Page 4 - Frequency and Transient Response; Specification
NI 5412 Specifications 4 ni.com Frequency and Transient Response Bandwidth 20 MHz –3 dB Digital Interpolation Filter Software-selectable Finite Impulse Response (FIR) filter. Available interpolation factors are 2, 4, or 8. — Passband Flatness Low-Gain and High-Gain Amplifiers Path — ±1.0 dB from DC ...
Page 7 - Sample Clock
© National Instruments Corporation 7 NI 5412 Specifications Sample Clock Spectral Characteristics (Continued) Average Noise Density Path Amplitude Range Average Noise Density Average Noise Density at small amplitudes is limited by a –148 dBm/Hz noise floor. V pk-pk dBm dBm/Hz dBFS/ Hz Low-Gain 2 10 ...
Page 9 - System Phase Noise and Jitter (10 MHz Carrier)
© National Instruments Corporation 9 NI 5412 Specifications System Phase Noise and Jitter (10 MHz Carrier) Sample Clock Source System Phase Noise Density (dBc/Hz) Offset System Output Jitter (Integrated from 100 Hz to 100 kHz) 1. High- Resolution specifications vary with Sample Rate. 2. All values a...
Page 11 - CLK IN; (Sample Clock and Reference Clock Input, Front Panel Connector)
© National Instruments Corporation 11 NI 5412 Specifications CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) Table 5. Specification Value Comments Connector SMB (jack) — Direction Input — Destinations 1. Sample Clock 2. PLL Reference Clock — Frequency Range 1 MHz to 105 MHz (S...
Page 12 - PFI 0 and PFI 1
NI 5412 Specifications 12 ni.com PFI 0 and PFI 1 (Programmable Function Interface, Front Panel Connectors) Table 6. Specification Value Comments Connectors Two SMB (jack) — Direction Bi-directional — Frequency Range DC to 105 MHz — As an Input (Trigger) Destinations Start Trigger — Maximum Input Ove...
Page 13 - Start Trigger
© National Instruments Corporation 13 NI 5412 Specifications Start Trigger V OH Minimum: 2.9 V (open load), 1.4 V (50 Ω load) Output drivers are +3.3 V TTL compatible. Measured with a 1 m cable. V OL Maximum: 0.2 V (open load), 0.2 V (50 Ω load) Rise/Fall Time(20% to 80%) ≤ 2.0 ns Load of 10 pF. Tab...
Page 15 - Markers
© National Instruments Corporation 15 NI 5412 Specifications Markers Table 8. Specification Value Comments Destinations 1. PFI<0..1> (SMB front panel connectors) 2. NI PXI-5412 —PXI_Trig<0..6> (backplane connector) NI PCI-5412 —RTSI<0..6> — Quantity One Marker per Segment. — Quantu...
Page 16 - Waveform and Instruction Memory Utilization
NI 5412 Specifications 16 ni.com Waveform and Instruction Memory Utilization Table 9. Specification Value Comments Memory Usage The NI 5412 uses the Synchronization and Memory Core (SMC) technology in which waveforms and instructions share onboard memory. Parameters, such as number of segments in se...
Page 17 - Memory Limits
© National Instruments Corporation 17 NI 5412 Specifications Loop Count 1 to 16,777,215.Burst trigger: Unlimited — Quantum Waveform size must be an integer multiple of four samples. — Memory Limits 8 MB Standard 32 MB Option 256 MB Option All trigger modes except where noted. Arbitrary Waveform Mode...
Page 19 - Software
© National Instruments Corporation 19 NI 5412 Specifications Software Table 12. Specification Value Comments Driver Software NI-FGEN 2.3 or later version. NI-FGEN is an IVI-compliant driver that allows you to configure, control, and calibrate the NI 5412. NI-FGEN provides application programming int...
Page 20 - Environment; Specifications
NI 5412 Specifications 20 ni.com Environment NI PXI-5412 Environment Note To ensure that the NI PXI-5412 cools effectively, follow the guidelines in the Maintain Forced-Air Cooling Note to Users included in the NI 5412 kit. The NI PXI-5412 is intended for indoor use only. Table 13. Specifications Va...
Page 23 - Notes
© National Instruments Corporation 23 NI 5412 Specifications EMC/EMI CE, C-Tick, and FCC Part 15 (Class A) Compliant Notes : 1. This device is not intended for, and is restricted from, use in residential areas. 2. For EMC compliance, operate this device with shielded cabling. 3. When connected to ot...
Page 24 - Physical
NI 5412 Specifications 24 ni.com Physical Table 16. Specification Value Comments Dimensions NI PXI-5412 NI PCI-5412 — Single 3U PXI slot. CompactPCI compatible. 2.0 × 13.0 × 21.6 cm (0.8 × 5.1 × 8.5 inches) 34.07 × 10.67 × 2.03 cm (13.4 × 4.20 × 0.8 inches) Weight 340 g (11 oz) 480 g (17 oz) — Front...
Page 25 - Where to Go for Support
© National Instruments Corporation 25 NI 5412 Specifications Where to Go for Support The National Instruments Web site is your complete resource for technical support. At ni.com/support you have access to everything from troubleshooting and application development self-help resources to email and ph...