Teledyne 3160 - Manuals
Teledyne 3160 – Manual in PDF format online.
Manuals:
Manual Teledyne 3160
Summary
ii TELEDYNE ANALYTICAL INSTRUMENTS Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in aretrieval system, or translated into any other language or computer language in whole or in part, in anyform or by a...
iv TELEDYNE ANALYTICAL INSTRUMENTS T T T T Tab ab ab ab able of Contents le of Contents le of Contents le of Contents le of Contents 1 Introduction 1.1 Features ....................................................................................... 1-1 1.1.1 Fixed Features .............................
1-1 Introduction 1 Introduction 1 Introduction 1 Introduction 1 Introduction 1 TELEDYNE ANALYTICAL INSTRUMENTS Introduction 1.1 Features 1.1.1 Fixed Features The Model 3160 is a microprocessor-based trace oxygen monitoring system that provides on-line monitoring of trace oxygen at parts-per-million ...
1-2 1 Introduction 1 Introduction 1 Introduction 1 Introduction 1 Introduction TELEDYNE ANALYTICAL INSTRUMENTS The 3160 features six full-scale linear ranges of analysis: • 0–1 PPM • 0–10 PPM • 0–100 PPM • 0–1000 PPM • 0–1% • 0–25% An AutoRanging feature automatically selects the range that is appro...
Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 2-1 TELEDYNE ANALYTICAL INSTRUMENTS Operational Theory 2.1 Principles of Operation The sub-systems that make up the analyzer are: 1. Micro-Fuel Cell Sensor 2. Sample System 3. Electronic Signal P...
Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 2-3 TELEDYNE ANALYTICAL INSTRUMENTS 2.1.1 Micro-Fuel Cell Sensor The Micro-Fuel Cell is a sealed plastic disposable oxygen transducer that mea- sures 1¼ inches in diameter and is ¾ inch thick. In...
2-4 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory TELEDYNE ANALYTICAL INSTRUMENTS • Electropolished 316L-type stainless steel components. To eliminate oxygen absorption and desorption from the internalwetted surfaces of the sample system com...
Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 2-7 TELEDYNE ANALYTICAL INSTRUMENTS The analog output printed circuit board (PCB) generates the two 0–1 volt and the two 4–20 mA analog signal outputs available on the rear panel of the analyzer....
2-8 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory 2 Operational Theory TELEDYNE ANALYTICAL INSTRUMENTS The digital printed circuit board (PCB) is a general purpose microcomputer used to control all functions of the analyzer. The analog input PCB and the analog o...
Installation 3 Installation 3 Installation 3 Installation 3 Installation 3 3-1 TELEDYNE ANALYTICAL INSTRUMENTS Installation Installation of the analyzer includes: 1. Unpacking the system. 2. Recognizing the necessary precautions when installing the system. 3. Hooking up the sample/span gas and air s...
3-2 3 Installation 3 Installation 3 Installation 3 Installation 3 Installation TELEDYNE ANALYTICAL INSTRUMENTS 3. Parts Per Million (PPM) and Percent Oxygen indicators: the PPM light will light while the instrument is measuring O 2 in units of PPM. The percent light will light while the instrument i...
Installation 3 Installation 3 Installation 3 Installation 3 Installation 3 3-3 TELEDYNE ANALYTICAL INSTRUMENTS Figure 3–1: Analyzer front and rear panels. LED DISPLAY PPM INDICATOR PERCENT INDICATOR LIQUID CRYSTAL DISPLAY (LCD) POWER SWITCH COMPUTER MODULE KEYPAD FLOW SET KNOB FLOWMETER SENSOR ACCES...
3-4 3 Installation 3 Installation 3 Installation 3 Installation 3 Installation TELEDYNE ANALYTICAL INSTRUMENTS 3.4 Gas Line Connections All of the gas lines to the system hook up at the back of the unit (see Figure 3–1). All of the fittings on the removable back panel are ¼ " male VCR type fitti...
3-6 3 Installation 3 Installation 3 Installation 3 Installation 3 Installation TELEDYNE ANALYTICAL INSTRUMENTS 1. The sensor compartment is located in the lower right-hand corner of the analyzer front panel. To remove the sensor compartment panel, loosen thethumbscrews located left and right center ...
Installation 3 Installation 3 Installation 3 Installation 3 Installation 3 3-7 TELEDYNE ANALYTICAL INSTRUMENTS Figure 3-3: 3160 Rear Panel. 3.6.1 Voltage Selection WARNING: Power cord must be disconnected before performing any voltageselection! The voltage setting and fuses of the analyzer can be ch...
3-8 3 Installation 3 Installation 3 Installation 3 Installation 3 Installation TELEDYNE ANALYTICAL INSTRUMENTS Figure 3-4: Removing the Voltage Card 4. Turn the card so that the desired voltage can be read at the bottom. 5. Slide the indicator pin around the card so that it is pointing up when the v...
Installation 3 Installation 3 Installation 3 Installation 3 Installation 3 3-9 TELEDYNE ANALYTICAL INSTRUMENTS WARNING: Power cord must be disconnected before performing anyvoltage selection! To change from North American to European fuses, perform the following: 1. Use a small blade screwdriver to ...
Operations 4 Operations 4 Operations 4 Operations 4 Operations 4 4-1 TELEDYNE ANALYTICAL INSTRUMENTS Operations 4.1 Front Panel Controls The front panel of the analyzer, shown in Figure 4-1, contains indicators and displays through which the computer module can be accessed. The upper left-hand side ...
4-2 4 Operations 4 Operations 4 Operations 4 Operations 4 Operations TELEDYNE ANALYTICAL INSTRUMENTS displaying the function of each button directly above it. The LCD screen displays thecurrent mode and any warning messages, instructions and button functions. 4.1.1 Spinning Wheel A “spinning wheel” ...
Operations 4 Operations 4 Operations 4 Operations 4 Operations 4 4-3 TELEDYNE ANALYTICAL INSTRUMENTS When waiting for manual sensor settling during calibration zeroing, wait until the oxygen concentration has reached its lowest value and has remainedso for 10–15 minutes. It is important to make sure...
4-10 4 Operations 4 Operations 4 Operations 4 Operations 4 Operations TELEDYNE ANALYTICAL INSTRUMENTS 13. If the O 2 cell current is too strong or too weak to span, an error message will appear. This will not occur if theuser-entered span gas concentration is correct. PressACK to retry, or ESC to ab...
4-11 Operations 4 Operations 4 Operations 4 Operations 4 Operations 4 TELEDYNE ANALYTICAL INSTRUMENTS 1. Enter set-up from the Analyze mode main menu by pressing SETUP.If only one cell block is installed, the TEST and INSTLoptions on the next screen are omitted. 2. Press SET to change the active sen...
4-13 Operations 4 Operations 4 Operations 4 Operations 4 Operations 4 TELEDYNE ANALYTICAL INSTRUMENTS 7. After pressing ACK in step 6, the new sensor is flushed for 10 minutes while the measured O 2 level is displayed. The 10-minute flush may be restarted by pressing RE-SET, or aborted by pressing S...
4-16 4 Operations 4 Operations 4 Operations 4 Operations 4 Operations TELEDYNE ANALYTICAL INSTRUMENTS • LATCHING actuates the alarm above a certainsetpoint even if the level falls back below thesetpoint. • RELAY indicates the relay (on the rear panel)actuated by the alarm, which is changeable. • FAI...
4-18 4 Operations 4 Operations 4 Operations 4 Operations 4 Operations TELEDYNE ANALYTICAL INSTRUMENTS Note: Note: Note: Note: Note: The above screens use T to indicate two values: seconds and minutes. The first value is The above screens use T to indicate two values: seconds and minutes. The first v...
4-20 4 Operations 4 Operations 4 Operations 4 Operations 4 Operations TELEDYNE ANALYTICAL INSTRUMENTS 7. If SET was pressed in step 1, the menu for setting the time and date will appear. Use the SELEC, UP andDOWN keys to set the time and date, and then pressACCPT. 8. If ZERO was pressed in step 1, t...
4-21 Operations 4 Operations 4 Operations 4 Operations 4 Operations 4 TELEDYNE ANALYTICAL INSTRUMENTS 1. Enter set-up from the Analyze mode by pressing SETUP. 2. Press SET to set the password(s). 3. Enter your authorization code by using the SELEC, UP and DOWN keys, and press ACCPT. You must enterth...
4-24 4 Operations 4 Operations 4 Operations 4 Operations 4 Operations TELEDYNE ANALYTICAL INSTRUMENTS 7. Status page five shows the time and date of the last time the alarm was triggered for each of the five program-mable alarms. 8. Status page six shows the set-up for the serial port, and allows yo...
5-1 Maintenance & Maintenance & Maintenance & Maintenance & Maintenance & T T T T Trrrrroub oub oub oub oubleshooting 5 leshooting 5 leshooting 5 leshooting 5 leshooting 5 TELEDYNE ANALYTICAL INSTRUMENTS Maintenance & Troubleshooting 5.1 Routine Maintenance 1. Calibrate the a...
A-1 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS Appendix A.1 Specifications (Subject to change without notice.) Ranges: (Six ranges with AutoRanging on all ranges) 1: 0–1 part per million (ppm) O 2 2: 0–10 ppm O 2 3: 0–100 ppm O 2 4: 0–1000 ppm O 2 5: 0–1 % O 2 6: 0–...
A-4 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS A.4 Material Safety Data Sheet I Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19.Mini-Micro-Fuel Cells, all classes. Manufac...
A-5 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fireconditions. No specific agents recommended. Wear NIOSH/OSHA approved self-contained breath...
A-6 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS Contact of electrolyte with skin or eyes will cause aburning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss of sleep,loss of appetite, metallic taste and fatigue. Carci...
A-7 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: Apron, face shield Ventilation: na Section IX – Disposal Both lead and potassium hydroxid...
A-9 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS Section III – Physical Hazards Potential for fire and explosion: The electrolyte in the Micro–Fuel Cells is not flam-mable. There are no fire or explosion hazards associated with Micro–Fuel Cells. Potential for reactivi...
A-10 Appendix Appendix Appendix Appendix Appendix TELEDYNE ANALYTICAL INSTRUMENTS Section V – Emergency and First Aid Procedures Eye Contact: Flush eyes with water for at least 15 minutes and get immediatemedical attention. Skin Contact: Wash affected area with plenty of water and remove contaminate...
Teledyne Manuals
-
Teledyne 4220
Manual
-
Teledyne 311-D
Manual
-
Teledyne HPM 4/6
Manual
-
Teledyne T802
Manual
-
Teledyne 318R
Manual
-
Teledyne 9700
Manual
-
Teledyne BDS-3960
Manual
-
Teledyne C-2000
Manual
-
Teledyne GFC 7001EM
Manual
-
Teledyne 2103
Manual
-
Teledyne 6400E
Manual
-
Teledyne AX300-I
Manual
-
Teledyne 6600
Manual
-
Teledyne 9950
Manual
-
Teledyne PEM 9002
Manual
-
Teledyne HPM-2002-OBE
Manual
-
Teledyne Infrared Gas Analyzer
Manual
-
Teledyne 102E
Manual
-
Teledyne OT-3
Manual
-
Teledyne LLG
Manual