Teledyne 3000PA - Manual

Teledyne 3000PA

Teledyne 3000PA – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
14 Page 14
15 Page 15
16 Page 16
17 Page 17
18 Page 18
19 Page 19
20 Page 20
21 Page 21
22 Page 22
23 Page 23
24 Page 24
25 Page 25
26 Page 26
27 Page 27
28 Page 28
29 Page 29
30 Page 30
31 Page 31
32 Page 32
33 Page 33
34 Page 34
35 Page 35
36 Page 36
37 Page 37
38 Page 38
39 Page 39
40 Page 40
41 Page 41
42 Page 42
43 Page 43
44 Page 44
45 Page 45
46 Page 46
47 Page 47
48 Page 48
49 Page 49
50 Page 50
51 Page 51
52 Page 52
53 Page 53
54 Page 54
55 Page 55
56 Page 56
57 Page 57
58 Page 58
59 Page 59
60 Page 60
61 Page 61
62 Page 62
63 Page 63
64 Page 64
65 Page 65
66 Page 66
67 Page 67
68 Page 68
69 Page 69
Page: / 69

Table of Contents:

  • Page 2 – ii; Teledyne Analytical Instruments; Copyright © 1999 Teledyne Analytical Instruments
  • Page 3 – iii; Percent Oxygen Analyzer; Enter Class Designation.
  • Page 4 – iv
  • Page 6 – COMBUSTIBLE GAS USAGE WARNING; DANGER
  • Page 7 – Introduction; Overview
  • Page 8 – Versatile analysis over a wide range of applications.; Model Designations
  • Page 10 – Analyze; CAUTION: The power cable must be unplugged to fully
  • Page 11 – LCD has GREEN background with BLACK characters. VFD has; Power Connection; Universal AC power source.; Gas Inlet and Outlet
  • Page 12 – Note: If you require highly accurate Auto-Cal timing, use external
  • Page 13 – Operational Theory; The analyzer is composed of three subsystems:; Micro-Fuel Cell Sensor; The active components of the Micro-Fuel Cell are a cathode, an anode,
  • Page 14 – The Micro-Fuel Cell is a cylinder only 11⁄4 inches in diameter and 1
  • Page 15 – Pb
  • Page 16 – cell at the time and (2) the amount of stored anode material.
  • Page 17 – Expected life in flue gas is 8 months.
  • Page 18 – Sample System
  • Page 19 – Figure 2-4: Piping Layout and Flow Diagram for Standard Model
  • Page 20 – Electronics and Signal Processing; The Model 3000P Percent Oxygen Analyzer uses an 8031 microcon-; Figure 2-6: Location of Electronic Components
  • Page 21 – Figure 2-7: Block Diagram of the Model 3000P Electronics
  • Page 22 – In the presence of oxygen the cell generates a current. A current to
  • Page 23 – Installation; Installation of the Model 3000PA Analyzer includes:; Unpacking the Analyzer; The Model 3000PA is for indoor use in a general purpose area. It is
  • Page 24 – Figure 3-1: Front Panel of the Model 3000PA; Rear Panel Connections; Figure 3-3 shows the Model 3000PA rear panel. It contains all of the
  • Page 25 – Figure 3-3: Rear Panel of the Model 3000PA; Remove the blue sticker from the restrictor before using.
  • Page 26 – SAMPLE IN; Note: If the unit is for vacuum service, the above numbers apply; EXHAUST OUT: Exhaust connections must be consistent with the; Note: If the unit is for vacuum service, see; ZERO IN and SPAN IN (Optional): These are additional input ports
  • Page 27 – The universal power supply requires a 85–250 V ac, 47-63 Hz power; Fuse Installation: The fuse block, at the right of the power cord
  • Page 28 – System
  • Page 29 – Zero; Remote Calibration Protocol: To properly time the Digital Remote
  • Page 30 – Raw Cell Output: 99
  • Page 31 – Figure 3-6: Remote Probe Connector Pinouts
  • Page 32 – Installing the Micro-Fuel Cell; The Micro-Fuel Cell is not installed in the cell block when the; Testing the System; Before plugging the instrument into the power source:
  • Page 33 – Operation; Once the analyzer has been installed, it can be configured for your
  • Page 34 – Enter
  • Page 35 – Figure 4-1: Hierarchy of Functions and Subfunctions; Monospaced; The; : Used to define an automatic calibration sequence; PSWD
  • Page 36 – : Logging out prevents an unauthorized tampering with; More; : The instrument performs a self-diagnostic test to; If you cannot read anything on the display after first powering up:
  • Page 37 – won’t be able to set; Auto—Cal
  • Page 38 – NOTE: If you use password security, it is advisable to keep a copy of; Entering the Password
  • Page 39 – Characters Available for Password Definition:
  • Page 40 – Note: If you log off the system using the logout function in the
  • Page 42 – Shut off the gas pressure before connecting it to the analyzer, and; Cell Failure; Version Self—Test
  • Page 47 – Range; Note: The ranges must be increasing from low to high. For example,
  • Page 49 – NOTE: When performing analysis on a fixed range, if the oxygen; Voltage Signal
  • Page 50 – The analog output signal has a voltage which depends on the oxygen; L O
  • Page 51 – Maintenance; Routine Maintenance; To have a replacement cell available when it is needed, TAI recom-
  • Page 52 – V e r s i o n S e l f — T e s t; Check Specific Model Information in the Front Matter in this
  • Page 55 – NOTE: Evidence of damage due to tampering or mishandling will
  • Page 56 – Fuse Replacement; Place small screwdriver in notch, and pry cover off, as shown in; Figure 5-2: Removing Fuse Block from Housing; To change between American and European fuses, remove the; System Self Diagnostic Test
  • Page 57 – Major Internal Components; WARNINGS: See warnings on the title page of this manual.
  • Page 58 – Cleaning
  • Page 59 – Troubleshooting; The analyzer may have been calibrated in an inaccurate fashion.
  • Page 61 – Appendix; Packaging: General Purpose
  • Page 64 – Final Assembly/Outline Drawing
  • Page 65 – A usable flow rate for a 3000 series analyzer is one which can be; CELL PRESSURE CONCERNS:
  • Page 66 – Example 2, A 3000PA is configured for vacuum service as follows. The
  • Page 67 – BLUE DOT
  • Page 68 – Zero Cal; arrow keys to toggle between; AUTO; Manual Mode Zeroing; keys
  • Page 69 – Man; Slope
Loading the manual

i

Percent Oxygen Analyzer

Teledyne Analytical Instruments

OPERATING INSTRUCTIONS FOR

Model 3000

PA

Percent Oxygen Analyzer

P/N M64573

08/06/99

ECO:#99-0323

H I G H L Y T O X I C A N D O R F L A M M A B L E L I Q U I D S O R G A S E S M A Y B E P R E S E N T I N T H I S
MONITORING SYSTEM.

PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.

HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PER-
SIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.

ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING.
BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED
S U P E R V I S O R / M A N A G E R .

DANGER

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - ii; Teledyne Analytical Instruments; Copyright © 1999 Teledyne Analytical Instruments

ii Model 3000PA Teledyne Analytical Instruments Copyright © 1999 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, tran- scribed, stored in a retrieval system, or translated into any other language or computerlanguage in whole or in part, in ...

Page 3 - iii; Percent Oxygen Analyzer; Enter Class Designation.

iii Percent Oxygen Analyzer Teledyne Analytical Instruments Specific Model Information The instrument for which this manual was supplied may incorporate one or more options not included with the standard instrument. Commonly availableoptions are listed below, with check boxes. Any that are incorpora...

Page 4 - iv

iv Model 3000PA Teledyne Analytical Instruments Table of Contents 1 Introduction 1.1 Overview ........................................................................ 1-11.2 Typical Applications ....................................................... 1-11.3 Main Features of the Analyzer ...............

Other Teledyne Models

All Teledyne Other