Page 2 - Manual Print History
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 2 of 16 Manual Print History The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which ar...
Page 3 - Table of Contents
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 3 of 16 Table of Contents 1. GENERAL INFORMATION............................................................................................................................................ 4 1.1. F EATURES ...............................................
Page 4 - General Information; Features
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 4 of 16 1. General Information This manual contains technical and general information relating to the installation, operation, and calibration of the Hastings 760 Plus vacuum gauge which is comprised of a piezo vacuum transducer (760s) and power supp...
Page 5 - Specifications
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 5 of 16 1.2. Specifications Piezo Vacuum Transducer (760s): Accuracy ......................................................................................................± 0.25% FS* (760 Torr) Operating pressure ........................................
Page 6 - Receiving Inspection
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 6 of 16 2. Installation This section contains the necessary instructions to assist in placing the HPM-760 Plus digital vacuum gauge into operation as quickly and easily as possible. Please read the following thoroughly before attempting to install or...
Page 7 - Power Supply/Display Installation; Power Supply/Display Cables and Wiring
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 7 of 16 2.3. Power Supply/Display Installation Cut a panel opening to the following dimensions; 3.62” (92mm) wide, 1.77” (45mm) tall. Insert the power supply/display through the panel cutout and secure in place using mounting hardware supplied. 2.3.1...
Page 8 - Operating Information; General
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 8 of 16 3. Operating Information This section contains information on operating the PIEZO Resistive Transducer in any of its various configurations. Refer to the appropriate section for information on the mode in use. 3.1. General The piezoresistive ...
Page 9 - Power Supply/Display Input Power
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 9 of 16 3.2. Piezo Transducer Input Power The 760s piezo transducer operates with a DC input from 15 to 30 volts. Use the 4-pin female connector supplied and makes necessary connections as outlined. See illustration in Section 2 of this manual. 3.3. ...
Page 10 - Rear Panel; Power Supply Relay Set Point Adjustment
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 10 of 16 3.5.2. Rear Panel Refer to page 19, 7.1.2 for rear view and specific pin outs and connector arrangements. AC Input Connector J1 Universal input power 110vac to 220 VAC. No hardware settings required. J3 Input/Output Connector Sensor Input (p...
Page 11 - Pressure Measurements
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 11 of 16 Auxiliary Relay Contacts The THPS-760 Power Supply has SP1 and SP2 relays for various external switching/monitoring applications. Each relay has a single set of Form “C” contacts. (Choice of normally open or normally closed contacts). The re...
Page 12 - Hastings 760 Plus Piezo Transducer Zero and Span Adjustment; Hastings 760s Option-5 Transducer Zero and Span Adjustment
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 12 of 16 4. Calibration 4.1. Hastings 760 Plus Piezo Transducer Zero and Span Adjustment 1. Install the transducer and power supply/display then apply operating power. Wait a mini mum of 15 minutes before proceeding. 2. Evacuate system to a pressure ...
Page 14 - Warranty Repair Policy; Non-Warranty Repair Policy; TOLL FREE
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 14 of 16 6. WARRANTY 6.1. Warranty Repair Policy Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship. This warranty does not apply to defects or failures re...
Page 15 - Diagrams and Drawings; Front Panel
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 15 of 16 7. Diagrams and Drawings 7.1. Front Panel The THPS-760 has a 4 1/2-digit display that displays continuous readout. 7.2. Rear Panel Decal is to be black ink printed on silver compucal, with adhesive on the back
Page 16 - Mounting; Panel Mounting
Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge Page 16 of 16 7.3. Mounting Cut out panel as shown in the diagram below. 7.4. Panel Mounting 1) Slide power supply into panel from the top. 2) Attach brackets to the side of the power supply. 3) Tighten screws against panel.