Page 3 - Date Rev; 707 T-Series initial release
i ABOUT THIS ADDENDUM This addendum, part number 07273, is to be used in conjunction with the Model T360 operation manual, part number 07272. Where operation of the Model T360U diverges from that of the Model T360, this addendum takes precedence. Please note that the T360 operation manual contains i...
Page 5 - iii; TABLE OF CONTENTS; Figures
iii TABLE OF CONTENTS About This Addendum ................................................................................................................ i Revision History .......................................................................................................................... i T...
Page 7 - CO; SPECIFICATIONS AND AGENCY APPROVALS; Table 2-1. T360U Specifications and Agency Approvals
5 1. OVERVIEW The Model T360U is a close derivative of the Model T360 CO Analyzer. The main differences are as follows: The optical bench is longer (14 m), making the instrument more sensitive at low CO 2 levels. The instrument has a user-selectable full scale range of 0-100 PPB to 0-100 PPM of ...
Page 8 - PARAMETER SPECIFICATION; AC Power; ELECTRICAL CONNECTIONS
6 PARAMETER SPECIFICATION AC Power 100 – 120V 50/60 Hz (120W) 220 – 240 V 50/60 Hz (144W) Environmental Conditions Installation Category (Over voltage Category) II Pollution Degree 2 Standard I/O 1 Ethernet: 10/100Base-T 2 RS-232 (300 – 115,200 baud) 2 USB device ports 8 opto-isolated digital status...
Page 9 - PNEUMATIC CONNECTIONS; Figure 4-1. T360U Pneumatic Flow Diagram
7 4. PNEUMATIC CONNECTIONS This section provides information on the pneumatic configurations. Figure 4-1 shows the basic pneumatic flow of the T360U. Figure 4-1. T360U Pneumatic Flow Diagram
Page 10 - Chassis; Figure 4-2. Pneumatic Connections–Basic Configuration
8 4.1. B ASIC AND Z ERO /S PAN V ALVE O PTION C ONFIGURATIONS CAUTION Sample and calibration gases should only come into contact with PTFE (Teflon), FEP, glass, stainless steel or brass. Do NOT operate without first removing dust plugs from rear panel fixtures. NOTE To prevent dust from entering the...
Page 11 - Table 4-1. Model T360U Pneumatic Connections
9 Needle valve, s et f low to 800 cc /min (+/- 10%) Chassis Sample Zero Exhaust Vent Span Pressure Span To Purge Sample gas Fro m Purge Fl owme ter Pre ssure re gula tor, set out put t o 5-10 psig 10-30 psig CO2 free ga s Sp an ga s CO2 free gas Pre ssure Re gula tor (5 -10 psig output ) Ne edle val...
Page 12 - CAUTION; pressure using maximum of 10 meters of 1⁄4” PTFE tubing.
10 4.2. M AKING THE P NEUMATIC C ONNECTIONS CAUTION Venting should be outside the shelter or immediate area surrounding the instrument. 1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long. 2. Attach sources of zero air and span gas 3. S...
Page 14 - This page intentionally left blank.
Page 15 - T360U SPARE PARTS LIST
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 PARTNUMBER DESCRIPTION 037600000 AKIT, EXP KIT, SODA LIME CH47 037860000 ORING, TEFLON, RETAINING RING, 47MM (KB) 039060100 ASSY, PRESSURE TRANSDUCER 039260501 DETECTOR, CO2, w/BANDPASS FILTER* 040010000 ASSY, FAN REAR PANEL (B/F) 040030300 PCA...