Page 3 - Important Information; Warranty; Patents; WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
Important Information Warranty The NI PXIe-1075 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defec...
Page 4 - Contents; About This Manual
© National Instruments Corporation v NI PXIe-1075 User Manual Contents About This Manual Conventions ................................................................................................................... viiRelated Documentation..............................................................
Page 5 - Glossary
Contents NI PXIe-1075 User Manual vi ni.com Installing Peripheral Modules........................................................................................ 2-9Power Inhibit Switch LED Indicator ............................................................................. 2-10Remote Voltage Mon...
Page 6 - Conventions; File»Page Setup»Options; bold
© National Instruments Corporation vii NI PXIe-1075 User Manual About This Manual The NI PXIe-1075 User Manual describes the features of the NI PXIe-1075 chassis and contains information about configuring the chassis, installing the modules, and operating the chassis. Conventions The following conve...
Page 7 - Related Documentation
About This Manual NI PXIe-1075 User Manual viii ni.com Related Documentation The following documents contain information that you might find helpful as you read this manual: • IEEE 1101.1-1991, IEEE Standard for Mechanical Core Specifications for Microcomputers Using IEC 603-2 Connectors • IEEE 1101...
Page 8 - Getting Started; Unpacking
© National Instruments Corporation 1-1 NI PXIe-1075 User Manual 1 Getting Started This chapter describes the key features of the NI PXIe-1075 chassis and lists the kit contents and optional equipment you can order from National Instruments. Unpacking Carefully inspect the shipping container and the ...
Page 9 - Key Features; Power Cable; Country
Chapter 1 Getting Started NI PXIe-1075 User Manual 1-2 ni.com For 100–120 VAC installation, use the NI cable part numbers listed in Table 1-2, which are rated for 125 V/15 A. Note NI PXI-1075 chassis ordered in North America or Japan ship with the proper cables by default. If you are missing any of ...
Page 10 - High Performance for Instrumentation Requirements
Chapter 1 Getting Started © National Instruments Corporation 1-3 NI PXIe-1075 User Manual The key features of the NI PXIe-1075 chassis include the following: High Performance for Instrumentation Requirements • Up to 1 GB/s (single direction) per PXI Express slot dedicated bandwidth (x4 PCIe) • 38 W ...
Page 11 - Chassis Description
Chapter 1 Getting Started NI PXIe-1075 User Manual 1-4 ni.com Chassis Description Figures 1-1 and 1-2 show the key features of the NI PXIe-1075 chassis front and back panels. Figure 1-1 shows the front view of the NI PXIe-1075. Figure 1-2 shows the rear view of the NI PXIe-1075. Figure 1-1. Front Vi...
Page 12 - Optional Equipment; EMC Filler Panels
Chapter 1 Getting Started © National Instruments Corporation 1-5 NI PXIe-1075 User Manual Figure 1-2. Rear View of the NI PXIe-1075 Chassis Optional Equipment Contact National Instruments to order the following options for the NI PXIe-1075 chassis. EMC Filler Panels Optional EMC filler panel kits ar...
Page 13 - Slot Blockers; NI PXIe-1075 Chassis Backplane Overview; Interoperability with CompactPCI
Chapter 1 Getting Started NI PXIe-1075 User Manual 1-6 ni.com Slot Blockers Optional slot blocker kits are available from National Instruments for improved thermal performance when all slots are not used. NI PXIe-1075 Chassis Backplane Overview This section provides an overview of the backplane feat...
Page 14 - System Controller Slot; Default; Hybrid Peripheral Slots
Chapter 1 Getting Started © National Instruments Corporation 1-7 NI PXIe-1075 User Manual System Controller Slot The system controller slot is Slot 1 of the chassis and is a 4-Link configuration system slot as defined by the CompactPCI Express and PXI Express specifications. It has three system cont...
Page 15 - PXI Express Peripheral Slots
Chapter 1 Getting Started NI PXIe-1075 User Manual 1-8 ni.com PXI Express Peripheral Slots There are eight (8) PXI Express peripheral slots: slots 6–9 and 11–14 (=8 slots). PXI Express peripheral slots can accept the following modules: • A PXI Express Peripheral with x4 or x1 PCI Express link to the...
Page 16 - PXI Local Bus
Chapter 1 Getting Started © National Instruments Corporation 1-9 NI PXIe-1075 User Manual Figure 1-4. PXIe_DSTAR and PXI Star Connectivity Diagram PXI Local Bus The PXI backplane local bus is a daisy-chained bus that connects each peripheral slot with adjacent peripheral slots to the left and right....
Page 17 - System Reference Clock
Chapter 1 Getting Started NI PXIe-1075 User Manual 1-10 ni.com The PXI trigger lines from adjacent PXI trigger bus segments can be routed in either direction across the PXI trigger bridges. Refer to Figure 1-5 for the connectivity diagram. This allows you to send trigger signals to, and receive trig...
Page 19 - System Timing Slot
Chapter 1 Getting Started NI PXIe-1075 User Manual 1-12 ni.com To synchronize the system to an external clock, you can drive PXI_CLK10 from an external source through the PXI_CLK10_IN pin on the System Timing Slot. Refer to Table B-5, XP4 Connector Pinout for the System Timing Slot , for the pinout....
Page 21 - Installation and Configuration; Safety Information; Chassis Grounding; Live Circuits; Explosive Atmosphere
© National Instruments Corporation 2-1 NI PXIe-1075 User Manual 2 Installation and Configuration This chapter describes how to prepare and operate the NI PXIe-1075 chassis. Before connecting the chassis to a power source, read this chapter and the Read Me First: Safety and Radio-Frequency Interferen...
Page 22 - Part Replacement; Modification; Chassis Cooling Considerations; Providing Adequate Clearance
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-2 ni.com • Part Replacement —Only service this equipment with parts that are exact replacements, both electrically and mechanically. Contact National Instruments for replacement part information. Installation of parts with those tha...
Page 24 - Chassis Ambient Temperature Definition
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-4 ni.com Figure 2-2. NI PXIe-1075 Vents Chassis Ambient Temperature Definition The chassis fan control system uses intake air temperature as the input for controlling fan speeds when in Auto Fan Speed mode. Because of this, the chas...
Page 25 - Setting Fan Speed; High; Installing Filler Panels; Rack Mounting
Chapter 2 Installation and Configuration © National Instruments Corporation 2-5 NI PXIe-1075 User Manual responsibility to ensure that this ambient temperature does not exceed the rated ambient temperature as stated in Appendix A, Specifications . If the temperature exceeds the stated spec the power...
Page 26 - Connecting to Power Source; Manual; Installing a PXI Express System Controller
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-6 ni.com If your power outlet does not have an appropriate ground connection, you must connect the premise safety ground to the chassis grounding screw located on the rear panel. Refer to Figure 1-2, Rear View of the NI PXIe-1075 Ch...
Page 29 - Installing Peripheral Modules
Chapter 2 Installation and Configuration © National Instruments Corporation 2-9 NI PXIe-1075 User Manual Installing Peripheral Modules Caution The NI PXIe-1075 chassis has been designed to accept a variety of peripheral module types in different slots. To prevent damage to the chassis, ensure that t...
Page 30 - Power Inhibit Switch LED Indicator
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-10 ni.com Figure 2-5. Installing PXI, PXI Express, or CompactPCI Peripheral Modules Power Inhibit Switch LED Indicator The chassis power inhibit switch has an integrated LED. Refer to Figure 1-1, Front View of the NI PXIe-1075 Chass...
Page 31 - Remote Voltage Monitoring and Control; Signal
Chapter 2 Installation and Configuration © National Instruments Corporation 2-11 NI PXIe-1075 User Manual Note If two system fans or both of the power supply fans fail the chassis will shut down automatically, preventing the chassis and modules from damage due to overheating. Remote Voltage Monitori...
Page 32 - Inhibit Mode Switch
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-12 ni.com You can use a digital voltmeter to ensure all voltage levels in the NI PXIe-1075 chassis are within the allowable limits. Referring to Table 2-2, connect one lead of the voltmeter to a supply pin on the remote voltage moni...
Page 33 - PXI_CLK10 Rear Connectors; PXI Express System Configuration with MAX
Chapter 2 Installation and Configuration © National Instruments Corporation 2-13 NI PXIe-1075 User Manual (pin 5) to a Logic Ground pin (pin 1 or 9). As long as this connection exists, the chassis will remain off (standby); when you remove this connection, the chassis turns on. Note For the Remote I...
Page 34 - PXI-1 System Configuration; Configuration; Identify As
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-14 ni.com Figure 2-6. Multichassis Configuration in MAX PXI-1 System Configuration 1. Launch MAX. 2. In the Configuration tree, click the Devices and Interfaces branch to expand it. 3. If the PXI system controller has not yet been c...
Page 35 - Trigger Configuration in MAX; Triggers
Chapter 2 Installation and Configuration © National Instruments Corporation 2-15 NI PXIe-1075 User Manual 4. Click the PXI system controller. The chassis (or multiple chassis, in a multichassis configuration) is listed below it. Identify each chassis by right-clicking its entry, then selecting the a...
Page 36 - PXI Trigger Bus Routing; Using System Configuration and Initialization Files
Chapter 2 Installation and Configuration NI PXIe-1075 User Manual 2-16 ni.com Complete the following steps to reserve these trigger lines in MAX. 1. In the Configuration tree, click on the PXI chassis branch you want to configure. 2. Then, in the right-hand pane, toward the bottom, click on the Trig...
Page 38 - Maintenance; Service Interval; Cleaning
© National Instruments Corporation 3-1 NI PXIe-1075 User Manual 3 Maintenance This chapter describes basic maintenance procedures you can perform on the NI PXIe-1075 chassis. Caution Disconnect the power cable prior to servicing a NI PXIe-1075 chassis. Service Interval Clean the chassis fan filters ...
Page 39 - Interior Cleaning; Cleaning the Fan Filters
Chapter 3 Maintenance NI PXIe-1075 User Manual 3-2 ni.com Caution Always disconnect the AC power cable before cleaning or servicing the chassis. Interior Cleaning Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a soft-bristle brush for cleaning around components. Exte...
Page 40 - Resetting the AC Mains Circuit Breaker; Replacing the Modular Power Supply Shuttle
Chapter 3 Maintenance © National Instruments Corporation 3-3 NI PXIe-1075 User Manual Resetting the AC Mains Circuit Breaker If the NI PXIe-1075 chassis is connected to an AC source and encounters an over-current condition, the circuit breaker on the rear panel will trip to prevent damage to the cha...
Page 41 - Removal
Chapter 3 Maintenance NI PXIe-1075 User Manual 3-4 ni.com Before connecting the power supply shuttle to a power source, read this section and the Read Me First: Safety and Radio-Frequency Interference document included with the kit. Removal The NI PXIe-1075 AC power supply shuttle is a replacement p...
Page 42 - Electrical; AC Input
© National Instruments Corporation A-1 NI PXIe-1075 User Manual A Specifications This appendix contains specifications for the NI PXIe-1075 chassis. Caution Specifications are subject to change without notice. Electrical AC Input Input voltage rating ................................ 100 to 120 VAC, ...
Page 43 - DC Output; Voltage
Appendix A Specifications NI PXIe-1075 User Manual A-2 ni.com Power disconnect ....................................The AC power cable provides main power disconnect. The front-panel power switch causes the internal chassis power supply to provide DC power to the CompactPCI/PXI Express backplane. You...
Page 44 - Slot
Appendix A Specifications © National Instruments Corporation A-3 NI PXIe-1075 User Manual Backplane pin current capacity Load regulation Maximum ripple and noise (20 MHz bandwidth) Over-current protection .......................... All outputs protected from short circuit and overload with automatic...
Page 45 - Chassis Cooling; Environmental
Appendix A Specifications NI PXIe-1075 User Manual A-4 ni.com Chassis Cooling Module cooling system ...........................Forced air circulation (positive pressurization) through three 165 cfm fans with High/Auto speed selector Slot airflow direction ..............................Bottom of modu...
Page 46 - Operating Environment; Acoustic Emissions
Appendix A Specifications © National Instruments Corporation A-5 NI PXIe-1075 User Manual Operating Environment Ambient temperature range.................... 0 to 55 °C (Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F C...
Page 47 - Sound Power; Safety; Electromagnetic Compatibility
Appendix A Specifications NI PXIe-1075 User Manual A-6 ni.com Sound Power Auto fan (up to ~30 °C ambient) ............55.5 dBA High fan ..................................................76.2 dBA Notes For EMC compliance, operate this device with shielded cabling. In addition, all covers and filler p...
Page 48 - CE Compliance; Environmental Management; Waste Electrical and Electronic Equipment (WEEE); ⬉ᄤֵᙃѻક∵ᵻࠊᄣ⧚ࡲ⊩ ˄Ё; RoHS
Appendix A Specifications © National Instruments Corporation A-7 NI PXIe-1075 User Manual CE Compliance This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows: • 2006/95/EC; Low-Voltage Directive (safety) • 2004/108/EC; Electromagnetic ...
Page 49 - Backplane
Appendix A Specifications NI PXIe-1075 User Manual A-8 ni.com Backplane Size .........................................................3U-sized; one system slot (with three system expansion slots) and 17 peripheral slots.Compliant with IEEE 1101.10 mechanical packaging.PXI Express Specification compl...
Page 51 - External Clock Source; PXI Star Trigger
Appendix A Specifications NI PXIe-1075 User Manual A-10 ni.com External Clock Source Frequency ...............................................10 MHz ±100 PPM Input amplitude Rear panel BNC...............................200 mV PP to 5 V PP square-wave or sine-wave System timing slotPXI_CLK10_IN ........
Page 52 - Mechanical
Appendix A Specifications © National Instruments Corporation A-11 NI PXIe-1075 User Manual Notes For PXIe slot to PXI_DSTAR mapping refer to the System Timing Slot section of Chapter 1, Getting Started . For other specifications, the NI PXIe-1075 complies with the PXI-5 PXI Express Hardware Specific...
Page 56 - Pinouts
© National Instruments Corporation B-1 NI PXIe-1075 User Manual B Pinouts This appendix describes the connector pinouts for the NI PXIe-1075 chassis backplane. Figure B-1 illustrates the types of PXIe connectors by providing a layout of the PXI Express system controller slot (slot 1). Table B-1 show...
Page 57 - System Controller Slot Pinouts
Appendix B Pinouts NI PXIe-1075 User Manual B-2 ni.com System Controller Slot Pinouts Figure B-1. PXI Express System Controller Slot Layout 1 XP4 Connector 2 XP3 Connector 3 XP2 Connector 4 XP1 Connector Table B-1. XP4 Connector Pinout for the System Controller Slot Pin Z A B C D E F 1 GND GA4 GA3 G...
Page 59 - System Timing Slot Pinouts
Appendix B Pinouts NI PXIe-1075 User Manual B-4 ni.com System Timing Slot Pinouts Figure B-2. PXI Express System Timing Slot Layout Table B-4. XP1 Connector Pinout for the System Controller Slot Pins Signals A GND B 12V C 12V D GND E 5V F 3.3V G GND 1 XP4 Connector 2 XP3 Connector 3 TP2 Connector 4 ...
Page 63 - Hybrid Slot Pinouts
Appendix B Pinouts NI PXIe-1075 User Manual B-8 ni.com Hybrid Slot Pinouts Figure B-3. PXI Express System Hybrid Slot Layout 1 XP4 Connector 2 XP3 Connector 3 P1 Connector Table B-9. XP4 Connector Pinout for the Hybrid Slot Pin Z A B C D E F 1 GND GA4 GA3 GA2 GA1 GA0 GND 2 GND 5Vaux GND SYSEN# WAKE#...
Page 68 - Symbol; Symbols
© National Instruments Corporation G-1 NI PXIe-1075 User Manual Glossary Symbol Prefix Value p pico 10 – 12 n nano 10 – 9 μ micro 10 – 6 m milli 10 – 3 k kilo 10 3 M mega 10 6 G giga 10 9 T tera 10 12 Symbols ° Degrees. ≥ Equal or greater than. ≤ Equal or less than. % Percent. A A Amperes. AC Altern...
Page 74 - Index
© National Instruments Corporation I-1 NI PXIe-1075 User Manual Index A AC power cables part numbers (table), 1-2 AC power cables (table), 1-2architecture, backplane, 1-6 B backplane architecture, 1-6hybrid peripheral slots, 1-7interoperability with CompactPCI, 1-6overview, 1-6PXI Express peripheral...