Page 3 - SAFETY MESSAGES; CAUTION
07303C DCN6646 i SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol, and placed throughout this manual and i...
Page 4 - ii; CONSIGNES DE SÉCURITÉ
ii 07303C DCN6646 CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est représentée par un pic...
Page 5 - iii; WARRANTY; CAUTION – Avoid Warranty Invalidation; Primer on Electro-Static Discharge
07303C DCN6646 iii WARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides that: Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment failure occur, TAPI assures its customers that...
Page 6 - iv
Model 200EU-NOy Option Manual Addendum iv 07303C DCN6646 This page intentionally left blank.
Page 7 - TABLE OF CONTENTS
Model 200EU-NOy Option Manual Addendum 07303C DCN6646 v TABLE OF CONTENTS 1.0 INTRODUCTION ...................................................................................................... 7 2.0 SPECIFICATIONS ........................................................................................
Page 8 - vi; List of Figures
Model 200EU-NOy Option Manual Addendum vi 07303C DCN6646 List of Figures Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Converter (bottom) ...................... 13 Figure 2. Rear Panel Pneumatic Connections ...........................................................................
Page 13 - GETTING STARTED; Unpacking and Inspection
07303C DCN6646 11 3.0 GETTING STARTED The NOy converter has been designed to operate with the T200U low level NOx analyzer, with some additional modification.The combination is referred to as the T200U-NOy analyzer. 3.1 Unpacking and Inspection The T200U- NOy comes in three boxes: The analyzer wit...
Page 14 - Electrical and Pneumatic Connections; NOTE; Figure 2
Model T200U-NOy Option Manual Addendum 12 07303C DCN6646 3.2 Electrical and Pneumatic Connections NOTE To maintain compliance with EMC standards, it is required that the cable length (at the rear of the analyzer) be no greater than 3 meters for all I/O connections, which include Analog In, Analog Ou...
Page 16 - Figure 2. Rear Panel Pneumatic Connections; between instruments.
Model T200U-NOy Option Manual Addendum 14 07303C DCN6646 Figure 2. Rear Panel Pneumatic Connections Note: Pneumatic tubes and rear panel labels include numbers to facilitate matching connections between instruments. CONNECTIONS: Analyzer to Pump to Converter and Calibrator to Pump Calibrator Analyze...
Page 17 - Initial Operation; C, indicating that the external Converter is at operating; Table 2 Final Test and Calibration Values
Model T200U-NOy Option Manual Addendum 07303C DCN6646 15 3.3 Initial Operation 1. After confirming proper supply voltage, turn on the power to the analyzer and the 501Y bypass pump, and plug in the external Pump. If you are unfamiliar with the T200U-NOy analyzer, we recommend that you read the overv...
Page 19 - Principle of Operation
07303C DCN6646 17 4.0 THE T200U-NOY EXTERNAL CONVERTER 4.1 Principle of Operation The T-API Model T200U-NOy is designed to measure the concentration of NO, NO 2 , and other compounds that are too unstable to be measured when taken in through the normal ambient air sample inlet system. Please refer t...
Page 21 - CALIBRATION AND CAL CHECKS; Do the zero calibration procedure
07303C DCN6646 19 5.0 CALIBRATION AND CAL CHECKS Unlike most NOx analyzers, the T200U-NOy does not have a sample inlet port on its rear panel. The sample inlet is located on the external Converter. Follow the steps in the T200U manual to: Enter the expected NO and NOy span gas concentrations in th...
Page 22 - Calibration or Cal Check Procedure; Table 3 Manual Zero Calibration Procedure
Model T200U-NOy Option Manual Addendum 20 07303C DCN6646 5.1 Calibration or Cal Check Procedure Calibration of the analyzer can be checked or adjusted using gas introduced through the CAL IN (3) port on the 501Y rear panel. The calibration gas is routed to a tee fitting (see CAL IN (6) in Figure 4 )...
Page 23 - Table 5 Span Calibration Procedure
Model T200U-NOy Option Manual Addendum 07303C DCN6646 21 Table 5 Span Calibration Procedure Step Action Comment 1. Press CAL The analyzer enters the calibrate mode. 2. Wait For Stability Type STB < 2 PPB Wait for reading to stabilize at span value. 3. Press SPAN If you change your mind after pres...
Page 25 - Maintenance Schedule; Qualified Maintenance Personnel Required
07303C DCN6646 23 6.0 MAINTENANCE 6.1 Maintenance Schedule The schedule for preventative maintenance is presented in Table 6 below. Qualified Maintenance Personnel Required Table 6 Preventative Maintenance Calendar Item Maintenance Interval Zero/Span Calibration Annually or after maintenance or repa...
Page 26 - Replacing the Sample Particulate Filters
Model T200U-NOy Option Manual Addendum 24 07303C DCN6646 6.2 Replacing the Sample Particulate Filters The particulate filter should be inspected often for signs of plugging or contamination. It is also common for dirt particles to cause instrument drift, and affect accuracy. To check and change the ...
Page 27 - Figure 3. Replacing the Particulate Filter
Model T200U-NOy Option Manual Addendum 07303C DCN6646 25 Figure 3. Replacing the Particulate Filter
Page 28 - Checking Analyzer Flow Rate; Replacing the Converter
Model T200U-NOy Option Manual Addendum 26 07303C DCN6646 6.3 Checking Analyzer Flow Rate The external Pump is capable of maintaining the reaction cell pressure at less than 5.0 In-Hg-A. If a higher pressure is noted, the Pump may need servicing. Check the Pump and pneumatic system for leaks or rebui...
Page 29 - Figure 4. NOy Converter Guts Assembly Location; Inspecting Pneumatic Lines
Model T200U-NOy Option Manual Addendum 07303C DCN6646 27 8. Turn the 501Y power back on. The insulation can emit a burnt odor for the first 24 hours; this is normal. Allow the Converter to burn-in for 24 hours, then re-calibrate the instrument. Figure 4. NOy Converter Guts Assembly Location 6.5 Insp...
Page 30 - Figure 5. Pneumatics
Model T200U-NOy Option Manual Addendum 28 07303C DCN6646 Figure 5. Pneumatics
Page 31 - Checking for Leaks; Figure 6. Location of O; it
Model T200U-NOy Option Manual Addendum 07303C DCN6646 29 6.6 Checking for Leaks If a leak checker is not available, it is possible to leak check the instrument using the sample pump plus a shut-off valve. 1. Turn off power to the analyzer and the 501Y, and unplug the external Pump. 2. Cap the sample...
Page 33 - Operation Verification – Diagnostic Techniques
07303C DCN6646 31 7.0 TROUBLESHOOTING, ADJUSTMENTS CAUTION! The operations outlined in this section must be performed by qualified maintenance personnel only! This section of the manual contains information on diagnosing and repairing instrument performance problems. It provides troubleshooting proc...
Page 34 - 501Y Bypass Pump Diagnostic Procedures; Setting the Converter Temperature; PV
Model T200U-NOy Option Manual Addendum 32 07303C DCN6646 7.1.3 501Y Bypass Pump Diagnostic Procedures The 501Y bypass pump is capable of maintaining a total bypass flow rate of ~2.2 lpm. If flow checks indicate a fall off in bypass flow, the pump may need servicing. Check the pump, inlet fittings, a...
Page 36 - PVOF; hold until
Model T200U-NOy Option Manual Addendum 34 07303C DCN6646 SECONDARY MENU (The fixed characteristics of the system). PRESS DISPLAY ACTION SEL hold until p-n1 SET TO 0 – REVERSE ACTION SEL hold until p-n2 SET TO 2 = TYPE J THERMOCOUPLE SEL p-dF DIGITAL FILTER SET TO 5 SEL P-SL LOWER LIMIT SET TO -40 (-...
Page 37 - Figure 7 Electrical Diagram
Model T200U-NOy Option Manual Addendum 07303C DCN6646 35 Figure 7 Electrical Diagram
Page 39 - T200U-NOY SPARE PARTS LIST; Note
07303C DCN6646 37 8.0 T200U-NOY SPARE PARTS LIST The following spare parts list supplements the spare parts list for the T200U analyzer. Note Use of replacement parts other than those supplied by API may result in non-compliance with European Standard EN 61010-1. Note Due to the dynamic nature of pa...