Page 3 - ABOUT THIS MANUAL; Please refer to our Website or Sales for more recent updates.; REVISION HISTORY; Date; Correct CE value
07271B DCN6646 i ABOUT THIS MANUAL This addendum is to be used in conjunction with the T200 NOx Analyzer manual, p/n 06858; for operating the T201 analyzer; it includes document(s) as listed below. Part No. Name/Description 07356 T201 Spare Parts List (in Section 7.0) Note: Please refer to our Websi...
Page 5 - SAFETY MESSAGES; CAUTION
Teledyne API Model T201 NH 3 Analyzer Operator Manual T201 Ammonia Analyzer 07271B DCN6646 iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is ...
Page 6 - iv; CONSIGNES DE SÉCURITÉ
iv 07271B DCN6646 CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est représentée par un pic...
Page 7 - TABLE OF CONTENTS; PMT; LIST OF FIGURES
07271B DCN6646 v TABLE OF CONTENTS MODEL T201 AMMONIA ANALYZER .............................................................I CONSIGNES DE SÉCURITÉ ......................................................................... IV 1.0 T201 AMMONIA ANALYZER ....................................................
Page 8 - vi; Calibration Procedure – NH
vi 07271B DCN6646 Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port ....................... 19 Table 2-5. Confirming Ammonia Converter Efficiency ................................................... 21 Table 2-6. NH 3 Calibration Procedure – NH 3 Gas through the SAMPLE Port ............
Page 11 - Pneumatic Flow; with Zero and Span Valve Options Pneumatic Flow
Teledyne API Model T201 NH 3 Analyzer Operator Manual T201 Ammonia Analyzer 07271B DCN6646 9 Figure 1-2. T201 and M501 NH 3 Pneumatic Flow M501NH3 TNX OUT TN OUT SPAN GAS INLET Zero Gas Valve Span Gas Valve Sample Gas Valve ZERO AIR INLET SAMPLE GAS INLET PUMP Optional Sample Filter P/N 016300800, 1...
Page 12 - HEORY OF; Note
T201 Ammonia Analyzer Teledyne API Model T201 NH 3 Analyzer Operator Manual 10 07271B DCN6646 1.1 T HEORY OF O PERATION The Teledyne-API Model T201 analyzer measures ammonia by oxidizing it to nitric oxide by the following reaction: 4NH 3 + 5O 2 4NO + 6H 2 O The resulting nitric oxide is then meas...
Page 13 - concentration is uncompensated. It is recommended that; Minimizing PMT Drift
Teledyne API Model T201 NH 3 Analyzer Operator Manual T201 Ammonia Analyzer 07271B DCN6646 11 The ammonia converter efficiency factor (NH 3 _CE_FACTOR1) is discussed in more detail in Section 2.3 of this manual. The conversion efficiency of the M501 NH 3 should be checked prior to starting long term...
Page 15 - NALYZER; Table 1-1. T201 Operating Specifications
Teledyne API Model T201 NH 3 Analyzer Operator Manual T201 Ammonia Analyzer 07271B DCN6646 13 1.4 T201 A NALYZER S PECIFICATIONS Table 1-1. T201 Operating Specifications Ranges Min: 0-50 ppb Full scale Max: 0-2000 ppb Full scale (selectable, independent NH 3 , NO, NO 2 , NOx ranges supported) Measur...
Page 16 - This page intentionally left blank.
07271B DCN6646 14 This page intentionally left blank.
Page 17 - PROCEDURE; DO NOT
07271B DCN6646 15 2.0 CALIBRATION PROCEDURE This section begins with a high-level overview of the calibration procedure for the T201 analyzer. Details are provided starting in Section 2.1. First, take the time to read the T200 manual to familiarize yourself with the Chemiluminescence process. It i...
Page 18 - EROING THE; Important; and NO; Action
Calibration Procedure Teledyne API Model T201 NH 3 Analyzer Operator Manual 16 07271B DCN6646 2.1 Z EROING THE A NALYZER The analyzer can be zeroed by either applying zero air straight to the sample port or through the optional Zero/Span valves, if installed. It is important to remember that if the ...
Page 19 - Table 2-2. Zero Calibration Procedure - Zero Gas through ZERO Port; PANNING THE; channel stabilizes. The process of
Teledyne API Model T201 NH 3 Analyzer Operator Manual Calibration Procedure 07271B DCN6646 17 Table 2-2. Zero Calibration Procedure - Zero Gas through ZERO Port Step No. Action Comment 1. Press CALZ The T201 enters the calibrate mode from sample mode. The zero gas is supplied through the ZEROinlet p...
Page 20 - Table 2-3. NO Calibration Procedure – NO Gas through the SAMPLE Port
Calibration Procedure Teledyne API Model T201 NH 3 Analyzer Operator Manual 18 07271B DCN6646 Table 2-3. NO Calibration Procedure – NO Gas through the SAMPLE Port Important: Each channel (TN x and NO x ) must be Spanned individually Step No. Action Comment 1. Press CAL The T201 enters the calibrate ...
Page 21 - Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port; CONC
Teledyne API Model T201 NH 3 Analyzer Operator Manual Calibration Procedure 07271B DCN6646 19 Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port Step No. Action Comment 1. Press CALS The T201 enters the calibrate mode from sample mode. The NO gas is supplied through the SPAN inlet po...
Page 22 - ammonia gas is “patience”. Ammonia is a very
Calibration Procedure Teledyne API Model T201 NH 3 Analyzer Operator Manual 20 07271B DCN6646 The most important criterion an individual can avail themselves of when spanning an analyzer with ammonia gas is “patience”. Ammonia is a very sticky gas and the response of the analyzer/calibration system ...
Page 23 - Table 2-5. Confirming Ammonia Converter Efficiency
Teledyne API Model T201 NH 3 Analyzer Operator Manual Calibration Procedure 07271B DCN6646 21 Table 2-5. Confirming Ammonia Converter Efficiency Step No. Action Comment 1. Press CAL The T201 enters the calibrate mode from Sample mode. 2. Channel Selection Press TN X then Press ENTR . 3. Range Select...
Page 25 - ammonia gas is supplied through the
Teledyne API Model T201 NH 3 Analyzer Operator Manual Calibration Procedure 07271B DCN6646 23 Table 2-7. NH 3 Calibration Procedure – NH 3 Gas through the SPAN Port Step No. Action Comment 1. Press CALS The T201 enters the calibrate mode from Sample mode. The ammonia gas is supplied through the SPAN...
Page 27 - CONFIGURABLE ANALOG OUTPUT OVERVIEW; DAS; ANALOG OUT; Table 3-1. Analog Output Data Default Settings; CHANNEL DEFAULT SETTING
07271B DCN6646 25 3.0 CONFIGURABLE ANALOG OUTPUT OVERVIEW There are three different methods to extract concentration data from the analyzer: internal data acquisition system (DAS), hyperlink, or strip chart. DAS The user can set up an internal DAS configuration either through the user display or usi...
Page 28 - CNC1 and output A2 to TNxCNC2. Gases with the
Configurable Analog Output Overview Teledyne API Model T201 NH 3 Analyzer Operator Manual 26 07271B DCN6646 Table 3-2. Analog Output Pin-Outs PIN ANALOG OUTPUT VOLTAGE SIGNAL CURRENT SIGNAL 1 V Out I Out + 2 A1 Ground I Out - 3 V Out I Out + 4 A2 Ground I Out - 5 V Out I Out + 6 A3 Ground I Out - 7 ...
Page 29 - SCHEDULE; The T201 requires the following additional maintenance items.; Table 4-1. Preventative Maintenance Schedule; Item; NH; AINTENANCE; external converter is operated at 825 oCelsius. Because of
07271B DCN6646 27 4.0 MAINTENANCE SCHEDULE The maintenance requirements of the T201 are the same as a standard T200 NO x analyzer but with additional maintenance items (Table 4-1). Please refer to Section 13 in the T200 Operator Manual for the T200 maintenance schedule and for maintenance and repair...
Page 30 - Converter Rebuild Parts List; INNER TUBE w/SCREEN; EPLACING THE; CAUTION – VERY HOT
Maintenance Schedule Teledyne API Model T201 NH 3 Analyzer Operator Manual 28 07271B DCN6646 Table 4-2. M501 NH 3 Converter Rebuild Parts List Part No. Description KIT000193 RETROFIT, M501 NH 3 INNER TUBE w/SCREEN KIT000139 REBUILD, CERAMIC BUSHINGS, M501 NH 3 HE0000007 CERAMIC HEATER, 220W@60V M501...
Page 31 - You must obtain the following replacement parts kit from TAPI:; converter. It can be distinguished by its wire colors: red
Teledyne API Model T201 NH 3 Analyzer Operator Manual Maintenance Schedule 07271B DCN6646 29 4.3 R EPLACING THE T HERMOCOUPLE The following instructions provide the necessary information to replace the existing thermocouple. Figure 4-2 is provided for reference. You must obtain the following replace...
Page 33 - The converter is now ready for operation.; UTO; This procedure should be performed with the sample pump running.; Note: Connect both the Teflon; “SAMPLE IN”; Note: Disconnect both the Teflon; Gas Port; “TO CONV”
Teledyne API Model T201 NH 3 Analyzer Operator Manual Maintenance Schedule 07271B DCN6646 31 18. After the Auto-Tune process is completed, verify that the “process” temperature is indicating that the desired temperature is stable and being regulated. The converter is now ready for operation. 4.4 A U...
Page 35 - LARM; Be aware that the
07271B DCN6646 33 5.0 ALARMS AND CAUTIONS Please refer to the T200 Operation Manual for a functional block diagram of the analyzer. 5.1 A LARM During initial warm up the internal Zero/Span factory option may exhibit a “Block Temperature” warning. This can be cleared once the block temperature stabil...
Page 37 - CALCULATING MOLY CONVERTER EFFICIENCIES; URPOSE; API T700 calibrator with O; ARTS; NONE; ROCEDURE; produced. For example, if the Ozone produced
07271B DCN6646 35 6.0 CALCULATING MOLY CONVERTER EFFICIENCIES 6.1 P URPOSE To provide instructions on how to calculate the efficiency of a Moly converter when using a GPT method of testing converters, by using the US EPA method, where the actual concentration of ozone is not a factor in the accuracy...
Page 39 - MOLY TEST DATA SHEET; RESULT; PPB; PPB; PPB; PPB
07271B DCN6646 37 MOLY TEST DATA SHEET Line # TEST RESULT 2 LEAK-CHECK (WHEN HOT) YES / NO 3 NO X RESPONSE (MOLY BYPASSED) __________ 4 NO X RESPONSE (MOLY IN-LINE) __________ 5 OUT-GASSING / EATING (NO – NO X ) __________ (>-5 PPB, <5 PPB) 6 (NOx ORIG ) (NO X mode, O 3 off) __________ PPB 7 (...
Page 41 - PARTNUMBER DESCRIPTION; ASSY, IZS VALVE HEATER
07271B DCN6646 39 7.0 T201 SPARE PARTS LIST This section lists the spare parts for the T201; however, due to the dynamic nature of spare parts numbers, it is recommended that you refer to our Website or contact Sales for recent updates. (Reference: 07356 29 March 2012, 12:17PM) PARTNUMBER DESCRIPTIO...
Page 43 - Teledyne API Model T201 NH3 Analyzer Operator Manual
Teledyne API Model T201 NH3 Analyzer Operator Manual 07271B DCN6646 41 (Reference: 07356 29 March 2012, 12:17PM) PARTNUMBER DESCRIPTION HW0000685 LATCH, MAGNETIC, FRONT PANEL KIT000095 AKIT, REPLACEMENT COOLER KIT000219 AKIT, 4-20MA CURRENT OUTPUT KIT000253 ASSY & TEST, SPARE PS37 KIT000254 ASSY...