Page 3 - GETTING STARTED
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 1 GETTING STARTED This Quickstart Guide is designed to get you set up and operating your Teledyne Analytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss...
Page 4 - Analyze
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 2 Function Keys : Six touch-sensitive membrane switches are used to change the specific function performed by the analyzer: • Analyze Perform analysis for oxygen content of a sample gas. • System Perform system-related tasks (des...
Page 6 - SETUP AND INSTALL; Mount the Analyzer; The analyzer is equipped with 1/4; Connect a vent line to
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 SETUP AND INSTALL Mount the Analyzer The Model 3000TA -EU is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. It is designed for flush panel mounting. Use the four mounting holes—one i...
Page 7 - Electrical Connections; Connect primary input power:; Insert the power cord into the power cord receptacle.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 5 Electrical Connections CAUTION: READ ALL SAFETY INFORMATION REGARDING ELECTRICAL HAZARDS ASSOCIATIED WITH THIS INSTRUMENT BEFORE MAKING ELECTRICAL CONNECTIONS. CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS TH...
Page 8 - Install the Micro-fuel Cell; Cell or as an option, Teledyne’s Insta Trace cell.; CAUTION: DO NOT TOUCH THE SENSING SURFACE OF THE CELL. IT IS COVERED
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 • Install 50-Pin Equipment Interface Connector Connect the 50 pin connector to the mating connector on the rear panel. • Connect the 9-pin D connector to the RS-232 serial port. See Section 3.3.2.3 in the Instruction Manual for...
Page 10 - BASIC OPERATION; Default Configuration; Using the Data Entry and Function Buttons; Enter
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 BASIC OPERATION Once the analyzer has been installed, it can be configured for your application. Prior to using the instrument you must: • Calibrate the instrument • Define the analysis ranges and choose whether to use autorang...
Page 11 - Available Functions; The six function buttons are used to control the analyzer.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 9 Available Functions The six function buttons are used to control the analyzer. Some functions have submenus that allow additional choices as shown below.
Page 12 - Password Protection; Entering the Password
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 Any function can be selected at any time by pressing the appropriate button (unless password restrictions apply). Password Protection If a password is assigned, then setting the following system parameters can be done only aft...
Page 13 - Calibration of the Analyzer; Zero Calibration; AUTO
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 11 ◄► arrow keys to scroll back and forth between letters, and the ▲▼ arrow keys to change the letters to the proper password. • Press Enter to enter the password. Change Password? <ENT>=Yes <ESC>=No • Press Escape . Ca...
Page 14 - Span Calibration
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode. Manual Mode Zeroing • Press Zero to enter the Zero fu...
Page 15 - Span; Setting Alarm Parameters
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 13 #### ppm Span Slope=#### ppm/s The instrument automatically returns to the Analyze mode at the conclusion of the auto span cycle.. Manual Mode Spanning: • Press Span to start the Span function. Span: Settling:MAN <ENT> For...
Page 16 - Alarm; Set The; Function; Range
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 • Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1 is blinking. AL—1 AL—2 Choose Alarm • Using the ◄► arrow keys to move the blinking over to AL–1. Then press Enter . AL—1 1000 ppm HI ...
Page 17 - The Analyze Function; Escape
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 15 • Use the ◄► arrow keys to blink the range to be set: low (L), medium (M), or high (H). • Use the ▲▼ arrow keys to enter the upper value of the range (all ranges begin at 0 ppm). Repeat for each range you want to set. • Press En...