Page 2 - Vertical; Specification; Impedance and Coupling
NI PXI/PCI-5114 Specifications 2 ni.com Power .......................................................................................................14Software ...................................................................................................15Environment ..............................
Page 3 - Voltage Levels; Accuracy
© National Instruments Corporation 3 NI PXI/PCI-5114 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable Vertical Offset 50 Ω 1 M Ω — Range (V pk-pk ) Vertical Offset Range (V) Range (V pk-pk ) Vertical Offset Range (V) 0.04 ±0.8 0.04 ±0.8 0.1 ±0.8 0.1 ±0.8 0.2 ±0.8 0.2 ±0.8 0...
Page 4 - Bandwidth and Transient Response
NI PXI/PCI-5114 Specifications 4 ni.com Figure 1. NI 5114 Frequency Response (Typical) Bandwidth and Transient Response Bandwidth (–3 dB) Range (V pk-pk ) Bandwidth Rise/Fall Time, Typical — All ranges except 0.04 125 MHz 2.8 ns 0.04 100 MHz 3.5 ns Bandwidth Limit Filter 20 MHz Noise Filter — AC Cou...
Page 6 - Horizontal; Sample Clock
NI PXI/PCI-5114 Specifications 6 ni.com Horizontal Sample Clock Specification Value Comments Sources Internal, Onboard Clock (internal VCXO)* External, CLK IN (front panel SMB connector) * Internal Sample Clock is locked to the Reference Clock or derived from the onboard VCXO. Onboard Clock (Interna...
Page 8 - Trigger
NI PXI/PCI-5114 Specifications 8 ni.com CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) Trigger Reference (Stop) Trigger Specification Value Comments Input Voltage Range Sine wave: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm to 13 dBm) Square wave: 0.2 V pk-pk to 2.8 V pk-pk — Maximum ...
Page 12 - TClk Specifications
NI PXI/PCI-5114 Specifications 12 ni.com TClk Specifications National Instruments TClk synchronization method and the NI-TClk driver are used to align the sample clocks on any number of SMC-based modules in a chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronizat...
Page 13 - Waveform Specifications
© National Instruments Corporation 13 NI PXI/PCI-5114 Specifications Waveform Specifications Specification Value Comments Onboard Memory Size 8 MB per Channel Standard 8 megasamples per channel — 64 MB per Channel Option 64 megasamples per channel 256 MB per Channel Option 256 megasamples per channe...
Page 15 - Software
© National Instruments Corporation 15 NI PXI/PCI-5114 Specifications Software Specification Value Comments Driver Software NI PXI-5114 : NI-SCOPE 2.9 or later. NI PCI-5114 : NI-SCOPE 3.1 or later. NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the NI 5114. N...
Page 16 - Environment
NI PXI/PCI-5114 Specifications 16 ni.com Environment NI PXI-5114 Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only. Specification Value Comment...
Page 18 - EU Customers
NI PXI/PCI-5114 Specifications 18 ni.com Safety, Electromagnetic Compatibility, and CE Compliance Specification Value Comments Safety The NI 5114 meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010...
Page 19 - Physical
© National Instruments Corporation 19 NI PXI/PCI-5114 Specifications Physical Specification Value Comments Dimensions NI PXI-5114 NI PCI-5114 — 3U, One slot, PXI/cPCI Module 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.) 35.5 × 2.0 × 11.3 cm (13.4 × 0.8 × 4.4 in.) Weight 455 g (16 oz) 421 g (14.8 oz) — ...
Page 20 - Where to Go for Support
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their resp...