Teledyne M5000B - Manual

Teledyne M5000B

Teledyne M5000B – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
14 Page 14
15 Page 15
16 Page 16
17 Page 17
18 Page 18
19 Page 19
20 Page 20
21 Page 21
22 Page 22
23 Page 23
24 Page 24
25 Page 25
26 Page 26
27 Page 27
28 Page 28
29 Page 29
30 Page 30
31 Page 31
32 Page 32
33 Page 33
34 Page 34
35 Page 35
36 Page 36
37 Page 37
38 Page 38
39 Page 39
40 Page 40
41 Page 41
42 Page 42
43 Page 43
44 Page 44
45 Page 45
46 Page 46
47 Page 47
48 Page 48
49 Page 49
50 Page 50
51 Page 51
52 Page 52
53 Page 53
54 Page 54
55 Page 55
56 Page 56
57 Page 57
58 Page 58
59 Page 59
60 Page 60
61 Page 61
62 Page 62
63 Page 63
64 Page 64
65 Page 65
66 Page 66
67 Page 67
68 Page 68
69 Page 69
70 Page 70
71 Page 71
72 Page 72
73 Page 73
74 Page 74
75 Page 75
76 Page 76
77 Page 77
78 Page 78
79 Page 79
80 Page 80
81 Page 81
82 Page 82
83 Page 83
84 Page 84
85 Page 85
86 Page 86
87 Page 87
88 Page 88
89 Page 89
90 Page 90
91 Page 91
92 Page 92
93 Page 93
94 Page 94
95 Page 95
96 Page 96
97 Page 97
98 Page 98
99 Page 99
100 Page 100
101 Page 101
102 Page 102
103 Page 103
104 Page 104
105 Page 105
106 Page 106
Page: / 106

Table of Contents:

  • Page 2 – i i; Teledyne Analytical Instruments; Copyright © 1999 Teledyne Analytical Instruments
  • Page 3 – iii; Photometric Analyzer; Table of Contents
  • Page 4 – i v
  • Page 5 – OPERATING INSTRUCTIONS; Model; NEMA 4 Bulkhead Mount
  • Page 9 – Introduction; Overview; WATER MONITORING; Background
  • Page 11 – NOTE: Range may be higher or lower per application.; OTHER NIR ABSORBERS (that can be measured)
  • Page 13 – Main Features of the Analyzer; use. The main features of the analyzer include:
  • Page 14 – Operator Interface; Escape
  • Page 15 – Digital Meter Display:; The meter display is a Light Emitting Diode
  • Page 16 – Control Section Interface Panel; Installation; Note: If you require highly accurate Auto-Cal timing, use external
  • Page 19 – Unpacking the Control/Analysis Unit
  • Page 20 – Figure 2-3: Interface Panel of the Model 6600 Control Section; For safe connections, ensure that no uninsulated wire extends; . Stripped wire ends must insert completely
  • Page 21 – Figure 2-4: Primary Input Power Connections; There are eight DC output signal connectors on
  • Page 22 – Concentration
  • Page 23 – Range; Range 1
  • Page 24 – Figure; Digital Remote Cal Inputs
  • Page 25 – Remote Calibration Protocol:
  • Page 27 – Testing the System
  • Page 29 – Operation; cautiously; Enter
  • Page 31 – Figure 3-1: Hierarchy of System Functions and Subfunctions
  • Page 32 – The System Function; Setting up an AUTO-CAL
  • Page 33 – AUTO—CAL
  • Page 34 – Password Protection; PWD; or
  • Page 35 – Installing or Changing the Password; Characters Available for Password Definition:
  • Page 36 – If you log off the system using the LOGOUT function in the; Logging Out
  • Page 37 – System Self-Diagnostic Test; System Self Diagnostic Test; the sampling cell at the time of the SELF-TEST.
  • Page 38 – Checking Linearity with ALGORITHM; AUTO
  • Page 39 – Digital Filter Setup; The 5000BF has the option of decreasing or increasing the amount; Setting
  • Page 40 – Filter or Solenoid Setup; Press the System key to start the System function:
  • Page 42 – 1 Analog 4 to 20 mA Output Calibration
  • Page 44 – The Zero and Span Functions; be sure to limit pressure to 40 psig or less when turning it back on.; Zero Cal; Auto Mode Zeroing; Zero
  • Page 45 – Manual Mode Zeroing
  • Page 47 – Span Cal
  • Page 48 – automatically; How to access the offset function:
  • Page 50 – The Alarms Function
  • Page 52 – The Range Select Function
  • Page 53 – Auto Screen
  • Page 54 – if; Precautions
  • Page 55 – Figure 3-2: Examples of Autoranging Schemes
  • Page 56 – The Analyze Function; Programming; CAUTION: The programming functions of the Set Range and Curve; rp
  • Page 57 – ALGORITHM; Now you will be able to select the APPLICATION and ALGO-; The Set Range Screen; The Set Range screen allows reprogramming of the three analysis
  • Page 58 – Fluid Name; st
  • Page 59 – The Curve Algorithm Screen
  • Page 60 – Note: The span fluid used to span the analyzer must be >90% of the
  • Page 61 – Maintenance; WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS; Fuse Replacement; The fuses are located inside the main housing on the Electrical
  • Page 62 – Figure 4-1: Removing Fuse Block Cap and Fuse from Housing; Power
  • Page 63 – Preamp; O K; Figure 4-3: Control Unit Major Internal Components
  • Page 64 – See the drawings in the Drawings section or addendum in back of
  • Page 65 – NEC or ATEX Type
  • Page 69 – Operations/Analysis Unit 4; Operations; During shipment, it is possible that components have been jarred
  • Page 70 – Operations/Analysis Unit; Control Functions; Analysis Section, fully explosion proof, Z or X; This function is hard wired into the
  • Page 71 – Preliminary Inspection
  • Page 72 – Power On Observations; customer to ON and make the following observations:
  • Page 73 – If the test procedure was normal, the devices should have been
  • Page 74 – Calibration; Standardization Fluids; into the cell at a flow rate of about 50 cc/minute to allow; Gas samples; can
  • Page 76 – The analyzer is now calibrated. It is often desirable to check; DIRECTIONS FOR USING TELEDYNE CALIBRATION OPTIONS
  • Page 77 – A zero may be established with a
  • Page 78 – as the off-line method used to verify the process variables.
  • Page 79 – good; ratio; This ratio is kept within the 1:2 to 2:1 values for; Zero calibration check after process on-line calibration.
  • Page 80 – zero offset check
  • Page 81 – ppm water in the 5020 meter or output. Check.
  • Page 83 – Zero in; Span in; G O
  • Page 85 – Operational Theory Cont'd; This energy is then fed through the sample, which is temperature
  • Page 86 – Source Module; In some applications where we have an abundance of energy due to
  • Page 87 – Optical System
  • Page 88 – sample temperature during analysis.
  • Page 89 – Power Module - Block Diagram
  • Page 90 – See Figure 4-3. The power to the analyzer unit, enables the switching; Detector compartment
  • Page 91 – Routine Maintenance; Automatic operation and routine operational duties
  • Page 92 – DAILY; NOTE: Be sure to wear UV filtering eye goggles.
  • Page 101 – Appendix; 000B Digital Control Module:; Operating Temperature:
  • Page 102 – Typical Analytical Performance Specifications:; will vary per application; Other Specifications:; Internal calibration:; Area Classification
  • Page 103 – * If ordered with purge option; General Installation Note:; Protect the instrument from
Loading the manual

i

Photometric Analyzer

Teledyne Analytical Instruments

OPERATING INSTRUCTIONS

Model

5000B

Photometric Analyzer

HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.

HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.

ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.

DANGER

M5000B

11/02/2006

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - i i; Teledyne Analytical Instruments; Copyright © 1999 Teledyne Analytical Instruments

i i Model 5000B Teledyne Analytical Instruments Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computerlanguage in whole or in part, in an...

Page 3 - iii; Photometric Analyzer; Table of Contents

iii Photometric Analyzer Teledyne Analytical Instruments Table of Contents Part I: Control Unit ................................ Part I: 1-1 Part II: Analysis Unit ............................ Part II: 4-1 Appendix ......................................................... A-1

Page 4 - i v

Other Teledyne Models

All Teledyne Other