Important Information Warranty The DAQCard E Series cards are warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves ...
National Instruments Corporation v DAQCard E Series User Manual Table of Contents About This Manual Organization of This Manual ........................................................................................xiConventions Used in This Manual ...................................................
Table of Contents DAQCard E Series User Manual vi National Instruments Corporation Digital I/O .....................................................................................................................3-12Timing Signal Routing ...............................................................
Table of Contents DAQCard E Series User Manual viii National Instruments Corporation Appendix FCustomer Communication Glossary Index
Table of Contents National Instruments Corporation ix DAQCard E Series User Manual Figures Figure 1-1. The Relationship between the Programming Environment, NI-DAQ, and Your Hardware ................................................................................1-4 Figure 2-1. A Typical Configura...
Table of Contents DAQCard E Series User Manual x National Instruments Corporation Figure 4-26. GPCTR0_OUT Signal Timing ................................................................4-39Figure 4-27. GPCTR1_SOURCE Signal Timing .........................................................4-40Figure 4...
National Instruments Corporation xi DAQCard E Series User Manual About This Manual This manual describes the electrical and mechanical aspects of each card in the DAQCard E Series product line and contains information concerning their operation and programming. Unless otherwise noted, text applies...
About This Manual DAQCard E Series User Manual xii National Instruments Corporation • Appendix B, Optional Cable Connector Descriptions , describes the connectors on the optional cables for the DAQCard E Series cards. • Appendix C, PC Card Questions and Answers , contains a list of common question...
About This Manual National Instruments Corporation xiii DAQCard E Series User Manual names, functions, operations, variables, filenames, and extensions, and for statements and comments taken from program code. NI-DAQ NI-DAQ refers to NI-DAQ software unless otherwise noted. PC Card PC Card refers t...
About This Manual DAQCard E Series User Manual xiv National Instruments Corporation • Accessory installation guides or manuals—If you are using accessory products, read the terminal block and cable assembly installation guides. They explain how to physically connect the relevant pieces of the syst...
National Instruments Corporation 1-1 DAQCard E Series User Manual Introduction Chapter 1 This chapter describes the DAQCard E Series cards, lists what you need to get started, describes the optional software and optional equipment, and explains how to unpack your DAQCard E Series card. About the D...
Chapter 1 Introduction DAQCard E Series User Manual 1-2 National Instruments Corporation What You Need to Get Started To set up and use your DAQCard E Series card, you will need the following: ❏ One of the following cards: DAQCard-AI-16E-4DAQCard-AI-16XE-50 ❏ DAQCard E Series User Manual ❏ One of ...
Chapter 1 Introduction National Instruments Corporation 1-3 DAQCard E Series User Manual LabWindows/CVI Data Acquisition Library, a series of functions for using LabWindows/CVI with National Instruments DAQ hardware, is included with the NI-DAQ software kit. The LabWindows/CVI Data Acquisition Lib...
Chapter 1 Introduction DAQCard E Series User Manual 1-4 National Instruments Corporation LabWindows/CVI, your application uses the NI-DAQ driver software, as illustrated in Figure 1-1. You can use your DAQCard E Series card, together with other PC, AT, EISA, DAQCard, and DAQPad Series DAQ and SCXI...
Chapter 1 Introduction National Instruments Corporation 1-5 DAQCard E Series User Manual Optional Equipment National Instruments offers a variety of products to use with your DAQCard E Series card, including cables, connector blocks, and other accessories, as follows: • Cables and cable assemblies...
Chapter 1 Introduction DAQCard E Series User Manual 1-6 National Instruments Corporation The following list gives recommended National Instruments cable assemblies that mate to your DAQCard I/O connector. ♦ DAQCard-AI-16E-4 PSHR68-68M, a shielded 68-position ribbon cable, with male-to-male connect...
National Instruments Corporation 2-1 DAQCard E Series User Manual Installation and Configuration Chapter 2 This chapter explains how to install and configure a DAQCard E Series card. Installation Note: You should install your driver software before installing your hardware. Refer to your NI-DAQ re...
Chapter 2 Installation and Configuration DAQCard E Series User Manual 2-2 National Instruments Corporation Figure 2-1 shows an example of a typical configuration. Configuration Your DAQCard is completely software-configurable. Refer to your software documentation to install and configure your soft...
Chapter 2 Installation and Configuration National Instruments Corporation 2-3 DAQCard E Series User Manual refer to the NI-DAQ release notes and follow the instructions given there for your operating system and LabVIEW. If you are using LabWindows/CVI, refer to your LabWindows/CVI release notes to...
National Instruments Corporation 3-1 DAQCard E Series User Manual Hardware Overview Chapter 3 This chapter presents an overview of the hardware functions on your DAQCard E Series card. Figure 3-1 shows the block diagram for the DAQCard-AI-16E-4. Figure 3-1. DAQCard-AI-16E-4 Block Diagram Timing PF...
Chapter 3 Hardware Overview DAQCard E Series User Manual 3-2 National Instruments Corporation Figure 3-2 shows a block diagram for the DAQCard-AI-16XE-50. Analog Input The analog input section of each DAQCard is software configurable. You can select different analog input configurations through ap...
Chapter 3 Hardware Overview National Instruments Corporation 3-3 DAQCard E Series User Manual single-ended channels. Table 3-1 describes the three input configurations. For more information about the three types of input configuration, refer to the Analog Input Signal Connections section in Chapte...
Chapter 3 Hardware Overview DAQCard E Series User Manual 3-4 National Instruments Corporation The software-programmable gain on these cards increases their overall flexibility by matching the input signal ranges to those that the ADC can accommodate. The DAQCard-AI-16E-4 has gains of 0.5, 1, 2, 5,...
Chapter 3 Hardware Overview DAQCard E Series User Manual 3-6 National Instruments Corporation Considerations for Selecting Input Ranges Which input polarity and range you select depends on the expected range of the incoming signal. A large input range can accommodate a large signal variation but r...
Chapter 3 Hardware Overview National Instruments Corporation 3-7 DAQCard E Series User Manual You cannot disable dither on the DAQCard-AI-16XE-50. This is because the ADC resolution is so fine that the ADC and the PGIA inherently produce more than 0.5 LSB rms of noise. This is equivalent to having...
Chapter 3 Hardware Overview DAQCard E Series User Manual 3-8 National Instruments Corporation impedances are low. Refer to Appendix A, Specifications , for a complete listing of settling times for each of the DAQCards. When scanning among channels at various gains, the settling times may increase....
Chapter 3 Hardware Overview National Instruments Corporation 3-9 DAQCard E Series User Manual Analog Trigger ♦ DAQCard-AI-16E-4 In addition to supporting internal software triggering and external digital triggering to initiate a data acquisition sequence, the DAQCard-AI-16E-4 also supports analog ...
Chapter 3 Hardware Overview National Instruments Corporation 3-11 DAQCard E Series User Manual In inside-region analog triggering mode, the trigger is generated when the signal value is between the lowValue and the highValue . In high-hysteresis analog triggering mode, the trigger is generated whe...
Chapter 3 Hardware Overview National Instruments Corporation 3-13 DAQCard E Series User Manual Timing Signal Routing The DAQ-STC provides a very flexible interface for connecting timing signals to other boards or external circuitry. Your DAQCard uses the Programmable Function Input (PFI) pins on t...
Chapter 3 Hardware Overview DAQCard E Series User Manual 3-14 National Instruments Corporation Programmable Function Inputs The 10 PFIs are connected to the signal routing multiplexer for each timing signal, and software can select one of the PFIs as the external source for a given timing signal. ...
National Instruments Corporation 4-1 DAQCard E Series User Manual Signal Connections Chapter 4 This chapter describes how to make input and output signal connections to your DAQCard E Series card via the DAQCard I/O connector. The I/O connector for the DAQCard E Series cards has 68 pins that you c...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-8 National Instruments Corporation Table 4-3 shows the I/O signal summary for the DAQCard-AI-16XE-50. Table 4-3. I/O Signal Summary, DAQCard-AI-16XE-50 Signal Name Drive Impedance Input/ Output Protection (Volts) On/Off Source (mA at V) S...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-10 National Instruments Corporation Analog Input Signal Connections The analog input signals are ACH<0..15>, AISENSE, and AIGND. The ACH<0..15> signals are tied to the 16 analog input channels of your DAQCard. In single-ended ...
Chapter 4 Signal Connections National Instruments Corporation 4-11 DAQCard E Series User Manual the PGIA in different ways. Figure 4-2 shows a diagram of your DAQCard PGIA. The PGIA applies gain and common-mode voltage rejection and presents high input impedance to the analog input signals connect...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-12 National Instruments Corporation Types of Signal Sources When configuring the input channels and making signal connections, you must first determine whether the signal sources are floating or ground-referenced. The following sections d...
Chapter 4 Signal Connections National Instruments Corporation 4-13 DAQCard E Series User Manual Figure 4-3 summarizes the recommended input configuration for both types of signal sources. Figure 4-3. Summary of Analog Input Connections Input Differential (DIFF) Single-Ended — Ground Referenced (RS...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-14 National Instruments Corporation Differential Connection Considerations (DIFF Input Configuration) A differential connection is one in which the DAQCard analog input signal has its own reference signal or signal return path. These conn...
Chapter 4 Signal Connections National Instruments Corporation 4-15 DAQCard E Series User Manual Differential Connections for Ground-Referenced Signal Sources Figure 4-4 shows how to connect a ground-referenced signal source to a channel on a DAQCard configured in DIFF input mode. With this type of...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-16 National Instruments Corporation Differential Connections for Nonreferenced or Floating Signal Sources Figure 4-5 shows how to connect a floating signal source to a channel on a DAQCard configured in DIFF input mode. Figure 4-5 shows t...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-18 National Instruments Corporation Single-Ended Connection Considerations A single-ended connection is one in which the DAQCard E Series card analog input signal is referenced to a ground that can be shared with other input signals. The ...
Chapter 4 Signal Connections National Instruments Corporation 4-19 DAQCard E Series User Manual Single-Ended Connections for Floating Signal Sources (RSE Configuration) Figures 4-6 shows how to connect a floating signal source to a channel on a DAQCard configured for RSE mode. Single-Ended Connect...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-20 National Instruments Corporation Figure 4-7 shows how to connect a grounded signal source to a channel on a DAQCard configured for NRSE mode. Common-Mode Signal Rejection Considerations Figures 4-4 and 4-7 show connections for signal s...
Chapter 4 Signal Connections National Instruments Corporation 4-21 DAQCard E Series User Manual Digital I/O Signal Connections The digital I/O signals are DIO<0..7> and DGND. The DIO<0..7> signals make up the DIO port, and DGND is the ground reference signal for this port. You can prog...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-22 National Instruments Corporation Figure 4-8 shows DIO<0..3> configured for digital input and DIO<4..7> configured for digital output. Digital input applications include receiving TTL signals and sensing external device stat...
Chapter 4 Signal Connections National Instruments Corporation 4-23 DAQCard E Series User Manual The data acquisition signals are explained in the Data Acquisition Timing Connections section later in this chapter. The general-purpose timing signals are explained in the General-Purpose Timing Signal...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-24 National Instruments Corporation Programmable Function Input Connections There are a total of 13 internal timing signals that you can externally control from the PFI pins. The source for each of these signals is software selectable fro...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-26 National Instruments Corporation SCANCLK Signal SCANCLK is an output-only signal that generates a pulse with the leading edge occurring approximately 50 to 100 ns after an A/D conversion begins. The polarity of this output is software-...
Chapter 4 Signal Connections National Instruments Corporation 4-27 DAQCard E Series User Manual EXTSTROBE* Signal EXTSTROBE* is an output-only signal that generates either a single pulse or a sequence of eight pulses in the hardware-strobe mode. An external device can use this signal to latch sign...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-28 National Instruments Corporation Figures 4-14 and 4-15 show the input and output timing requirements for the TRIG1 signal. The DAQCard also uses the TRIG1 signal to initiate pretriggered data acquisition operations. In most pretriggere...
Chapter 4 Signal Connections National Instruments Corporation 4-29 DAQCard E Series User Manual TRIG2 Signal Any PFI pin can externally input the TRIG2 signal, which is available as an output on the PFI1/TRIG2 pin. Refer to Figure 4-11 for the relationship of TRIG2 to the data acquisition sequence...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-30 National Instruments Corporation Figures 4-16 and 4-17 show the input and output timing requirements for the TRIG2 signal. STARTSCAN Signal Any PFI pin can externally input the STARTSCAN signal, which is available as an output on the P...
Chapter 4 Signal Connections National Instruments Corporation 4-31 DAQCard E Series User Manual scan. The sample interval counter is started if you select internally triggered CONVERT*. As an output, the STARTSCAN signal reflects the actual start pulse that initiates a scan. This is true even if t...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-32 National Instruments Corporation The CONVERT* pulses are masked off until the DAQCard generates the STARTSCAN signal. If you are using internally generated conversions, the first CONVERT* will appear when the onboard sample interval co...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-34 National Instruments Corporation The ADC switches to hold mode within 60 ns of the selected edge. This hold-mode delay time is a function of temperature and does not vary from one conversion to the next. Separate the CONVERT* pulses by...
Chapter 4 Signal Connections National Instruments Corporation 4-35 DAQCard E Series User Manual mode, the first active edge disables the STARTSCAN signal, and the second active edge enables STARTSCAN. The AIGATE signal can neither stop a scan in progress nor continue a previously gated-off scan; i...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-36 National Instruments Corporation UISOURCE Signal Any PFI pin can externally input the UISOURCE signal, which is not available as an output on the I/O connector. The UI counter uses the UISOURCE signal as a clock to time the generation ...
Chapter 4 Signal Connections National Instruments Corporation 4-37 DAQCard E Series User Manual General-Purpose Timing Signal Connections The general-purpose timing signals are GPCTR0_SOURCE, GPCTR0_GATE, GPCTR0_OUT, GPCTR0_UP_DOWN, GPCTR1_SOURCE, GPCTR1_GATE, GPCTR1_OUT, GPCTR1_UP_DOWN, and FREQ_...
Chapter 4 Signal Connections National Instruments Corporation 4-39 DAQCard E Series User Manual input (High-Z) at startup. Figure 4-26 shows the timing of the GPCTR0_OUT signal. GPCTR0_UP_DOWN Signal This signal can be externally input on the DIO6 pin and is not available as an output on the I/O c...
Chapter 4 Signal Connections National Instruments Corporation 4-41 DAQCard E Series User Manual Figure 4-28 shows the timing requirements for the GPCTR1_GATE signal. GPCTR1_OUT Signal This signal is available only as an output on the GPCTR1_OUT pin. The GPCTR1_OUT signal monitors the TC board gene...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-42 National Instruments Corporation GPCTR1_UP_DOWN Signal This signal can be externally input on the DIO7 pin and is not available as an output on the I/O connector. General-purpose counter 1 counts down when this pin is at a logic low an...
Chapter 4 Signal Connections National Instruments Corporation 4-43 DAQCard E Series User Manual DAQCard. Figure 4-30 shows the GATE signal referenced to the rising edge of a source signal. The gate must be valid (either high or low) for at least 10 ns before the rising or falling edge of a source ...
Chapter 4 Signal Connections DAQCard E Series User Manual 4-44 National Instruments Corporation Field Wiring Considerations Environmental noise can seriously affect the accuracy of measurements made with your DAQCard if you do not take proper care when running signal wires between signal sources a...
National Instruments Corporation 5-1 DAQCard E Series User Manual Calibration Chapter 5 This chapter discusses the calibration procedures for your DAQCard E Series card. If you are using the NI-DAQ device driver, that software includes calibration functions for performing all of the steps in the c...
Chapter 5 Calibration DAQCard E Series User Manual 5-2 National Instruments Corporation the CalDACs with values either from the original factory calibration or from a calibration that you subsequently performed. This method of calibration is not very accurate because it does not take into account ...
Chapter 5 Calibration National Instruments Corporation 5-3 DAQCard E Series User Manual perform an external calibration very often. You can externally calibrate your DAQCard by calling the NI-DAQ calibration function. To externally calibrate your DAQCard, use a very accurate external reference. Th...
National Instruments Corporation A-1 DAQCard E Series User Manual Specifications Appendix A This appendix lists the specifications of each DAQCard in the DAQCard E Series. These specifications are typical at 25 ° C unless otherwise noted. DAQCard-AI-16E-4 Analog Input Input Characteristics Number ...
Appendix A Specifications for DAQCard-AI-16E-4 DAQCard E Series User Manual A-2 National Instruments Corporation Input coupling ................................... DC Max working voltage (signal + common mode) ................ Each input should remain within ± 11 V of ground Overvoltage protection...
Appendix A Specifications for DAQCard-AI-16E-4 National Instruments Corporation A-3 DAQCard E Series User Manual Transfer Characteristics Relative accuracy ............................. ± 0.5 LSB typ dithered, ± 1.5 LSB max undithered DNL .................................................. ± 0.5 LS...
Appendix A Specifications for DAQCard-AI-16E-4 DAQCard E Series User Manual A-4 National Instruments Corporation Crosstalk ........................................... -80 dB, DC to 100 kHz Stability Recommended warm-up time ............ 15 min Offset temperature coefficient Pregain ..................
Appendix A Specifications for DAQCard-AI-16E-4 National Instruments Corporation A-5 DAQCard E Series User Manual Digital I/O Number of channels .......................... 8 input/output Compatibility .................................... TTL/CMOS Power-on state ................................... I...
Appendix A Specifications for DAQCard-AI-16E-4 DAQCard E Series User Manual A-6 National Instruments Corporation Triggers Analog Trigger Source ............................................... ACH<0..15>, external trigger (PFI0/TRIG1) Level ................................................. ± ...
Appendix A Specifications for DAQCard-AI-16E-4 National Instruments Corporation A-7 DAQCard E Series User Manual Note: These power usage figures do not include the power used by external devices that are connected to the fused supply present on the I/O connector. Note also that under ordinary oper...
National Instruments Corporation A-8 DAQCard E Series User Manual DAQCard-AI-16XE-50 Analog Input Input Characteristics Number of channels ...........................16 single-ended or 8 differential (software-selectable) Type of ADC .....................................Successive approximation Re...
Appendix A Specifications for DAQCard-AI-16XE-50 DAQCard E Series User Manual A-10 National Instruments Corporation Dynamic Characteristics Bandwidth Gain = 1, 2 .................................. 69 kHz Gain = 10 .................................... 66 kHz Gain = 100 ................................
National Instruments Corporation B-1 DAQCard E Series User Manual Optional Cable ConnectorDescriptions Appendix B This appendix describes the connectors on the optional cables for the DAQCard E Series cards. Figure B-1 shows the pin assignments for the 68-pin AI connector. This connector is availa...
National Instruments Corporation C-1 DAQCard E Series User Manual PC Card Questions and Answers Appendix C This appendix contains a list of common questions and answers relating to PC Card (PCMCIA) operation. The questions are grouped according to the type of information requested. You may find th...
Appendix C PC Card Questions and Answers DAQCard E Series User Manual C-2 National Instruments Corporation Operation 1. My PC Card works when inserted before power-on time, but it does not work when hot inserted. What is wrong? You may have an interrupt conflict. If you have a utility such as MSD....
Appendix C PC Card Questions and Answers National Instruments Corporation C-3 DAQCard E Series User Manual by Card Services and/or the memory manager. Second, you can attempt to determine all of the memory that Card Services can possibly use and then exclude all but that memory from use by Card Se...
Appendix C PC Card Questions and Answers DAQCard E Series User Manual C-4 National Instruments Corporation managers often consume an enormous amount of memory, and you will need to determine what memory is really usable by Card Services. When you have determined what memory is available for Card S...
National Instruments Corporation D-1 DAQCard E Series User Manual Common Questions Appendix D This appendix contains a list of commonly asked questions and their answers relating to usage and special features of your DAQCard E Series card. General Information 1. What are the DAQCard E Series cards...
Appendix D Common Questions DAQCard E Series User Manual D-2 National Instruments Corporation channel at 250 kS/s or two channels at 125 kS/s per channel illustrates the relationship. Notice, however, that some DAQCard E Series cards have settling times that vary with gain and accuracy. See Append...
Appendix D Common Questions National Instruments Corporation D-3 DAQCard E Series User Manual Analog Input and Output 1. I’m using my DAQCard in differential analog input mode and I have connected a differential input signal, but my readings are random and drift rapidly. What’s wrong? Check your g...
Appendix D Common Questions DAQCard E Series User Manual D-4 National Instruments Corporation 3. What is the difference in timebases between the Am9513 counter/timer and the DAQ-STC? The DAQ-STC-based boards have a 20 MHz timebase. The Am9513-based boards have a 1 or 5 MHz timebase. 4. The counter...
Appendix D Common Questions National Instruments Corporation D-5 DAQCard E Series User Manual 7. How does NI-DAQ treat bogus missed data transfer errors that can arise during DMA-driven GPCTR buffered-input operations? When doing buffered transfers using GPCTR function calls with DMA, you can call...
National Instruments Corporation E-1 DAQCard E Series User Manual Power-Management Modes Appendix E This appendix describes the power-management modes of the DAQCard E Series cards. • Normal Mode—This is the normal operating mode of the DAQCard E Series cards in which all the circuits are fully fu...
Appendix E Power-Management Modes DAQCard E Series User Manual E-2 National Instruments Corporation Table E-1. DAQCard E Series Power-Management Modes Normal Mode Power-Down Mode Analog Input Functional. Protected to ± 25 V. Impedance > 1 G Ω . Nonfunctional. Draws negligible power. Protected t...
National Instruments Corporation F-1 DAQCard E Series User Manual Customer Communication Appendix F For your convenience, this appendix contains forms to help you gather the information necessary to help us solve your technical problems and a form you can use to comment on the product documentatio...
FaxBack is a 24-hour information retrieval system containing a library of documents on a wide range of technical information. You can access FaxBack from a touch-tone telephone at (512) 418-1111. You can submit technical support questions to the appropriate applications engineering team through e-ma...
Technical Support Form Photocopy this form and update it each time you make changes to your software or hardware, and use the completed copy of this form as a reference for your current configuration. Completing this form accurately before contacting National Instruments for technical support helps ...
Documentation Comment Form National Instruments encourages you to comment on the documentation supplied with our products. This information helps us provide quality products to meet your needs. Title: DAQCard E Series User Manual Edition Date: June 1996 Part Number: 321138A-01 Please comment on the ...
National Instruments Corporation G-1 DAQCard E Series User Manual Glossary Symbols ˚ degrees – negative of, or minus Ω ohms / per % percent ± plus or minus + positive of, or plus square root of +5 V +5 VDC source signal Prefix Meaning Value p- pico- 10 - 12 n- nano- 10 - 9 µ - micro- 10 - 6 m- mil...
Glossary DAQCard E Series User Manual G-8 National Instruments Corporation V in volts in V O volts, output V OH volts, output high V OL volts, output low V ref reference voltage W W watts WFTRIG waveform generation trigger signal
National Instruments Corporation I -1 DAQCard E Series User Manual Index Numbers +5 V signal description, 4-3power connections, 4-22 A ACH<0..15> signal analog input connections, 4-10DAQCard-AI-16E-4 (table), 4-5DAQCard-AI-16XE-50 (table), 4-8description, 4-3 AIGATE signal, 4-34 to 4-35AIGND...
Index National Instruments Corporation I -3 DAQCard E Series User Manual ground-referenced signal sources, 4-15 illustration, 4-15 nonreferenced or floating signal sources, 4-16 to 4-17 illustration, 4-16 recommended configuration (figure), 4-13 single-ended connections, 4-18 floating signal sourc...
Index DAQCard E Series User Manual I -4 National Instruments Corporation G general-purpose timing signal connections, 4-37 to 4-43 FREQ_OUT signal, 4-43GPCTR0_GATE signal, 4-38GPCTR0_OUT signal, 4-38 to 4-39GPCTR0_SOURCE signal, 4-37 to 4-38GPCTR0_UP_DOWN signal, 4-39GPCTR1_GATE signal, 4-40 to 4-...
Index National Instruments Corporation I -5 DAQCard E Series User Manual differential connections DIFF input configuration, 4-14floating signal sources, 4-16 to 4-17ground-referenced signal sources, 4-15 nonreferenced signal sources, 4-16 to 4-17 recommended configuration (figure), 4-13 single-end...
Index DAQCard E Series User Manual I -6 National Instruments Corporation PFI0/TRIG1 signal DAQCard-AI-16E-4 (table), 4-6DAQCard-AI-16XE-50 (table), 4-8description, 4-3 PFI1/TRIG2 signal DAQCard-AI-16E-4 (table), 4-6DAQCard-AI-16XE-50 (table), 4-8description, 4-4 PFI2/CONVERT* signal DAQCard-AI-16E...
Index National Instruments Corporation I -7 DAQCard E Series User Manual programmable function inputs (PFIs). See PFIs (programmable function inputs) . programmable gain instrumentation amplifier. See PGIA (programmable gain instrumentation amplifier) . Q questions and answers DAQCard E Series que...
Index National Instruments Corporation I -9 DAQCard E Series User Manual T technical support, F-1 to F-2temperature DAQCard-AI-16E-4, A-7DAQCard-AI-16XE-50, A-12 theory of operation. See hardware overview . timebases, 3-14timing connections, 4-22 to 4-43 data acquisition timing connections, 4-24 t...
Index DAQCard E Series User Manual I -10 National Instruments Corporation U UISOURCE signal, 4-36unipolar input DAQCard-AI-16E-4, 3-3DAQCard-AI-16XE-50, 3-4 to 3-5mixing bipolar and unipolar channels (note), 3-5 unpacking DAQCard E Series, 1-6 V VCC signal DAQCard-AI-16E-4 (table), 4-6DAQCard-AI-1...
National Instruments Manuals
-
National Instruments SCC-AI01
Manual
-
National Instruments USB-6008/6009
Manual
-
National Instruments 7334
Manual
-
National Instruments NI 9403E
Manual
-
National Instruments DIO 6533
Manual
-
National Instruments NI PCI-5114
Manual
-
National Instruments 1000DC
Manual
-
National Instruments 321518A-01
Manual
-
National Instruments NI 5672
Manual
-
National Instruments USB-9221
Manual
-
National Instruments PCI
Manual
-
National Instruments NI 9229
Manual
-
National Instruments 6211
Manual
-
National Instruments NI 9217
Manual
-
National Instruments NI USB-6009
Manual
-
National Instruments NI sbRIO-960x
Manual
-
National Instruments 6024E
Manual
-
National Instruments 3.x
Manual
-
National Instruments NI 784xR
Manual
-
National Instruments NI-Motion
Manual