Page 4 - Manual Print History; Warranty
Manual Print History The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subse- quent updates. Addenda, which are released between revisions, contain important change i...
Page 5 - Table of Contents
Table of Contents 1 General Information ................................................................ 1-2 Introduction .............................................................................................. 1-2Specifications ....................................................................
Page 7 - General Information; SECTION 1
Page 8 - Introduction; counts; Ordering Information; Ordering options for the HCM-300 series power supply:
1-2 General Information Introduction The Hastings Capacitance Manometer Power Supply (HCM-300) supplies stable ±15 VDC specifically for operating the Hastings Capacitance Manometer Trans- ducer. The transducer returns a signal to the HCM-300 which displays an accurate pressure reading. Specification...
Page 9 - Installation; SECTION 2
Page 10 - Receiving Inspection; Cut out panel as shown in the diagram below.
2-2 Installation Receiving Inspection Carefully unpack the HCM-300. Inspect all items received at this time for any obvious damage that may have occurred during shipment. Immediately advise the shipping agent if there is any damage to items in your shipment. Identify each itemand the quantity receiv...
Page 11 - ) Slide power supply into panel from the front.
Installation 2-3 To panel mount the HCM-300: 1) Slide power supply into panel from the front. 2) Attach brackets to the side of the power supply. 3) Tighten screws against panel. See Figure below:
Page 12 - Hastings Capacitance Manometer Power Supply; back of the instrument in the socket marked AC Power.
2-4 Installation Hastings Capacitance Manometer Power Supply Cable Connections The HCM-300 is supplied with a power cable. The 3-wire cable plugs into the back of the instrument in the socket marked AC Power. A 5-wire cable connects the instrument to the transducer. The diagram belowshows the prop...
Page 13 - Operating Information; SECTION 3
Page 14 - ) Connect the interface cable from the HCM-300 to the transducer.; Caution; which is not covered by the warranty.; Analog Output; displayed pressure. The analog output signal is given by:
3-2 Operating Information Normal Start-up 1) Connect the interface cable from the HCM-300 to the transducer. (See figure on page 2-4 of this manual) 2) Verify that the AC power to be applied to the HCM-300 is correct (115 VAC). 3) Apply AC power to the HCM-300. 4) The HCM transducer has a local zero...
Page 15 - Internal Adjustments; Full Scale; instrument. Apply a calibrated +10 VDC signal to the; Notes on internal adjustments:
Operating Information 3-3 Conversely, if the analog output voltage is measured, then the pressure reading is given by: Internal Adjustments The internal adjustments are factory set and should never need to be changed. However, if the need arises and suitable standards are available the following adj...
Page 16 - Display Configuration; configuration jumpers; Jumper Location
Display Configuration The jumper position on the power supply circuit board determines the location of the decimal on the display.The jumper is factory set to match the configuration specified by the customer. Should the need arise to change the display configura- tion, this jumper can be easily mov...
Page 17 - Warranty and Repair; SECTION 4
Page 18 - Warranty Repair; RMA number before returning any product for repair.; Teledyne Brown Engineering
4-2 Maintainance and Repair Warranty Repair Hastings Instruments will repair or replace, at their option, any product covered under the warranty expressed in the front of this manual, that is returned transpor- tation prepaid to their plant in Hampton, Virginia. All products returned must be accompa...