i A BOUT T ELEDYNE ADVANCED P OLLUTION I NSTRUMENTATION (TAPI) Teledyne Advanced Pollution Instrumentation, Inc. (TAPI) is a worldwide market leader in the design and manufacture of precision analytical instrumentation used for air quality monitoring, continuous emissions monitoring, and specialty p...
ii This page intentionally left blank. 07276B DCN6418
iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol, and are placed throughout this manual; the safety sy...
Teledyne API T803 CO2/O2 Analyzer Operation Manual iv CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de...
v WARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides that: Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment failure occur, TAPI assures its customers that prompt service a...
vii ABOUT THIS MANUAL This T803 Operation Manual, PN 07276, is comprised of multiple documents in PDF format, as listed below. Part No. Rev Name/Description 06759 B T803 Operation Manual, top level assy 06763 C Appendix A Menu Trees, software version A.2 (E-Series), 1.0.0 (T-Series) 07269 A Spare Pa...
Teledyne API T803 CO2/O2 Analyzer Operation Manual viii REVISION HISTORY T803 Operation and Maintenance Manual, PN07276 Date Rev DCN Description 2013 January 14 B 6418 Administrative updates and Specs corrections 2011 March 11 A 6006 Initial Release 07276B DCN6418
ix TABLE OF CONTENTS A BOUT T ELEDYNE ADVANCED P OLLUTION I NSTRUMENTATION (TAPI) ...................................................................... i Safety Messages ....................................................................................................................................
Teledyne API T803 CO2/O2 Analyzer Operation Manual Table of Contents xi 9.3. Assessing Calibration Quality ..................................................................................................................... 166 9.4. Calibration of the T803’s Electronic Subsystems .....................
Teledyne API T803 CO2/O2 Analyzer Operation Manual xii 12.3. Operation within the T803 Analyzer ......................................................................................................... 208 12.4. Pneumatic Operation .......................................................................
Teledyne API T803 CO2/O2 Analyzer Operation Manual Table of Contents xiii Figure 7-1: Default DAS Channel Setup ......................................................................................... 133 Figure 7-2: APICOM Remote Control Program Interface .............................................
Teledyne API T803 CO2/O2 Analyzer Operation Manual xiv Table 5-7: Test Channels Functions available on the T803’s Analog Output ................................. 96 Table 5-8: Concentration Alarm Default Settings .......................................................................... 100 Table 6-...
17 1. INTRODUCTION, FEATUERES, AND OPTIONS 1.1. T803 OVERVIEW The Model T803 Carbon Dioxide/Oxygen Analyzer (T803 CO2/O2 Analyzer) is a microprocessor-controlled analyzer that determines the concentration of molecular carbon dioxide (CO2) and oxygen (O2) in the sample gas drawn through the instrumen...
Introduction, Featueres, and Options Teledyne API T803 CO2/O2 Analyzer Operation Manual 18 HAZARD OXYGEN IS A STRONG OXIDIZER. This is a general purpose instrument designed for usage in nonhazardous areas. Ensure that all safety precautions related combustible gases are followed. Before working with...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Introduction, Featueres, and Options 19 Option Option Number Description/Notes Reference Carrying Strap/Handle Side-mounted strap for hand-carrying analyzer 29 Extends from “flat” position to accommodate hand for carrying. Recesses to 9mm (3/8”) dim...
21 2. SPECIFICATIONS, APPROVALS & COMPLIANCE 2.1. SPECIFICATIONS Table 2-1: T803 Specifications Parameter Description CO 2 Ranges Min: 0-1% Full scale Max: 0-20% Full scale (user selectable, dual ranges and auto-ranging supported) Zero Noise 1 <0.02% (RMS) Span Noise 1 < 0.1% of reading (R...
Specifications, Approvals & Compliance Teledyne API T803 CO2/O2 Analyzer Operation Manual 22 Parameter Description Standard I/O 1 Ethernet: 10/100Base-T 2 RS-232 (300 – 115,200 baud) 2 USB device ports 8 opto-isolated digital status outputs 6 opto-isolated digital control inputs 4 analog outputs...
23 3. GETTING STARTED This section addresses the procedures for unpacking the instrument and inspecting for damage, presents clearance specifications for proper ventilation, introduces the instrument layout, then presents the procedures for getting started: making electrical and pneumatic connection...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 25 3.1.1. VENTILATION CLEARANCE Whether the analyzer is set up on a bench or installed into an instrument rack, be sure to leave sufficient ventilation clearance. Table 3-1: Ventilation Clearance AREA MINIMUM REQUIRED CLEARANCE Back ...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 26 Figure 3-2: Display Screen and Touch Control The front panel liquid crystal display screen includes touch control. Upon analyzer start-up, the screen shows a splash screen and other initialization indicators before the main displa...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 27 Table 3-2: Display Screen and Touch Control Description Field Description/Function LEDs indicating the states of Sample, Calibration and Fault, as follows: Name Color State Definition SAMPLE Green Off On Blinking Unit is not opera...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 28 Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts 07276B DCN6418
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 29 3.2.2. REAR PANEL Figure 3-4: Rear Panel Layout 07276B DCN6418
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 31 3.2.3. INTERNAL CHASSIS LAYOUT Figure 3-5: Internal Layout 07276B DCN6418
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 32 3.3. CONNECTIONS AND SETUP This section presents the electrical (Section 3.3.1) and pneumatic (Section 0) connections for setup and preparing for instrument operation. 3.3.1. ELECTRICAL CONNECTIONS CAUTION E LECTRICAL S HOCK H AZA...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 33 CAUTION G ENERAL S AFETY H AZARD The T803 analyzer can be configured for both 100-130 V and 210-240 V at either 47 or 63 Hz. To avoid damage to your analyzer, make sure that the AC power voltage matches the voltage indicated on th...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 34 3.3.1.3. ANALOG OUTPUT CONNECTIONS The T803 is equipped with several analog output channels accessible through the ANALOG OUT connector on the rear panel of the instrument. The standard configuration for these outputs is VDC. An o...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 35 Figure 3-8: Current Loop Option Installed CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD VOLTAGE OUTPUTS CAUTION Servicing or handling of circuit components requires electrostatic discharge (ESD) protection, i.e. ESD grounding...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 36 a) Remove the set screw located at the top center of the rear panel b) Remove the screws fastening the top cover to the unit (two per side). c) Slide cover back. d) Lift the cover straight up. 4. Remove the screw holding the curre...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 37 STATUS 1 2 3 4 5 6 7 8 D + SY S T E M O K CONC VAL ID CAL M O DE / M E A S MOD E S P A N CAL / ZERO MODE CAL M O DE – RA NG E 2 CO 2 CAL Figure 3-9: Status Output Connector Table 3-6: Status Output Signals Rear Panel Label Status ...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 38 left). However, if full isolation is required, an external 5 VDC power supply should be used (Figure 3-10, right). CONTROL IN A B C D E F U + SPA N /Z ER O C A L CA L/ M E AS M O D E CONTROL IN A B C D E F U + - + 5 VDC Power Supp...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 39 3.3.1.7. CONCENTRATION ALARM RELAY (OPTION 61) The Teledyne API T-Series analyzers have an option for four (4) “dry contact” relays on the rear panel of the instrument. This relay option is different from and in addition to the “C...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 40 The software for this instrument is flexible enough to allow you to configure the alarms so that you can have 2 alarm levels for each gas. CO 2 Alarm 1 = 20 % CO 2 Alarm 2 = 100 % O 2 Alarm 1 = 20 % O 2 Alarm 2 = 100 % In this exa...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 41 I MPORTANT I MPACT ON R EADINGS OR D ATA Cables that appear to be compatible because of matching connectors may incorporate internal wiring that makes the link inoperable. Check cables acquired from sources other than Teledyne API...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 42 Figure 3-13: CPU Connector Pin-Outs for RS-232 Mode To assist in properly connecting the serial ports to either a computer or a modem, there are activity indicators just above the RS-232 port. Once a cable is connected between the...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 43 ATTENTION COULD DAMAGE INSTRUMENT AND VOID WARRANTY Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges too small to be felt by the human nervous system. Failure to use ESD protection when working with ele...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 44 Figure 3-14: Jumper and Cables for Multidrop Mode (Note: If you are adding an instrument to the end of a previously configured chain, remove the shunt between Pins 21 22 of JP2 on the Multidrop/LVDS PCA in the instrument that wa...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 45 Analyzer Analyzer Analyzer Last Analyzer Female DB9 Male DB9 RS-232 COM2 RS-232 COM2 RS-232 COM2 RS-232 COM2 Host RS-232 port Ensure jumper is installed between JP2 pins 21 22 in last instrument of multidrop chain. Figure 3-15: ...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 46 RS-485 C ONNECTION As delivered from the factory, COM2 is configured for RS-232 communications. This port can be reconfigured for operation as a non-isolated, half-duplex RS-485 port. Using COM2 for RS-485 communication will disab...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 47 CAUTION G ENERAL S AFETY H AZARD Rapid release of pure N 2 gas into an enclosed space can displace oxygen, and therefore represents an asphyxiation hazard. This may happen with few warning symptoms. Do not vent calibration gases i...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 48 Calib rated N 2 100 % C once ntratio n Calib rated g a s ( O 2 at 20 .95% or CO 2 at 16% Span C oncentration) VENT Figure 3-16: Pneumatic Connections, Using Bottled Span Gas 3.3.2.6. SAMPLE GAS SOURCE Attach a sample inlet line to...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 49 3.3.2.9. EXHAUST OUTLET Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line should be: PTEF tubing; minimum O.D ¼” A maximum of 10 meters long. Vented outside the T803 analyzer’s enclosure, pref...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 50 3.4. STARTUP, FUNCTIONAL CHECKS, AND INITIAL CALIBRATION I MPORTANT I MPACT ON R EADINGS OR D ATA The analyzer’s cover must be installed to ensure that the temperatures of the internal components are properly controlled. If you ar...
Getting Started Teledyne API T803 CO2/O2 Analyzer Operation Manual 52 To view and clear warning messages: Suppresses the warning messages Press CLR to clear the current message. If more than one warning is active, the next message will take its place. Once the last warning has been cleared, the RANG...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Getting Started 53 3.5.2. INITIAL CALIBRATION To perform the calibration you must have sources for zero air and span gas available for input into the SAMPLE port on the back of the analyzer. See Section 0 for instructions for connecting these gas so...
57 4. BASIC OPERATION The T803 analyzer is a computer-controlled analyzer with a dynamic menu interface that allows all major operations to be controlled from the front panel touchscreen through user-friendly menus (A complete set of menu trees is located in Appendix A of this manual) This section i...
Basic Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 58 Figure 4-1: Front Panel Display The mode field of the front panel display indicates to the user which operating mode the unit is currently running. In addition to SAMPLE and SETUP , other modes the analyzer can be operated in are:...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Basic Operation 59 A variety of TEST functions are available for viewing at the front panel whenever the analyzer is at the MAIN MENU . These functions provide information about the various functional parameters related to the analyzers operation an...
Basic Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 60 Table 4-2: Test Functions Defined PARAMETER DISPLAY TITLE UNITS DEFINITION Range RNG RN1 RN2 % The full scale limit at which the reporting range of the analyzer is currently set. THIS IS NOT the Physical Range of the instrument. S...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Basic Operation 61 4.4. SETUP MODE The SETUP mode is used to configure the analyzer’s hardware and software features, perform diagnostic procedures, gather information on the instrument’s performance and configure or access data from the internal da...
Basic Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 62 Table 6-5: Secondary Setup Mode Features and Functions MODE OR FEATURE MENU ITEM DESCRIPTION MANUAL SECTION External Communication Channel Configuration COM Used to set up and operate the analyzer’s various external I/O channels i...
63 5. SETUP MENU The SETUP menu is used to set instrument parameters for performing configuration, calibration, reporting and diagnostics operations. 5.1. SETUP CFG: CONFIGURATION INFORMATION Pressing the CFG button displays the instrument’s configuration information. This display lists the analyz...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 64 5.4. SETUP RNGE: ANALOG OUTPUT REPORTING RANGE CONFIGURATION 5.4.1. PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING RANGES Functionally, the T803 analyzers have one hardware PHYSICAL RANGE that is capable of determining concentrations ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 65 5.4.2. ANALOG OUTPUT RANGES FOR CO 2 AND O 2 CONCENTRATION The analyzer has several active analog output signals accessible through a connector on the rear panel (see Figure 3-4). O 2 concentration output ANALOG OUT A1 A2 A3 A4 + - + -...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 66 5.4.3. REPORTING RANGE MODES The T803 provides three analog output range modes to choose from. Single range ( SNGL ) mode sets a single maximum range for the analog output. If single range is selected all three outputs are slaved tog...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 67 5.4.3.1. RNGE MODE SNGL: CONFIGURING THE T803 ANALYZER FOR SINGLE RANGE MODE When the single range mode is selected ( SNGL ), all concentration outputs ( A1, A2 and A3 ) are slaved together and set to the same reporting range limit...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 70 5.4.3.3. RNGE MODE AUTO: CONFIGURING THE T803 ANALYZER FOR AUTO RANGE MODE In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges (low and high). The unit will switch from low ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 71 (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 5.4.4. SETUP RNGE DIL: USING THE OPTIONAL DILUTION RATIO FEATURE This feature is an optional software utility that is used to compensate for ...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 72 2. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1 part of sample gas): 3. Calibrate the analyzer. Make sure that the calibration span gas is either supplied through the same dilution system as the...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 73 Table 5-6: Password Levels PASSWORD LEVEL MENU ACCESS ALLOWED Null (000) Operation All functions of the main menu (top level, or Primary, menu). 101 Configuration/Maintenance Access to Primary and Secondary SETUP Menus when PASSWORD is...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 74 Example: If all passwords are enabled, the following touchscreen button sequence would be required to enter the SETUP menu: SYSTEM ENTER PASSWORD:101 1 0 1 ENTR EXIT SYSTEM ENTER PASSWORD:0 0 0 0 ENTR EXIT EXAMPLE: This password enable...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 75 5.6. SETUP CLK: SETTING THE T803 ANALYZER’S INTERNAL CLOCK 5.6.1. SETTING THE INTERNAL CLOCK’S TIME AND DAY The T803 has a time of day clock that supports the time of day TEST function, the time stamps for the DAS feature and most CO...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 79 To access and navigate the VARS menu, use the following button sequence. SETUP X.X 6) CLOCK_ADJUST=0 Sec/Day PREV NEXT JUMP EDIT ENTR EXIT SETUP X.X CLOCK_ADJUST=0 Sec/Day + 0 0 ENTR EXIT Enter sign and number of seconds per day the cl...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 80 5.9. SETUP MORE DIAG: USING THE DIAGNOSTICS FUNCTIONS A series of diagnostic tools is grouped together under the SETUP MORE DIAG menu. These tools can be used in a variety of troubleshooting and diagnostic procedures and are re...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 81 5.9.1. ACCESSING THE DIAGNOSTIC FEATURES To access the DIAG functions press the following buttons: <TST TST> CAL MSG SETUP X.X CFG DAS ACAL RNGE PASS CLK EXIT SETUP X.X COMM VARS EXIT SETUP X.X EXIT PREV returns to the Activates ...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 82 5.10. USING THE T803 ANALYZER’S ANALOG OUTPUTS The T803 analyzer comes equipped with four analog outputs. The first two analog output ( A1 & A2) signals represent the currently measured CO 2 concentration (see Section 5.4.2). T...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 83 To access the ANALOG I/O CONFIGURATION sub menu, press: DIAG AIO AIN CALIBRATED: YES <SET CAL EXIT DIAG AIO TEST_OUTPUT: 5V,OVR, CAL <SET SET> EDIT EXIT DIAG AIO CONC_OUT_3: 5V, OVR, CAL <SET SET> EDIT EXIT SAMPLE CO2 RN...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 84 5.10.2. ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION In its standard configuration the analog outputs are set to output a 0 – 5 VDC signals. Several other output ranges are available. Each range is usable from -5% to + 5% of the rat...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 85 To change the output type and range, select the ANALOG I/O CONFIGURATION submenu (see Section 5.10.1) then press, 5.10.3. CALIBRATION OF THE ANALOG OUTPUTS Analog output calibration should to be carried out on first startup of the anal...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 86 5.10.3.1. ENABLING OR DISABLING THE AUTOCAL FOR AN INDIVIDUAL ANALOG OUTPUT To enable or disable the AutoCal feature for an individual analog output, elect the ANALOG I/O CONFIGURATION submenu (see Section 5.10.1) then press: 07276B DC...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 87 5.10.3.2. AUTOMATIC CALIBRATION OF THE ANALOG OUTPUTS I MPORTANT I MPACT ON R EADINGS OR D ATA Manual calibration should be used for any analog output set for a 0.1V output range or in cases where the outputs must be closely matched to...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 88 5.10.3.3. INDIVIDUAL (SINGLE-CHANNEL) CALIBRATION OF THE ANALOG OUTPUTS To use the AUTO CAL feature to initiate an automatic calibration for an individual analog output, select the ANALOG I/O CONFIGURATION submenu (see Section 5.10.1) ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 89 5.10.3.4. MANUAL CALIBRATION OF THE ANALOG OUTPUTS CONFIGURED FOR VOLTAGE RANGES For highest accuracy, the voltages of the analog outputs can be manually calibrated. Note The menu for manually adjusting the analog output signal level w...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 91 5.10.3.5. MANUAL ADJUSTMENT OF CURRENT LOOP OPTION OUTPUT SPAN AND OFFSET A current loop option may be purchased for the A1 , A2 and A3 analog outputs of the analyzer. This option places circuitry in series with the output of the A-to-...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 92 To adjust the zero and span signal levels of the current outputs, select the ANALOG I/O CONFIGURATION submenu (see Section 5.10.1) then press: An alternative method for measuring the output of the Current Loop converter is to connect a...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 93 current meter (see Figure 3-7 for pin assignments and diagram of the analog output connector). This allows the use of a voltmeter connected across the resistor to measure converter output as VDC or mVDC. V +DC Gnd Figure 5-4: Alternati...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 94 5.10.4. TURNING AN ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF In its default configuration, a ± 5% over-range is available on each of the T803’s analog outputs. This over-range can be disabled if your recording device is sensitive to exce...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 95 5.10.5. ADDING A RECORDER OFFSET TO AN ANALOG OUTPUT Some analog signal recorders require that the zero signal be significantly different from the baseline of the recorder in order to record slightly negative readings from noise around...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 96 5.10.6. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4 The test functions available to be reported are: Table 5-7: Test Channels Functions available on the T803’s Analog Output TEST CHANNEL DESCRIPTION ZERO FULL SCALE NONE Test Channe...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 97 To activate the TEST Channel and select a function (in this example SAMPLE PRESSURE ), press: SAMPLE CO2 RNG=20.00 % CO2=XXX.XX <TST TST> CAL SETUP DIAG TEST CHAN:NONE PREV NEXT ENTR EXIT DIAG SIGNAL I/O PREV NEXT ENTR EXIT DIAG ...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 98 5.10.7. AIN CALIBRATION This is the sub-menu to conduct a calibration of the T803 analyzer’s analog inputs. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. To pe...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 99 5.10.8. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGURATION To configure the analyzer’s optional analog inputs, define for each channel : gain (number of units represented by 1 volt) offset (volts) engineering units to be represented ...
Setup Menu Teledyne API T803 CO2/O2 Analyzer Operation Manual 100 5.11. SETUP MORE ALRM: USING THE GAS CONCENTRATION ALARMS (OPTION 61) The T803 includes two concentration alarms. Each alarm has a user-settable limit, and is associated with an opto-isolated TTL relay accessible via the status ou...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Setup Menu 101 5.11.1. SETTING THE T803 OPTION 61 CONCENTRATION ALARM LIMITS To enable concentration alarms and set the Limit points, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07276B DCN6418
103 6. COMMUNICATIONS SETUP AND OPERATION The T803 is equipped with an Ethernet port, a USB port and two serial communication (COM) ports (RS232 and COM2) located on the rear panel (see Figure 3-2). Both COM ports operate similarly and give the user the ability to communicate with, issue commands to...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 104 Enabled mode (32) are selected, the analyzer would display a combined MODE ID of 35 . Table 6-1: COM Port Communication Modes MODE 1 ID DESCRIPTION QUIET 1 Quiet mode suppresses any feedback from the analyzer (...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 105 Press the following buttons to select communication modes for one of the COM Ports, such as the following example where RS-485 mode is enabled: (The ACAL submenu in the Primary Setup Menu is a special configura...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 106 6.2.2. COM PORT BAUD RATE To select the baud rate of either one of the COM Ports, press: 07276B DCN6418
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 107 6.2.3. COM PORT TESTING The serial ports can be tested for correct connection and output in the COM menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, the ...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 108 (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 6.2.4. MACHINE ID Each type of Teledyne API’s analyzer is configured with a default ID code. The default ID code for the...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 109 6.3. REMOTE ACCESS VIA THE ETHERNET For network or Internet communication, the analyzer can be connected to any standard 10BaseT or 100BaseT Ethernet network via low-cost network hubs, switches or routers. The ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 111 To view the above properties list in Table 6-3 press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07276B DCN6418
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 112 6.3.2. MANUALLY CONFIGURING THE NETWORK IP ADDRESSES There are several circumstances when you may need to configure the interface settings of the analyzer’s Ethernet card manually: your LAN is not running a D...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 114 6.3.3. CHANGING THE ANALYZER’S HOSTNAME The HOSTNAME is the name by which the analyzer appears on your network. The default name for all Teledyne API T803 analyzers is T803 . To change this name (particularly...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 115 6.4. USB PORT FOR REMOTE ACCESS The analyzer can be operated through a personal computer by downloading the TAPI USB driver and directly connecting their respective USB ports. 1. Install the Teledyne T-Series U...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 117 6.5. COMMUNICATIONS PROTOCOLS This section presents MODBUS and HESSEN informtion. 6.5.1. MODBUS SETUP The following set of instructions assumes that the user is familiar with MODBUS communications, and provides...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 118 Example Read/Write Definition window: Example Connection Setup window: Example MODBUS Poll window: 6.5.2. HESSEN The Hessen protocol is a multidrop protocol, in which several remote instruments are connected vi...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 119 host computer and only when they receive a command containing their own unique ID number. The Hessen protocol is designed to accomplish two things: to obtain the status of remote instruments, including the conc...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 120 Press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 6.5.2.3. SELECTING A HESSEN PROTOCOL TYPE Currently there are two versions of Hessen Protocol in use. The origina...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 121 TYPE 2 that has more flexibility when operating with instruments that can measure more than one type of gas. For more specific information about the difference between TYPE 1 and TYPE 2 download the Manual Adde...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 122 6.5.2.4. SETTING THE HESSEN PROTOCOL RESPONSE MODE The Teledyne API implementation of Hessen Protocol allows the user to choose one of several different modes of response for the analyzer. Table 6-5: Teledyne A...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 123 6.5.2.5. GAS LIST ENTRY FORMAT AND DEFINITIONS The T803 analyzer keeps a list of available gas types. Each entry in this list is of the following format. [GAS TYPE],[RANGE],[GAS ID],[REPORTED] WHERE: GAS TYPE =...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 124 6.5.2.6. EDITING OR ADDING HESSEN GAS LIST ENTRIES To add or edit an entry to the Hessen Gas List, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07276B DCN6418
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 125 6.5.2.7. DELETING HESSEN GAS LIST ENTRIES To delete an entry from the Hessen Gas list, press: (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07276B DCN6418
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 126 6.5.3. SETTING HESSEN PROTOCOL STATUS FLAGS Teledyne API’s implementation of Hessen protocols includes a set of status bits that the instrument includes in responses to inform the host computer of its condition...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Communications Setup and Operation 127 To assign or reset the status flag bit assignments, press: SAMPLE CO2 RNG=20.00 % CO2=XXX.XX <TST TST> CAL SETUP DEL deletes the character currently inside the cursor brackets. Press the [?] key repeatedl...
Communications Setup and Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 128 6.5.4. INSTRUMENT ID CODE Each instrument on a Hessen Protocol network must have a unique ID code. If more than one T803 analyzer is on the Hessen network, you will have to change this code for all but one of t...
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 130 7.1.1. DAS STATUS The green SAMPLE LED on the instrument front panel, which indicates the analyzer status, also indicates certain aspects of the DAS status: Table 7-1: SAMPLE LED Status Indicators for DA...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 131 7.1.2.1. DAS CHANNELS The key to the flexibility of the DAS is its ability to store a large number of combinations of triggering events and data parameters in the form of data channels. Users may create ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 133 Triggering Events and Data Parameters/Functions for these default channels are: Figure 7-1: Default DAS Channel Setup 07276B DCN6418
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 134 7.1.4. SETUP DAS VIEW: VIEWING DAS CHANNELS AND INDIVIDUAL RECORDS DAS data and settings can be viewed on the front panel display through the following touchscreen button sequence: Continue pressing ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 135 7.1.5. SETUP DAS EDIT: ACCESSING THE DAS EDIT MODE DAS configuration is most conveniently done through the APICOM remote control program. The following list of touchscreen buttons shows how to edit u...
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 136 7.1.5.1. EDITING DAS DATA CHANNEL NAMES To edit the name of a DAS data channel, follow the instruction shown in Section 7.1.5 then press: Starting at the EDIT CHANNEL MENU SETUP X.X 0) CONC: ATIMER,3,403...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 137 instantaneous data points, e.g., for the new slope and offset (concentration response) values at the end of a calibration. Zero and slope values are valuable to monitor response drift and to document whe...
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 138 concentration measurements, temperatures of the various heaters placed around the analyzer, pressures and flows of the pneumatic subsystem and other diagnostic measurements as well as calibration data su...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 141 To define the REPORT PERIOD , follow the instruction shown in Section 7.1.5 then press: The SAMPLE PERIOD and REPORT PERIOD intervals are synchronized to the beginning and end of the appropriate interval...
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 142 7.1.5.5. REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF If the instrument is powered off in the middle of a REPORT PERIOD , the samples accumulated so far during that period are lost. Once the...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 143 To set the NUMBER OF RECORDS , follow the instruction shown in Section 7.1.5 then press: 07276B DCN6418
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 144 7.1.5.7. RS-232 REPORT FUNCTION The DAS can automatically report data to the communications ports, where they can be captured with a terminal emulation program or simply viewed by the user using the APIC...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 145 7.1.5.8. ENABLING / DISABLING THE HOLD OFF FEATURE The DAS HOLD OFF feature prevents data collection during calibration operations. To enable or disable the HOLD OFF , follow the instruction shown in Sec...
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 146 7.1.5.9. THE COMPACT REPORT FEATURE When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each parameter in one channel on a separate line, up to five param...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 147 7.1.6. DISABLING/ENABLING DATA CHANNELS Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-module. To disable a data channel, follow the instruction shown in S...
Data Acquisition System (DAS & APICOM Teledyne API T803 CO2/O2 Analyzer Operation Manual 148 7.2. REMOTE DAS CONFIGURATION The DAS can be configured and operated remotely via either the APICOM interface or a terminal emulation program. (Refer to Section 8 of this manual for details on remote ope...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Data Acquisition System (DAS & APICOM 149 Figure 7-3: APICOM User Interface for Configuring the DAS Once a DAS configuration is created, it is conveniently saved to your computer and can be uploaded to any instrument; it can be stored on a compu...
151 8. REMOTE OPERATION This section provides information needed when using external digital and serial I/O for remote operation. It assumes that the electrical connections have been made as described in Section 3.3.1. Your analyzer can be remotely configured, calibrated or queried for stored data t...
Remote Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 152 8.2. INTERACTIVE MODE Interactive mode is used with a terminal emulation program such as HyperTerminal or a “dumb” computer terminal. 8.2.1.1. HELP COMMANDS IN INTERACTIVE MODE Table 8-1: Terminal Mode Software Commands COMMAND ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Remote Operation 153 COMMAND is the command designator: This string is the name of the command being issued (LIST, ABORT, NAME, EXIT, etc.). Some commands may have additional arguments that define how the command is to be executed. Press ? <CR>...
Remote Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 154 Text strings : Used to represent data that cannot be easily represented by other data types, such as data channel names, which may contain letters and numbers. They consist of a quotation mark, followed by one or more printabl...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Remote Operation 155 8.2.1.6. COM PORT PASSWORD SECURITY In order to provide security for remote access of the T803, a LOGON feature can be enabled to require a password before the instrument will accept commands. This is done by turning on the SECU...
Remote Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 156 The DTE-DCE switch on the rear panel is in the DCE position. The T803 COM port is set for a baud rate that is compatible with the modem, The Modem is designed to operate with an 8-bit word length with one stop bit. The M...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Remote Operation 157 To initialize the modem press: SAMPLE CO2 RNG=20.00 % CO2=XXX.XX <TST TST> CAL SETUP If there is a problem initializing the modem the message, “MODEM NOT INITIALIZED” will appear. SETUP X.X COMMUNICATIONS MENU ID INET COM1...
159 9. CALIBRATION PROCEDURES This section contains a variety of information regarding the various methods for calibrating a T803 as well as other supporting information This section is organized as follows: Section 9.1 – Before Calibration This section contains general information you should know b...
Calibration Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 160 9.1. BEFORE CALIBRATION The calibration procedures in this section assume that the range mode, analog range and units of measure have already been selected for the analyzer. If this has not been done, please do so before c...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Calibration Procedures 161 Cylinders of both calibrated O 2 and CO 2 gas traceable to NIST-Standard Reference Material specifications (also referred to as SRMs or EPA protocol calibration gases) are commercially available. Table 9-1: NIST SRM's Avai...
Calibration Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 162 9.2. MANUAL CALIBRATION CHECKS AND CALIBRATION OF THE T803 ANALYZER I MPORTANT I MPACT ON R EADINGS OR D ATA ZERO/SPAN CALIBRATION CHECKS VS. ZERO/SPAN CALIBRATION Pressing the ENTR button during the following procedure re...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Calibration Procedures 163 9.2.2. PERFORMING A MANUAL CALIBRATION CHECK Note If the ZERO or SPAN menu buttons are not displayed, the measurement made during this cal check is out of the allowable range allowed for a reliable calibration. See Section...
Calibration Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 164 9.2.3.1. SETTING THE EXPECTED SPAN GAS CONCENTRATION NOTE When setting expected concentration values, consider impurities in your span gas. The expected CO 2 span gas concentration should be 80% of the reporting range of t...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Calibration Procedures 165 9.2.3.2. ZERO/SPAN POINT CALIBRATION PROCEDURE To perform the zero/span calibration procedure: 07276B DCN6418
Calibration Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 166 9.3. ASSESSING CALIBRATION QUALITY After completing one of the calibration procedures described above, it is important to evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear re...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Calibration Procedures 167 9.4. CALIBRATION OF THE T803’S ELECTRONIC SUBSYSTEMS 9.4.1. PRESSURE CALIBRATION A sensor in the sample path continuously measures the pressure of the sample gas. This data is used to compensate the measured CO 2 and O 2 c...
Calibration Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 168 To cause the analyzer to measure and record a value for PRES, press. (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). 07276B DCN6418
Teledyne API T803 CO2/O2 Analyzer Operation Manual Calibration Procedures 169 9.4.2. FLOW CALIBRATION The flow calibration allows the user to adjust the values of the sample flow rates as they are displayed on the front panel and reported through COM ports to match the actual flow rate measured at t...
173 10. MAINTENANCE SCHEDULE & PROCEDURES The T803 Analyzer utilizes technologies that are non-depleting and require very little maintenance. However, there are a minimal number of simple procedures that when performed regularly will ensure that the analyzer continues to operate accurately and r...
Maintenance Schedule & Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 177 10.2. PREDICTIVE DIAGNOSTICS The Test Functions can be used to predict failures by looking at how their values change over time. Initially it may be useful to compare the state of these Test Functions to the...
Maintenance Schedule & Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 178 To change the filter: 1. Turn OFF the analyzer to prevent drawing debris into the instrument. 2. Open the T803’s hinged front panel and unscrew the knurled retaining ring on the filter assembly. Figure 10-1:...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Maintenance Schedule & Procedures 179 10.3.3. PERFORMING LEAK CHECKS HAZARD S TRONG O XIDIZER OXYGEN IS A STRONG OXIDIZER. ONLY Perform Leak Checks using N 2 gas and after thoroughly purging the analyzer’s internal pneumatics. Leaks are the most...
Maintenance Schedule & Procedures Teledyne API T803 CO2/O2 Analyzer Operation Manual 180 CAUTION G ENERAL S AFETY H AZARD Do not use bubble solution with vacuum applied to the analyzer. The solution may contaminate the instrument. Do not exceed 15 PSI pressure. 1. Turn OFF power to the instrumen...
181 11. TROUBLESHOOTING AND SERVICE This section describes a variety of methods for identifying the source of performance problems with the analyzer. Also included here are procedures that are used to repair the instrument. HAZARD S TRONG O XIDIZER OXYGEN IS A STRONG OXIDIZER. Before working with th...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 182 1. Note any WARNING MESSAGES and take corrective action as necessary. 2. Examine the values of all TEST functions and compare them to factory values. Note any major deviations from the factory values and take correcti...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 184 To view or clear the various warning messages press: Suppresses the warning messages Press CLR to clear the current message. If more than one warning is active, the next message will take its place. Once the last warn...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 185 WARNING MESSAGE FAULT CONDITION POSSIBLE CAUSES RELAY BOARD WARN The CPU cannot communicate with the Relay Board. I 2 C bus failure Failed relay board Loose connectors/wiring SAMPLE FLOW WARN Sample flow rate is < ...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 186 Table 11-2: Test Functions - Indicated Failures TEST FUNCTIONS ( As Displayed) INDICATED FAILURE(S) TIME Time of day clock is too fast or slow To adjust See Section 5.6. Battery in clock chip on CPU board may be dead....
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 187 Figure 11-2: Example of Signal I/O Function (The ACAL submenu in the Primary Setup Menu is a special configuration; consult factory). Note Any I/O signals changed while in the signal I/O menu will remain in effect ONL...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 188 11.2. USING THE INTERNAL ELECTRONIC STATUS LEDS Several LEDs are located inside the instrument to assist in determining if the analyzer’s CPU, I 2 C bus and relay board are functioning properly. 11.2.1. CPU STATUS IND...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 189 11.2.2.2. RELAY PCA STATUS LED S D6 (Yellow) O 2 Sensor Heater D5 (Yellow) – CO 2 Sensor Heater (only with CO 2 option) D1 (RED) Watchdog Indicator Figure 11-4: Relay PCA Status LEDs Used for Troubleshooting Table 11-...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 190 11.3. GAS FLOW PROBLEMS When troubleshooting flow problems, it is a good idea to first confirm that the actual flow and not the analyzer’s flow sensor and software are in error, or the flow meter is in error. Use an i...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 191 11.3.2. TYPICAL SAMPLE GAS FLOW PROBLEMS 11.3.2.1. FLOW IS ZERO The unit displays a SAMPLE FLOW warning message on the front panel display or the SAMPLE FLOW test function reports a zero or very low flow rate. Confirm...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 192 11.3.2.4. DISPLAYED FLOW WARNINGS This warning means that there is inadequate gas flow. There are four conditions that might cause this: A leak upstream or downstream of the flow sensor A flow obstruction upstream...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 193 11.4.2. NON-REPEATABLE ZERO AND SPAN As stated earlier, leaks both in the T803 and in the external system are a common source of unstable and non-repeatable readings. 1. Check for leaks in the pneumatic systems as des...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 194 11.5.1. TEMPERATURE PROBLEMS Individual control loops are used to maintain the set point of the temperatures to both sensors.. If any of these temperatures are out of range or are poorly controlled, the T803 will perf...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 195 +5V 2 Red Agnd 3 Green +15V 4 Blue -15V 5 Yellow +12V Ret (ground) 6 Purple +12V 7 Orange A voltmeter should be used to verify that the DC voltages are correct per the values in the table below, and an oscilloscope, i...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 196 remote commands and the display changes accordingly, the touchscreen interface may be faulty. 11.6.5. LCD DISPLAY MODULE Verify the functioning of the front panel display by observing it when power is applied to the i...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 198 11.6.9. MOTHERBOARD 11.6.9.1. A/D FUNCTIONS The simplest method to check the operation of the A-to-D converter on the motherboard is to use the Signal I/O function under the DIAG menu to check the two A/D reference vo...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 199 11.6.9.3. STATUS OUTPUTS The procedure below can be used to test the Status outputs: 1. Connect a jumper between the “D“ pin and the “ ” pin on the status output connector. 2. Connect a 1000 ohm resistor between the...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 200 11.6.11. RS-232 COMMUNICATIONS 11.6.11.1. GENERAL RS-232 TROUBLESHOOTING Teledyne API analyzers use the RS-232 communications protocol to allow the instrument to be connected to a variety of computer-based equipment. ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Troubleshooting and Service 201 11.6.12. CO 2 SENSOR STATUS LED’S There are Two LEDs located on the CO 2 sensor PCA. Figure 11-6: Location of Diagnostic LEDs on CO 2 Sensor PCA Normal Operation: V8 is not lit – V9 is Blinking Error State: Both L...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 202 9. After reconnecting the power and pneumatic lines, flow check the instrument as described in the Section 10.3.4. Figure 11-7: Critical Flow Restrictor Assembly / Disassembly 11.7.2. DISK-ON-MODULE REPLACEMENT PROCED...
Troubleshooting and Service Teledyne API T803 CO2/O2 Analyzer Operation Manual 204 QUESTION ANSWER How do I measure the sample flow? Sample flow is measured by attaching a calibrated rotameter, wet test meter, or other flow-measuring device to the sample inlet port when the instrument is operating. ...
205 12. PRINCIPLES OF OPERATION The T803 analyzer is a microprocessor-controlled analyzer that determines the percent concentration of carbon dioxde (CO 2 ) and molecular oxygen (O 2 ) in a sample gas drawn through the instrument. It uses a paramagnetic sensor that relies on the relatively high reac...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 206 Figure 12-1: Paramagnetic O 2 Sensor Design When sample gas containing oxygen flows into the magneto-mechanical sensor, the O2 molecules are drawn toward the strong point of the magnetic field. This causes the N2 filled s...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 207 This current is passed to a feedback loop, which generates a second current to a wire winding (in effect, a small DC electric motor) mounted on the suspended mirror. The more O 2 present, the more the dumbbell and its att...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 208 chamber. This dual wavelength method of measuring CO 2 allows the instrument to compensate for ancillary effects like sensor aging and contamination. 12.3. OPERATION WITHIN THE T803 ANALYZER Operationally, both the CO 2 a...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 209 12.4. PNEUMATIC OPERATION CAUTION G ENERAL S AFETY H AZARD I T IS IMPORTANT THAT THE SAMPLE AIRFLOW SYSTEM IS BOTH LEAK - TIGHT AND NOT PRESSURIZED OVER AMBIENT PRESSURE . Regular leak checks should be performed on the an...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 210 12.5. FLOW RATE CONTROL To maintain a constant flow rate of the sample gas through the instrument, the T803 uses a special flow control assembly located in the exhaust gas line just before the optional internal pump or co...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 211 SPRING O-RINGS FILTER CRITICAL FLOW ORIFICE AREA OF LOW PRESSURE AREA OF HIGH PRESSURE Sonic Shockwave Figure 12-6: Flow Control Assembly & Critical Flow Orifice The actual flow rate of gas through the orifice (volume...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 212 sensor is mounted to a printed circuit board with the Sample Pressure sensor on the sample chamber; see the previous section and Figure 3-5. 12.6. ELECTRONIC OPERATION 12.6.1. OVERVIEW Figure 10-9 shows a block diagram of...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 213 Flow/Pressure Sensor PCA CO 2 Sensor Heater Analog Outputs Aout 1 Aout 4 Analog Outputs (D/A) External Digital I/O Power Up Circuit PC 104 Bus PC 104 CPU Card Disk on Module Flash Chip Aout 3 Aout 2 TEST CHANNEL OUTPUT St...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 214 Figure 12-8. CPU Card The CPU includes two types of non-volatile data storage: an embedded 2MB flash chip and a Disk on Module (DOM). 12.6.2.1. DISK-ON-MODULE (DOM) The DOM is a 44-pin IDE flash disk with a storage capaci...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 215 12.6.3. RELAY BOARD The CPU issues commands via a series of relays and switches located on a separate printed circuit assembly, called the relay PCA, to control the function of key electromechanical devices such as heater...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 216 AC Relay Retainer Plate Retainer Mounting Screws Figure 12-10: Relay PCA with AC Relay Retainer in Place 12.6.3.1. STATUS LEDS LEDs located on the Analyzer’s relay PCA, show the current status of various control functions...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 217 D6 (Yellow) O 2 Sensor Heater D5 (Yellow) –CO2 Sensor D1 (RED) Watchdog Indicator Figure 12-11: Status LED Locations – Relay PCA 12.6.3.2. WATCHDOG CIRCUITRY The most important of the status LEDs on the relay board is the...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 218 12.6.5. MOTHERBOARD This printed circuit assembly provides a multitude of functions including, A/D conversion, digital input/output, PC-104 to I 2 C translation, temperature sensor signal processing, and serves as a pass-...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 219 current temperature of the chamber housing to the CPU as part of the bench heater control loop. Another thermistor, attached to the motherboard, measures the analyzer’s inside temperature (box temperature). This informati...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 220 12.6.7. POWER SUPPLY / CIRCUIT BREAKER The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at either 50Hz or 60Hz. Individual units are set up at the factory to accept any combination of these five attributes. As i...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 221 12.6.8. FRONT PANEL TOUCHSCREEN/DISPLAY INTERFACE Users can input data and receive information directly through the front panel touchscreen display. The LCD display is controlled directly by the CPU board. The touchscreen...
Principles of Operation Teledyne API T803 CO2/O2 Analyzer Operation Manual 222 12.6.9. SOFTWARE OPERATION The T803 Analyzer is at its heart a high performance, 386-based microcomputer running MS-DOS. Inside the DOS shell, special software developed by Teledyne API interprets user commands via the va...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Principles of Operation 223 12.6.11. CALIBRATION - SLOPE AND OFFSET Calibration of the analyzer is performed exclusively in software. During instrument calibration the user enters expected values for span via the front panel touchscreen; values are ...
225 13. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne API considers the prevention of damage caused by the discharge of static electricity to be extremely important part of making sure that your analyzer continues to provide reliable service for a long time. This section describes how static electri...
A Primer on Electro-Static Discharge Teledyne API T803 CO2/O2 Analyzer Operation Manual 226 Walking across nylon carpet 1,500V 35,000V Walking across vinyl tile 250V 12,000V Worker at bench 100V 6,000V Poly bag picked up from bench 1,200V 20,000V Moving around in a chair padded with urethane foam 1,...
Teledyne API T803 CO2/O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 227 When static charges accumulated on a sensitive device discharges from the device to another surface such as packaging materials, work surfaces, machine surfaces or other device. In some cases, charged device ...
A Primer on Electro-Static Discharge Teledyne API T803 CO2/O2 Analyzer Operation Manual 228 As long as my analyzer is properly installed, it is safe from damage caused by static discharges : It is true that when properly installed the chassis ground of your analyzer is tied to earth ground and its e...
Teledyne API T803 CO2/O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 229 Always store sensitive components and assemblies in anti-ESD storage bags or bins: Even when you are not working on them, store all devices and assemblies in a closed anti-Static bag or bin. This will prevent...
A Primer on Electro-Static Discharge Teledyne API T803 CO2/O2 Analyzer Operation Manual 230 3. Open the casing of the analyzer and begin work. Up to this point, the closed metal casing of your analyzer has isolated the components and assemblies inside from any conducted or induced static charges. 4....
Teledyne API T803 CO2/O2 Analyzer Operation Manual A Primer on Electro-Static Discharge 231 If you are at an anti-ESD workbench, lay the container down on the conductive work surface. In either case wait several seconds. 10. Place the item in the container. 11. Seal the container. If using a bag, fo...
A Primer on Electro-Static Discharge Teledyne API T803 CO2/O2 Analyzer Operation Manual 232 CAUTION ESD Hazard DO NOT use pink-poly bags. NEVER allow any standard plastic packaging materials to touch the electronic component/assembly directly. This includes , but is not limited to, plastic bubbl...
233 GLOSSARY Note: Some terms in this glossary may not occur elsewhere in this manual. Term Description/Definition 10BaseT an Ethernet standard that uses twisted (“T”) pairs of copper wires to transmit at 10 megabits per second (Mbps) 100BaseT same as 10BaseT except ten times faster (100 Mbps) APICO...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Index 237 INDEX A AC Power 60 Hz, 35 AIN, 99 ALRM, 64, 101 ANALOG CAL WARNING, 52 Analog Inputs, 99 Analog Outputs, 36, 37, 63, 64, 66, 67, 81 Ain Calibration, 99 CONC1, 54 CONC2, 54 Configuration & Calibration, 64, 86, 87, 88, 89, 91, 93, 95, 9...
Index Teledyne API T803 CO2/O2 Analyzer Operation Manual 238 Control Inputs Electrical Connections, 39 Control Inputs Pin Assignments, 40 Control Inputs, 60 Control Inputs, 219 CPU, 182, 185, 186, 188, 199, 200, 212, 218, 219, 222 AIN Calibration, 99 Analog to Digital Converter, 52, 83 and Relay Boa...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Index 239 Environmental Protection Agency(EPA), 161 Ethernet, 108, 109, 110 Configuration, 109–14 Property Defaults, 110 using DHCP, 109 DHCP, 54, 109, 110 HOSTNAME, 114 Exhaust Gas, 32, 210 Exhaust Gas Outlet , 32, 50 EXIT Key, 64 EXITZR, 136 Exter...
Index Teledyne API T803 CO2/O2 Analyzer Operation Manual 240 O2 CELL TEMP, 62, 186 O2 CELL TEMP WARNING, 53, 184 O2 CONC ALRM1 WARNING, 53 O2 CONC ALRM2 WARNING, 53 O2 M-P CAL, 60 O2 OFFSET, 62, 186 O2 RANGE #1 AUTO, 72 O2 RANGE #2 AUTO, 72 O 2 sensor, 62, 160 zero cal, 48 O2 SLOPE, 62, 186 OFFSET, ...
Teledyne API T803 CO2/O2 Analyzer Operation Manual Index 241 SNGL, 68, 69 SPAN, 40 SPAN CAL, 176, 199 SPAN CAL [type], 60 Span Gas Concentration, 161, 164 Flow Rate, 50 Initial Cal, 54 Pressure Leak Check, 179 Troubleshooting, 185, 186, 191, 192, 193 Venting, 50 with Alarm Options, 101 Span Inlet , ...
Index Teledyne API T803 CO2/O2 Analyzer Operation Manual 242 SYSTEM RESET, 53, 126, 185 Warranty, 21 Watchdog Circuit, 188 Status LED, 188, 189, 216 Z Zero Air, 48, 159, 160 Flow Rate, 50 Initial Cal, 54 Troubleshooting, 177, 186, 191, 192, 193 Venting, 50 ZERO AIR Inlet , 32 ZERO CAL, 176, 199 ZERO...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A A-1 APPENDIX A – Version Specific Software Documentation APPENDIX A-1: Menu Trees, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) ................................................... 3 APPENDIX A-2: Setup Variables For ...
Appendix A Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) A-3 APPENDIX A-1: Menu Trees, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) PRIMARY SETUP MENU SAMPLE MSG 1 CLR 1 SETUP <TST TST> CAL TEST 1 CO2 CONC ZERO SPAN DAS CFG ACAL CLK RNGE PASS MORE 1 Only appears when...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A A-4 SETUP PASS DAS RNGE CLK MORE ACAL CFG <TST TST> MODEL TYPE AND NUMBER PART NUMBER SERIAL NUMBER SOFTWARE REVISION LIBRARY REVISION iCHIP SOFTWARE REVISION HESSEN PROTOCOL REVISION 2 CPU TYPE & OS REVISION DA...
Appendix A Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) A-7 SETUP PASS DAS RNGE CLK MORE ACAL CFG SAMPLE COMM VARS DIAG Go to DIAG Menu Tree HESN 2 INET 1 ID ENTER PASSWORD: 818 ENTER PASSWORD: 818 ENTER PASSWORD: 818 1 E-Series: only appears if Ethernet Option is installed. 2...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-9 APPENDIX A-2: Setup Variables For Serial I/O, Software Versions 1.0.3 (T-Series)/A3 (E-Series) Table A-1: Setup Variables Setup Variable Num...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-17 APPENDIX A-3: Warnings and Test Measurements, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) Table A-2: Warning Messages Name 1 Message ...
Appendix A Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) A-18 Error! Unknown document property name.Error! Unknown document property name.Error! Unknown doc Table A-3: Test Measurements Name 1 Message Text Description Test Measurements O2RANGE 5 CO2RANGE 4 O2 RNG=500.0 % CO2 RN...
Appendix A Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) A-20 Error! Unknown document property name.Error! Unknown document property name. Error! Unknown document property name.Error! Unknown document property name. APPENDIX A-4: Signal I/O Definitions, Software Versions 1.0.3 ...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-21 Signal Name Bit or Channel Number Description A status outputs, U24, J1017, pins 1 – 8 = bits 0 – 7, default I/O address 323 hex ST_SYSTEM_...
Appendix A Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) A-22 Error! Unknown document property name.Error! Unknown document property name. Error! Unknown document property name.Error! Unknown document property name. Signal Name Bit or Channel Number Description CO2_SPAN_VALVE 2...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-23 Signal Name Bit or Channel Number Description External analog input board, default I 2 C address 5C hex XIN1 7 0 External analog input 1 XI...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-27 APPENDIX A-6: Terminal Command Designators, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) Table A-7: Terminal Command Designators COMMA...
Appendix A Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) A-28 Error! Unknown document property name.Error! Unknown document property name.Error! Unknown docu Table A-8: Terminal Key Assignments TERMINAL KEY ASSIGNMENTS ESC Abort line CR (ENTER) Execute command Ctrl-C Switch to ...
Models T803, 803E Appendix A Menu Trees (Reference: 06763C DCN6418) Appendix A Error! Unknown document property name.Error! Unknown document property name. A-29 APPENDIX A-7: MODBUS ® Register Map, Software Versions 1.0.3 (T-Series)/A.3 (E-Series) MODBUS Register Address (dec., 0-based) Description ...
APPENDIX B - Spare Parts Note Use of replacement parts other than those supplied by Teledyne Advanced Pollution Instrumentation (TAPI) may result in non-compliance with European standard EN 61010-1. Note Due to the dynamic nature of part numbers, please refer to the TAPI Website at
T80X Spare Parts List (Ref: 072690000 A DCN6431, 2012 April 12) PARTNUMBER DESCRIPTION 000940700 CD, ORIFICE, .005 YELLOW 001763500 ASSY, FLOW CTL, 110CC, 1/4" ELBOW ‐ B 003290000 THERMISTOR, BASIC (VENDOR ASSY)(KB) 009690200 AKIT, TFE FLTR ...
Appendix C Warranty/Repair Questionnaire T80X, M80XE (06532C DCN 5798) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 C-1 CUSTOMER: _______________________________ PHONE: _____________________________________ ...
Appendix C Warranty/Repair Questionnaire T80X, M80XE (06532C DCN 5798) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 C-2 ____________________________________________________________________________________ If...
APPENDIX D – Wire List and Electronic Schematics 07276B DCN6418 D-1
Advanced Pollution Instrumentation A Teledyne Technologies Company TELEDYNE ' & % $ % 7KHLQIRUPDWLRQKHUHRQLVWKHSURSHUW\RI $3,DQGLVVXEPLWWHGLQVWULFWHVW FRQILGHQFHIRUUHIHUHQFHRQO\ 8QDXWKRUL]HGXVHE\DQ\RQHIRUDQ\ RWKHUSXUSRVHVLVSURKLWHGWKLV GRFXPHQWRUDQLQIRUPDWLRQFRQWDLQHG LQLWPD\QRWEHGXSOLFDWHGZLWKR...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P1.R3.schdoc Drawn By: GUI Interface 06698 D 1 4 2 3 4 5 6 7 8 1 51 52 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 ...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P2.R3.schdoc Drawn By: 06698 D 2 4 22uH L1 Vin 5 SHDN 4 SW 1 GN D 2 FB 3 U1 CAT4139TD-GT3 4.7uF/16V C9 1K R12 2.0 R14 5V-GND +5V 9.76 R13 3.9uH L2 487K R9 309K R8 80.6K R18 66.5K R1...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P3.R3.schdoc Drawn By: 06698 D 4 3 12345 To old TScreen 70553-004 NI J11 LLRLSDRTLT +5V CHASSIS 1 2 NI J12 BUS +5 12345 To new TScreen 70553-004 J10 LT 1 RT 2 SH L D 3 RL 4 LL 5 GN ...
1 1 2 2 3 3 4 4 5 5 6 6 D D C C B B A A Title Number Revision Size B Date: 6/24/2010 Sheet of File: N:\PCBMGR\..\06696.P4.R3.schdoc Drawn By: 06698 D 4 4 TOUCH SCREEN INTERFACE CIRCUITRY ( TBD) GUI Interface R40 10K O p tion CLKOUT_P CLKOUT_N 3.3V C49 0.1 R41100 FB6 FB14 Vcc PIN 28 C47 0.01 22uF/6.3...
1 1 2 2 3 3 4 4 D D C C B B A A Title Number Revision Size A Date: 5/7/2010 Sheet of File: N:\PCBMGR\..\06882-P1-R0.SchDoc Drawn By: LVDS, Transmitter Board B 1 1 RT 06882 VAD6VAD7VAD8VAD9VAD10VAD11VBD10VBD11VAD0VAD1VAD2VAD3VBD2VBD3VBD4VBD5VBD6VBD7 VBDE Y0_NY0_PY1_NY1_PY2_NY2_P CLKOUT_NCLKOUT_P R2 2...
1 1 2 2 3 3 4 4 D D C C B B A A Title Number Revision Size A Date: 5/6/2011 Sheet of File: N:\PCBMGR\..\06731-1_ETHERNET.SchDoc Drawn By: Auxiliary I/O Board (PWR-ETHERNET) B 3 1 RT 06731 STRAIGHT THROUGH ETHERNET GND +5V SCL SDA +5V-ISO + C17 100uF GND DS3 GRN 12345678 P2 Header 8 1 2 3 4 5 6 U6 SP...
Teledyne Manuals
-
Teledyne 4220
Manual
-
Teledyne 311-D
Manual
-
Teledyne HPM 4/6
Manual
-
Teledyne T802
Manual
-
Teledyne 318R
Manual
-
Teledyne 9700
Manual
-
Teledyne BDS-3960
Manual
-
Teledyne C-2000
Manual
-
Teledyne GFC 7001EM
Manual
-
Teledyne 2103
Manual
-
Teledyne 6400E
Manual
-
Teledyne AX300-I
Manual
-
Teledyne 6600
Manual
-
Teledyne 9950
Manual
-
Teledyne PEM 9002
Manual
-
Teledyne HPM-2002-OBE
Manual
-
Teledyne Infrared Gas Analyzer
Manual
-
Teledyne 102E
Manual
-
Teledyne OT-3
Manual
-
Teledyne LLG
Manual