Leica DMI3000B - Manuals
Leica DMI3000B – Manual in PDF format online.
Manuals:
Manual Leica DMI3000B
Summary
3 Leica DMI3000 BLeica DMI4000 BLeica DMI6000 B Operating Manual
4 Copyrights Copyrights All rights to this documentation are held byLeica Microsystems Wetzlar GmbH. Reproduc-tion of text or illustrations (in whole or in part) byprint, photocopy, microfilm or other method (in-cluding electronic systems) is not allowed with-out express written permission from Leic...
5 Contents 6.11 Installation of Lamp Housing Mount and Mirror Housing ................................... 42 6.12 Installation and Replacement of Incident-Light Lamps ............................ 44 6.13 Equipping the Incident Light Turret Disk ......................... 48 6.14 Inserting the Front M...
7 1. Important Notes about this Manual (1.2) → p. 20 ! * Numbers in parentheses, such as “(1.2)”, corre-spond to illustrations (in the example, Figure 1,Item 2). Numbers with pointer arrows (for example → p. 20), point to a certain page of this manual. Caution!Special safety instructions within this...
8 2. Intended Purpose of the Microscope 2. Intended Purpose of the Microscope The microscopes of the Leica DMI series cov-ered in this manual are designed for biologicalroutine and research applications. This includesthe examination of samples taken from the hu-man body with a view to providing info...
9 3. Safety Notes 3. Safety Notes 3.1 General Safety Notes This safety class 1 device is constructed andtested in accordance withEN 61010-2-101:2002, EN 61010-1:2001,IEC 1010-1:2001,Safety regulations for electrical measuring, con-trol, and laboratory devices. Caution! In order to maintain this cond...
12 4. Overview of the Instrument 4.1 Specifications 4. Overview of the Leica DMI Series Contrast Methods Transmitted Light Axis Leica DMI Series• transmitted light (DL): BF, DF, PH, DIC, Pol • intermediate pupil: IMC (integrated modulation contrast)IPH (Integrated phase contrast) Leica DMI4000 B and...
16 4. Overview of the Instrument Observation ports Controls Electronics box Leica DMI3000 B(a manual side port is a standard feature of the Leica DMI3000 Bstand)• manual• left side port (80% transmission) Leica DMI4000 B and Leica DMI6000 B• 7 fixed control buttons for illumination and apertures• 7 ...
17 4. Overview of the Instrument Leica DMI4000 B and Leica DMI6000 B• 2 x RS232C• 2 x USB• 4 x external/internal peripherals• CTR boxes• SmartMove Leica DMI4000 B and Leica DMI6000 B• Leica Application Suite (LAS) for Windows TM 2000, XP with plug-ins for:• microscope and camera configuration• micro...
26 5. Unpacking the Microscope The microscope is delivered in several pack-ages. The stand package contains the following com- ponents: • Stand with integrated incident-light axis, objective turret and tube • Illumination arm • Specimen stage • CD with Leica Application Suite (LAS) soft- ware packag...
27 5. Unpacking the Microscope Please carefully compare the contents of the de-livery to the packing slip, delivery note or in-voice. We urgently recommend storing a copy ofthese documents with the manual to ensure thatyou have information on the time and scope ofdelivery handy for subsequent orders...
29 6. Assembly 6. Assembling the Microscope The microscope components* are logically as-sembled in this order: • Transmitted-light illumination carrier• DIC module and DIC objective prisms• Condenser with condenser head• Eyepieces• Objectives• Transmitted-light lamps• Lamp housing mount (mirror hous...
30 6. Assembly 6.2 Installation of the transmitted-light illumination carrier (DL) Wipe the installation surface on the microscope(8.3) with a dry cloth. Tip the illumination carrier(8.1) back slightly and install it so that the pin(8.2) engages the groove in the support surface(8.4). Set the DL ill...
31 6. Assembly Fig. 12 IC objective prism 1 Objective prism in frame 2 Screw and washer Fig. 10b DIC objective prism turret (coded and motorized) 1 IC objective prism in frame 2 Identification letter (ID) 3 Orientation pin Fig. 11 Front cover, DIC prism disk 6.3 Installation of the DIC Moduleand DIC...
32 6. Assembly 6.4 Installation of Specimen Stages A wide range of specimen stages are available.The most important are the following: • Fixed stage (248 mm x 204 mm) (Fig. 13): normal, heating and temperature-controlled, with and without attachable mechanicalstage • Fixed micromanipulation stage (2...
33 6. Assembly Fig. 17 Fixed micromanipulation stage Fig. 18 Attachable mechanical stage for fixedmicromanipulation stage The assembly of these stages is identical. Thestages are solidly attached to the microscope bythree screws. In the case of fixed stages, an at-tachable mechanical stage may be in...
34 6. Assembly Fig. 19 a, b Inserts for attachable mechanical stage (fixed stage) • Next, firmly tighten the two rear screws. • Finally, give the front screw a final firm tight- ening. Fixed stage Attachable mechanical stages designed to ac-cept a variety of culture dishes are also avail-able for fi...
35 6. Assembly Manual fixed micromanipulation stage To install the attachable mechanical stage forthe manual fixed micromanipulation stage(Fig. 24), proceed as you would for the attach-able mechanical stage of the standard stage. The insert frames (Fig. 20a to c) differ at thispoint. These are held ...
36 6. Assembly Motorized 3-plate or scanning stages 3-plate stages and scanning stages: after in-stalling the stage, connect the included stagecable (for motorized stages) first to the socketon the stage, then to the CRT6000 or CTR6500box. The correct place on the box is called“XY Stage”. A variety ...
37 6. Assembly Fig. 311 Insert frame for coverslips 2 Leaf springs Rotating Stage and Insert Frames for Coverslips The rotating stage (Fig. 30) is also mounted with3 screws (30.2). Rotate the stage to make all ofthe threaded holes accessible. Insert thescrews (30.2). Caution: Use additional washers ...
38 6. Assembly 6.5 Installation of Condensers All condensers of the Leica DMI series areequipped with a 7-position turret disk that canbe equipped with light rings phase contrast (PH)or dark field (DF), IC prisms for transmitted-lightinterference contrast (DIC) or slit illuminators for integrated mo...
39 6. Assembly • Next, take the special condenser tool (Fig. 39.1). • If possible, install the light rings 0 to 3 in as- cending order. The numbering of the openingsis located at the edge of the crown gear(4 large openings: 1-4; 3 small openings: 5-7). • Grasp the light ring to be installed with the...
40 6. Assembly Please continue reading if you also have to in-stall IC prisms. Otherwise, skip to the next sec-tion. Installation of IC prisms • Switch the microscope off. • Remove the condenser cover (38.1). Insert the prism in one of the condenser disk’s large re-ceptacles with guide grooves. • Tu...
41 6. Assembly Caution: Do not press the spring clip down under anycircumstances. This can destroy the clip orresult in an unstable position of the prism. Turn the prism to ensure that it snaps into po-sition and release the tool.Remove fingerprints or dust from the prismwith care. • Use the left ce...
42 6. Assembly Fig. 42 Installation of condenser heads S1 1 Condenser base 2 Spacer ring 3 Condenser head Fig. 43 Installation of condenser head S28 Condenser heads Four different condenser heads are available: 1) S1/1.40 oil 2) S1/0.90 dry 3) S23/0.53 4) S28/0.55 Condenser heads 3 and 4 are screwed...
44 6. Assembly 6.8 Installation of filters in the illumination arm The Leica DMI series is equipped with a filtermagazine to accommodate two 40 mm dia. filtersas a standard feature. The filters are installed atthe factory. To change filters yourself, proceedas follows:• Release the screw (46.1) and ...
45 6. Assembly 6.10 Installation and replacement of the transmitted-light lamps: 107 or 107/2 lamphousing This lamp housing is used with a 12V 100W halo-gen lamp, which is already mounted.In case the lamp has to be removed: Changing the 12 V 100 W halogen lamp Caution! Ensure that the lamp housing h...
46 6. Assembly 6.11 Installing the lamp housing mount and mirror housing (Leica DMI4000 B andDMI6000 B) Place lamp housing mount (Fig. 53) or mirrorhousing on rear wall. Mount from front withsocket head screws. Fig. 54 Lamp housing 106z 1 Collector adjustment 2 Vertical lamp adjustment 3 Horizontal ...
47 6. Assembly Fig. 55 Booster lens Fig. 56 Booster lens in stand 1 Booster lens If a booster lens is included in the scope of de-livery, insert it into the rear stand opening at theleft or right, depending on the stand model. The booster slide has several positions:1. Slide pulled out: no effect 2....
48 6. Assembly 6.12 Installation and replacement of incident-light lamps Caution! Leica DMI4000 B and Leica DMI6000 B:Light sources pose a potential irradiationrisk (glare, UV-radiation, IR-radiation).Therefore, lamps have to be operated inclosed housings. Ensure that the lamp housing has been dis-c...
49 6. Assembly Fig. 60 Lamp mounts for Hg 100 gas discharge lamp 1 Uper clamping system 2 Lower clamping system 3 Cooling element a b Hg 100 Xe 75 3 1 2 3 1 2 4 Inserting gas discharge lamps (Hg and Xe) inthe 106z lamp housing Hg and Xe lamps are powered by separate sup-ply units.Please also read th...
52 6. Assembly 6.13 Equipping the Incident Light Turret Disc Caution: Please read this section completely before be-ginning with the assembly of the turret disk.Leica DMI4000 B and Leica DMI6000 B:The fluorescence drawer is located on the rightside of the stand. Before opening this drawer,remove the...
54 6. Assembly Fig. 70 Mechanical polarizer holder 1 Manual polarizer 2 Manual analyzer Fig. 71 Condenser with motorized polarizer 6.14 Inserting the Front Module Slider If your microscope is prepared for integratedmodulation contrast or integrated phase con-trast, a front module (possibly in conjun...
55 6. Assembly Fig. 73 Inserting the analyzer 1 Slot 2 Analyzer Analyzer for incident light and transmitted light. • Remove the cap (Fig. 72) on the right side of the stand (under the fluorescence drawer). • Insert the analyzer into the receptacle until it latches in place (Fig. 73.1). Fig. 72 Analy...
57 6. Assembly Fig. 78 Rear view of stand 1 RS232 ports 2 2 x USB 3 4 x EXT. 4 XYZ control for SmartMove 5 Electronic box connection 6 Condenser cable 7 Lamp power cable Fig. 77 Rear view of CTR6000 1 AC power socket 2 XY Stage socket for motorized stage 3 Direct interface socket optional 4 Z Contro...
58 6. Assembly CTR6000 and CTR6500 electronics box: Note: These electronics boxes must not be used withother stands. The serial number of the associatedstand has been recorded on the back of the elec-tronics box. A 3-axis control unit for focus and 3-plate crossstages is integrated in the CTR6000. A...
59 7. Start-up 7. Start-up 7.1 Functional Principle (Leica DMI4000 B and Leica DMI6000 B) 1. Intelligent automation • Switching between contrast methods at the touch of a button. Light rings, DIC prisms, etc. are automatically positioned in the beam path. • The microscope recognizes the selected obj...
63 7. Start-up If a component has not been installed correctly,the LeicaScreen will display an error message.See Troubleshooting chapter, → p. 102. Components such as diaphragms, condensers,light and phase rings have been pre-centered atthe factory. It may be necessary to correct thecentering after ...
65 7. Start-up Note: Changes to the light intensity as well as apertureand field diaphragm settings are stored for theindividual objectives and contrast methods. Fig. 84 Fixed function buttons (left side of stand) 1 Toggle transmitted light/incident light 2 Aperture diaphragm 3 light intensity 4 fie...
66 7. Start-up Variable function buttons on the stand The variable function buttons are assigned func-tions at the factory that are appropriate to thefeatures of your microscope. They are labeledaccordingly. For details on button assignments,please refer to the included identification sheet.For info...
67 7. Start-up Function buttons on the front panel (Fig. 87) 100% of the light goes to the eye-piece (87.1). Toggle function for the side ports(87.2). This function depends on theindividual microscope configuration.Note:Switching to the bottom port:via the variable function buttons(Leica DMI6000 B o...
68 7. Start-up Fig. 89 SmartMove remote control module 1 travel in x 2 Travel in y 3 Focus 4 Individual adjustment of button height 5 Variable function buttons (factory preset) 1 2 3 4 5 SET + Cube buttons 1-6(Leica DMI4000 B and DMI6000 B only) The SET command can be assigned to a functionbutton of...
71 7. Start-up field of view. The black edges of the imageshould have the same distance to the outeredge of the field of view on all sides. If not,recenter the image with the centering screws.Adjust the height of the condenser until theedges are sharp. Now open the field dia-phragm until the image f...
73 7. Start-up 7.7 Checking Phase Contrast Rings If your microscope is equipped for phase con-trast, light rings to match your objectives will beinstalled in the condenser.The light rings are already centered in the fac-tory. However, the centering should berechecked. Note: Each objective has its ow...
83 8. Operation The fluorescence filter turret will automati-cally go to an empty position or to the “A-TL”filter cube. • Insert a transmitted light specimen. • Rotate an appropriate objective into place. • Focus the image with the knob on the SmartMove or the focusing wheel and adjustthe intensity ...
99 8. Operation 8.11 Magnification Changer Leica DMI3000 B:A mechanical magnification changer can beused optionally. The following magnification fac-tors are available: 1.5x, 1.6x and 2x A slider switches between 1x and the magnifica-tion factor. The mechanical magnificationchanger affects the eyepi...
100 8. Operation Adjust intensity via fixed function buttons on stand variable function buttons on stand and SmartMove Leica Application Suite (LAS) Software Fig. 118 Fixed function buttons, left side of stand 1 Aperture diaphragm 2 Transmitted light/incident light 3 Field diaphragm 4 Light intensit...
107 10. Pflege des Mikroskops Caution! Unplug the power supply before performingcleaning and maintenance work!Protect electrical components from mois-ture! Microscopes in warm and warm-damp climaticzones require special care in order to preventfungus contamination.The microscope should be cleaned af...
108 10. Pflege des Mikroskops Cleaning Glass Surfaces Remove dust on glass surfaces with a fine, dryand fat-free hair brush, by blowing with a blowbag or vacuum suction. Remove stubborn dirt on glass surfaces with aclean cloth dampened with distilled water. If thedirt still can not be removed, use p...
109 11. Major Consumable and Replacement Parts Order No.Material No. Name Used for Replacement Lamp11 500 974 Halogen lamp 12V 100 W 107/2 lamp housing 11 500 137 High-pressure mercury burner 50 W 106 z lamp housing 11 500 138 High-pressure mercury burner 100 W 106 z lamp housing 11 500 321 High-pre...
111 13. Abbreviations and Icons Contrasting method Magnification Illumination Ports/Eyepiece Focus Lower focus stop not set Lower focus stop set Focus position not set Focus position set Shutter open Shutter closed Transmitted-light filter Field diaphragm, rectangular Field diaphragm, round Aperture...
116 15. EU Declaration of Conformity 15. EU Declaration of Conformity We hereby declare that the device described below, both in its basic design and construction and inthe version marked by us, conforms to the relevant safety- and health-related requirements of theappropriate EU directives.This dec...
3 Leica DMI3000 BLeica DMI4000 BLeica DMI6000 B Bedienungsanleitung
5 Inhalt 6.11 Montage von Lampenhausaufnahme und Spiegelhaus ........................................ 42 6.12 Montage und Wechsel der Auflichtlampen .................................... 44 6.13 Bestückung der Auflicht-Revolverscheibe ........................ 48 6.14 Einsetzen des Front-Modul Schiebe...
7 1. Wichtige Hinweise zur Anleitung (1.2) → S. 20 ! * Ziffern in Klammern, z.B. (1.2), beziehen sich aufAbbildungen, im Beispiel Abb. 1, Pos. 2. Ziffern mit Hinweispfeil, z.B. → S. 20, weisen auf eine bestimmte Seite dieser Anleitung hin. Achtung!Besondere Sicherheitshinweise in dieserAnleitung sin...
8 2. Zweckbestimmung des Mikroskops 2. Zweckbestimmung des Mikroskops Die Mikroskope der Leica DMI-Serie, zu denendiese Bedienungsanleitung gehört, sind für bio-logische Routine- und Forschungsanwendungenvorgesehen. Dies schließt die Untersuchungvon aus dem menschlichen Körper stammendenProben zum Z...
9 3. Sicherheitshinweise 3. Sicherheitshinweise 3.1 Allgemeine Sicherheitshinweise Dieses Gerät der Schutzklasse 1 ist gemäßEN 61010-2-101:2002, EN 61010-1:2001, IEC 1010-1:2001,Sicherheitsbestimmungen für elektrischeMess-, Steuer-, Regel- und Laborgeräte gebautund geprüft. Achtung! Um diesen Auslie...
12 4. Geräteübersicht 4.1 Spezifikationen 4. Geräteübersicht Leica DMI-Serie Kontrastverfahren Durchlichtachse Leica DMI-Serie• Durchlicht (DL): BF, DF, PH, DIC, Pol • Zwischenpupille: IMC (Integrierter Modulationskontrast)IPH (Integrierter Phasenkontrast) Leica DMI4000 B und DMI6000 B • Auflicht (I...
15 4. Geräteübersicht Kondensoren Z-Fokus Beobachtungsausgänge Leica DMI4000 B und Leica DMI6000 B(identisch für Leica DMI3000 B jedoch manuell)• motorisiert und kodiert oder manuell und kodiert• motorisierte oder manuelle Aperturblende• Kontrastiermethoden: BF, DF, PH, DIC, Pol, IMC, IPH• automatis...
16 4. Geräteübersicht Beobachtungsausgänge Bedienelemente Elektronikbox Leica DMI3000 B(Beim Leica DMI3000 B ist ein manueller Seitenport immer Be-standteil des Stativs)• manuell• Linker Seitenport (80% Transmission) Leica DMI4000 B und Leica DMI6000 B• 7 feste Bedientasten für die Beleuchtung und A...
17 4. Geräteübersicht Leica DMI4000 B und Leica DMI6000 B• 2 x RS232C• 2 x USB• 4 x externe/interne Peripheriegeräte• CTR-Boxen• SmartMove Leica DMI4000 B und Leica DMI6000 B• Leica Application Suite (LAS) für Windows TM 2000, XP mit Plug-ins für:• Mikroskop- und Kamera-Konfiguration• Mikroskop- und...
27 5. Auspacken Bitte vergleichen Sie die Lieferung sorgfältig mitdem Packzettel, Lieferschein oder der Rech-nung. Wir empfehlen dringend, eine Kopiedieser Dokumente mit der Anleitung aufzube-wahren, um z.B. bei späteren Nachbestellungenoder Servicearbeiten Informationen über Liefer-zeitpunkt und Li...
28 5. Auspacken Transport Für den Versand oder Transport des Mikroskopsund seiner Zubehörkomponenten sollte dieOriginalverpackung verwendet werden. Um Beschädigungen durch Erschütterungen zuvermeiden, sollten vorsorglich folgende Kom-ponenten demontiert und gesondert verpacktwerden: • Schrauben Sie ...
29 6. Montage 6. Montage des Mikroskops Die Mikroskopkomponenten* werden sinnvol-lerweise in dieser Reihenfolge montiert: • Durchlicht-Beleuchtungsträger• DIC-Modul und DIC-Objektivprismen• Kondensor mit Kondensorkopf• Okulare• Objektive• Durchlichtlampen• Lampenhausaufnahme (Spiegelhäuser)• Auflich...
31 6. Montage Abb. 12 IC-Objektivprisma 1 Objektivprisma in Fassung 2 Unterlegscheibe und Schraube Abb. 10b DIC-Objektivprismenrevolver (kodiert und motorisch) 1 IC-Objektivprisma in Fassung 2 Kennbuchstabe (ID) 3 Orientierungsstift Abb. 11 Frontabdeckung DIC-Prismenscheibe 6.3 Montage des DIC-Modul...
32 6. Montage 6.4 Montage der Objekttische Es stehen eine Vielzahl von Objekttischen zurVerfügung. Die wichtigsten sind die folgenden: • Fester Tisch (248 mm x 204 mm) (Abb. 13): normal, beheizbar und temperierbar mit und ohne Objektführer • Fester Mikromanipulationstisch (248 mm x 204/112 mm) (Abb....
33 6. Montage Abb. 17 Fester Mikromanipulationstisch Abb. 18 Objektführer für festen Mikromanipulationstisch Die Montage dieser Tische ist jeweils identisch.Durch 3 Schrauben werden die Tische fest mitdem Mikroskop verbunden. Im Falle der festenTische ist wahlweise rechts oder links einObjektführer ...
34 6. Montage Abb. 19 a, b Inserts für Objektführer (fester Tisch) • Anschließend drehen Sie die hinteren beiden Schrauben fest an. • Zum Schluss ziehen Sie die vordere Schraube noch einmal fest nach. Fester Tisch Für die festen Tischplatten werden auch wahl-weise Objektführer zur Aufnahme von Halte...
35 6. Montage Manueller fester Mikromanipulationstisch Zur Montage des Objektführers für den manuel-len festen Mikromanipulationstisch (Abb. 24) ge-hen Sie genauso vor wie für den Objektführerdes normalen Tisches. Die Einlegerahmen (Abb. 20a bis c) sind hier un-terschiedlich. Diese werden durch 2 Sc...
36 6. Montage Motorische 3-Platten oder Scanningtische 3-Platten-Tische und Scanningtische: Nach derMontage des Tisches verbinden Sie (im Fallemotorischer Tische) das mitgelieferte Tisch-kabel zunächst mit der Buchse des Tischesund anschließend mit der CRT6000 oderCTR6500 Box. Die entsprechende Mark...
37 6. Montage Abb. 311 Einlegerahmen für Deckgläser 2 Blattfedern Drehtisch und Einlegerahmen für Deckgläser Der Drehtisch (Abb. 30) wird ebenfalls durch 3Schrauben (30.2) befestigt. Um alle Gewinde-bohrungen zugänglich zu machen, ist der Dreh-aufsatz zu bewegen. Schrauben (30.2) ansetzen. Achtung: ...
39 6. Montage • Nehmen Sie nun die spezielle Kondensor- zange zu Hand (Abb. 39.1). • Montieren Sie, wenn möglich, die Lichtringe 0..3 in aufsteigender Reihenfolge. Sie findeneine Nummerierung der Öffnungen am Randdes Zahnkranzes (4 große Öffnungen: 1-4; 3kleine Öffnungen: 5-7). • Greifen Sie mit der...
40 6. Montage Wenn Sie auch IC Kondensor Prismen einzubau-en haben lesen, Sie weiter, sonst fahren Sie imnächsten Abschnitt fort. Montage IC-Kondensorprismen • Schalten Sie das Mikroskop aus. • Entfernen Sie die Kondensorklappe (38.1). Das Prisma wird in eine der großen mit Führungs-nuten versehenen...
42 6. Montage Abb. 42 Einsetzen der Kondensorköpfe S1 1 Kondensorbasis 2 Zwischenring 3 Kondensorkopf Abb. 43 Einsetzen des Kondensorkopfes S28 Kondensorköpfe Es stehen 4 verschiedene Kondensorköpfe zurVerfügung: 1) S1/1.40 oil 2) S1/0.90 dry 3) S23/0.53 4) S28/0.55 Die Kondensorköpfe 3 und 4 werden...
44 6. Montage 6.8 Montage der Filter im Beleuchtungsarm Die Leica DMI-Serie ist generell mit einem Filter-magazin zur Aufnahme von 2 Filtern mit Ø 40 mmausgestattet. Die Filter sind in der Regel bereitswerkseitig in den Halter eingesetzt. Falls Sie Fil-ter nachrüsten und selbst montieren:• Schraube ...
45 6. Montage 6.10 Montage und Wechsel der Durchlicht-lampen: Lampenhaus 107 oder 107/2 Dieses Lampenhaus wird mit einer 12V 100WHalogenglühlampe verwendet, die bereits ein-gebaut ist.Soll die Lampe ausgewechselt werden, gehenSie folgendermaßen vor: Wechseln der 12V 100W Halogen-Glühlampe Achtung! A...
47 6. Montage Abb. 55 Booster-Linse Abb. 56 Booster-Linse im Stativ 1 Booster-Linse Sollte eine Booster-Linse zum Lieferumfang ge-hören, kann diese je nach Stativ nun in die hin-tere seitliche Stativöffnung rechts oder links ge-schoben werden. Der Booster-Schieber hat mehrere Stellungen:1. Schieber ...
48 6. Montage 6.12 Montage und Wechsel der Auflichtlampen Leica DMI4000 B und Leica DMI6000 B:Es besteht generell bei den Lichtquelleneine Gefährdung durch Strahlung (Blen-dung, UV-Strahlung, IR-Strahlung). Lampenmüssen daher in geschlossenen Gehäusenbetrieben werden. Achten Sie darauf, dass das Lam...
49 6. Montage Abb. 60 Lampenfassungen für Gasentladungslampe Hg 100 1 Obere Klemmung 2 Untere Klemmung 3 Kühlelement a b Hg 100 Xe 75 3 1 2 3 1 2 4 Einsetzen der Gasentladungslampen (Hg undXe) in das Lampenhaus 106z Hg- und Xe-Lampen werden mit separaten Vor-schaltgeräten betrieben.Bitte unbedingt d...
52 6. Montage 6.13 Bestückung der Auflicht-Revolverscheibe Achtung: Diesen Abschnitt zunächst komplett lesen, be-vor Sie mit der Bestückung der Revolverscheibebeginnen.Leica DMI4000 B und Leica DMI6000 B:Auf der rechten Seite des Stativs befindet sichdie Fluoreszenzschublade. Bevor Sie dieseSchublad...
54 6. Montage Abb. 70 Mechanischer Polarisationshalter 1 Manueller Polarisator 2 Manueller Analysator Abb. 71 Kondensor mit motorischem Polarisator 6.14 Einsetzen des Front-Modul Schiebers Ist Ihr Mikroskop für den Integrierten Modu-lationskontrast oder für den Integrierten Pha-senkontrast vorbereit...
55 6. Montage Abb. 73 Montage des Analysators 1 Aufnahme 2 Analysator Durchlicht- und Auflichtanalysator • Entfernen Sie die Steckkappe (Abb. 72) auf der rechten Seite des Stativs (unter der Fluo-Schublade). • Schieben Sie den Analysator bis zur Rastung in die Aufnahme (Abb. 73.1). Abb. 72 Steckkapp...
56 6. Montage 6.16 Optionales Zubehör Kamera Anschluss einer KameraEine Kamera kann über einen c-Mount oderVario-Mounts adaptiert werden. • Setzen Sie den c-Mount bzw. den Vario- Mount auf einen der Kameraausgänge desMikroskops auf und befestigen Sie ihn mit derseitlichen Klemmschraube. • Schrauben ...
57 6. Montage Abb. 78 Stativrückseite 1 RS232-Schnittstellen 2 2 x USB 3 4 x EXT. 4 XYZ-Control für Smart Move 5 Anschluss Elektronikbox 6 Kondensorkabel 7 Lampenversorgungskabel des Stativs Abb. 77 Rückseite CTR6000 1 Buchse für Netzkabel 2 Buchse XY-Stage für motorisierten Tisch 3 Buchse Direct in...
59 7. Inbetriebnahme 7. Inbetriebnahme 7.1 Funktionsprinzip (Leica DMI4000 B und Leica DMI6000 B) 1. Intelligente Automatisierung • Umschalten zwischen verschiedenen Kontrastverfahren auf Knopfdruck. Lichtringe, DIC- Prismen, etc. werden automatisch in den Strahlengang gebracht. • Das Mikroskop erke...
63 7. Inbetriebnahme Ist eine der Komponenten nicht ordnungsgemäßmontiert, erscheint eine Fehlermeldung auf demLeicaDisplay.Siehe Kapitel Trouble Shooting → S. 102. Die mikroskopischen Komponenten wie Blen-den, Kondensor, Licht-und Phasenringe sind be-reits werkseitig vorzentriert. Durch Transportun...
65 7. Inbetriebnahme Hinweis: Änderungen der Lichtintensität sowie der Ein-stellung von Apertur- und Leuchtfeldblendewerden für das jeweilige Objektiv und Kontrast-verfahren abgespeichert. Abb. 84 Fest definierte Funktionstasten (linke Stativseite) 1 Wechsel Durchlicht/Auflicht 2 Aperturblende 3 Lic...
66 7. Inbetriebnahme Variable Funktionstasten am Stativ Passend zu Ihrer Mikroskopausrüstung erfolgtwerkseitig eine Vorbelegung der variablenFunktionstasten. Die Tasten sind entsprechendbeschriftet. Die Tastenbelegung entnehmen Siebitte dem „Identification Sheet“.Die Bedeutung der Abkürzungen entneh...
67 7. Inbetriebnahme Funktionstasten am Frontbedienfeld (Abb. 87) 100% des Lichtes geht zum Okular(87.1). Toggelfunktion für die seitlichenPorts (87.2). Die Funktion ist abhän-gig von der jeweiligen Mikroskop-konfiguration.Hinweis:Umschaltung auf Bottom Port:Über die variablen Funktionstasten(nur Le...
68 7. Inbetriebnahme Abb. 89 Fernsteuermodul SmartMove 1 Verfahren in X-Richtung 2 Verfahren in Y-Richtung 3 Fokuseinstellung 4 Individuelle Einstellung der Knopfhöhenposition 5 Variable Funktionstasten (werkseitig vorbelegt) 1 2 3 4 5 SET + Cube-Taste 1-6 (nur Leica DMI4000 B undLeica DMI6000 B) Be...
71 7. Inbetriebnahme die Mitte des Sehfeldes (91.c). Anschließendöffnen Sie die Feldblende, bis das Bild annä-hernd das gesamte Bildfeld ausfüllt. Dieschwarzen Ränder des Bildes solltenrundherum den gleichen Abstand zum äuße-ren Rand des Sehfeldes haben. Falls nicht,zentrieren Sie das Bild erneut mi...
73 7. Inbetriebnahme 7.7 Phasenkontrastringe überprüfen Ist Ihr Mikroskop für die Verwendung vonPhasenkontrast ausgerüstet, sind im Kondensordie zu den Objektiven passenden Lichtringe ein-gebaut.Die Lichtringe sind bereits werkseitig zentriert.Die Zentrierung sollte jedoch noch einmal über-prüft wer...
91 8. Bedienung 8.4 Kombi-Verfahren (Leica DMI4000 und DMI6000 B) Je nach Ausstattung Ihres Mikroskops sind biszu zwei Kombi-Verfahren möglich: FLUO/PH und FLUO/DIC • Wählen Sie das Kombi-Verfahren. Drücken Sie dazu die variable Taste COMBI . Alternativ: Drücken Sie die variable Taste CHANGE COMBI ....
92 8. Bedienung 8.5 Fokussierung Leica DMI3000 B und Leica DMI4000 B:Über die linken Fokusräder können sowohlGrob- als auch Feinfokussierung vorgenommenwerden über das rechte Fokusrad kann eineFeinfokussierung vorgenommen werden (essteht auch eine rechts und links vertauschteVersion des DMI3000 B zu...
99 8. Bedienung 8.11 Vergrößerungswechsler Leica DMI3000 B:Optional kann ein mechanischer Vergröße-rungswechsler eingesetzt werden. Die folgen-den Vergrößerungsfaktoren können wahlweiseeingestellt werden: 1,5x ; 1,6x oder 2x Über einen Schieber kann zwischen 1x und demVergrößerungsfaktor geschaltet ...
100 8. Bedienung Licht einstellen über Feste Funktionstasten am Stativ Variable Funktionstasten am Stativ und SmartMove Software Leica Application Suite (LAS) Abb. 118 Feste Funktionstasten, linke Stativseite 1 Aperturblende 2 Durchlicht/Auflicht 3 Leuchtfeldblende 4 Lichtintensität 1 2 3 4 8.11 Lic...
109 11. Wichtigste Verschleiß- und Ersatzteile Bestell-NummerSach-Nummer Bezeichnung Verwendung für Ersatzlampen11 500 974 Halogenglühlampe 12 V 100 W Lampenhaus 107/2 11 500 137 Hg-Höchstdrucklampe 50 W Lampenhaus 106 z 11 500 138 Hg-Höchstdrucklampe 100 W Lampenhaus 106 z 11 500 321 Hg-Höchstdruck...
111 13. Abkürzungen und Piktogramme Kontrastverfahren Vergrößerung Beleuchtung Ports/Okular Fokus Untere Fokusschwelle nicht gesetzt Untere Fokusschwelle gesetzt Fokusposition nicht gesetzt Fokusposition gesetzt Shutter auf Shutter zu Durchlichtfilter Feldblende, rechteckig Feldblende, rund Aperturb...
113 14. Index 14. Index A bkürzungen 99 Aktive Ports 60Analysator 50, 51Analysatoraufnahme 19Anwahl der Ports 84Aperturblende 16, 18, 61, 76, 90 Auflicht-Fluoreszenz 68Auflicht-Lampenhaus 19Auflicht-Revolverscheibe 48Auflichtachse 12Aufstellungsort 23Augenabstand 84 B edienelemente 15, 55 Beleuchtun...
Leica Manuals
-
Leica VT1200S
User Manual
-
Leica VT1200S
Manual
-
Leica HER2 FISH System - 30 Test
User Manual
-
Leica Leica DISTO A6 - DISTO transfer manuals for PC
User Manual
-
Leica M7
User Manual
-
Leica EG1130
User Manual
-
Leica Laser class 2 products - Safety Manual
User Manual
- Leica DISTO + Bluetooth – getting started on iOS User Manual
-
Leica SNLL121
User Manual
-
Leica on letter m
User Manual
-
Leica SP9000
Manual
-
Leica V-LUX 1
User Manual
-
Leica V-LUX 1
Manual
-
Leica Leica DISTO pro /pro4a - interface manual
User Manual
-
Leica Disto D410
User Manual
-
Leica Leica DISTO D410 - Quickstart
User Manual
-
Leica TPS800 DXF Export
User Manual
- Leica DISTO + Bluetooth – getting started on Windows 8 User Manual
-
Leica DISTO X310
User Manual
-
Leica CM1950
User Manual