Page 3 - TELEDYNE; MANUAL ADDENDUM; CHEMILUMINESCENCE OZONE
TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company MANUAL ADDENDUM MODEL 265A CHEMILUMINESCENCE OZONE ANALYZER (ADDENDUM TO M200A INSTRUCTION MANUAL #02246) TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 921...
Page 5 - SAFETY MESSAGES
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 i PRINTED DOCUMENTS ARE UNCONTROLLED SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safe...
Page 6 - TABLE OF CONTENTS
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 ii PRINTED DOCUMENTS ARE UNCONTROLLED TABLE OF CONTENTS SAFETY MESSAGES ....................................................................................................... I TABLE OF CONTENTS ........................
Page 9 - GETTING STARTED
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 1 PRINTED DOCUMENTS ARE UNCONTROLLED 1 GETTING STARTED 1.0 Overview The Model 265A is a close derivative of the Model 200A Chemiluminescence NOx Analyzer. Because of this, the Model 200A Instruction manual continues t...
Page 10 - Electrical and Pneumatic Connections
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 2 PRINTED DOCUMENTS ARE UNCONTROLLED 1.2 Electrical and Pneumatic Connections 1. See Figure 1-2 and Figure 1-3 to locate the rear panel electrical and pneumatic connections. 2. Attach the pump to the “Exhaust Out” por...
Page 15 - Initial Operation
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 7 PRINTED DOCUMENTS ARE UNCONTROLLED 1.3 Initial Operation 1. Turn on the instrument power. The display should immediately light, showing the computer’s memory configuration, then the instrument type - M265A. A diagra...
Page 19 - SPECIFICATIONS, WARRANTY
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 11 PRINTED DOCUMENTS ARE UNCONTROLLED 2 SPECIFICATIONS, WARRANTY 2.1 Specifications Table 2.1: M265A Specifications Ranges In 1 ppb increments from 100 ppb to 20,000 ppb Measurement Mode Single range or AutoRange Meas...
Page 21 - Test and Calibration Values
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 13 PRINTED DOCUMENTS ARE UNCONTROLLED 2.3 Test and Calibration Values Table 2-2: Final Test and Calibration Values TEST Values Observed Value Units Nominal Range Reference Section RANGE ppb 10-20000 NOISE ppb 0.0 – 7....
Page 23 - THE M265A OZONE ANALYZER; Principle of Operation
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 15 PRINTED DOCUMENTS ARE UNCONTROLLED 3 THE M265A OZONE ANALYZER 3.1 Principle of Operation The Teledyne API Model 265A analyzer is designed to measure the concentration of ozone using the chemiluminescence reaction b...
Page 27 - SOFTWARE FEATURES
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 19 PRINTED DOCUMENTS ARE UNCONTROLLED 4 SOFTWARE FEATURES The M265A software has two major operating modes. The SAMPLE mode is the normal mode when the instrument is taking data. The software menu that covers the SAMP...
Page 31 - MAINTENANCE; Maintenance Schedule
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 23 PRINTED DOCUMENTS ARE UNCONTROLLED 5 MAINTENANCE 5.1 Maintenance Schedule NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. Table 5-1: Preventative Maintenanc...