Page 3 - Operating Manual
3 Leica DMI3000 BLeica DMI4000 BLeica DMI6000 B Operating Manual
Page 4 - Copyrights
4 Copyrights Copyrights All rights to this documentation are held byLeica Microsystems Wetzlar GmbH. Reproduc-tion of text or illustrations (in whole or in part) byprint, photocopy, microfilm or other method (in-cluding electronic systems) is not allowed with-out express written permission from Leic...
Page 5 - Installation of Lamp Housing Mount; Contents
5 Contents 6.11 Installation of Lamp Housing Mount and Mirror Housing ................................... 42 6.12 Installation and Replacement of Incident-Light Lamps ............................ 44 6.13 Equipping the Incident Light Turret Disk ......................... 48 6.14 Inserting the Front M...
Page 7 - Numbers with pointer arrows (for example; Text symbols, pictograms and their meanings:; Important Notes about this Manual
7 1. Important Notes about this Manual (1.2) → p. 20 ! * Numbers in parentheses, such as “(1.2)”, corre-spond to illustrations (in the example, Figure 1,Item 2). Numbers with pointer arrows (for example → p. 20), point to a certain page of this manual. Caution!Special safety instructions within this...
Page 8 - Intended Purpose of the Microscope
8 2. Intended Purpose of the Microscope 2. Intended Purpose of the Microscope The microscopes of the Leica DMI series cov-ered in this manual are designed for biologicalroutine and research applications. This includesthe examination of samples taken from the hu-man body with a view to providing info...
Page 9 - General Safety Notes
9 3. Safety Notes 3. Safety Notes 3.1 General Safety Notes This safety class 1 device is constructed andtested in accordance withEN 61010-2-101:2002, EN 61010-1:2001,IEC 1010-1:2001,Safety regulations for electrical measuring, con-trol, and laboratory devices. Caution! In order to maintain this cond...
Page 12 - Overview of the Instrument; Overview of the Leica DMI Series; Contrast Methods; • intermediate pupil:
12 4. Overview of the Instrument 4.1 Specifications 4. Overview of the Leica DMI Series Contrast Methods Transmitted Light Axis Leica DMI Series• transmitted light (DL): BF, DF, PH, DIC, Pol • intermediate pupil: IMC (integrated modulation contrast)IPH (Integrated phase contrast) Leica DMI4000 B and...
Page 16 - and four additional variable function buttons
16 4. Overview of the Instrument Observation ports Controls Electronics box Leica DMI3000 B(a manual side port is a standard feature of the Leica DMI3000 Bstand)• manual• left side port (80% transmission) Leica DMI4000 B and Leica DMI6000 B• 7 fixed control buttons for illumination and apertures• 7 ...
Page 17 - Interfaces
17 4. Overview of the Instrument Leica DMI4000 B and Leica DMI6000 B• 2 x RS232C• 2 x USB• 4 x external/internal peripherals• CTR boxes• SmartMove Leica DMI4000 B and Leica DMI6000 B• Leica Application Suite (LAS) for Windows TM 2000, XP with plug-ins for:• microscope and camera configuration• micro...
Page 26 - Unpacking the Microscope
26 5. Unpacking the Microscope The microscope is delivered in several pack-ages. The stand package contains the following com- ponents: • Stand with integrated incident-light axis, objective turret and tube • Illumination arm • Specimen stage • CD with Leica Application Suite (LAS) soft- ware packag...
Page 27 - Relative humidity
27 5. Unpacking the Microscope Please carefully compare the contents of the de-livery to the packing slip, delivery note or in-voice. We urgently recommend storing a copy ofthese documents with the manual to ensure thatyou have information on the time and scope ofdelivery handy for subsequent orders...
Page 29 - Assembling the Microscope; The order may be vary when using; climate
29 6. Assembly 6. Assembling the Microscope The microscope components* are logically as-sembled in this order: • Transmitted-light illumination carrier• DIC module and DIC objective prisms• Condenser with condenser head• Eyepieces• Objectives• Transmitted-light lamps• Lamp housing mount (mirror hous...
Page 30 - Installation of the
30 6. Assembly 6.2 Installation of the transmitted-light illumination carrier (DL) Wipe the installation surface on the microscope(8.3) with a dry cloth. Tip the illumination carrier(8.1) back slightly and install it so that the pin(8.2) engages the groove in the support surface(8.4). Set the DL ill...
Page 31 - • Release the two socket screws and remove
31 6. Assembly Fig. 12 IC objective prism 1 Objective prism in frame 2 Screw and washer Fig. 10b DIC objective prism turret (coded and motorized) 1 IC objective prism in frame 2 Identification letter (ID) 3 Orientation pin Fig. 11 Front cover, DIC prism disk 6.3 Installation of the DIC Moduleand DIC...
Page 32 - Installation of Specimen Stages; • Fixed micromanipulation stage (248 mm x 204/
32 6. Assembly 6.4 Installation of Specimen Stages A wide range of specimen stages are available.The most important are the following: • Fixed stage (248 mm x 204 mm) (Fig. 13): normal, heating and temperature-controlled, with and without attachable mechanicalstage • Fixed micromanipulation stage (2...
Page 33 - • If the screws for the stage are already in the
33 6. Assembly Fig. 17 Fixed micromanipulation stage Fig. 18 Attachable mechanical stage for fixedmicromanipulation stage The assembly of these stages is identical. Thestages are solidly attached to the microscope bythree screws. In the case of fixed stages, an at-tachable mechanical stage may be in...
Page 34 - • Next, firmly tighten the two rear screws.; Fixed stage
34 6. Assembly Fig. 19 a, b Inserts for attachable mechanical stage (fixed stage) • Next, firmly tighten the two rear screws. • Finally, give the front screw a final firm tight- ening. Fixed stage Attachable mechanical stages designed to ac-cept a variety of culture dishes are also avail-able for fi...
Page 35 - Manual fixed micromanipulation stage
35 6. Assembly Manual fixed micromanipulation stage To install the attachable mechanical stage forthe manual fixed micromanipulation stage(Fig. 24), proceed as you would for the attach-able mechanical stage of the standard stage. The insert frames (Fig. 20a to c) differ at thispoint. These are held ...
Page 36 - Motorized 3-plate or scanning stages
36 6. Assembly Motorized 3-plate or scanning stages 3-plate stages and scanning stages: after in-stalling the stage, connect the included stagecable (for motorized stages) first to the socketon the stage, then to the CRT6000 or CTR6500box. The correct place on the box is called“XY Stage”. A variety ...
Page 37 - Rotating Stage and Insert Frames for Coverslips
37 6. Assembly Fig. 311 Insert frame for coverslips 2 Leaf springs Rotating Stage and Insert Frames for Coverslips The rotating stage (Fig. 30) is also mounted with3 screws (30.2). Rotate the stage to make all ofthe threaded holes accessible. Insert thescrews (30.2). Caution: Use additional washers ...
Page 38 - • Switch the microscope off.; maximum of 1 mm; in the location holes of the turret
38 6. Assembly 6.5 Installation of Condensers All condensers of the Leica DMI series areequipped with a 7-position turret disk that canbe equipped with light rings phase contrast (PH)or dark field (DF), IC prisms for transmitted-lightinterference contrast (DIC) or slit illuminators for integrated mo...
Page 39 - not
39 6. Assembly • Next, take the special condenser tool (Fig. 39.1). • If possible, install the light rings 0 to 3 in as- cending order. The numbering of the openingsis located at the edge of the crown gear(4 large openings: 1-4; 3 small openings: 5-7). • Grasp the light ring to be installed with the...
Page 40 - Installation of IC prisms; • Grasp the prism to be installed with the con-
40 6. Assembly Please continue reading if you also have to in-stall IC prisms. Otherwise, skip to the next sec-tion. Installation of IC prisms • Switch the microscope off. • Remove the condenser cover (38.1). Insert the prism in one of the condenser disk’s large re-ceptacles with guide grooves. • Tu...
Page 41 - • Note the number of the opening and the prism; Installation of Condensers
41 6. Assembly Caution: Do not press the spring clip down under anycircumstances. This can destroy the clip orresult in an unstable position of the prism. Turn the prism to ensure that it snaps into po-sition and release the tool.Remove fingerprints or dust from the prismwith care. • Use the left ce...
Page 42 - Condenser heads; Four different condenser heads are available:
42 6. Assembly Fig. 42 Installation of condenser heads S1 1 Condenser base 2 Spacer ring 3 Condenser head Fig. 43 Installation of condenser head S28 Condenser heads Four different condenser heads are available: 1) S1/1.40 oil 2) S1/0.90 dry 3) S23/0.53 4) S28/0.55 Condenser heads 3 and 4 are screwed...
Page 44 - Installation of filters in the illumination arm; carrier and fasten with the locking screw.; Installing the transmitted-light lamp; • Place the lamp housing in the transmitted light
44 6. Assembly 6.8 Installation of filters in the illumination arm The Leica DMI series is equipped with a filtermagazine to accommodate two 40 mm dia. filtersas a standard feature. The filters are installed atthe factory. To change filters yourself, proceedas follows:• Release the screw (46.1) and ...
Page 45 - • Remove the fastener screw on the housing
45 6. Assembly 6.10 Installation and replacement of the transmitted-light lamps: 107 or 107/2 lamphousing This lamp housing is used with a 12V 100W halo-gen lamp, which is already mounted.In case the lamp has to be removed: Changing the 12 V 100 W halogen lamp Caution! Ensure that the lamp housing h...
Page 46 - Installing the lamp housing mount and
46 6. Assembly 6.11 Installing the lamp housing mount and mirror housing (Leica DMI4000 B andDMI6000 B) Place lamp housing mount (Fig. 53) or mirrorhousing on rear wall. Mount from front withsocket head screws. Fig. 54 Lamp housing 106z 1 Collector adjustment 2 Vertical lamp adjustment 3 Horizontal ...
Page 47 - no effect
47 6. Assembly Fig. 55 Booster lens Fig. 56 Booster lens in stand 1 Booster lens If a booster lens is included in the scope of de-livery, insert it into the rear stand opening at theleft or right, depending on the stand model. The booster slide has several positions:1. Slide pulled out: no effect 2....
Page 48 - Installation and replacement of
48 6. Assembly 6.12 Installation and replacement of incident-light lamps Caution! Leica DMI4000 B and Leica DMI6000 B:Light sources pose a potential irradiationrisk (glare, UV-radiation, IR-radiation).Therefore, lamps have to be operated inclosed housings. Ensure that the lamp housing has been dis-c...
Page 49 - Type; * Please observe the data sheets of the lamp manufacturer.
49 6. Assembly Fig. 60 Lamp mounts for Hg 100 gas discharge lamp 1 Uper clamping system 2 Lower clamping system 3 Cooling element a b Hg 100 Xe 75 3 1 2 3 1 2 4 Inserting gas discharge lamps (Hg and Xe) inthe 106z lamp housing Hg and Xe lamps are powered by separate sup-ply units.Please also read th...
Page 52 - Equipping the Incident Light Turret Disc
52 6. Assembly 6.13 Equipping the Incident Light Turret Disc Caution: Please read this section completely before be-ginning with the assembly of the turret disk.Leica DMI4000 B and Leica DMI6000 B:The fluorescence drawer is located on the rightside of the stand. Before opening this drawer,remove the...
Page 54 - Inserting the Front Module Slider
54 6. Assembly Fig. 70 Mechanical polarizer holder 1 Manual polarizer 2 Manual analyzer Fig. 71 Condenser with motorized polarizer 6.14 Inserting the Front Module Slider If your microscope is prepared for integratedmodulation contrast or integrated phase con-trast, a front module (possibly in conjun...
Page 55 - Analyzer for incident light and transmitted light.; • Insert the analyzer into the receptacle until it
55 6. Assembly Fig. 73 Inserting the analyzer 1 Slot 2 Analyzer Analyzer for incident light and transmitted light. • Remove the cap (Fig. 72) on the right side of the stand (under the fluorescence drawer). • Insert the analyzer into the receptacle until it latches in place (Fig. 73.1). Fig. 72 Analy...
Page 57 - Connection to the Electronics Box
57 6. Assembly Fig. 78 Rear view of stand 1 RS232 ports 2 2 x USB 3 4 x EXT. 4 XYZ control for SmartMove 5 Electronic box connection 6 Condenser cable 7 Lamp power cable Fig. 77 Rear view of CTR6000 1 AC power socket 2 XY Stage socket for motorized stage 3 Direct interface socket optional 4 Z Contro...
Page 58 - Microscope; Connection to the Computer
58 6. Assembly CTR6000 and CTR6500 electronics box: Note: These electronics boxes must not be used withother stands. The serial number of the associatedstand has been recorded on the back of the elec-tronics box. A 3-axis control unit for focus and 3-plate crossstages is integrated in the CTR6000. A...
Page 59 - Intelligent automation
59 7. Start-up 7. Start-up 7.1 Functional Principle (Leica DMI4000 B and Leica DMI6000 B) 1. Intelligent automation • Switching between contrast methods at the touch of a button. Light rings, DIC prisms, etc. are automatically positioned in the beam path. • The microscope recognizes the selected obj...
Page 63 - do not; turn on the power supply unit; Switching on the Microscope; If a PC is connected, switch on; the electronics
63 7. Start-up If a component has not been installed correctly,the LeicaScreen will display an error message.See Troubleshooting chapter, → p. 102. Components such as diaphragms, condensers,light and phase rings have been pre-centered atthe factory. It may be necessary to correct thecentering after ...
Page 65 - • Light intensity: the transmitted-light intensity
65 7. Start-up Note: Changes to the light intensity as well as apertureand field diaphragm settings are stored for theindividual objectives and contrast methods. Fig. 84 Fixed function buttons (left side of stand) 1 Toggle transmitted light/incident light 2 Aperture diaphragm 3 light intensity 4 fie...
Page 66 - * See page 62 for abbreviations
66 7. Start-up Variable function buttons on the stand The variable function buttons are assigned func-tions at the factory that are appropriate to thefeatures of your microscope. They are labeledaccordingly. For details on button assignments,please refer to the included identification sheet.For info...
Page 67 - Sets the upper focus stop.
67 7. Start-up Function buttons on the front panel (Fig. 87) 100% of the light goes to the eye-piece (87.1). Toggle function for the side ports(87.2). This function depends on theindividual microscope configuration.Note:Switching to the bottom port:via the variable function buttons(Leica DMI6000 B o...
Page 68 - • Set the condenser to the bright field position.
68 7. Start-up Fig. 89 SmartMove remote control module 1 travel in x 2 Travel in y 3 Focus 4 Individual adjustment of button height 5 Variable function buttons (factory preset) 1 2 3 4 5 SET + Cube buttons 1-6(Leica DMI4000 B and DMI6000 B only) The SET command can be assigned to a functionbutton of...
Page 71 - • The last step is the; adaptation of the contrast; • Set the aperture diaphragm to cover 2/3 to 4/5
71 7. Start-up field of view. The black edges of the imageshould have the same distance to the outeredge of the field of view on all sides. If not,recenter the image with the centering screws.Adjust the height of the condenser until theedges are sharp. Now open the field dia-phragm until the image f...
Page 73 - BF; • Instead of an eyepiece, place a focusing tel-
73 7. Start-up 7.7 Checking Phase Contrast Rings If your microscope is equipped for phase con-trast, light rings to match your objectives will beinstalled in the condenser.The light rings are already centered in the fac-tory. However, the centering should berechecked. Note: Each objective has its ow...
Page 83 - INT
83 8. Operation The fluorescence filter turret will automati-cally go to an empty position or to the “A-TL”filter cube. • Insert a transmitted light specimen. • Rotate an appropriate objective into place. • Focus the image with the knob on the SmartMove or the focusing wheel and adjustthe intensity ...
Page 99 - Magnification Changer; Change magnification via; fixed function buttons on stand
99 8. Operation 8.11 Magnification Changer Leica DMI3000 B:A mechanical magnification changer can beused optionally. The following magnification fac-tors are available: 1.5x, 1.6x and 2x A slider switches between 1x and the magnifica-tion factor. The mechanical magnificationchanger affects the eyepi...
Page 100 - Adjust intensity via
100 8. Operation Adjust intensity via fixed function buttons on stand variable function buttons on stand and SmartMove Leica Application Suite (LAS) Software Fig. 118 Fixed function buttons, left side of stand 1 Aperture diaphragm 2 Transmitted light/incident light 3 Field diaphragm 4 Light intensit...
Page 107 - Pflege des Mikroskops
107 10. Pflege des Mikroskops Caution! Unplug the power supply before performingcleaning and maintenance work!Protect electrical components from mois-ture! Microscopes in warm and warm-damp climaticzones require special care in order to preventfungus contamination.The microscope should be cleaned af...
Page 108 - Follow safety instructions for immersion oil!; Handling Acids and Bases
108 10. Pflege des Mikroskops Cleaning Glass Surfaces Remove dust on glass surfaces with a fine, dryand fat-free hair brush, by blowing with a blowbag or vacuum suction. Remove stubborn dirt on glass surfaces with aclean cloth dampened with distilled water. If thedirt still can not be removed, use p...
Page 109 - Major Consumable and Replacement Parts; Major Consumable and Replacement
109 11. Major Consumable and Replacement Parts Order No.Material No. Name Used for Replacement Lamp11 500 974 Halogen lamp 12V 100 W 107/2 lamp housing 11 500 137 High-pressure mercury burner 50 W 106 z lamp housing 11 500 138 High-pressure mercury burner 100 W 106 z lamp housing 11 500 321 High-pre...
Page 111 - Abbreviations and Icons; Abbreviations and Pictograms
111 13. Abbreviations and Icons Contrasting method Magnification Illumination Ports/Eyepiece Focus Lower focus stop not set Lower focus stop set Focus position not set Focus position set Shutter open Shutter closed Transmitted-light filter Field diaphragm, rectangular Field diaphragm, round Aperture...
Page 116 - EU Declaration of Conformity
116 15. EU Declaration of Conformity 15. EU Declaration of Conformity We hereby declare that the device described below, both in its basic design and construction and inthe version marked by us, conforms to the relevant safety- and health-related requirements of theappropriate EU directives.This dec...
Page 117 - Bedienungsanleitung
3 Leica DMI3000 BLeica DMI4000 BLeica DMI6000 B Bedienungsanleitung
Page 119 - Montage von Lampenhausaufnahme; Inhalt
5 Inhalt 6.11 Montage von Lampenhausaufnahme und Spiegelhaus ........................................ 42 6.12 Montage und Wechsel der Auflichtlampen .................................... 44 6.13 Bestückung der Auflicht-Revolverscheibe ........................ 48 6.14 Einsetzen des Front-Modul Schiebe...
Page 121 - eine bestimmte Seite dieser Anleitung hin.; Textsymbole, Piktogramme und ihre Bedeutung:; Wichtige Hinweise zur Anleitung
7 1. Wichtige Hinweise zur Anleitung (1.2) → S. 20 ! * Ziffern in Klammern, z.B. (1.2), beziehen sich aufAbbildungen, im Beispiel Abb. 1, Pos. 2. Ziffern mit Hinweispfeil, z.B. → S. 20, weisen auf eine bestimmte Seite dieser Anleitung hin. Achtung!Besondere Sicherheitshinweise in dieserAnleitung sin...
Page 122 - Zweckbestimmung des Mikroskops
8 2. Zweckbestimmung des Mikroskops 2. Zweckbestimmung des Mikroskops Die Mikroskope der Leica DMI-Serie, zu denendiese Bedienungsanleitung gehört, sind für bio-logische Routine- und Forschungsanwendungenvorgesehen. Dies schließt die Untersuchungvon aus dem menschlichen Körper stammendenProben zum Z...
Page 123 - Sicherheitshinweise; Allgemeine Sicherheitshinweise
9 3. Sicherheitshinweise 3. Sicherheitshinweise 3.1 Allgemeine Sicherheitshinweise Dieses Gerät der Schutzklasse 1 ist gemäßEN 61010-2-101:2002, EN 61010-1:2001, IEC 1010-1:2001,Sicherheitsbestimmungen für elektrischeMess-, Steuer-, Regel- und Laborgeräte gebautund geprüft. Achtung! Um diesen Auslie...
Page 126 - Fluo
12 4. Geräteübersicht 4.1 Spezifikationen 4. Geräteübersicht Leica DMI-Serie Kontrastverfahren Durchlichtachse Leica DMI-Serie• Durchlicht (DL): BF, DF, PH, DIC, Pol • Zwischenpupille: IMC (Integrierter Modulationskontrast)IPH (Integrierter Phasenkontrast) Leica DMI4000 B und DMI6000 B • Auflicht (I...
Page 129 - scheibe
15 4. Geräteübersicht Kondensoren Z-Fokus Beobachtungsausgänge Leica DMI4000 B und Leica DMI6000 B(identisch für Leica DMI3000 B jedoch manuell)• motorisiert und kodiert oder manuell und kodiert• motorisierte oder manuelle Aperturblende• Kontrastiermethoden: BF, DF, PH, DIC, Pol, IMC, IPH• automatis...
Page 130 - x,y,z und 4 zusätzliche variable Funktionstasten
16 4. Geräteübersicht Beobachtungsausgänge Bedienelemente Elektronikbox Leica DMI3000 B(Beim Leica DMI3000 B ist ein manueller Seitenport immer Be-standteil des Stativs)• manuell• Linker Seitenport (80% Transmission) Leica DMI4000 B und Leica DMI6000 B• 7 feste Bedientasten für die Beleuchtung und A...
Page 131 - Schnittstellen
17 4. Geräteübersicht Leica DMI4000 B und Leica DMI6000 B• 2 x RS232C• 2 x USB• 4 x externe/interne Peripheriegeräte• CTR-Boxen• SmartMove Leica DMI4000 B und Leica DMI6000 B• Leica Application Suite (LAS) für Windows TM 2000, XP mit Plug-ins für:• Mikroskop- und Kamera-Konfiguration• Mikroskop- und...
Page 141 - Aufstellungsort; Relative Luftfeuchtigkeit
27 5. Auspacken Bitte vergleichen Sie die Lieferung sorgfältig mitdem Packzettel, Lieferschein oder der Rech-nung. Wir empfehlen dringend, eine Kopiedieser Dokumente mit der Anleitung aufzube-wahren, um z.B. bei späteren Nachbestellungenoder Servicearbeiten Informationen über Liefer-zeitpunkt und Li...
Page 142 - Transport
28 5. Auspacken Transport Für den Versand oder Transport des Mikroskopsund seiner Zubehörkomponenten sollte dieOriginalverpackung verwendet werden. Um Beschädigungen durch Erschütterungen zuvermeiden, sollten vorsorglich folgende Kom-ponenten demontiert und gesondert verpacktwerden: • Schrauben Sie ...
Page 143 - Montage des Mikroskops; Bei Verwendung von; Klimakammern oder ande-; Systemen und erweitertem optischen Zube-
29 6. Montage 6. Montage des Mikroskops Die Mikroskopkomponenten* werden sinnvol-lerweise in dieser Reihenfolge montiert: • Durchlicht-Beleuchtungsträger• DIC-Modul und DIC-Objektivprismen• Kondensor mit Kondensorkopf• Okulare• Objektive• Durchlichtlampen• Lampenhausaufnahme (Spiegelhäuser)• Auflich...
Page 145 - Die; • Lösen Sie die beiden Imbusschrauben und
31 6. Montage Abb. 12 IC-Objektivprisma 1 Objektivprisma in Fassung 2 Unterlegscheibe und Schraube Abb. 10b DIC-Objektivprismenrevolver (kodiert und motorisch) 1 IC-Objektivprisma in Fassung 2 Kennbuchstabe (ID) 3 Orientierungsstift Abb. 11 Frontabdeckung DIC-Prismenscheibe 6.3 Montage des DIC-Modul...
Page 146 - Montage der Objekttische; normal, beheizbar und temperierbar; • Fester Mikromanipulationstisch (248 mm x; mit und ohne Objektführer
32 6. Montage 6.4 Montage der Objekttische Es stehen eine Vielzahl von Objekttischen zurVerfügung. Die wichtigsten sind die folgenden: • Fester Tisch (248 mm x 204 mm) (Abb. 13): normal, beheizbar und temperierbar mit und ohne Objektführer • Fester Mikromanipulationstisch (248 mm x 204/112 mm) (Abb....
Page 147 - • Sollten die Schrauben für den Tisch bereits im
33 6. Montage Abb. 17 Fester Mikromanipulationstisch Abb. 18 Objektführer für festen Mikromanipulationstisch Die Montage dieser Tische ist jeweils identisch.Durch 3 Schrauben werden die Tische fest mitdem Mikroskop verbunden. Im Falle der festenTische ist wahlweise rechts oder links einObjektführer ...
Page 148 - Fester Tisch
34 6. Montage Abb. 19 a, b Inserts für Objektführer (fester Tisch) • Anschließend drehen Sie die hinteren beiden Schrauben fest an. • Zum Schluss ziehen Sie die vordere Schraube noch einmal fest nach. Fester Tisch Für die festen Tischplatten werden auch wahl-weise Objektführer zur Aufnahme von Halte...
Page 149 - Manueller fester Mikromanipulationstisch
35 6. Montage Manueller fester Mikromanipulationstisch Zur Montage des Objektführers für den manuel-len festen Mikromanipulationstisch (Abb. 24) ge-hen Sie genauso vor wie für den Objektführerdes normalen Tisches. Die Einlegerahmen (Abb. 20a bis c) sind hier un-terschiedlich. Diese werden durch 2 Sc...
Page 150 - Motorische 3-Platten oder Scanningtische
36 6. Montage Motorische 3-Platten oder Scanningtische 3-Platten-Tische und Scanningtische: Nach derMontage des Tisches verbinden Sie (im Fallemotorischer Tische) das mitgelieferte Tisch-kabel zunächst mit der Buchse des Tischesund anschließend mit der CRT6000 oderCTR6500 Box. Die entsprechende Mark...
Page 151 - Drehtisch und Einlegerahmen für Deckgläser
37 6. Montage Abb. 311 Einlegerahmen für Deckgläser 2 Blattfedern Drehtisch und Einlegerahmen für Deckgläser Der Drehtisch (Abb. 30) wird ebenfalls durch 3Schrauben (30.2) befestigt. Um alle Gewinde-bohrungen zugänglich zu machen, ist der Dreh-aufsatz zu bewegen. Schrauben (30.2) ansetzen. Achtung: ...
Page 153 - nem
39 6. Montage • Nehmen Sie nun die spezielle Kondensor- zange zu Hand (Abb. 39.1). • Montieren Sie, wenn möglich, die Lichtringe 0..3 in aufsteigender Reihenfolge. Sie findeneine Nummerierung der Öffnungen am Randdes Zahnkranzes (4 große Öffnungen: 1-4; 3kleine Öffnungen: 5-7). • Greifen Sie mit der...
Page 154 - Montage IC-Kondensorprismen; • Schalten Sie das Mikroskop aus.; maximal 1 mm; in die Öffnung hinein-
40 6. Montage Wenn Sie auch IC Kondensor Prismen einzubau-en haben lesen, Sie weiter, sonst fahren Sie imnächsten Abschnitt fort. Montage IC-Kondensorprismen • Schalten Sie das Mikroskop aus. • Entfernen Sie die Kondensorklappe (38.1). Das Prisma wird in eine der großen mit Führungs-nuten versehenen...
Page 156 - Kondensorköpfe
42 6. Montage Abb. 42 Einsetzen der Kondensorköpfe S1 1 Kondensorbasis 2 Zwischenring 3 Kondensorkopf Abb. 43 Einsetzen des Kondensorkopfes S28 Kondensorköpfe Es stehen 4 verschiedene Kondensorköpfe zurVerfügung: 1) S1/1.40 oil 2) S1/0.90 dry 3) S23/0.53 4) S28/0.55 Die Kondensorköpfe 3 und 4 werden...
Page 158 - Montage der Filter im Beleuchtungsarm; mit der Klemmschraube befestigen.; • Setzen Sie das Lampenhaus an die Durch-
44 6. Montage 6.8 Montage der Filter im Beleuchtungsarm Die Leica DMI-Serie ist generell mit einem Filter-magazin zur Aufnahme von 2 Filtern mit Ø 40 mmausgestattet. Die Filter sind in der Regel bereitswerkseitig in den Halter eingesetzt. Falls Sie Fil-ter nachrüsten und selbst montieren:• Schraube ...
Page 159 - Wechseln der 12V 100W Halogen-Glühlampe
45 6. Montage 6.10 Montage und Wechsel der Durchlicht-lampen: Lampenhaus 107 oder 107/2 Dieses Lampenhaus wird mit einer 12V 100WHalogenglühlampe verwendet, die bereits ein-gebaut ist.Soll die Lampe ausgewechselt werden, gehenSie folgendermaßen vor: Wechseln der 12V 100W Halogen-Glühlampe Achtung! A...
Page 161 - keine Wirkung
47 6. Montage Abb. 55 Booster-Linse Abb. 56 Booster-Linse im Stativ 1 Booster-Linse Sollte eine Booster-Linse zum Lieferumfang ge-hören, kann diese je nach Stativ nun in die hin-tere seitliche Stativöffnung rechts oder links ge-schoben werden. Der Booster-Schieber hat mehrere Stellungen:1. Schieber ...
Page 162 - Montage und Wechsel der Auflichtlampen
48 6. Montage 6.12 Montage und Wechsel der Auflichtlampen Leica DMI4000 B und Leica DMI6000 B:Es besteht generell bei den Lichtquelleneine Gefährdung durch Strahlung (Blen-dung, UV-Strahlung, IR-Strahlung). Lampenmüssen daher in geschlossenen Gehäusenbetrieben werden. Achten Sie darauf, dass das Lam...
Page 163 - * Bitte beachten Sie die Datenblätter der Lampenhersteller.
49 6. Montage Abb. 60 Lampenfassungen für Gasentladungslampe Hg 100 1 Obere Klemmung 2 Untere Klemmung 3 Kühlelement a b Hg 100 Xe 75 3 1 2 3 1 2 4 Einsetzen der Gasentladungslampen (Hg undXe) in das Lampenhaus 106z Hg- und Xe-Lampen werden mit separaten Vor-schaltgeräten betrieben.Bitte unbedingt d...
Page 166 - Bestückung der Auflicht-Revolverscheibe
52 6. Montage 6.13 Bestückung der Auflicht-Revolverscheibe Achtung: Diesen Abschnitt zunächst komplett lesen, be-vor Sie mit der Bestückung der Revolverscheibebeginnen.Leica DMI4000 B und Leica DMI6000 B:Auf der rechten Seite des Stativs befindet sichdie Fluoreszenzschublade. Bevor Sie dieseSchublad...
Page 168 - Einsetzen des Front-Modul Schiebers; beiden Anschlägen drehen lassen.
54 6. Montage Abb. 70 Mechanischer Polarisationshalter 1 Manueller Polarisator 2 Manueller Analysator Abb. 71 Kondensor mit motorischem Polarisator 6.14 Einsetzen des Front-Modul Schiebers Ist Ihr Mikroskop für den Integrierten Modu-lationskontrast oder für den Integrierten Pha-senkontrast vorbereit...
Page 169 - Durchlicht- und Auflichtanalysator; • Schieben Sie den Analysator bis zur Rastung
55 6. Montage Abb. 73 Montage des Analysators 1 Aufnahme 2 Analysator Durchlicht- und Auflichtanalysator • Entfernen Sie die Steckkappe (Abb. 72) auf der rechten Seite des Stativs (unter der Fluo-Schublade). • Schieben Sie den Analysator bis zur Rastung in die Aufnahme (Abb. 73.1). Abb. 72 Steckkapp...
Page 170 - • Schrauben Sie die Kamera auf.
56 6. Montage 6.16 Optionales Zubehör Kamera Anschluss einer KameraEine Kamera kann über einen c-Mount oderVario-Mounts adaptiert werden. • Setzen Sie den c-Mount bzw. den Vario- Mount auf einen der Kameraausgänge desMikroskops auf und befestigen Sie ihn mit derseitlichen Klemmschraube. • Schrauben ...
Page 171 - Anschluss an die Elektronikbox
57 6. Montage Abb. 78 Stativrückseite 1 RS232-Schnittstellen 2 2 x USB 3 4 x EXT. 4 XYZ-Control für Smart Move 5 Anschluss Elektronikbox 6 Kondensorkabel 7 Lampenversorgungskabel des Stativs Abb. 77 Rückseite CTR6000 1 Buchse für Netzkabel 2 Buchse XY-Stage für motorisierten Tisch 3 Buchse Direct in...
Page 173 - Intelligente Automatisierung
59 7. Inbetriebnahme 7. Inbetriebnahme 7.1 Funktionsprinzip (Leica DMI4000 B und Leica DMI6000 B) 1. Intelligente Automatisierung • Umschalten zwischen verschiedenen Kontrastverfahren auf Knopfdruck. Lichtringe, DIC- Prismen, etc. werden automatisch in den Strahlengang gebracht. • Das Mikroskop erke...
Page 177 - noch nicht; ein. Arbeiten Sie zunächst; zuerst die Elektronikbox; und; danach den
63 7. Inbetriebnahme Ist eine der Komponenten nicht ordnungsgemäßmontiert, erscheint eine Fehlermeldung auf demLeicaDisplay.Siehe Kapitel Trouble Shooting → S. 102. Die mikroskopischen Komponenten wie Blen-den, Kondensor, Licht-und Phasenringe sind be-reits werkseitig vorzentriert. Durch Transportun...
Page 179 - Die Funktionstasten am Stativ; • Licht Intensität über das Potentiometer (links
65 7. Inbetriebnahme Hinweis: Änderungen der Lichtintensität sowie der Ein-stellung von Apertur- und Leuchtfeldblendewerden für das jeweilige Objektiv und Kontrast-verfahren abgespeichert. Abb. 84 Fest definierte Funktionstasten (linke Stativseite) 1 Wechsel Durchlicht/Auflicht 2 Aperturblende 3 Lic...
Page 180 - * Abkürzungen siehe Seite 62
66 7. Inbetriebnahme Variable Funktionstasten am Stativ Passend zu Ihrer Mikroskopausrüstung erfolgtwerkseitig eine Vorbelegung der variablenFunktionstasten. Die Tasten sind entsprechendbeschriftet. Die Tastenbelegung entnehmen Siebitte dem „Identification Sheet“.Die Bedeutung der Abkürzungen entneh...
Page 181 - Setzen der Fokusschwelle.
67 7. Inbetriebnahme Funktionstasten am Frontbedienfeld (Abb. 87) 100% des Lichtes geht zum Okular(87.1). Toggelfunktion für die seitlichenPorts (87.2). Die Funktion ist abhän-gig von der jeweiligen Mikroskop-konfiguration.Hinweis:Umschaltung auf Bottom Port:Über die variablen Funktionstasten(nur Le...
Page 182 - • Wählen Sie am Kondensor die Hellfeld Position.
68 7. Inbetriebnahme Abb. 89 Fernsteuermodul SmartMove 1 Verfahren in X-Richtung 2 Verfahren in Y-Richtung 3 Fokuseinstellung 4 Individuelle Einstellung der Knopfhöhenposition 5 Variable Funktionstasten (werkseitig vorbelegt) 1 2 3 4 5 SET + Cube-Taste 1-6 (nur Leica DMI4000 B undLeica DMI6000 B) Be...
Page 185 - • Der letzte Schritt ist die; Anpassung der; • Stellen Sie die Aperturblende so ein, dass sie
71 7. Inbetriebnahme die Mitte des Sehfeldes (91.c). Anschließendöffnen Sie die Feldblende, bis das Bild annä-hernd das gesamte Bildfeld ausfüllt. Dieschwarzen Ränder des Bildes solltenrundherum den gleichen Abstand zum äuße-ren Rand des Sehfeldes haben. Falls nicht,zentrieren Sie das Bild erneut mi...
Page 187 - • Setzen Sie anstelle eines Okulars das Einstell-
73 7. Inbetriebnahme 7.7 Phasenkontrastringe überprüfen Ist Ihr Mikroskop für die Verwendung vonPhasenkontrast ausgerüstet, sind im Kondensordie zu den Objektiven passenden Lichtringe ein-gebaut.Die Lichtringe sind bereits werkseitig zentriert.Die Zentrierung sollte jedoch noch einmal über-prüft wer...
Page 205 - COMBI
91 8. Bedienung 8.4 Kombi-Verfahren (Leica DMI4000 und DMI6000 B) Je nach Ausstattung Ihres Mikroskops sind biszu zwei Kombi-Verfahren möglich: FLUO/PH und FLUO/DIC • Wählen Sie das Kombi-Verfahren. Drücken Sie dazu die variable Taste COMBI . Alternativ: Drücken Sie die variable Taste CHANGE COMBI ....
Page 206 - Stativseite benutzt wer-
92 8. Bedienung 8.5 Fokussierung Leica DMI3000 B und Leica DMI4000 B:Über die linken Fokusräder können sowohlGrob- als auch Feinfokussierung vorgenommenwerden über das rechte Fokusrad kann eineFeinfokussierung vorgenommen werden (essteht auch eine rechts und links vertauschteVersion des DMI3000 B zu...
Page 213 - Vergrößerungswechsler; Vergrößerung wechseln über; Feste Funktionstasten am Stativ
99 8. Bedienung 8.11 Vergrößerungswechsler Leica DMI3000 B:Optional kann ein mechanischer Vergröße-rungswechsler eingesetzt werden. Die folgen-den Vergrößerungsfaktoren können wahlweiseeingestellt werden: 1,5x ; 1,6x oder 2x Über einen Schieber kann zwischen 1x und demVergrößerungsfaktor geschaltet ...
Page 214 - Licht einstellen über
100 8. Bedienung Licht einstellen über Feste Funktionstasten am Stativ Variable Funktionstasten am Stativ und SmartMove Software Leica Application Suite (LAS) Abb. 118 Feste Funktionstasten, linke Stativseite 1 Aperturblende 2 Durchlicht/Auflicht 3 Leuchtfeldblende 4 Lichtintensität 1 2 3 4 8.11 Lic...
Page 223 - Wichtigste Verschleiß- und Ersatzteile
109 11. Wichtigste Verschleiß- und Ersatzteile Bestell-NummerSach-Nummer Bezeichnung Verwendung für Ersatzlampen11 500 974 Halogenglühlampe 12 V 100 W Lampenhaus 107/2 11 500 137 Hg-Höchstdrucklampe 50 W Lampenhaus 106 z 11 500 138 Hg-Höchstdrucklampe 100 W Lampenhaus 106 z 11 500 321 Hg-Höchstdruck...
Page 225 - Abkürzungen und Piktogramme
111 13. Abkürzungen und Piktogramme Kontrastverfahren Vergrößerung Beleuchtung Ports/Okular Fokus Untere Fokusschwelle nicht gesetzt Untere Fokusschwelle gesetzt Fokusposition nicht gesetzt Fokusposition gesetzt Shutter auf Shutter zu Durchlichtfilter Feldblende, rechteckig Feldblende, rund Aperturb...
Page 227 - bkürzungen 99
113 14. Index 14. Index A bkürzungen 99 Aktive Ports 60Analysator 50, 51Analysatoraufnahme 19Anwahl der Ports 84Aperturblende 16, 18, 61, 76, 90 Auflicht-Fluoreszenz 68Auflicht-Lampenhaus 19Auflicht-Revolverscheibe 48Auflichtachse 12Aufstellungsort 23Augenabstand 84 B edienelemente 15, 55 Beleuchtun...