Page 2 - Hot Surface; Safety Guidelines for Hazardous Voltages; or
NI USB-9229/9239 User Guide and Specifications 2 ni.com Dimensions Figure 2 shows the NI USB-9229/9239 device dimensions. Figure 2. NI USB-9229/9239 Device in Millimeters (Inches) Safety Guidelines Operate the NI USB-9229/9239 only as described in these operating instructions. Hot Surface This icon ...
Page 3 - Caution; Related Documentation; DAQ Getting Started Guide; Start»All
© National Instruments Corporation 3 NI USB-9229/9239 User Guide and Specifications Caution Make sure that devices and circuits connected to the module are properly insulated from human contact. Caution When module terminals are hazardous voltage LIVE (>42.4V pk /60 VDC), you must ensure that dev...
Page 4 - LabVIEW
NI USB-9229/9239 User Guide and Specifications 4 ni.com The NI-DAQmx Help contains general information about measurement concepts, key NI-DAQmx concepts, and common applications that are applicable to all programming environments. Select Start»All Programs»National Instruments»NI-DAQ»NI-DAQmx Help ....
Page 5 - Measurement Studio; DAQ; ANSI C without NI Application Software; NI-DAQmx C Reference Help; .NET Languages without NI Application Software
© National Instruments Corporation 5 NI USB-9229/9239 User Guide and Specifications Measurement Studio If you program your NI-DAQmx-supported device in Measurement Studio using Visual C++, Visual C#, or Visual Basic .NET, you can interactively create channels and tasks by launching the DAQ Assistant...
Page 6 - Device Documentation and Specifications; Note; Training Courses; Installing the Software
NI USB-9229/9239 User Guide and Specifications 6 ni.com The installed documentation contains the NI-DAQmx API overview, measurement tasks and concepts, and function reference. This help is fully integrated into the Visual Studio .NET documentation. To view the NI-DAQmx .NET documentation, go to Star...
Page 7 - Installing Other Software; for USB Devices Getting Started Guide; Start»All Programs»National Instruments»
© National Instruments Corporation 7 NI USB-9229/9239 User Guide and Specifications Installing Other Software If you are using other software, refer to the installation instructions that accompany your software. Example Programs The NI-DAQmx CD contains example programs that you can use to get start...
Page 8 - for more information.
NI USB-9229/9239 User Guide and Specifications 8 ni.com 5. Press firmly on the connector side of the NI 9229/9239 module until the latches lock the module into place, as shown in Figure 5. Figure 5. Locking Module into Place Mounting the NI USB-9229/9239 to a Panel Threaded inserts are located in th...
Page 9 - LED Indicator; LED State; Specifications; Accessory
© National Instruments Corporation 9 NI USB-9229/9239 User Guide and Specifications LED Indicator The NI USB-9229/9239 device has a green LED next to the USB connector. The LED indicator indicates device status, as listed in Table 1. When the device is connected to a USB port, the LED blinks steadil...
Page 10 - Wiring for High-Vibration Applications; Module
NI USB-9229/9239 User Guide and Specifications 10 ni.com source to the AI– terminal. Refer to Table 3 for the terminal assignments for each channel. Wiring for High-Vibration Applications If an application is subject to high vibration, National Instruments recommends that you either use ferrules to ...
Page 13 - Passband; Stopband
© National Instruments Corporation 13 NI USB-9229/9239 User Guide and Specifications Passband The signals within the passband have frequency-dependent gain or attenuation. The small amount of variation in gain with respect to frequency is called the passband flatness. The digital filters of the NI U...
Page 14 - Input Characteristics; Operating voltage ranges
NI USB-9229/9239 User Guide and Specifications 14 ni.com Specifications The following specifications are typical for the range 0 to 60 °Cunless otherwise noted. All voltages are relative to the AI– signal on each channel unless otherwise noted. Input Characteristics Number of channels .................
Page 15 - Input noise; Stability
© National Instruments Corporation 15 NI USB-9229/9239 User Guide and Specifications Accuracy, NI USB-9229 Accuracy, NI USB-9239 Input noise NI USB-9229 .................................. 320 μ V rms NI USB-9239 .................................. 70 μ V rms Stability Gain drift ........................
Page 16 - Input delay; Power Requirements
NI USB-9229/9239 User Guide and Specifications 16 ni.com NI USB-9229 ...................................0.045°/kHz + 360°· f in /M clk 1 NI USB-9239 ...................................0.075°/kHz + 360°· f in /M clk 1 Phase nonlinearity ( f s = 50 kS/s).............0.11° max Input delay NI USB-9229 ....
Page 17 - Physical Characteristics; not; Safety Standards
© National Instruments Corporation 17 NI USB-9229/9239 User Guide and Specifications Physical Characteristics Screw-terminal wiring............................ 16 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped from the end Torque for screw terminals .................... ...
Page 18 - link in the Certification column.; Hazardous Locations; noncondensing; Electromagnetic Compatibility; EN 61326 EMC requirements Industrial Immunity
NI USB-9229/9239 User Guide and Specifications 18 ni.com Note For UL and other safety certifications, refer to the product label or visit ni.com/ certification , search for the module number or product line, and click the appropriate link in the Certification column. Hazardous Locations The NI USB-9...
Page 19 - CE Compliance; Environmental Management; NI and the; Waste Electrical and Electronic Equipment (WEEE); EU Customers; Calibration; ⬉ᄤֵᙃѻક∵ᵻࠊᄣ⧚ࡲ⊩ ˄Ё; RoHS
© National Instruments Corporation 19 NI USB-9229/9239 User Guide and Specifications CE Compliance This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows: • 2006/95/EC; Low-Voltage Directive (safety) • 2004/108/EEC; Electromagnetic Comp...
Page 20 - Where to Go for Support; you have access to everything from
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their resp...