Page 3 - GETTING STARTED
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 3 GETTING STARTED This Quickstart Guide is designed to get you set up and operating your Teledyne Analytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss. This ...
Page 4 - Escape; Analyze; Digital Meter Display:
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 • Left & Right Arrows Select between functions currently displayed on the VFD screen. • Up & Down Arrows Increment or decrement values of functions currently displayed. • Enter Advances VFD display to the next screen in...
Page 6 - SETUP AND INSTALL; Mount the Analyzer
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 SETUP AND INSTALL Mount the Analyzer The Model 3000TA is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. It is designed for flush panel mounting. Use the four mounting holes—one in ea...
Page 7 - If using remote calibration feature:; Electrical Connections
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 7 4. If using remote calibration feature: • Connect zero gas to the ZERO IN port. • Connect span gas to the SPAN IN port. 5. Otherwise, calibration gases must be Tee'd into the sample inlet with appropriate valves. 6. Regulate the sample ...
Page 8 - Insert the power cord into the power cord receptacle.; Connect Analog Outputs
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 • Connect primary input power: Insert the power cord into the power cord receptacle. • Install the fuse: Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. Place small screwdrive...
Page 9 - Install the Micro-fuel Cell; Remove power from the instrument.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 9 • Connect Range ID Contacts: There are four dedicated Range ID relay contacts on the rear panel. The first three ranges are assigned to relays in ascending order. Range 1 ID— Low range Range 2 ID— Medium range Range 3 ID— High range Ran...
Page 10 - Purge the system using sample or zero gas.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 9. With the O-ring in place, align the guide pin with the hole on the cell holder. Then, with the holder, lift cell into the cell block. 10. Push the gate on the cell block down so that the slots on the side of the gate engage...
Page 11 - BASIC OPERATION; Default Configuration; Using the Data Entry and Function Buttons; Enter
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 11 BASIC OPERATION Once the analyzer has been installed, it can be configured for your application. Prior to using the instrument you must: • Calibrate the instrument • Define the analysis ranges and choose whether to use autoranging or a...
Page 12 - Available Functions; The six function buttons are used to control the analyzer.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 Available Functions The six function buttons are used to control the analyzer. Some functions have submenus that allow additional choices as shown below. Any function can be selected at any time by pressing the appropriate but...
Page 13 - Password Protection
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 13 Password Protection If a password is assigned, then setting the following system parameters can be done only after the password is entered: span and zero settings, alarm setpoints, analysis range definitions, switching between autorang...
Page 14 - Calibration of the Analyzer; Zero Calibration
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 TRAK/HLD Auto—Cal PSWD Logout More • Use the ◄► arrow keys to scroll the blinking over to PSWD , • Press Enter to select the password function. Either the default TETAI password or AAAAA place holders for an existing password ...
Page 15 - Zero
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 15 Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Use the ▲▼ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears blinking. Zero: Settling: AUTO <ENT> To Begin • Press Enter to ...
Page 16 - Span Calibration
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 16 Span Calibration Make sure an appropriate span gas is connected to the sample port. Auto Mode Spanning: • Press Span to enter the span function. • Use the arrow keys to toggle between AUTO and MAN span settling. Stop when AUTO...
Page 17 - Setting Alarm Parameters; Entering the Password; AL—1 AL—2; Dft—N Fs—N Ltch—N
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 17 Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val: 000008.00 <ENT>Span <UP>Mod # • Press ▲ (<UP>) to permit modification ( Mod # ) of span value. • Use the arrow keys to enter the oxygen conc...
Page 18 - Alarms; Dft–; Set The; Function; Range; AUTO
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 18 • Latching? Ltch–Y/N (Yes/No). • To change any of the alarm parameters, use the ◄► arrow keys to move the blinking over to the specific parameter. • Use the ▲▼ arrow keys to change the value of the parameter. Holding down the ...
Page 19 - The Analyze Function
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 19 • Use the ▲▼ arrow keys to switch from AUTO to FX/LO , FX/MED , or FX/HI to set the instrument on the desired fixed range (low, medium, or high). L—100 M—1000 H—10000 Mode—FX/LO —or— L—1 M—10 H—100 Mode—FX/MED —or— L—1 M—10 H—100 Mode—...