Page 2 - Series; ii; Teledyne; Copyright © 2010 Teledyne Analytical Instruments
Series 4060 ii Teledyne Analytical Instruments Copyright © 2010 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in an...
Page 3 - Safety Messages; GENERAL; Technician Symbol
Series 4060 Teledyne Analytical Instruments iii Safety Messages Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others...
Page 4 - iv; Occasionally, some instruments are customized for a particular
Series 4060 iv Teledyne Analytical Instruments CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL. IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS ...
Page 5 - Additional Safety Information; DANGER; COMBUSTIBLE GAS USAGE
Series 4060 Teledyne Analytical Instruments v Additional Safety Information DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for use in a non-hazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed si...
Page 7 - Discharge the anode supply voltage.; Work with one hand only.
Series 4060 Teledyne Analytical Instruments vii 3. External hazardous voltages which may be connected to the Analyzer alarm relay connections. Procedure for Removal of Internal Inaccessible Shock Hazards CAUTION: SERVICING OR MAINTENANCE OF THE ANALYZER SHOULD ONLY BE DONE BY SUITABLE TRAINED PERSON...
Page 8 - viii
Series 4060 viii Teledyne Analytical Instruments WARNING: THIS INSTRUMENT IS DESIGNED TO BE OPERATED IN A NONHAZARDOUS AREA. THE ANALYZER USES HYDROGEN GAS AND/OR OTHER COMBUSTIBLE GASES IN ITS OPERATION. THIS EQUIPMENT, IF NOT USED AND MAINTAINED PROPERLY CAN BE AN EXPLOSION HAZARD. THE ANALYZER, D...
Page 9 - Table of Contents
Series 4060 Teledyne Analytical Instruments ix Table of Contents Safety Messages .......................................................................... iii Additional Safety Information ...................................................... v List of Figures ........................................
Page 10 - User Connections
Series 4060 x Teledyne Analytical Instruments 3.3 User Connections 28 3.3.1 Electrical Power Connections 28 3.3.2 Electronic Connections 28 3.3.2.1 Primary Input Power 29 3.3.2.2 Fuse Installation 30 3.3.2.3 50-Pin Equipment Interface Connector 30 3.3.2.4 Analog Output 30 3.3.2.5 Alarm Relays 32 3.3...
Page 11 - Analyzer Operation
Series 4060 Teledyne Analytical Instruments xi 4.5 Sample Pump 44 4.6 Analyzer Operation 44 4.6.1 Default Parameters 46 4.6.2 Style Conventions 46 4.6.3 Navigation and Data Entry 47 4.6.3.1 Arrow Keys 47 4.6.3.2 ENTER 47 4.6.3.3 ESCAPE 48 4.6.4 Menu Structure 48 4.6.4.1 Main Menus 49 4.6.5 Span 50 4...
Page 12 - xii
Series 4060 xii Teledyne Analytical Instruments 5.1.4 Collector Cable 68 5.2 Temperature Control Electronic Check 68 5.3 Ignition and/or Flame Guard Circuit Checks 69 5.4 Sampling System 70 5.5 Printed Circuit Board Descriptions 71 5.5.1 Flame Guard and Anode Power Supply PCB 71 5.5.2 Proportional T...
Page 13 - List of Figures
Series 4060 Teledyne Analytical Instruments xiii List of Figures Figure 1-1: Model 4060 Front Panel Interface ............................... 17 Figure 1-2: Model 4060 Rear Panel .............................................. 17 Figure 2-1: Internal Temperature Controllers .........................
Page 14 - xiv; List of Tables
Series 4060 xiv Teledyne Analytical Instruments List of Tables Table 3-1: Analog Output Connections ......................................... 31 Table 3-2: Analog Concentration Output—Example ..................... 31 Table 3-3: Analog Range ID Output—Example ............................. 32 Table 3-...
Page 15 - Introduction; Teledyne Analytical Instruments Series 4060, Trace Aromatic; Main Features of the Analyzer; The Analyzer is sophisticated yet simple to use. A dual display on
Series 4060 Introduction Teledyne Analytical Instruments 15 Introduction Teledyne Analytical Instruments Series 4060, Trace Aromatic Hydrocarbon Analyzer, is a versatile instrument designed to measure the quantity of aromatic hydrocarbons, primarily benzene, present in a positive pressure sample gas...
Page 16 - Principle of Operation; Analyzer Description
Introduction Series 4060 16 Teledyne Analytical Instruments Extensive self-diagnostic testing at startup and on demand with continuous power supply monitoring. RS-232 serial digital port for use with a computer or other digital communication device. Analog outputs for concentration and range i...
Page 17 - Figure 1-1: Model 4060 Front Panel Interface
Series 4060 Introduction Teledyne Analytical Instruments 17 Figure 1-1: Model 4060 Front Panel Interface Figure 1-2: Model 4060 Rear Panel
Page 18 - Gas purity certification.
Introduction Series 4060 18 Teledyne Analytical Instruments 1.4 Applications Monitoring the benzene concentration of carbon dioxide gas for use in the beverage industry. Monitoring aromatic contamination in air liquefaction and other gas production processes. Gas purity certification. Detect...
Page 19 - Operational Theory; Series 4060 Analyzer for Benzene uses a FID detector and a Gas; Modes of Operation; The analyzer has 2 modes of operation depending on the position of
Series 4060 Operational Theory Teledyne Analytical Instruments 19 Operational Theory 2.1 Introduction Series 4060 Analyzer for Benzene uses a FID detector and a Gas Chromatograph (GC) Column. Using a carrier gas and a microprocessor actuated switching valve, a fixed volume of sample is pushed into t...
Page 20 - Addendum and Testing Results; Analysis Mode; Analyzer Subsystems
Operational Theory Series 4060 20 Teledyne Analytical Instruments 1. Valve Position A— Sampling Mode In this mode the analyzer configures the operational valve to back flush the column and charge the sample loop. The Sample Mode is programmed to continue for a 7 minute period by factory default; how...
Page 21 - Adjustments are made using the appropriate control on the front
Series 4060 Operational Theory Teledyne Analytical Instruments 21 analyzer chassis. They are accessible by opening the front door of the analyzer . Adjustments are made using the appropriate control on the front panel. The analyzer contains three separate isothermal chambers ‘SAMPLE’, ‘FID’, and ‘CO...
Page 22 - The Piping Diagram for the standard instrument is shown in Figure
Operational Theory Series 4060 22 Teledyne Analytical Instruments supplied at constant pressures using two stage stainless steel diaphragm gas regulators. The recommended pressure range is 30 to 8 0 psig, however, the span gas should be supplied at a pressure of 20 psig to the restrictor fitting eit...
Page 23 - Figure 2-2: Typical Piping Diagram for the Model 4060
Series 4060 Operational Theory Teledyne Analytical Instruments 23 Figure 2-2: Typical Piping Diagram for the Model 4060
Page 24 - The thermistor acts as the sensor in the flame guard circuit. Its
Operational Theory Series 4060 24 Teledyne Analytical Instruments 2.4 Detection Cell The upper section of the stainless steel flame ionization cell houses the cylindrical collector electrode, the high voltage (+260 VDC) anode-igniter coil, and the sensing thermistor of the flame guard circuit (see c...
Page 25 - Electrometer-Amplifier PC Board
Series 4060 Operational Theory Teledyne Analytical Instruments 25 2.4.1 Electrometer-Amplifier The collector cable is coupled directly to a coaxial fitting located on the electrometer-amplifier PC board. The PC board is located on the side panel next to but outside of the isothermal chamber. It cons...
Page 26 - The high voltage anode power supply components are mounted on
Operational Theory Series 4060 26 Teledyne Analytical Instruments 2.4.2 Anode Power Supply The high voltage anode power supply components are mounted on the anode power supply printed circuit board. High voltage regulation is achieved through the use of series-connected zener diodes. The simplicity ...
Page 27 - Installation; Unpacking the Analyzer; Note: To completely slide the analyzer out of the enclosure, pull
Series 4060 Installation Teledyne Analytical Instruments 27 Installation Installation of the Series 4060 Analyzer includes: 1. Unpacking 2. Mounting 3. Gas connections 4. Electrical connections 5. Testing the system. 3.1 Unpacking the Analyzer Although the analyzer is shipped with all the materials ...
Page 28 - All user connections are made on the rear panel. Consult the input-
Installation Series 4060 28 Teledyne Analytical Instruments levers found almost at the end of the sliders, both sides at the same time. Regardless of configuration, the analyzer/system must be installed on a level surface with sufficient space allocated on either side for personnel and test equipmen...
Page 29 - For safe connections, no uninsulated wiring should be able to come; The power cord receptacle and fuse block are located in the same
Series 4060 Installation Teledyne Analytical Instruments 29 For safe connections, no uninsulated wiring should be able to come in contact with fingers, tools or clothing during normal operation. CAUTION: USE SHIELDED CABLES. ALSO, USE PLUGS THAT PROVIDE EXCELLENT EMI/RFI PROTECTION. THE PLUG CASE MU...
Page 30 - The fuse block, at the right of the power cord receptacle, accepts US
Installation Series 4060 30 Teledyne Analytical Instruments 3.3.2.2 F USE I NSTALLATION The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. 3.3.2.3 50-P IN E QUIPMENT I NTERFACE C ONNECTOR ...
Page 31 - Channel 1, negative ground; Ppb
Series 4060 Installation Teledyne Analytical Instruments 31 Table 3-1: Analog Output Connections Pin Function 3 Channel 2, + 4-20 mA, floating 4 Channel 2, – 4-20 mA, floating 5 Channel 1, + 4-20 mA, floating 6 Channel 1, – 4-20 mA, floating 8 Channel 2, 0-1 VDC 23 Channel 2, 0-1 VDC, negative groun...
Page 32 - Table 3-3: Analog Range ID Output—Example; Range; Threshold Alarm 1; Can be configured as failsafe or non-failsafe
Installation Series 4060 32 Teledyne Analytical Instruments 40 0.4 10.4 50 0.5 12.0 60 0.6 13.6 70 0.7 15.2 80 0.8 16.8 90 0.9 18.4 100 1.0 20.0 Table 3-3: Analog Range ID Output—Example Range Voltage Signal Current Signal Application Output (VDC) Output (mA DC) Range 1 0.20 7.2 0–100 ppb Benzene Ra...
Page 33 - Note: Reset by pressing the STANDBY button to remove power.; Pin Contact
Series 4060 Installation Teledyne Analytical Instruments 33 Threshold Alarm 2: Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold). Can be configured as failsafe or non-failsafe Can be configured as non-latching ...
Page 34 - Remote Calibration; Pin; Remote Calibration Protocol:
Installation Series 4060 34 Teledyne Analytical Instruments 3.3.2.6 D IGITAL R EMOTE C AL I NPUTS The digital remote calibration input accepts 0 V (off) or 24 VDC (on) for remote control of calibration. (See Remote Calibration Protocol below.) See Table 3-5 for pin connections. Span: Floating input....
Page 35 - When the CRC closes, remove the span command.; A serial digital input/output for local network protocol. At this
Series 4060 Installation Teledyne Analytical Instruments 35 2. When the CRC closes, remove the span command. When CRC opens again, the span calibration is done, and the sample is being analyzed. 3.3.2.7 R ANGE ID R ELAYS There are three dedicated Range ID relay contacts. They are assigned to relays ...
Page 36 - The input functions using RS-232 that have been implemented
Installation Series 4060 36 Teledyne Analytical Instruments Table 3-7: Pin out of 50 pin D-Sub Connector pin # Description pin # Description 1 26 2 27 3 + Output 4-20 ma - Channel 2 28 Alarm 1 C Contact 4 - Output 4-20 ma - Channel 2 29 5 + Output 4-20 ma – Channel 1 30 6 - Output 4-20 ma – Channel ...
Page 37 - The RS-232 protocol allows some flexibility in its
Series 4060 Installation Teledyne Analytical Instruments 37 Table 3-8: Commands via RS-232 Input Command Description AS <enter> Immediately starts an autospan. AL <enter> Immediately revert to Analyze (CH4 Mode) RL <enter> Change to Manual Range LOW RM <enter> Change to Manua...
Page 38 - Gas connections to the instrument are made at the 1/8”or 1/4”
Installation Series 4060 38 Teledyne Analytical Instruments Figure 3-3: Gas Connections Gas connections to the instrument are made at the 1/8”or 1/4” stainless steel tube fittings provided on the rear panel. Note that the Purge and Sensor Vent fittings are 1/4” while all other gas connections are 1/...
Page 39 - CARRIER GAS: NITROGEN GAS, ULTRA ZERO GRADE WITH THC; Normally, four supporting gases of different composition (see
Series 4060 Installation Teledyne Analytical Instruments 39 not representative of the current contents of the line, or which might keep contaminants from being purged out of the system. CAUTION: THE GASES USED MUST BE OF THE HIGHEST QUALITY, ULTRA ZERO GRADES, AS SHOWN BELOW. FAILURE TO DO SO WILL R...
Page 40 - The sample bypassed by the back-pressure regulation system vents; Placing the System in Operation
Installation Series 4060 40 Teledyne Analytical Instruments Addenda included with this manual to determine if special conditions apply. 3.3.3.1 E FFLUENT All the gases introduced into the detection cell vent from one fitting at the rear of the analyzer. TAI recommends that the cell be permitted to v...
Page 41 - Operation; Addendum and; Fuel; Gas Connections
Series 4060 Operation Teledyne Analytical Instruments 41 Operation This section of the manual describes how to setup and operate the Model 4060 Analyzer. Sections 4.1 through 4.5 describe preliminary steps and equipment needed to operate the analyzer. Beginning with Section 4.6, the actual operation...
Page 42 - Make the following checks of the installation before proceeding; Activating the Support Gases; All gas controls are located on the front panel as shown in Figure 1-
Operation Series 4060 42 Teledyne Analytical Instruments CAUTION: THE GASES USED MUST BE OF THE HIGHEST QUALITY, ULTRA ZERO GRADES, AS SHOWN BELOW. FAILURE TO DO SO WILL RESULT IN CONTAMINATION AND FAILURE TO DETECT LOW PPB BENZENE. AIR: USE WATER PUMPED AIR WITH THC LESS THAN 0.1 PPM. DO NOT USE OI...
Page 43 - Note: Adjust fuel settings only when the red LED (flame failure
Series 4060 Operation Teledyne Analytical Instruments 43 4.3.1 Air 1. Set the air source regulator to 40 psig. 2. Adjust the analyzer air regulator until the air pressure gauge reads the recommended air pressure of 7.0 psig. After the air is flowing through the sensor and warm-up time has been compl...
Page 44 - Verification of the Flame Guard Circuit; Although the Model 4060 has been programmed for your application
Operation Series 4060 44 Teledyne Analytical Instruments factory. If the ignition process fails, the instrument will attempt to ignite the flame 4 more times. If it continues to fail after the fifth attempt, a flame failure message will appear on the display. If this occurs refer to Section 5. 4.4.1...
Page 46 - Auto; ANALYZE
Operation Series 4060 46 Teledyne Analytical Instruments 4.6.1 Default Parameters The versatility of this analyzer usually results in significant changes being made to parameters over the course of time to better suit a particular application. Occasionally processes change requiring alteration to al...
Page 47 - ANALYSIS SCREEN
Series 4060 Operation Teledyne Analytical Instruments 47 particular key it will be typed as text using all caps. Example: this is the ENTER key. Screen Modes: Times New Roman 12 pt. italic. Example: Analysis Mode or Setup Mode. Screen Readout: Arial Narrow, 12 pt bold. Example: AUTOCAL. 4.6.3 Naviga...
Page 48 - ESCAPE
Operation Series 4060 48 Teledyne Analytical Instruments 1. When the selected option is a function on the MAIN MENU or any SUB MENU screen, the function name appears with an arrow next to it . In some cases, the item in the menu is the parent of another SUB MENU , in other cases, the item in the men...
Page 49 - MAIN MENU; AUTOCAL is an optional feature not
Series 4060 Operation Teledyne Analytical Instruments 49 2. The MAIN MENU and SUB MENU screens have a flashing arrow on the left side indicating which item is being pointed to. 3. MODIFIABLE ITEM screens have a flashing value somewhere in the line. 4.6.4.1 M AIN M ENUS The MAIN MENU consists of 15 f...
Page 50 - Enter; Span Begin
Operation Series 4060 50 Teledyne Analytical Instruments GROUP SETUP : Determines compound names, associated timing and measurement ranges. Do not change these. PSWD: Used to establish password protection or change the existing password. LOGOUT: Logging out prevents unauthorized tampering with...
Page 51 - Pressing enters the SPAN mode in the Sampling Phase.
Series 4060 Operation Teledyne Analytical Instruments 51 ________________________________________________ Begin Span: ENTER-YES ESCAPE-NO ________________________________________________ Pressing <Enter> enters the SPAN mode in the Sampling Phase. This is similar to the Analysis Mode both in t...
Page 52 - Analyze Mode
Operation Series 4060 52 Teledyne Analytical Instruments To initiate a self-test: < UP/DOWN >From the MAIN MENU scroll to the SELF-TEST function. < Enter > Activate the SELF-TEST function by pressing <Enter>. This brings up the SELF-TEST initialization screen. _____________________...
Page 53 - Note: For the purpose of example, Alarm 2 has been set
Series 4060 Operation Teledyne Analytical Instruments 53 -> AL-1 DEFEATED AL-1 HIGH AL-1 NON-FAILSAFE AL-1 NON-LATCHING AL-1 20.00 ppb AL-2 ACTIVE AL-2 LOW AL-2 FAILSAFE AL-2 LATCHING AL-2 50.00 ppb ________________________________________________ Note: For the purpose of example, Alarm 2 has bee...
Page 54 - The latching property configures the alarm such that the; RELEASING A LATCHED ALARM:; When alarm conditions are
Operation Series 4060 54 Teledyne Analytical Instruments LATCHING: The latching property configures the alarm such that the user must manually relieve the alarm condition even though the concentration no longer violates the trip point of the alarm. So, if an alarm is NON-LATCHING, and the analysis c...
Page 56 - CH1 Gain; Not applicable to the 4060 configuration.
Operation Series 4060 56 Teledyne Analytical Instruments ______________________________________________ -> CH1 Offset CH1 Gain CH1 (Wave, Id for Ch2, or Group Output) CH1 (Select Groups) CH2 (Wave, Id for Ch2, or Group Output) CH2 (Select Groups) ______________________________________________ 4.6...
Page 57 - Note: If you require highly accurate AUTOCAL timing, use; PASSWORD; AUTO SPAN Enabled
Series 4060 Operation Teledyne Analytical Instruments 57 the Span functions as described in Section 4.6.5, and follow the precautions given there. Auto Span performs a span calibration using the target span value set in the Span feature menu. Note: If you require highly accurate AUTOCAL timing, use ...
Page 58 - Note: For instruction on how to set the parameters, please refer
Operation Series 4060 58 Teledyne Analytical Instruments Note: For instruction on how to set the parameters, please refer to the Data Entry section. The first line tells how soon the next automatic span will occur. Pressing <Enter> on this line has no effect. The second line Enables/Disables t...
Page 59 - Note: If you use password security, it is strongly advised to keep a; ENTERING A PASSWORD; AUTOCAL
Series 4060 Operation Teledyne Analytical Instruments 59 If user has LOGGED OUT, then only the following SUB MENU procedures may be executed: 1. PASSWORD 2. LOGOUT (which only repeats the logout) 3. MODEL Note however, that the instrument can still be used for analysis without entering the password....
Page 60 - Enter a Password
Operation Series 4060 60 Teledyne Analytical Instruments ENTER PASSWORD SCREEN ________________________________________________ Enter a Password 'T' 'A' 'I' ________________________________________________ < UP/DOWN >Use the UP/DOWN keys to change the letters to the proper password. < Enter...
Page 61 - Reenter a new Password; Note: If you log off the system using the
Series 4060 Operation Teledyne Analytical Instruments 61 < UP/DOWN >Use the UP/DOWN keys to change the letters to the new password. < Enter > Press <Enter> to advance to the next letter < Enter > The last <Enter> moves to the REPEAT PASSWORD ENTRY screen. < Escape &g...
Page 63 - >From the MAIN MENU, scroll to the STANDBY
Series 4060 Operation Teledyne Analytical Instruments 63 < UP/DOWN >From the MAIN MENU, scroll to the STANDBY function using the UP/DOWN keys. < Enter > Pressing <Enter> places the instrument in STANDBY. To exit STANDBY, scroll again to the STANDBY function and press <Enter> ...
Page 65 - Maintenance & Troubleshooting
Series 4060 Maintenance Teledyne Analytical Instruments 65 Maintenance & Troubleshooting WARNING: DANGEROUS HIGH VOLTAGES EXIST INSIDE THIS INSTRUMENT. THERE ARE NO USER SERVICEABLE PARTS WITHIN THE COVER ON THE INSIDE OF THE DOOR, INSIDE THE ISOTHERMAL CHAMBER, (SAMPLE SYSTEM), AND ON THE ELECT...
Page 66 - Measuring Circuit Electrical Checks; negative
Maintenance Series 4060 66 Teledyne Analytical Instruments attempt to evaluate performance on sample gas. If analyzer sensitivity is questionable, use the span gas. For all other evaluations, use the zero gas and low range for maximum sensitivity. The important consideration is to control as many va...
Page 67 - If the checks outlined above indicate that conditions are normal,
Series 4060 Maintenance Teledyne Analytical Instruments 67 2. If no reading is obtained, disconnect the anode-igniter cable and check for the voltage on pin “J4-1” of the connector located at PC board part number B74671. If the proper voltage is still not present, replace the flame guard and anode p...
Page 68 - Whenever the flame guard and anode power supply printed circuit; Temperature Control Electronic Check; Consult the 4060 schematic and assembly drawings, as well as the
Maintenance Series 4060 68 Teledyne Analytical Instruments Whenever the flame guard and anode power supply printed circuit board have been replaced the analyzer must be recalibrated. If the instrument performs as outlined in this section, the problem is not related to the measuring circuit electroni...
Page 69 - Note: If any of the components located inside the isothermal; Ignition and/or Flame Guard Circuit Checks; If the flame guard circuit will not hold the flame-out lamp off when
Series 4060 Maintenance Teledyne Analytical Instruments 69 measure resistance between that wire and the remaining undisturbed terminal. Resistance of the thermistor varies with its surrounding temperature. A reading of between 10KOhms and 16Kohms at around 25°C may be measured. (Under very cold cond...
Page 71 - Printed Circuit Board Descriptions; Flame Guard and Anode Power Supply PCB; Anode Power Supply; : The high voltage anode power supply; Flame Guard Circuit:; A thermistor-controlled, comparator circuit is
Series 4060 Maintenance Teledyne Analytical Instruments 71 5.5 Printed Circuit Board Descriptions The electronics circuitry of the analyzer is designed with the latest integrated circuit technology. The individual circuits which are required to process the incoming signal and condition it to provide...
Page 72 - Proportional Temperature Controller PCB
Maintenance Series 4060 72 Teledyne Analytical Instruments 5.5.2 Proportional Temperature Controller PCB Schematic No. B-30974 Assembly Dwg. No. B-30927 The temperature of the chamber to be controlled is regulated by a thermistor-directed electronic circuit. The thermistor and heating element are lo...
Page 73 - When the temperature reaches the selected control point, pin 13
Series 4060 Maintenance Teledyne Analytical Instruments 73 the thyristor gating circuit is continuously delivering gate current from pin 4 of A1, thus maintaining constant full wave AC power to the heater. When the temperature reaches the selected control point, pin 13 voltage is about the same as p...
Page 75 - Appendix; A.1 Specifications and Initial Settings:; Signal Output
Series 4060 Appendix Teledyne Analytical Instruments 75 Appendix A.1 Specifications and Initial Settings: Range(s): Refer to Addendum and Testing Results in Appendix B. Power Requirements: Refer to Addendum and Testing Results in Appendix B. Signal Output : Two analog channels: Channel 1: 0-1 VDC an...
Page 76 - A.2 Recommended Spare Parts List; Analytical Instruments
Appendix Series 4060 76 Teledyne Analytical Instruments A.2 Recommended Spare Parts List Qty. P/N Description 1 B74671A PC board, Flame guard & anode power supply 1 B79153A PC board, Electrometer 1 C75825A Micro-processor PC board 1 CP2540 Coaxial cable 1 B79154 Sensor Assy. 1 C62371A Display PC...
Page 77 - PC Board Assemblies
Series 4060 Appendix Teledyne Analytical Instruments 77 A.3 Drawing List (Refer to the Addendum and Testing Results section of Appendix B for actual drawing list for your application.) D77394 Outline Diagram D77488 Wiring Diagram C76068 Piping Diagram B79159 Schematic, Electrometer PCB B74672 Schema...
Page 79 - Appendix B; B1 Addendum and Testing Results
Series 4060 Appendix Teledyne Analytical Instruments 79 Appendix B B1 Addendum and Testing Results