HP HP 8568B - Manuals
HP HP 8568B – Manual in PDF format online.
Manuals:
Manual HP HP 8568B
Summary
Hewlett-Packard Company certifies that this product met itspublished specifications at the time of shipment from the factory.Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by...
Assistance Product maintenance agreements and other customer assistance agreements are available for products. any assistance, contact your nearest Sales and Service
Safety Symbols Caution Warning The following safety symbols are used throughout this manual.Familiarize yourself with each of the symbols and its meaning before operating this instrument. The caution sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhere...
HP Spectrum Analyzer Documentation Outline HP Installation and Verification Manual HP Operating and Programming Manual HP Performance and Adjustments Manual HP RF Section Troubleshooting and Repair Manual HP IF-Display Section Troubleshooting and Repair Manual Included with the HP Model Spectrum Ana...
1. General Information Introduction . . . . . . . . . . . . . . . . . . . . . Instruments Covered by this Manual . . . . . . . . . Operation Verification . . . . . . . . . . . . . . . .Option 462 Instruments . . . . . . . . . . . . . . .Option 857 Instruments . . . . . . . . . . . . . . . 2. Perform...
3. Test 11. Log Scale Switching Uncertainty Test . . . . Test 12. Amplitude Fidelity Test . . . . . . . . . . .Test 13. Average Noise Level Test . . . . . . . . . .Test 14. Residual Responses Test . . . . . . . . . . .Test 15. Spurious Responses Test . . . . . . . . . . Test 16. Residual FM Test . ....
Crystal Filter Bypass Network Configuration . . . . . 3-151 4. Option 462 Introduction . . . . . . . . . . . . . . . . . . . . . 4. 6 Resolution Bandwidth Accuracy Test . . . . . 4-2 4. Impulse and Resolution Bandwidth Accuracy Test . 4-4 5. 6 Resolution Bandwidth Selectivity Test . . . . 4-10 5. Im...
Introduction Warning Instruments Covered by this Manual This HP Tests and Adjustments Manual contains two sections: Performance Tests and Adjustments Procedures. The Performance Tests provided should be performed for the following reasons: If the test equipment for the Operation Verification Program...
l-l. Recommended Test Equipment (1 of 5) Critical Specifications for Equipment Substitution Model Frequency: 10 MHz to 1500 MHz Output Power: + 10 maximum (leveled) Aging Rate: x Spurious Signals: HP Amplitude Modulation: dc to 100 Leveling: Internal, External Power Meter Frequency: 20 MHz to 450 MH...
Introduction The procedures in this section test the instrument’s electricalperformance using the Specifications in the Installation and Verification Manual as the performance standards. None of the tests require access to the interior of the instrument. The manual Performance Tests provided in this...
Equipment Required Equipment required for the manual performance tests andadjustments is listed in 2-1, Recommended Test Equipment, at the beginning of this manual. Any equipment that satisfies thecritical specifications given in the list may be substituted for therecommended model. Test Record The ...
1. Center Frequency Readout Accuracy 1. Center Frequency Readout Accuracy Related Adjustments Specification (uncorrected) Description Frequency Control Adjustments Time Base Adjustment Step Gain and 18.4 MHz Local Oscillator Adjustments 50 MHz Voltage-Tuned Oscillator Adjustments of frequency span +...
1. Center Frequency Readout Accuracy Test Equipment Synthesized Sweeper . . . . . . . . . HP Frequency Standard . 10 MHz standard, within 1 part in e.g. HP Adapter, Type N (m) to SMA (f) . . . HP1250-1250 61 cm (24 in.) Cable Assembly, SMA Male Connectors HP 5061-1086 Procedure 1. 2 . 3 . 4 . 5 . 6 ...
1. Center Frequency Readout Accuracy Note Spectrum analyzer center frequency readout can fall outside ofspecified limits if 10 MHz frequency reference has not been calibratedwithin the past year. To eliminate “frequency reference error xtune frequency” term, substitute spectrum analyzer 10 MHz FREQR...
2. Frequency Span Accuracy Related Adjustments Frequency Control Adjustments 50 MHz Voltage-Tuned Oscillator Adjustments Specification Span MHz MHz Uncertainty of the actual frequency separation between two points of span setting) of the actual frequency separation between two points of span setting...
Equipment Procedure y 100.1 1.01 MHz 20.1 MHz 1.5 Analyzer Center y 100 MHz 100 MHz 100 MHz 100 MHz 100 MHz 100 MHz 900 MHz 2. Frequency Span Accuracy Test Synthesized Sweeper ..................................................................... AdapterTypeN(m) to SMA(f)................................
2. Frequency Span Accuracy Freq Span 1 . 5 Hz Hz 2-4. Span Error Error ADUT-ASyn from 2-3 Min Max - 1 0 H z 1 0 H z - 5 0 0 0 H z 5000 H z - 5 , 0 0 5 H z 5,005 H z - 5 0 , 0 0 0 Hz 50,000 H z - 2 3 , 2 3 0 H z 23,230 H z - 4 6 0 , 0 0 0 H z 460,000 H z- 4 6 2 , 3 0 0 H z 462,300 H z Note The specif...
3. Sweep Time Accuracy Test ms) 3. Sweep Time Accuracy Test Related Adjustment Frequency Control Adjustments Specification for sweep times seconds for sweep times seconds Description Preferred Procedure This test is for sweep times ms. For faster sweep times, refer to Fast Sweep Time Accuracy Test (...
3. Sweep Time Accuracy ms) Equipment Universal Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP Digital Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP Procedure Sweep Times ms 1. Connect equipment as sh...
3. Sweep Time Accuracy Test ms) 2-6. Sweep Time Accuracy, Sweep Times s [ S W E E P T I M E ) Marker A Time Min Measured Max 20 s 3.6 s 4.4 s 200 s 32 s 48 s Sweep Times ms 14. Sweep times ms are tested without external test equipment (Alternate Procedure) by the following procedure. 15. Press Start...
4. Resolution Bandwidth Accuracy 4. Resolution Bandwidth Accuracy Test Related Adjustment Specification Description Equipment Procedure (For instruments with Option 462, refer to Chapter 4.) Bandwidth Adjustments 3 MHz 3 to 1 MHz 10 Hz to 1 30 and 100 bandwidth accuracy figures apply only with Relat...
4. Resolution Bandwidth Accuracy Test Figure 2-6. Resolution Bandwidth Measurement 8. Vary spectrum analyzer settings according to 2-8. Measure the 3 bandwidth for each resolution bandwidth setting by the procedure of steps 6 and 7 and record the value in 2-8. The measured bandwidth should fall betw...
5. Resolution Bandwidth Selectivity Test 6. Read the 60 bandwidth for the 3 MHz resolution bandwidth setting from the MARKER A frequency readout (see Figure 2-7)and record the value in 2-9. 7. Vary spectrum analyzer settings according to Table 2-9. Measure the 60 bandwidth for each resolution bandwi...
5. Resolution Bandwidth Selectivity Test 2-9. Resolution Bandwidth Selectivity Spectrum Analyzer Measured Measured Bandwidth Maximum ( F R E Q U E N C Y S P A N ) 3 Selectivity Selectivity Ratio Bandwidth Bandwidth (60 BW 3 MHz 20 MHz 100 Hz 1 MHz 15 MHz 300 Hz 300 5 MHz AUTO 100 2 MHz AUTO 30 500 A...
6. Resolution Bandwidth Switching Uncertainty Test Related Adjustments Specification Description Equipment Procedure (For instruments with Option 462, refer to Chapter 4.) 3 MHz Bandwidth Filter Adjustments 21.4 MHz Bandwidth Filter Adjustments Down/Up ConverterAdjustments (uncorrected; referenced t...
7. Input Attenuator Switching Uncertainty 6. Press MARKER 7. Set LEVEL ], and frequency synthesizer amplitude according to 1. At each setting, press MARKER ( PEAK SEARCH ) and record the deviation from the 10 setting from the MARKER A amplitude readout (see Figure 2-10). The deviationshould not exce...
8. Frequency Response Test Related Adjustment Specification Description SYNTHESIZED SWEEPER Slope Compensation Adjustment SIGNAL INPUT 1 100 Hz to 1.5 100 Hz to 500 MHz SIGNAL INPUT 2 100 to 1.5 Frequency response at both analyzer inputs is tested by slowlysweeping a flat signal source over the freq...
8. Frequency Response Test Note Equipment listed is for three test setups, Figure 2-11, Figure 2-13, and Figure 2-15. Equipment Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....
8. Frequency Response Test 7. 8. 9 . Key in the following spectrum analyzer settings: 10. Press TRACE A on the analyzer. 11. Press SWEEP SINGLE on the synthesized sweeper. Adjust POWER LEVEL on synthesized sweeper (using data knob)to place peak of 20 MHz signal near reference level (top) graticuleli...
8. Frequency Response Test Option 001: Set [ REFERENCE LEVEL ] TO -6.0 15. Repeat steps 6 through 11. Press DISPLAY LINE on the spectrum analyzer. Use the Display Line to measure the maximumand minimum points on the trace. Record measurements below. SIGNAL INPUT 1 (20 MHz to 1.5 Maximum Minimum 16. ...
8. Frequency Response 25. After completion of sweep, press DISPLAY LINE on the spectrum analyzer. Use the Display Line to measure the maximumand minimum points on the trace. Record the measurements below. SIGNAL INPUT 1 (100 to 20 MHz) Maximum Minimum 26. Measure and record signal level at start of ...
8. Frequency Response 42. Press DISPLAY LINE on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points onthe trace. (Disregard LO Feedthrough at 1 Record the measurements below. SIGNAL INPUT 1 (1 to 100 Maximum Minimum 43. Set Display Line to peak of trace at 1 44. Key...
9. RF Gain Uncertainty Test h Figure 2-16. RF Gain Uncertainty Measurement 2 - 3 2 P e r f o r m a n c e T e s t s
10. IF Gain Uncertainty 10. IF Gain Uncertainty Test Related Adjustments Step Gain and 18.4 MHz Local Oscillator Adjustments 21.4 MHz Bandwidth Filter Adjustments Specification Assuming the internal calibration signal is used to calibrate thereference level at -10 and the input attenuator is fixed a...
10. IF Gain Uncertainty Test Equipment Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP Adapter, Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . HP 1250-0780 P r o c e d u r e 1 . 2. 3. 4. 10 Gain Steps 5. 6. 7. 8 9 10. 11. 12. ...
10. IF Gain Uncertainty Test 2-12. IF Gain Uncertainty, 10 Steps [ R E F E R E N C E L E V E L ) 0 -10 - 2 0 - 3 0 - 4 0 - 5 0 - 6 0 - 7 0 - 8 0 - 3 2 - 9 0 - 4 2 - 1 0 0 - 5 2 - 1 1 0 - 6 2 - 1 2 0 - 7 2 Frequency Synthesizer Amplitude - 2 - 1 2 - 2 2 - 3 2 - 4 2 - 5 2 - 6 2 - 7 2 100 100 100 100 1...
10. IF Gain Uncertainty Test 2 Gain Steps 13. Press 7. 14. Set [ REFERENCE to -1.9 15. Press MARKER (OFF). Set to 100 Hz. 16. Set the frequency synthesizer for an output power level of -3.9 Set the amplitude increment for 2 steps. 17. Press MARKER [ PEAK 18. Set the analyzer LEVEL ) and the frequenc...
10. IF Gain Uncertainty 0.1 Gain Steps 19. Set [ REFERENCE to 0 20. Set the frequency synthesizer for an output power level of -2.00 Set the amplitude increment for 0.1 steps. 21. Press MARKER [ PEAK 22. Set the analyzer and the frequency synthesizer amplitude according to 2-14. At each setting, not...
10. IF Gain Uncertainty 2-14. IF Gain Uncertainty, 0.1 Steps [ R E F E R E N C E L E V E L ) F r e q u e n c y Deviation Synthesizer (MKR A Amplitude Amplitude 0.0 - 2 . 0 0 0 (ref) - 0 . 1 - 2 . 1 0 - 0 . 2 - 2 . 2 0 - 0 . 3 - 2 . 3 0 - 0 . 4 - 2 . 4 0 - 0 . 5 - 2 . 5 0 - 0 . 6 - 2 . 6 0 - 0 . 7 - ...
12. Amplitude Fidelity 12. Amplitude Fidelity Test (For instruments with Option 857, refer to Chapter 5.) Related Adjustment Log Amplifier Adjustments Specification Log: Incremental over 0 to 80 display Cumulative 3 MHz to 30 Hz Resolution Bandwidth max over 0 to 80 display (20 1.5 max over 0 to 90 ...
12. Amplitude Fidelity 8. The fidelity error for amplitude steps from -10 to -80 should be 1.0 9. The fidelity error at the -90 setting should be L I I I I Figure 2-22. Amplitude Fidelity Measurement Linear Fidelity 10. Key in analyzer settings as follows: ..............................................
12. Amplitude Fidelity Test 2-17. Linear Amplitude Fidelity Frequency MARKER A Allowable Range Synthesizer Amplitude % of Reference Level) Amplitude Min Max 0 - 1 0 . 8 7 - 9 . 2 1 - 1 0 - 2 3 . 1 0 - 1 7 . 7 2 2 - 4 4 P e r f o r m a n c e T e s t s
13. Average Noise Level Test 13. Average Noise Level Specification Description Equipment Procedure for frequencies MHz, for frequencies MHz but Hz with 10 Hz resolution bandwidth, 0 input attenuation, 1 Hz video filter. Option 001: for frequencies MHz, for frequencies MHz but Hz with 10 Hz resolutio...
13. Average Noise Level Figure 2-23. Average Noise Level Measurement 9. Read the average noise level from the DISPLAY LINE readout. The value should be 10. Change [ CENTER FREQUENCY ) to 1.001 MHz. Follow the procedure to steps 7 through 9 to determine the average noise level. Thevalue should be 11....
14. Residual Responses 14. Residual Responses Test Specification Description Equipment Procedure for frequencies Hz with 0 input attenuation (no signal present at input) Option 100: for frequencies Hz with 0 input attenuation (SIGNAL INPUT 1 only). Option 400: for frequencies Hz with 0 input attenua...
15. Spurious Responses Test 15. Spurious Responses Test Related Adjustment Second Converter Adjustments Specification For total signal power of at the input mixer of the analyzer, all image and out-of-band mixing responses, harmonic and intermodulation distortion products are below the total signal ...
15. Spurious Responses Test SPECTRUM ANALYZER SYNTHESIZED ATTENUATOR 300 MHZ LPF Figure 2-25. Harmonic Distortion Test Setup Note Equipment listed is for two test setups, Figure 2-25 and Figure 2-26. Equipment Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . . HP Frequency Synthesiz...
15. Spurious Responses Test signal for a marker indication of -20.00 (-30.0 at the input mixer with 10 of input attenuation). 6 On the spectrum analyzer, key in MARKER FREQUENCY ) 560 MHz, MARKER SEARCH ) to position a second marker on the peak of the second harmonic distortion product of the 280 MH...
15. Spurious Responses Test Intermodulation Distortion Figure 2-26. Intermodulation Distortion Setup 11. Connect equipment as shown in Figure 2-26. 12. Set the controls of the spectrum analyzer as follows: CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . 29.5 MHzFREQUENCY SPAN ...
15. Spurious Responses - 0 - 1 0 - 2 0 - 3 0 3 - 4 0 - 5 0 - 6 0 - 7 0 Note If unable to locate intermodulation distortion products, temporarily increase output power level of frequency synthesizer and synthesized sweeper by 10 Return the output power level of both signal sources to the previous set...
15. Spurious Responses 24. On the frequency synthesizer, readjust the signal amplitude as necessary to position the peak of the displayed 29.99 MHz signal at the top CRT graticule line. 25. On the spectrum analyzer, key in MARKER (CENTER FR E QUENCY ) 30.01 MHz, MARKER ( PEAK SEARCH ) to position a ...
15. Spurious Responses 35. On the spectrum analyzer, key in [ CENTER FREQUENCY ] 59 MHz, MARKER [ PEAK SEARCH ) to position a second marker at the peak of the 59 MHz second-order intermodulation distortion product. The response should be below the display line below the total input power). Distortio...
16. Residual FM Test Specification Hz peak-to-peak in s; frequency span resolution bandwidth Hz, video bandwidth Hz. Description The spectrum analyzer CAL OUTPUT is used to supply a stable 20MHz signal to the analyzer. The analyzer is tuned in zero span to a point on the 30 Hz bandwidth response for...
16. Residual FM Test Figure 2-28. Bandwidth Filter Slope Measurement 8. Compute the detection slope of the 30 Hz filter between the markers by dividing the MARKER A amplitude by the MARKER Afrequency: filter slope = MARKER A amplitude/MARKER Afrequency = 9. Press SWEEP 10. Change SPAN ] to 0 Hz. Rea...
16. Residual FM Test Figure 2-29. Slope Detected Residual FM 11. Press SWEEP and wait for completion of the sweep. 12. Press MARKER [ PEAK SEARCH _). Press DISPLAY LINE and position the display line at the lowest point on the trace. Figure 2-30. Peak-to-Peak Amplitude Measurement 2 - 5 8 P e r f o r...
17. Line-Related Sidebands Tests Specification below the peak of a CW signal. Option below the peak of a CW signal. Description The spectrally pure calibrator signal of the spectrum analyzer is applied to the analyzer input and the line related sidebands near thesignal are measured. Equipment None r...
O p t i o n 4 0 0 2. 3. Press 4. Key in the following analyzer settings: 5. 6. 7. 17. Line-Related Sidebands Figure 2-31. Line Related Sidebands Measurement Press PRESET ]. Connect CAL OUTPUT to SIGNAL INPUT 2. Press 8 and adjust AMPTD CAL for a MARKER amplitude of -10. 00 l FREQUENCY] . . . . . . ....
18. Calibrator Amplitude Accuracy Related Adjustment 20 MHz Reference Adjustments Specification -10 Description The output level of the calibrator signal is measured with a power meter. Figure 2-32. Calibrator Amplitude Accuracy Setup Equipment Power Meter . . . . . . . . . . . . . . . . . . . . . ....
19. Fast Sweep Time Accuracy Test ms) 19. Fast Sweep Time Accuracy Test ms) Related Adjustment None Specification for sweep times seconds Description The triangular wave output of a function generator is used tomodulate a 500 MHz signal which is applied to the spectrum analyzer SIGNAL INPUT. The sig...
19. Fast Sweep Time Accuracy Test ms) 5. Press MARKER ( PEAK 6. Set [ FREQUENCY SPAN ) to 0 Hz, to 3 MHz, to 3 MHz, and press TRIGGER 7. Set synthesized sweeper for an amplitude-modulated output. 8. Set function generator controls as follows: FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . ...
19. Fast Sweep Time Accuracy Test ms) 2-18. Fast Sweep Time Accuracy ms) [ SWEEP TIME ] Function Generator Frequency Sweep Time Error (divisions) 5 ms 2.00 2 ms 5.00 1 ms 10.0 200 50.0 100 100 P e r f o r m a n c e T e s t s 2 - 6 5
20. LO Output Amplitude Test Specification from 2.0 to 3.7 Description The power level at the LO OUTPUT connected is measured as the first L.O. is swept over its 2.0 to 3.1 range. SPECTRUM ANALYZER POWER METER Figure 2-35. LO Output Amplitude Test Setup Equipment Power Meter . . . . . . . . . . . . ...
21. Frequency Reference Error 21. Frequency Reference Error Test Related Adjustment Time Base Adjustment Specification Aging Rate x and x year; attained after 30 days warmup from cold start at Temperature Stability x 0” to Frequency is within 1 x of final stabilized frequency within 30 minutes. Desc...
2-19. Performance Record Hewlett-Packard Company Model HP Serial No. IF-Display Section RF Section Tested by Report No. Date Performance Tests 2-69
1. Center Frequency Readout Accuracy Test Step 8. Center Frequency Readout Error Test Record Comb Spectrum Analyzer Generator Comb Frequency [ C E N T E R 100 MC EXT TRIG (1, 2, 5, or 10 MHz) trigger signal 1 100 MHz 100 100 MHz 500 100 MHz 1000 10 MHz 100 10 MHz 500 10 MHz 1000 10 MHz 1500 1 MHz 10...
2. Frequency Span Accuracy 2. Frequency Span Accuracy Test r 1 Spectrum Span 200 Hz 100.1 1.01 MHz 20 MHz 20.1 MHz 1.5 Steps 7, 9, and 11. Wide Span Error Analyzer Synthesized Sweeper DUT Measured Center Freq. A Freq. B A Synth Freq. C Freq. D A DUT Frequency span cf + span 100 MHz 99.999 910 MHz 10...
Test 3. Sweep Time Accuracy Step 6. Sweep Time Accuracy, Sweep Times ms [ S W E E P T I M E ) Marker A Time Min Measured Max 20 ms 18 ms 22 ms 50 ms 45 ms 55 ms 100 ms 90 ms 110 ms 500 ms 450 ms 550 ms 900 ms 1.10 Step 12. Sweep Time Accuracy, Sweep Times s Step 19. Sweep Time Accuracy, Sweep Times ...
4. Resolution Bandwidth Accuracy 4. Resolution Bandwidth Accuracy Step 8. Bandwidth Accuracy [ F R E Q U E N C Y S P A N ) 2 MHz 2 500 Hz MARKER A Readout of 3 Min Measured 2.400 MHz 270.0 90.0 27.00 9.00 2.700 d 3 Bandwidth Max 3.600 MHz 1.100 MHz 330.0 110.0 33.00 11.00 3.300 1.200 36 Hz 12 Hz Per...
Test 5. Resolution Bandwidth Selectivity Steps 7, 8 and 9. Resolution Bandwidth Selectivity Spectrum Measured (VIDEOBW] 6 0 Bandwidth Measured Bandwidth Maximum 3 Selectivity Selectivity Ratio Bandwidth (60 BW 3 BW) 3 MHz 20 MHz 300 5 MHz 2 MHz 3 0 500 10 200 3 50 1 10 3 0 0 H z 5 100 Hz 2 3 0 Hz 50...
Test 6. Resolution Bandwidth Switching Uncertainty Test 6. Resolution Bandwidth Switching Uncertainty Test Step 6. Bandwidth Switching Uncertainty 1 MHz 3 MHz 300 100 30 10 3 1 300 Hz 100 Hz 30 Hz 10 Hz ‘FREQUENCY SPAN] 5 MHz 5 MHz 5 MHz 500 500 50 50 10 1 1 200 Hz 100 Hz Deviation (MKR A Readout, 0...
Test 7. Input Attenuator Switching Uncertainty Test 1 0 20 30 4 0 50 6 0 7 0 Step 7. Input Attenuator Switching Uncertainty ( R E F E R E N C E L E V E L ] - 5 0 - 4 0 - 3 0 - 2 0 - 1 0 0 Frequency Synthesizer Amplitude - 5 2 - 4 2 - 3 2 - 2 2 - 1 2 - 2 8 Deviation (MARKER A Amplitude 0 (ref) Correc...
Test 8. Frequency Response 8. Frequency Respons 12 1 5 16 2 5 2 6 3 1 4 2 48 49 50 Signal Input SIGNAL INPUT 2 (20 MHz to 1.5 SIGNAL INPUT 1 (20 MHz to 1.5 SIGNAL INPUT 1 (20 MHz to 500 MHz) SIGNAL INPUT 1 (100 to 20 MHz) SIGNAL INPUT 1 (100 SIGNAL INPUT 2 (100 to 20 MHz) SIGNAL INPUT 1 (1 to 100 SI...
Test 9. RF Gain Uncertainty Test Step 6. LO Shift Min Measured Max - 1 . 0 + 1.0 2-78 Performance Tests
Test 10. IF Gain Uncertainty Test Test 10. IF Gain Uncertainty Test Step 12. Step IF Gain Uncertainty, 10 Steps 0 - 2 - 1 0 - 1 2 - 2 0 - 2 2 - 3 0 - 3 2 - 4 0 - 4 2 - 5 0 - 5 2 - 6 0 - 6 2 - 7 0 - 7 2 - 8 0 - 3 2 - 9 0 - 4 2 - 1 0 0 - 5 2 - 1 1 0 - 6 2 - 1 2 0 - 7 2 Frequency Synthesizer Amplitude ...
11. Log Scale Switching Uncertainty Step 6. Log Scale Switching Uncertainty SCALE MKR Amplitude (dB/DIV) 1 2 5 1 0 Deviation Allowable Deviation 0 (ref) 0 (ref) 2 - 8 2 P e r f o r m a n c e T e s t s
Test 12. Amplitude Fidelity Test 12. Amplitude Fidelity Step 6. Log Amplitude Fidelity Frequency 1 2 Fidelity Error Synthesizer Calibrated MARKER A Amplitude (Column 2 Column A m p l i t u d e A m p l i t u d e Step 0 (ref) 0 (ref) 0 (ref) 0 - 1 0 - 1 0 - 2 0 - 2 0 - 3 0 - 3 0 - 4 0 - 4 0 - 5 0 - 5 ...
Test 13. Average Noise Level Test 2-84 Performance Tests
14. Residual Responses Test 14. Residual Responses Test Step 11. Maximum Residual Response Frequency Range 500 Hz to 1500 MHz Option 400: 500 Hz to 2.5 2.5 to 1500 MHz Measured Measured Max Max Amplitude Frequency - 1 0 5 - 9 5 - 1 0 5 Performance Tests 2-85
15. Spurious Responses 2 1 2 5 2 6 3 4 3 5 Description Second Harmonic Third Harmonic Third Order Intermodulation Distortion 30 MHz input signals, 1 MHz separation Third Order Intermodulation Distortion 30 MHz input signals, 1 MHz separation Third Order Intermodulation Distortion 30 MHz input signal...
Test 16. Residual FM Test 16. Residual FM Test Step 14. Residual FM Performance Tests 2-87
17. Line-Related Sidebands Test Step 7 120 Hz (100 Hz)180 Hz (150 Hz) 240 Hz (200 Hz) 7. Option 4 0 0 400 Hz 800 Hz 1200 Hz Min M e a s u r e d M a x - 8 5 - 8 5 - 8 5 - 7 5 - 7 5 - 7 5 P e r f o r m a n c e T e s t s
Test 18. Calibrator Amplitude Accuracy Test Test 18. Calibrator Amplitude Accuracy Step 2. CAL OUTPUT Amplitude Min M e a s u r e d M a x - 1 0 . 3 - 9 . 7 0 P e r f o r m a n c e T e s t s
19. Fast Sweep Time Accuracy Test Step 11. Sweep Time Accuracy ms) [ SWEEP Function Generator Frequency 5 ms 2.00 2 ms 5.00 1 ms 10.0 200 50.0 100 100 P e r f o r m a n c e T e s t s
Test 20. LO Output Amplitude Test 20. LO Output Amplitude Step 4. LO Output Level P e r f o r m a n c e T e s t s
Test 21. Frequency Reference Error Test Min Measured Max 1 4. Frequency (initial) 10. MHz 5. Frequency (after 24 hours) 10. MHz 6. Difference b e t w e e n 4 and 5 H z 0.01 H z 2 - 9 2 P e r f o r m a n c e T e s t s
Introduction The procedures in this section are for the adjustment of the instrument’s electrical performance characteristics. Warning The procedures require access to the interior of the instrument and therefore should only be performed by qualified service personnel. Refer to Safety Considerations...
Safety Considerations Although this instrument has been designed in accordance withinternational safety standards, this manual contains information,cautions, and warnings which must be followed to ensure safe operations and to retain the instrument in safe condition. Service andadjustments should be...
Adjustment Cross Reference Function Adjusted Low VoltageHigh Voltage CRT Display (Standard) CRT Display (Digital Storage)IF Gains Log ScalesBandwidth Amplitudes 3 Bandwidth 10 MHz Internal Time Base CAL OUTPUT LevelPhase Lock Loops Adjustment Procedure 1. Low Voltage Power Supply Adjustments 2. High...
Related Adjustments Any adjustments which interact with, or are related to, other adjustments are indicated in the adjustments procedures. It is important that adjustments so noted are performed in the orderindicated to ensure that the instrument meets specifications. Location of Points and Adjustme...
3-2. Adjustable Components Reference Designator Adjustment Name ZHF GAIN INT GAIN FOCUS COMP T/B FOC T/B CTRR/L FOCR/L CTR ORTHO INTENSITY LIMIT ASTIG FOCUS LIMIT X POSN X GAIN XHF GAIN Y POSN Y GAIN YHF GAIN + 15 ADJ HV ADJUST Adjustment Number 333 3 333 3 3 3 3 3 3 3 3 1 2 Adjustment Function Adju...
3-2. Adjustable Components (continued) Reference Designator Adjustment Name CTR 10 Hz AMPTD Adjustment Number 7 7 Adjustment Function Centers 3 MHz bandwidth filter pole Adjusts 3 MHz bandwidth filter 10 Hz bandwidth amplitude. 1 10 Hz AMPTD 7 Adjusts 3 MHz bandwidth filter 10 Hz bandwidth amplitude...
3-2. Adjustable Components (continued) Reference Designator 1 1 Adjustment Name IO VEVD v c VB G A GBG C GDG E GF A l B l C l E l GAIN D l 5.8 2 Adjustment Number 21 2 1 2121212 12 12 12 12 12 12 12 12 12 12 12 121 2 1 2 1 2 1 212 12 12 12 12 12 12 12 1 2 12 12 12 12 1 Adjustment Function Adjusts 3....
3-2. Adjustable Components (continued) Reference Adjustment Adjustment Adjustment Function Designator Name Number ORTHO 3 Sets orthogonality of CRT. PATTERN 3 Adjusts for optimum rectangular shape of CRTdisplay. INTENSITY 3 Sets adjustment range of front-panel INTENSITYcontrol. LIMIT ASTIG 3 Adjusts...
3-3. Factory-Selected Components Reference Designator 1 1 1 Adjustment Procedure Range of Values 2.87 K to 6.19 K 19.6 K to 42.2 K 121 K to 162 K 10.0 K to 26.1 K Open or 1.0-10.0 1.0 to 10.0 5.0 K to 12.5 K5.0 K to 12.5 K 562 to 1.33 K 56.2 K to 825 K 68.1 to 178 1.96 K to 5.11 K 1 K to 31.6 K 90.9...
3-3. Factory-Selected Components (continued) Reference Designator 1 1 Adjustmenl Procedure 1 0 1 1101 01 0 777 7 7 Range of Values 1.78 K to 13.3 K 8.25 K to 82.5 K 1.0 K to 6.81 K 1 K-OPEN 1.0 to 8.2 180 to 270 1.0 to 8.21.0 to 8.2 180 to 270 180 to 270 180 to 270 13 13 ‘4 to 13 0 to 9.09 3.16 K to...
3-3. Factory-Selected Components (continued) Reference Designator 14 1 Adjustment Procedure 10 Range of Values 10.0 K to 17.8 K10.0 K to 17.8 K10.0 K to 17.8 K10.0 K to 17.8 K10.0 K to 17.8 K 10.0 K to 17.8 K10.0 K to 17.8 K 7.50 K to 13.3 K7.50 K to 13.3 K 38.3 to 68.138.3 to 68.1 100 to 178 383 to...
3-3. Factory-Selected Components (continued) Reference Designator Adjustment Procedure 1 819 18 2 12 12 12 12 12 1 2 1 21 2 12 12 12 1 1 41 4 14 1 4 1 4 1 17 17 1 71 7 Range of Values 23.7 to 180 909 to 1.21 K Function of Component Sets sampler drive levelSets adjustment range of CAL OUTPUT 56.2 K S...
3-3. Factory-Selected Components (continued) Reference Adjustment Range of Values Function of Component Designator Procedure Al 348 to 562 Sets YTO loop gain crossover to 20 Al 1 6 130 to 220 Sets YTO loop response MHz. 1 6 2.2 to 3.3 Sets YTO loop response. Al 16 15 to 51.1 Sets YTO loop 20 to 30 M...
3-3. Factory-Selected Components (continued) Reference Adjustment Range of Values Designator Procedure Function of Component Option 462 5.62 K to 7.5 K5.62 K to 7.5 K5.62 K to 7.5 K5.62 K to 7.5 K5.62 K to 7.5 K 5.62 K to 7.5 K 5.11 K to 6.81 K5.11 K to 6.81 K5.11 K to 6.81 K5.11 K to 6.81 K 99 to 1...
3-5. Standard Value Replacement 0.125 Resistors Resistors Type: Fixed-Film Range: 10 to Ohms Wattage: 0.125 at Tolerance: Value Fart Number 10.0 0757-0346 11.012.1 0757-0379 13.3 0698-3427 14.7 0698-3428 16.2 0757-0382 17.8 0757-0294 19.6 0698-3429 21.5 0698-3430 23.7 0698-3431 26.1 0698-3432 28.7 0...
3-5. Standard Value Replacement 0.125 Resistors (continued) Resistors Type: Fixed-Film Range: 10 to Ohms Wattage: 0.125 at Value Part Number 0698-3136 8 0698-3157 3 0757-0199 3 0698-3158 4 0698-3159 5 0698-3449 6 0698-3160 8 0757-0123 3 0698-3161 9 0698-3450 9 0698-3162 0 7 0757-0459 8 0757-0460 1 0...
3-6. Standard Value Replacement 0.5 Resistors Resistors Type: Fixed-Film Range: 10 to Ohms Wattage: 0.5 at Value Part Number Value Fart Number 10.0 0757-0984 4 383 0698-3404 3 11.0 0575-0985 5 422 0698-3405 4 12.1 0757-0986 6 464 0698-0090 7 13.3 0757-0001 6 511 0757-0814 9 14.7 0698-3388 2 562 0757...
3-6. Standard Value Replacement 0.5 Resistors (continued) Resistors Value 38.313 46.413 Type: Fixed-Film Range: 10 to Ohms Wattage: 0.5 at Fart Number e * a n Value Fart Number 0757-0844 5 0757-0130 2 0698-0025 8 0757-0129 9 0698-3415 6 0757-0063 0 0698-3416 7 0757-0127 7 0698-3417 8 0698-3424 7 069...
1. Low-Voltage Power Supply Adjustments 1. Low-Voltage Power Supply Adjustments Reference IF-Display Section: V Regulator + 120 V, V Regulator (Serial Number Prefix and above) + 100 V, V Regulator (Serial Number Prefix and below) RF Section: Voltage Regulator Description The + 15 V supply is adjuste...
1. Low-Voltage Power Supply Adjustments 4. Connect the DVM to on the IF-Display Section. DVM indication should be + 15.000 V dc. If the voltage is out of tolerance, adjust + 15 V ADJ for the specified voltage. , 5 Figure 3-2. IF-Display Section Low-Voltage Adjustments (SN and Below) Figure 3-3. A l ...
1. Low-Voltage Power Supply Adjustments 7. Verify that the + 120 V indicator (yellow LED) is lit. Note On IF-Display Sections serial prefixed and below, indicator is a + 100 V indicator. 8. Connect the DVM to DVM indication should be + 120.0 V dc. The + 120 V supply is referenced to the + 15 V suppl...
1. Low-Voltage Power Supply Adjustments supply, therefore, if the supply is out of tolerance, a circuit malfunction is indicated. 17. The indicator (yellow LED) should be lit. 18. Connect the DVM to The DVM indication should be -5.200 V The supply is referenced to the supply, therefore, if the suppl...
2. High-Voltage Adjustment (SN and Below) D I G I T I Z I N G OSCILLOSCOPE HI-VOLTAGE SIGNAL ANALYZER Figure 3-5. High Voltage Adjustment Setup Equipment Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC High-Voltage Probe (1000: 1 divider) . . . . . . ...
2. High-Voltage Adjustment (SN and Below) Warning With the protective cover removed in the following step, do not place hands near the High-Voltage assembly. High voltage (approximately -4000 V dc) can be present even when the ac line cord is disconnected. 10. Wait at least one minute for capacitors...
2. High-Voltage Adjustment (SN and Below) If the calibration factor calculated in step 8 is 0.00099, and is labeled for -3875 V, then adjust HV ADJ for a DVM indication of: 0.00099 x (-3875 V) = -3.836 V dc 16. With the front-panel INTENSITY control fully counterclockwise, wait approximately 30 minu...
2. High-Voltage Adjustment (SN and Below) 27. On the oscilloscope press 28. Connect the oscilloscope channel 1 probe to using a long probe extension. See Figure 3-7 for the location of 29. Reconnect the ac line cords to each instrument section. Adjust the front-panel INTENSITY control fully counter-...
2. High-Voltage Adjustment (SN and Below) 34. On the oscilloscope, adjust the channel 1 offset voltage as necessary to measure the peak-to-peak CRT cut-off voltage, V,,, at See Figure 3-9. This peak-to-peak voltage should be between 45-75 Note this voltage for use in step 39. 1 1 0 . 0 6 0 . 0 0 v 1...
2. High-Voltage Adjustment (SN and Below) 39. Slowly adjust the front-panel INTENSITY control through its entire range while monitoring the peak-to-peak voltage at As the INTENSITY control is turned clockwise, the peak-to-peak voltage at will drop. To prevent long-term CRT damage, this voltage shoul...
2. High-Voltage Adjustment (SN and Below) 5. While holding the insulated handle of the screwdriver, touch the grounded blade to the following connections: a. Both brown wires going to the rear of the CRT from via cable harness b. The yellow, blue, and orange wires in the same cable as “a. above. c. ...
2. High-Voltage Adjustment (SN and Above) 11. Remove the protective cover from the High-Voltage Regulator Assembly. A label should be visible on the High Voltage Assembly. is mounted on the non-component side of the High-Voltage Regulator Assembly asshown in Figure 3-14.) Record the voltage listed o...
2. High-Voltage Adjustment (SN and Above) 3. Connect one end of a jumper wire (made of insulated wire and two alligator clips) to the blade of the screwdriver. Connect the other end of the jumper wire to the metal chassis of the IF DisplaySection. This grounds the screwdriver. 4. Slide the screwdriv...
3. Preliminary Display Adjustments (SN and Below) 3. Preliminary Display Adjustments (SN and Below) Reference Keyboard Z-Axis Amplifier A1 X-Deflection Amplifier Y-Deflection Amplifier Note Note Adjustment 2, “High-Voltage Adjustment,” should be performed beforeperforming the following adjustment pr...
3. Preliminary Display Adjustments (SN and Below) Procedure X and Y Deflection 1. Connect a 10:1 (10 divider probe to the oscilloscope’s channel Amplifier Pulse 1 input and a 10: 1 divider probe to the channel 4 input. Response Adjustments 2. On the oscilloscope, press to perform a soft reset. On th...
3. Preliminary Display Adjustments (SN and Below) Figure 3-17. Location of and HF GA H F G A I N , Figure 3-18. and Adjustment Locations 10. Set the Pulse/Function Generator controls as follows: MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . NORM Waveform ....
3. Preliminary Display Adjustments (SN and Below) 16. Three waveforms should be displayed on the oscilloscope, as shown in Figure 3-20. The lower composite waveform represents the combined X deflection voltage applied to the CRT. Use the oscilloscope’s front-panel knob to adjust waveform fl sensitiv...
3. Preliminary Display Adjustments (SN and Below) 19. Connect the oscilloscope’s channel 1 probe to and the channel 4 probe to See Figure 3-18 for the location of the test points. Connect the output of the pulse/function generator to (Y input) on the Display Adjustment PC board in the slot. 20. The ...
3. Preliminary Display Adjustments (SN and Below) 28. Set the oscilloscope controls as follows: Press Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . onamplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Press 29. Adjust X POS...
3. Preliminary Display Adjustments (SN and Above) Reference Keyboard Al X, Z Axis Amplifier Note Adjustment Procedure 2, “High-Voltage Adjustment,” should beperformed before performing the following adjustment procedure. Note Perform this adjustment only if components have been replaced onthe X, Y, ...
3. Preliminary Display Adjustments (SN and Above) Procedure X and Y Deflection 1. Connect a (10 divider probe to the oscilloscope’s channel Amplifier Pulse 1 input and a divider probe to the channel 4 input. Response Adjustments 2. On the oscilloscope, press (RECALL) to perform a soft reset. On the ...
3. Preliminary Display Adjustments (SN and Above) Figure 3-24. Location of and Figure 3-25. Adjustment Locations 10. Set the Pulse/Function Generator controls as follows: MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM Waveform . . . . . . . . . . . . ...
3. Preliminary Display Adjustments (SN and Above) 12. Set the oscilloscope controls as follows: Press Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . onamplitude scale offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....
3. Preliminary Display Adjustments (SN and Above) the combined X deflection voltage applied to the CRT. Use theoscilloscope’s front-panel knob to adjust waveform fl sensitivityfor approximately 8 vertical divisions. r u n n i n g 1 2 0 . 0 o f f s e t : 2 5 . 0 0 V - 1 2 5 . 0 0 0 1 2 5 . 0 0 0 3 7 ...
19. 20. Pulse Response of 21. Control Gate Z Amplifier to BLANK Input 22. 23. 24. Set the oscilloscope controls as follows: Note 25. 3. Preliminary Display Adjustments (SN and Above) Connect the oscilloscope’s channel 1 probe to and the channel 4 probe to See Figure 3-25 for the location of the test...
3. Preliminary Display Adjustments (SN and Above) 29. Set the oscilloscope controls as follows: Press Channel 1 . . . . . . . . . . onamplitude scale . . . . . . . Press 30. Adjust the spectrum analyzer’s front-panel INTENSITY control for peak-to-peak (8 divisions) as indicated on the oscilloscope. ...
4. Final Display Adjustments (SN and Below) 4. Final Display Adjustments (SN and Below) Reference Description Note Procedure Note Keyboard Z Axis Amplifier X Deflection Amplifier Y Deflection Amplifier This procedure is used to optimize the appearance of the CRT display during routine maintenance or...
4. Final Display Adjustments (SN and Below) 5. For best overall focusing of the display, adjust the following potentiometers in the sequence listed below: a. FOCUS LIMIT for best focus of graticule lines (long vectors) b. ASTIG c. FOCUS GAIN for best focus of annotation (short vectors) 6. Adjust ORT...
4. Final Display Adjustments (SN and Above) 4. Final Display Adjustments (SN and Above) Reference Description Equipment Procedure Note Keyboard X, Y, Z Axis Amplifiers This procedure is used to optimize the appearance of the CRT display during routine maintenance or after CRT replacement or minor re...
4. Final Display Adjustments (SN and Above) Figure 3-32. Location of Final Display Adjustments on 3-6. Initial Adjustment Positions Adjustment Y GAIN Y POSN X GAIN X POSN INT GAIN FOCUS COMP T/B FOC Al T/B CTR Al R/L FOC A R/L CTR ORTHO INT LIM ASTIG Front-panel INTENSITYFront-panel FOCUSFront-panel...
4. Final Display Adjustments (SN and Above) 17. Adjust ORTHO and the front-panel ALIGN control to optimize the orientation and appearance of the rectangulargraticule pattern on the CRT display. 18. Repeat steps 13 through 17 as needed to optimize overall display focus and appearance. 3 - 6 4 A d j u...
5. Log Amplifier Adjustments 5. Log Amplifier Adjustments Reference Related Performance Tests Note IF-Display Section Log Amplifier-Filter Log Amplifier-Detector Scale Fidelity Test The Log Amplifier-Filter and Log Amplifier Detector are temperature compensated as a matched set at the factory. In th...
5. Log Amplifier Adjustments 4. Connect DVM to and DVM ground to the IF casting. Connect the frequency synthesizer to the RF INPUT. Key in 80 MHz and [ A M P L I T U D E ) -86.98 The frequency synthesizer will now provide a load. Offset Adjustment Check 5. Adjust ZERO for 0.0000 V dc. See Figure 3-3...
5. Log Amplifier Adjustments Filter Amplitude Adjustment 9. Connect one end of a jumper wire to Connect the other end of the jumper to (+ Connecting the jumper to first reduces the chance of shorting the + to ground. Note DVM indication. V dc 10. Remove the short from between and 11. Adjust AMPTD fo...
6. Video Processor Adjustments 6. Video Processor Adjustments Reference Related Performance Description IF-Display Section 1 Video Processor Log Scale Switching Uncertainty Test The CAL OUTPUT signal is connected to the RF INPUT through a step attenuator. The instrument is placed in zero frequency s...
6. Video Processor Adjustments 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. Set step attenuator to 120 DVM indication should be 0.000 V dc. (If DVM indication is out of tolerance, adjust ZERO on the log amplifier-detector board..) Set step attenuator to 0 Key in and adjust DATA knob for DVM indication as...
7. 3 MHz Bandwidth Filter Adjustments Reference IF-Display Section 3 MHz Bandwidth Filter Related Performance Resolution Bandwidth Switching Uncertainty TestResolution Bandwidth Selectivity Test Description With the CAL OUTPUT signal connected to the RF INPUT, the 18.4 MHz oscillator can be adjusted...
P r o c e d u r e Position instrument upright as shown in Figure 3-37 and removetop cover. 2. Set LINE switch to ON and press PRESET ). 7. 3 MHz Bandwidth Filter Adjustments Frequency Zero Check 3. Connect CAL OUTPUT signal to RF INPUT 4. Key in 5. Adjust front panel FREQ ZERO control for maximum si...
7. 3 MHz Bandwidth Filter Adjustments 25. Adjust PK for maximum peak-to-peak signal on Channel 2 display. See Figure 3-39 for location of adjustment. If unable toachieve a “peak” in signal amplitude, increase or decrease valueof Refer to 3-3 for range of values. 3 M H z B a n d w i d t h F i l t e r...
8. 21.4 MHz Bandwidth Filter Adjustments Equipment Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . .HP 10 Step Attenuator . . . . . . . . . . . . . . HP Option 1 Step Attenuator . . . . . . . . . . . . . . . HP Option Crystal Filter Bypass Network (2 required) . . . . Refer to ...
8. 21.4 MHz Bandwidth Filter Adjustments 10. Key in 1 MHz, and 1 MHz. 11. Press MARKER SEARCH ], MARKER 12. Key in 100 200 and MARKER [ PEAK SEARCH). 13. Adjust LC to align markers on display. MARKER A level should indicate 1.00 X. See Figure 3-41 for location of adjustment. 14. Repeat steps 10 thro...
8. 21.4 MHz Bandwidth Filter Adjustments 22. Adjust CTR to center signal on center graticule line. Adjust SYM for best symmetry of signal. See Figure 3-42 for location of adjustments. If unable to adjust SYM for satisfactory signal symmetry, increase or decrease value of Refer to 3-3 for range of va...
8. 21.4 MHz Bandwidth Filter Adjustments 45. 46. 4 7 . 48. 49. 50. 51. 52. 53. 54. 55. 56. 57. 58. 59. 60. 61. Adjust XTAL to align markers on display. MARKER A level should indicate 1.00 X. See Figure 3-44 for location of adjustment. Dip Adjustments Refer to the Resolution Bandwidth Switching Uncer...
62. 63. 64. 65. 66. 67. 8. 21.4 MHz Bandwidth Filter Adjustments approximately -17 (to the left). If unable to achieve a “dip” in signal amplitude, increase or decrease value of Refer to 3-3 for range of values. Remove short from and short to ground. Adjust LC DIP for minimum amplitude of signal pea...
9. 3 Bandwidth Adjustments IF CONTROL \ Figure 3-45. Location of 3 Bandwidth Adjustments 9. Press MARKER Adjust marker to 3 point on opposite side of signal (CRT MKR A annotation indicates 1.00 X). There are now two markers; one on each side of the signal at the 3 points. 10. CRT MKR A annotation no...
10. Step Gain and 18.4 MHz Local Oscillator Adjustments 10. Step Gain and 18.4 MHz Local Oscillator Adjustments Reference Related Performance Tests Description IF-Display Section 3 MHz Bandwidth Filter Step Gain Resolution Bandwidth Selectivity TestIF Gain Uncertainty TestCenter Frequency Readout Ac...
10. Step Gain and 18.4 MHz Local Oscillator Adjustments Equipment Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . HP Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...
10. Step Gain and 18.4 MHz Local Oscillator Adjustments 10. If CAL adjustment does not have sufficient range to adjust trace to the top CRT graticule line, increase or decrease the value of as necessary to achieve the proper adjustment range of CAL adjustment. See Figure 3-39 for the location of Ref...
10. Step Gain and 18.4 MHz Local Oscillator Adjustments S T E P G A I N F R E O A D J Z E R O C O A R S E “ R Figure 3-49. Location of Gain Step, 18.4 MHz LO, and + Adjustments 36. Key in [ FREQUENCY 1 100 Hz, and [ PEAK SEARCH] 37. Adjust front-panel FREQ ZERO control fully clockwise. Press SEARCH ...
12. Time Base Adjustment (SN and Below, also and Above) 12. Time Base Adjustment (SN and Below, also and Above) Reference Related Performance Test Description Equipment RF Section: 10 MHz Quartz Crystal Oscillator Center Frequency Readout Accuracy Test The frequency of the internal 10 MHz Frequency ...
13. 20 MHz Reference Adjustments 13. 20 MHz Reference Adjustments Reference Related Performance Description Equipment RF Section: 20 MHz Reference Calibrator Amplitude Accuracy Test The 20 MHz output is peaked and amplitude checked for proper level.The INTERNAL REFERENCE output level is then checked...
13. 20 MHz Reference Adjustments Procedure 1. Position instrument on right side as shown in Figure 3-56 and remove bottom cover. Remove 20 MHz Reference and install on extenders. See Figure 3-57 for the location of components. 2. Set LINE switch to ON and press PRESET ). 3. Set rear-panel FREQ REFER...
14. 249 MHz Phase Lock Oscillator Adjustments 14. 249 MHz Phase Lock Oscillator Adjustments Reference Description Equipment RF Section: 249 MHz Phase Lock Oscillator Two center frequencies are chosen: one which will tune the 249 MHzOscillator to its low-end frequency and one which will tune the 249M...
14. 249 MHz Phase Lock Oscillator Adjustments P r o c e d u r e Place instrument on right side with IF-Display Section facing rightas shown in Figure 3-59. 2.3. 4. 5. 6. 7. 8. 9. 10. 11. Key in 2 and adjust for + 13.0 fO.l V dc at 12. 13. 14. adjustment. Press 1 (RECALL 1) and adjust for V dc. Repea...
15. 275 MHz Phase Lock Oscillator Adjustment 2 7 5 P H A S E L O C K O S C I L L A T O R 0 U 1 Figure 3-62. Location of 275 MHz PLO Adjustment 7. Disconnect test equipment from instrument. Adjustments 11
16. Second IF Amplifier and Third Converter Adjustment Reference RF Section: Second IF Amplifier Third Converter Description A synthesized sweeper is used to inject a signal of 301.4 MHz at -20 in to the Second IF Amplifier. The output of the amplifier is displayed on a scalar network analyzer. The ...
Procedure Second IF Amplifier Adjustments 16. Second IF Amplifier and Third Converter Adjustment Adapters: Type N (f) to (f) . . . . . . . . . . . . . . . . . . . . . . 1250-1745 Type N (m) to BNC (f) required) . . . . . . . . . . . . . 1250-0780Type N (f) to BNC (f) required) . . . . . . . . . . . ...
16. Second IF Amplifier and Third Converter Adjustment 14. 15. 16. 17. See Figure 3-65 for the typical response when the filter is properly adjusted. On the scalar network analyzer, press MAX. Press cursor A, ON and set the cursor to the -3 point on the low side of the filter response Press cursor A...
16. Second IF Amplifier and Third Converter Adjustment 17. Pilot Second IF Adjustments Reference RF Section: Pilot Second IF Amplifier Pilot Third Converter Description A synthesized sweeper is used to inject a signal of 269 MHz at -20 into the Pilot Second IF Amplifier. The output of the amplifier ...
17. Pilot Second IF Amplifier Adjustments Procedure 1. Position instrument on right side as shown in Figure 3-67, with bottom cover removed. 2. Set LINE switch to ON and press PRESET ) on HP (DUT), HP and HP 3. Connect 20 Attenuator and power splitter to RF OUTPUT of synthesized sweeper. Connect one...
17. Pilot Second IF Amplifier Adjustments P I L O T P I L O T I F A M P L I F I E R C O N V E R T E R \ F I L T E R Figure 3-68. Location of 269 MHz Filter Adjustments 3 2 1 Figure 3-69. 269 MHz Filter Adjustments Waveforms 15. On the scalar network analyzer, press MAX. Press cursor A, ON and set th...
18. Frequency Control Adjustments 18. Frequency Control Adjustments Reference Related Performance Tests Description Equipment Procedure RF Section: Frequency Control Sweep Time Accuracy TestFrequency Span Accuracy Test Center Frequency Readout Accuracy Test The sweep reference voltage is adjusted an...
18. Frequency Control Adjustments START and STOP Adjustments FM SPAN Adjustment 30. 22. 23. Key in STEP 1023 Hz. Adjust LSD for DVM indication of V dc. If not using an HP DVM, adjust for specified voltage plus the DVM indication in step 20. See Figure 3-71 forlocation of adjustment. 24. On the HP pr...
19. Second Converter Adjustments 19. Second Converter Adjustments Reference RF Section: RF Converter Related Performance RF Gain Uncertainty Test T e s t Spurious Responses Test Description First, the second LO frequency is adjusted for proper frequency and then the LO shift is adjusted by using the...
19. Second Converter Adjustments Procedure Second LO Frequency 3. Set HP Spectrum Analyzer LINE to ON and press and Shift Adjustments PRESET]. Note The second LO and pilot second LO output power is typically -35 or less. An HP amplifier is used in steps 1 through 26 to amplify the LO power to a leve...
19. Second Converter Adjustments Second Converter 27. Key in [ FREQUENCY SPAN ] 0 Hz. Filter Adjustments 28. On the synthesized sweeper, key in 240 MHz, 50 MHz, and 10 29. Connect the synthesized sweeper’s SWEEP OUTPUT (rear panel), Z-AXIS BLANK/MKRS (rear panel), and PULSE MODULATIONINPUT (front pa...
19. Second Converter Adjustments 48. Disconnect cable connected to and connect to IF IN). Reconnect semi-rigid cable to that was disconnected in step 36. 49. Set the synthesized sweeper’s for 2052.5 MHz MHz. Adjust to center the signal. 50. Adjust and for best shape and flatness at maximum amplitude...
20. 50 MHz Voltage-Tuned Oscillator Adjustments Reference Related Performance Test Description Equipment Procedure RF Section: All 50 MHz Voltage-Tuned Oscillator (VTO) Frequency Span Accuracy Test Center Frequency Readout Accuracy Test First, the voltage reference for the Shaping network is set by ...
20. 50 MHz Voltage-Tuned Oscillator Adjustments DACS Accuracy Check 3. Connect DVM to and ground lead to 4. Key in [SHIFT) STEP 0 Hz. If using an DVM, then If not using an HP DVM, note voltage indication for reference later. 5. Key in STEP SIZE ) 1023 6. If using an HP DVM, voltage indication should...
20. 50 MHz Voltage-Tuned Oscillator Adjustments 14. Adjust All POS SUPPLY for a DVM indication the same as that noted in step 12. See Figure 3-79 for location of adjustment. High-Frequency Key in STEP SIZE ] and End Adjustment 16. Adjust All OFFSET for VTP frequency indication 28.000 MHz MHz. Key in...
2 1. Slope Compensation Adjustments 21. Slope Compensation Adjustments Reference RF Section: Frequency Control Related Performance Test Frequency Response Test Description The HP Spectrum Analyzer is swept between 10 MHz and 1500 MHz, using a synthesized sweeper which has been power-meter leveled. T...
21. Slope Compensation Adjustments Note At this sweep time, some trace discontinuities are common. 14. Adjust TILT for best flatness (clockwise rotation increases the power slope), and trigger two sweeps on the synthesized sweeper. See Figure 3-81 for the location of Compare the resultant trace with...
22. Comb Generator Adjustments Reference RF Section: RF Converter 20 MHz Reference Description The output of the Pilot First Converter is connected to the signal input of the Second Converter. This allows the comb teeth from the Comb Generator to be displayed on the CRT display. The phase lock flags...
22. Comb Generator Adjustments Equipment Cable, SMA (m) to SMA (m) . . . . . . . . . . . . . . . HP 85680-20094 Procedure 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. Set instrument LINE switch to ON and press PRESET ). Connect CAL OUTPUT to SIGNAL INPUT 2. Key in FREQUENCY ) 20 MHz, [ FREQUENCY SPAN ...
23. Analog-To-Digital Converter Adjustments 23. Analog-To-Digital Converter Adjustments Reference Description Equipment Analog-to-Digital Converter The Analog-to-Digital Ramp Converter is adjusted at zero and full-scale by injecting a 0 V dc input and + 10 V dc input and adjustingthe OFFS and GAIN c...
23. Analog-To-Digital Converter Adjustments Procedure 1. Position instrument upright as shown in Figure 3-85 and remove top cover. 2. Set LINE switch to ON and press PRESET ). Standard Procedure 3. Procedure using Low-Noise DC Supply is illustrated in Figure 3-93. a. Key in TRACE A and SWEEP b. Disc...
24. Track and Hold Adjustments Reference Track and Hold Description The CAL OUTPUT signal is connected to the RF INPUT. Theinstrument is placed in zero frequency span to produce a dc level output from the IF-Video section and this dc level is regulated byadjusting the reference level. The Offsets an...
24. Track and Hold Adjustments AS ES C A L A D J I 10. 11. ‘ L E R M A T C H C A L O U T P U T Figure 3-88. Location of Track and Hold Adjustments Key in TRACE A Adjust OFFS POS until MARKER A level indication as indicated by CRT annotation flickers back and forth between .OOand 12. Key in TRACE A 1...
24. Track and Hold Adjustments 24. Adjust GNEG for MARKER A level indication as indicated by CRT annotation of 100 25. Repeat steps 4 through 24 until no further adjustments are required. 3-144 Adjustments
25. Digital Storage Display Adjustments 25. Digital Storage Display Adjustments Reference Description Equipment Procedure Preliminary Graticule Adjustments 1 Trigger Intensity Control Line Generator First, preliminary CRT graticule adjustments are performed to positionthe graticule on the CRT. These...
2 5 . Digital Storage Display Adjustments I N T E N S I T Y L I N E T R I G G E R C O N T R O L G E N E R A T O R Y H BAL X H S W P O F F S E T Figure 3-90. Location of Digital Storage Display Adjustments 5. Adjust LL THRESH fully clockwise. See Figure 3-90 for location of adjustment. 6. Adjust XLL ...
25. Digital Storage Display Adjustments Sample and Hold 13. Set LINE switch to STANDBY. Balance Adjustments 14. Place Line Generator on extender boards. 15. Set LINE switch to ON. Press PRESET ]. 16. Key in (RECORDER LOWER LEFT) 0 [Hz). Press (RECORDER UPPER RIGHT) 1028 17. Connect oscilloscope to 1...
25. Digital Storage Display Adjustments 20. 21. 22. 23. 24. Set LINE switch to ON. X and Y Offset and 25. Gain Adjustments 27. Press Key in [ FREQUENCY SPAN ] 0 Hz, [ SWEEP T I M E ) 100 Disconnect cable 9 (white) from and connect to test connector on Intensity Control; the other end of the cable re...
25. Digital Storage Display Adjustments 41. Adjust Y GAIN to align the top graticule line with the fast sweep signal trace. Final Graticule 42. Press PRESET ], TRACE A Adjustments 43. Set LL THRESH fully clockwise. 44. Adjust XLL and YLL to align horizontal and vertical lines so that each line meets...
4. 6 Resolution Bandwidth Accuracy Test Related Adjustment Specification Description Equipment Procedure 6 Bandwidth Adjustments 3 MHz bandwidth 30 Hz to 1 MHz bandwidths + -0%, 10 Hz bandwidth 30 and 100 bandwidth accuracy figures only applicable Relative Humidity, C. The 6 bandwidth for each resol...
4. Impulse and Resolution Bandwidth Accuracy 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. 32. to 30 On the Set the frequency synthesizer spectrum analyzer key in [ FREQUENCY 0.5 seconds, SWEEP MA MARKER amplitude in 4 S W E E P T I M E ] RKER ( P E A K S E A R C H ) . Record -2. Calcu...
7 . 8 . 9 . 10. 5. 6 Resolution Bandwidth Selectivity Vary spectrum analyzer settings according to Table 4-4. PressSWEEP and measure the 60 bandwidth for each resolution bandwidth setting by the procedure of steps 4 through6. Record the value in 4-4. Record the 6 bandwidths from Table in 4-4. Calcul...
5. Impulse and Resolution Bandwidth Selectivity 0 . 0 - 0 . 1 0 Figure 4-5. 60 Bandwidth Measurement 4. Press MARKER In] and position the positive movable marker 60 down from the signal peak on the negative-going edge of the signaltrace (the MARKER amplitude readout should be 0.00 5. Read the 60 ban...
6. Impulse and Resolution Bandwidth Switching Uncertainty A I I I I I I I M A R 0 \ Figure 4-6. Bandwidth Switching Uncertainty Measurement 4-6. Bandwidth Switching Uncertainty Res B W 1 MHz (i) 3 MHz (i) 300 (i) 100 (i) 30 (i) 10 (i) 3 (i) 1 (i) 300 Hz (i) 100 Hz (i) 30 Hz (i) 10 Hz (i) Frequency S...
3 MHz (i) 1 MHz (i) 300 (i) 100 (i) 30 (i) 10 (i) 3 (i) 1 (i) 3 MHz 3 MHz 3 MHz 1 MHz 300 100 3 0 1 0 Test 4. Impulse and Resolution Bandwidth Accuracy Test (p/o 2-19, Performance Record) Test 4. Impulse and Resolution Bandwidth Accuracy Test (p/o 2-19, Performance Test Record) Steps 1 through 38. I...
9. 6 Resolution Bandwidth Adjustments 13. Using the DATA knob, adjust the marker down one side of the display signal to the 6 point; CRT MKR A annotation indicates x. I I F C O N T R O L \ ~000000000000000000 000000000000000000 Figure 4-7. Location of Bandwidth Adjustments 14. Adjust 1 MHz for MKR A...
9. Impulse Bandwidth Adjustments 9. Impulse Bandwidth Adjustments Reference Related Performance Test Description Equipment Procedure IF-Display Section IF Control Impulse Bandwidth Accuracy Test The CAL OUTPUT signal is connected to the SIGNAL INPUT 1. Each of the adjustable resolution bandwidths is...
9. Impulse Bandwidth Adjustments 13. Using the DATA knob, adjust the marker down one side of the display signal to the 7.3 point; CRT MKR A annotation indicates 0.430 X. I F C O N T R O L \ 1 1 Figure 4-8. Location of Bandwidth Adjustments 14. Adjust 1 MHz for MKR A indication of 500 while maintaini...
Option 857 Introduction This chapter contains a modified performance test for Option 857instruments. When working on Option 857 instruments, substitutethe procedure in this chapter for the standard version contained in Chapter 2. The procedure included in this chapter is listed below: Performance Te...
12. Option 857 Amplitude Fidelity Test Related Adjustment Log Amplifier Adjustments Specification Log: Incremental fO.l over 0 to 80 display Cumulative 3 MHz to 30 Hz Resolution Bandwidth: max over 0 to 70 display (20 max over 0 to 90 display 10 Hz Resolution Bandwidth: max over 0 to 70 display (20 ...
12. Option 857 Amplitude Fidelity Equipment Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP Adapter, Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . HP 1250-0780 (2) BNC to BNC cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...
12. Option 857 Amplitude Fidelity Frequency 1 Synthesizer Calibrated Amplitude Amplitude Step 0 (ref) 0 - 1 0 - 1 0 - 2 0 - 2 0 - 3 0 - 3 0 - 4 0 - 4 0 - 5 0 - 5 0 - 6 0 - 6 0 - 7 0 - 7 0 - 8 0 - 8 0 - 9 0 20. Subtract the greatest negative fidelity error from the greatest positive fidelity error fo...
Major Assembly and Component Locations IF-Display Section Figure Index Assembly See Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4, 6-5 . . . . . . . . ...
HP Manuals
-
HP CM1312 MFP
Manual
-
HP DV2700
Manual
-
HP 203A
Manual
-
HP 13-2000
Manual
-
HP 1320NW
Manual
-
HP D4360
Manual
-
HP W19Q
Manual
-
HP FC2143
Manual
- HP ZV6000 Manual
-
HP 7500
Manual
-
HP C4591A
Manual
-
HP E4411B
Manual
-
HP N6310
Manual
- HP nx7000 Manual
-
HP DL120
Manual
- HP w1707 Manual
-
HP CV136A#B1H
Manual
-
HP X585z
Manual
-
HP 686714-S01
Manual
- HP 1200 series Manual