Page 2 - S9 Escape indications for use; English
English 1 S9 Escape indications for use The S9 Escape CPAP system is indicated for the treatment of obstructive sleep apnoea (OSA) in patients weighing more than 30 kg. The S9 Escape CPAP system is intended for home and hospital use. Contraindications Positive airway pressure therapy may be contrain...
Page 3 - Control panel; dial allows you to enter into a menu and confirm your choice.; Navigating the menus; until the parameter you require is displayed in blue.; Getting started; Daily
English 2 Control panel Refer to illustration B.The control panel of your S9 device includes the following items: 1. Start/Stop button: Starts or stops treatment2. Home menu 3. LCD screen 4. Info menu*: Allows you to view your sleep statistics5. Setup menu*: Allows you to make changes to settings6. ...
Page 4 - Replacing the air filter:; Troubleshooting; Problem/Possible cause; No display
English 3 Weekly: 1. Remove the air tubing from the S9 device and the mask.2. Wash the air tubing in warm water using mild detergent.3. Rinse thoroughly, hang, and allow to dry.4. Before next use, reconnect the air tubing to the air outlet and mask. Monthly: 1. Wipe the exterior of the S9 with a dam...
Page 5 - Pressure rises inappropriately; High temperature fault, refer to user manual; Check ResMed 30/90W Power Supply Unit and fully insert the; Displays error message:; Tube blocked, please check your tube
English 4 Mask and headgear are not positioned correctly. Adjust position of mask and headgear. Incorrect air tubing selected. If you are using the SlimLine or Standard air tubing ensure that you have the correct air tubing selected via the menu. Pressure rises inappropriately Talking, coughing or b...
Page 6 - SD card; High leak, please check system setup and all connections; This message only appears once. If you re-insert the; Settings updated successfully, press any key
English 5 SD card An SD card has been supplied to gather therapy data from your S9 device and provide settings updates from your clinician. When instructed to do so, disconnect the S9 device from the power outlet, remove your SD card, insert it in the protective folder and send it to your clinician....
Page 7 - The air flow for breathing produced by this therapy
English 6 DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871: Sound pressure level 24 dBA as measured according to ISO 17510-1:200226 dBA with uncertainty of 2 dBA as measured according to EN ISO 17510-1:2009 Sound power level 34 dBA with uncertainty of 2 dBA as measured accordin...
Page 8 - are not
English 7 Notes: • The manufacturer reserves the right to change these specifications without notice. • The temperature and relative humidity settings displayed for ClimateLine or ClimateLine MAX are not measured values. • Check with your clinician/service provider before using the SlimLine air tubi...
Page 9 - Pneumatic flow path; Environmental information; Servicing; S9 Escape
English 8 Pneumatic flow path Symbols The following symbols may appear on your S9, power supply unit, air tubing or packaging. Caution; Read instructions before use; Protection against insertion of fingers and against vertically dripping water; Not drip proof; Type BF equipment; Class II equipment; ...
Page 10 - Limited warranty; Product
English 9 authorised ResMed Service Centre if there is any sign of wear or concern with device function. Otherwise, service and inspection of the devices generally should not be required during the five year design life of the device. Limited warranty ResMed Ltd (hereafter ’ResMed’) warrants that yo...
Page 11 - WARNINGS; Read the entire manual before using the device.
English 10 ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation ma...
Page 12 - Do not use electrically conductive or antistatic air tubings.; CAUTIONS
English 11 • Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an open flame. • Always ensure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off, so that unu...
Page 13 - User Guide; POSITIVE AIRWAY PRESSURE DEVICES
Global leaders in sleep and respirator y medicine w w w.resmed.com User Guide English • Deutsch • Français • Italiano • Español • Português • Nederlands • Ελληνικά • Polski • Magyar • Русский • Türkçe • Česky • Slovensky S9 Escape ™ POSITIVE AIRWAY PRESSURE DEVICES Manufacturer: ResMed Ltd 1 Elizabe...