Page 3 - CONTENTS
© Adam Equipment Company 2013 1 CONTENTS 1.0 INTRODUCTION ...................................................................................................... 3 2.0 SPECIFICATIONS .................................................................................................... 4 3.0 INSTALLATIO...
Page 6 - SPECIFICATIONS
© Adam Equipment Company 2013 4 2.0 SPECIFICATIONS EC Type Approved Models (Type Approval Number T7476/TC7477) WBW 3M WBW 6M WBW 15M Kilograms Max 3 kg 6 kg 15 kg e = 0.001 kg 0.002 kg 0.005 kg Grams Max 3000 g 6000 g 15000 g e = 1 g 2 g 5 g NTEP Type Approved Models (Type Approavl number 10-082) WB...
Page 8 - Stainless steel top pan
© Adam Equipment Company 2013 6 3.0 INSTALLATION 3.1 UNPACKING Remove the scale from the packing carefully. Inside the box you will find everything needed to start using the scale- Pre-assembled scale Stainless steel top pan Power Supply Module Operators Manual 3.2 LOCATING • The scales should not b...
Page 10 - KEYS; NUMERIC ENTRY METHOD; To set a value when required, use the keys as given below-
© Adam Equipment Company 2013 8 4.0 KEYPAD KEYS PRIMARY FUNCTION SECONDARY FUNCTION [O/I] Switches the scale on or off [Zero] Sets the zero point for all subsequent weighing. The display shows zero. Escapes from a parameter or an option without changing the value. [Tare] Tares the scale and stores t...
Page 11 - SYMBOLS AND INDICATORS; Net; Weight
© Adam Equipment Company 2013 9 5.0 DISPLAY The LCD will display a decimal point in the WBW..a and WBW..aM models or a comma as shown in the WBW and WBW..M models. 6.0 SYMBOLS AND INDICATORS The LCD has unique symbols to indicate the following: 0 The display is at Zero The scale is Stable Net Net ...
Page 12 - BATTERY OPERATION; S3 AoF
© Adam Equipment Company 2013 10 7.0 BATTERY OPERATION • The scales can be operated from the battery, if desired. The battery life is approximately 50 hours. • When the battery needs charging a symbol on the display will turn on. The battery should be charged when the symbol is on. The scale will st...
Page 13 - OPERATIONS; You can press the [Zero] key at any time to set the zero point; The scale has an automatic re-zeroing function to account for; Press [Tare] when the reading is stable. The weight that was; As a product is added only the weight of the product will be
© Adam Equipment Company 2013 11 10.0 OPERATIONS 10.1 ZEROING • You can press the [Zero] key at any time to set the zero point from which all other weighing and counting is measured. This will usually be necessary when the platform is empty. When the zero point is obtained the display will show the ...
Page 14 - Mass on the balance is above the high limit
© Adam Equipment Company 2013 12 NOTE: When the container is removed a negative value will be shown. If the scale was tared just before removing the container, this value is the gross weight of the container plus all products which were removed. The zero indicator will also be on as the platform is ...
Page 16 - SECURITY SEALS; Method of sealing
© Adam Equipment Company 2013 14 12.0 CALIBRATION 12.1 WBW..M and WBW..aM APPROVED SCALES, CALIBRATION The scales are sealed to prevent unauthorised access to the circuit boards and load cell. One method calibration of the WBW-M scale is accomplished by breaking the security seals to gain access to ...
Page 17 - CALIBRATION COUNTER; ” using the numeric entry method (see section; Remove all weight from the pan and then press the [Tare] key; when the scale is stable.
© Adam Equipment Company 2013 15 WBW..aM SECURITY SEALS: The device is equipped with a Category 1 Audit Trail. In addition a lead &wire and/or a paper security seal may be installed. To access the audit trail press the [Zero] button for 3 seconds. The display will flash the calibration counter a...
Page 18 - The scale will calibrate to the mass and then return to weighing.; NOTE: If an error message “; FAIL H
© Adam Equipment Company 2013 16 • Press the [Tare] key when the stable indicator is on. • The scale will calibrate to the mass and then return to weighing. • Remove the calibration weight as soon as calibration is complete. NOTE: If an error message “ FAIL H ” or “ FAIL L ” is shown during calibrat...
Page 19 - PARAMETER SETTING; CHECK WEIGHING PARAMETERS; ” in this section. Enter; Setting the LED
© Adam Equipment Company 2013 17 13.0 PARAMETER SETTING This section allows the user to access the parameters for customising the scale. The parameters are split into 2 groups- 1. Check weighing Parameters 2. Scale Parameters In the following sections use the [Limit] key to scroll through the option...
Page 20 - These parameters are used to control the operation of the scale.
© Adam Equipment Company 2013 18 “ donE ”. If the old parameter is other than “0000”, enter the old one when “ P _ _ _ _ ” is displayed and then enter the new password twice when asked. W hen complete, it will display “ donE ”. 13.2 SCALE PARAMETERS To enter this section press and hold the [Unit] ke...
Page 21 - DESCRIPTION; Err 4; High limit input error; FAIL H; or; FAIL L; Calibration error
© Adam Equipment Company 2013 19 14.0 ERROR CODES During the initial power-on testing or during operation, the scale may show an error message. The meaning of the error messages is described below. If an error message is shown, repeat the step that caused the message. If the error message is still s...
Page 22 - REPLACEMENT PARTS AND ACCESSORIES; Power Supply Module; SERVICE INFORMATION
© Adam Equipment Company 2013 20 15.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact your supplier or Adam Equipment. A partial list of such items is as follows- • Power Supply Module • Replacement Battery • Stainless Steel Pan • 16.0 SERVICE INFORMAT...
Page 23 - WARRANTY INFORMATION
© Adam Equipment Company 2013 21 17.0 WARRANTY INFORMATION Adam Equipment offers Limited W arranty (Parts and Labour) for the components failed due to defects in materials or workmanship. W arranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the pur...
Page 24 - Adam Equipment
© Adam Equipment Company 2013 22 Adam Equipment ADAM EQUIPMENT, BOND AVENUE, DENBIGH EAST INDUSTRIAL ESTATE, MILTON KEYNES, MK1 1SW, U.K. Tel: (01908) 274545 Fax: (01908) 641339 Intl Tel: -44 1908 -274545 Intl Fax: -44 1908 641339 E-Mail Address: [email protected] Declaration of Conformity Ve...
Page 25 - APPENDIX
Page 27 - FCC COMPLIANCE
© Adam Equipment Company 2013 Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following the provisions of the below stated directives: 2011/65/EC (RoHS) Restriction of the use of certain hazardous substances in electri...