Page 63 - Replacing the UV Lamp; Always grasp the main body of the lamp.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) UV Lamp Bracket Mounting Screws Shutter Housing Shutter Assy UV Filter Retainer & Lens Housing UV Lamp Power Supply Wires Thumb Screw DO NOT use Lamp Cap to adjust Lamp position Reaction Cell Adjust Lamp Pos...
Page 64 - Replacing the UV filter/lens; affect the accuracy of the analyzer’s SO; Unplug the J4 connector from the motherboard to allow tool access.; Figure 10-2: Disassembling the Shutter Assembly
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9. Finger tighten the thumbscrew. NOTE: DO NOT over-tighten the thumbscrew. 10. Perform a lamp calibration procedure (see Section 6.9.7 of the M101E Manual - P/N 04740 Rev A) and a zero point and span point cali...
Page 65 - Replacing the PMT, HVPS or TEC; pneumatic and electrical connections from the sensor assembly.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 5. Carefully remove the UV filter. 6. Install the UV filter. Handle carefully and never touch the filter’s surface. The UV filter’s wider ( ring ) side should be facing out. Install the UV filter retainer and ti...
Page 67 - M102E PMT Hardware Calibration (; FACTORY CAL; SNGL
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 15. Re-insert the PMT/HVPS subassembly. • Don’t forget the gasket between HVPS and PMT. • Use new plastic screws to mount the PMT assembly on the PMT cold block. 16. Insert the new desiccant bags. 17. Carefully ...
Page 68 - STABIL; Feed 400 ppb span gas into the analyzer.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 6. Locate the Preamp board (see Figure 3-1). 7. Locate the Following Components On the Preamp board (see Figure 10-4): • HVPS coarse adjustment switch (Range 0-9, then A-F) • HVPS fine adjustment switch (Range 0...
Page 69 - Replacing the TRS Converter Heating Tube
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) If a reporting range other than 500 ppb is used in this procedure: Use a span gas equal to 80% of the reporting range and adjust the PMT to a target NORM PMT value of twice the ppb value of the span gas. EXAMPLE...
Page 70 - The ceramic bobbins at each end of the heater assembly are fragile.; Manually Programming the M501-TRS
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 2. Allow it to cool to room temperature. This may take up to 30 minutes. 3. Remove the instruments top cover (see Section 3.2 of this addendum) 4. Remove the four screws holding the converter cover in place. 5. ...
Page 71 - Temperature Controller Primary Menu Parameters; To set these parameters:
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) DO NOT manually alter the PID parameters of the M501-TRS temperature Controller unless directed to do so by Teledyne Instruments customer service personnel. In the event that the temperature controller of the M5...
Page 72 - Table 10-3 – Temperature Controller – Primary Parameter Settings
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Table 10-3 – Temperature Controller – Primary Parameter Settings DISPLAY PARAMETER NAME SET TO COMMENTS Proportional Band 11 Sets the bandwidth of the proportional control function to ± 11% of full scale Integra...
Page 73 - Table 10-4 – Temperature Controller – Primary Parameter Settings; ON; Technical Assistance
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) Table 10-4 – Temperature Controller – Primary Parameter Settings DISPLAY PARAMETER NAME SET TO COMMENTS OFF 0 Sets the controller to reverse action (hotter temperatures = less heating; lower temperatures = more ...
Page 75 - APPENDIX A - Version Specific Software Documentation
M102E/M501 TRS APPENDIX A - Version Specific Software Documentation (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A - Version Specific Software Documentation APPENDIX A-1: Model 102E Software Menu Trees NOTE The menu tree structure for the M102E is nearly identical to that of the M101E. Incl...
Page 76 - LOW; Figure A-1: Basic Sample Display Menu
APPENDIX A-1: M102E Software Menu Trees, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual) APPENDIX A-1: M102E Software Menu Trees, Revision A.2 ZERO SPAN CONC SAMPLE <TST TST> RANGE STABIL PRES SAMP FL PMT NORM PMT UV LAMP LAMP RATIO STR. LGT DARK PMT DARK LAMP SLOPE OFFSET HVPS RCELL TE...
Page 78 - ACAL
APPENDIX A-1: M102E Software Menu Trees, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual) . YES NO SAMPLE FLOW WARNING BENCH TEMP WARNING SOURCE WARNING BOX TEMP WARNING WHEEL TEMP WARNING SAMPLE TEMP WARNING SAMPLE PRESSURE WARNING INVALID CONC INSTRUMENT OFF IN MANUAL CALIBRATION MODE IN ZER...
Page 79 - APPENDI
X A-1: M102E Software Menu Trees, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual) 515 Rev A.2 A-5 32 INTERNAL ANALOG ↓ VOLTAGE SIGNALS 56 (see Appendix A) CFG ACAL 1 CLK PASS RNGE DAS MORE COMM VARS 2 Only relevant to analyzers with IZS options installed ANALOG OUTPUT LAMP CALIBRATION 0) EXT ...
Page 82 - Rear board primary MUX analog inputs
APPENDIX A-4: M102E Signal I/O Definitions, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A-4: M102E Signal I/O Definitions, Revision A.2 NOTE Signal I/O Definitions for the M102E are the same as those for the M102E (see Appendix A-4 of the M101E Manual - P/N 0474...
Page 85 - APPENDIX B - M102E Spare Parts List
M102E/M501 TRS APPENDIX B - M102E Spare Parts List (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX B - M102E Spare Parts List NOTE Spare parts for the M102E are the same as those for the M101E (see Appendix B of the M101E Manual - P/N 04740120 Rev A) Spare parts available for the M501-TRS are ...
Page 87 - INSTRUMENTS
M102E/M501 TRS Warranty/Repair TELEDYNE Addendum to Appendix C M101E Manual (P/N 04740) Questionnaire Model 102E INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company Company: _________________________ Contact Name: _____________________________ Phone Number: ___________ Fax...
Page 88 - TELEDYNE; Questionnaire
M102E/M501 TRS Warranty/Repair TELEDYNE Addendum to Appendix C M101E Manual (P/N 04740) Questionnaire Model 102E INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company Notes and further information: _____________________________________________________ _______________________...
Page 89 - APPENDIX D - ELECTRONIC SCHEMATICS; Table D-1: List of Included Electronic Schematics; Rev A
M102E/M501 TRS APPENDIX D - ELECTRONIC SCHEMATICS (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX D - ELECTRONIC SCHEMATICS The following drawing(s) are relevant to the M501-TRS. All drawings relevant to the M102E are listed in Appendix D of the M101E Manual. Table D-1: List of Included Electr...
Page 90 - User Notes
APPENDIX D - ELECTRONIC SCHEMATICS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) User Notes D-2 05518 Rev A