National Instruments NI PXI-5422 - Manual

National Instruments NI PXI-5422

National Instruments NI PXI-5422 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
14 Page 14
15 Page 15
16 Page 16
17 Page 17
18 Page 18
19 Page 19
20 Page 20
21 Page 21
22 Page 22
23 Page 23
24 Page 24
25 Page 25
26 Page 26
27 Page 27
28 Page 28
29 Page 29
30 Page 30
31 Page 31
32 Page 32
33 Page 33
34 Page 34
35 Page 35
36 Page 36
37 Page 37
38 Page 38
39 Page 39
40 Page 40
41 Page 41
42 Page 42
43 Page 43
44 Page 44
45 Page 45
46 Page 46
47 Page 47
48 Page 48
49 Page 49
50 Page 50
51 Page 51
52 Page 52
53 Page 53
54 Page 54
55 Page 55
56 Page 56
57 Page 57
58 Page 58
59 Page 59
60 Page 60
61 Page 61
62 Page 62
63 Page 63
Page: / 63

Table of Contents:

  • Page 2 – Conventions; File»Page Setup»Options; Read Me First: Safety and Radio-Frequency; bold; italic; Introduction
  • Page 3 – External Calibration; Caution
  • Page 4 – External Calibration Options; Software and Documentation Requirements; Software; Self-Calibration Procedures
  • Page 5 – File Name and Location
  • Page 6 – Documentation; NI Signal Generators Getting Started Guide; Start»Programs»National Instruments»NI-FGEN»; Verifying the DC Gain and Offset; MAX; My System»Devices and Interfaces»NI-DAQmx Devices
  • Page 7 – FGEN Soft Front Panel
  • Page 8 – calibration test limits; Complete Calibration
  • Page 10 – Optional Calibration; section for more information.
  • Page 12 – External Calibration Requirements; Test Equipment
  • Page 13 – External Calibration Procedures; External; Writing Your Calibration Procedure; Functions»; Calibration Procedures in LabWindows/CVI
  • Page 14 – Verifying NI PXI-5422 Specifications; Note
  • Page 16 – Verifying the Oscillator Frequency Accuracy
  • Page 18 – Verifying the DC Gain and Offset Accuracy; Verifying the Main Analog Path Gain
  • Page 21 – to set the offset (niFgen; Arbitrary Waveform Output»Arbitrary Waveform; : The session handle returned from; channelName; Table 4, changing the
  • Page 22 – Table
  • Page 25 – Verifying the Main Analog Path Offset
  • Page 26 – Table 5
  • Page 29 – Verifying the Gain of the Direct Path
  • Page 31 – Gain
  • Page 33 – resourceName; action
  • Page 35 – Initializing the External Calibration Session; Adjusting the Analog Output; Initializing Analog Output Calibration
  • Page 37 – Adjusting the Main Path Pre-Amplifier Offset
  • Page 38 – Analog Filter Enable
  • Page 39 – Pre-Amplifier Attenuation
  • Page 41 – Current Configuration; Adjusting the Main Path Pre-Amplifier Gain
  • Page 45 – Adjusting the Main Path Post-Amplifier Gain and Offset; Analog Path
  • Page 46 – Post-Amplifier Attenuation
  • Page 48 – Adjusting the Direct Path Gain
  • Page 52 – Adjusting the Oscillator Frequency
  • Page 55 – Adjusting the Calibration ADC
  • Page 60 – Closing the External Adjustment Session
  • Page 61 – Calibration Utilities; To retrieve data using MAX, complete the following steps:; Select the; Calibration Tab; on the lower right corner.; FGEN SFP; To retrieve data using the FGEN SFP, complete the following steps:; Edit»Device Configuration; Open the Calibration dialog box (; Utility»Calibration
  • Page 62 – NI Signal Generators Help
  • Page 63 – Where to Go for Support; you have access to everything from
Loading the manual

CALIBRATION PROCEDURE

NI PXI-5422

This document contains step-by-step instructions for writing a calibration
procedure for the NI PXI-5422 arbitrary waveform generator.

Contents

Conventions ............................................................................................ 1
Introduction ............................................................................................. 2

What Is Calibration? ........................................................................ 2

Self-Calibration ........................................................................ 2
External Calibration .................................................................. 2

Why Should You Calibrate? ............................................................ 2
How Often Should You Calibrate? .................................................. 3

Software and Documentation Requirements........................................... 3

Software ........................................................................................... 3
Documentation ................................................................................. 5

Self-Calibration Procedures .................................................................... 5

MAX ................................................................................................ 5
FGEN Soft Front Panel .................................................................... 6
NI-FGEN ......................................................................................... 6

External Calibration Options .................................................................. 7

Complete Calibration ....................................................................... 7
Optional Calibration ........................................................................ 9

External Calibration Requirements ......................................................... 11

Test Equipment ................................................................................ 11
Test Conditions ................................................................................ 11

External Calibration Procedures ............................................................. 12

Writing Your Calibration Procedure................................................ 12

Calibration Procedures in LabVIEW ........................................ 12
Calibration Procedures in LabWindows/CVI ........................... 12
Calibration Procedures in C ...................................................... 12

Verifying NI PXI-5422 Specifications ............................................ 13

Verifying the Oscillator Frequency Accuracy .......................... 15
Verifying the DC Gain and Offset Accuracy ........................... 17

Adjusting the NI PXI-5422 .............................................................. 31

Initializing the External Calibration Session ............................ 35

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Conventions; File»Page Setup»Options; Read Me First: Safety and Radio-Frequency; bold; italic; Introduction

NI PXI-5422 Calibration Procedure 2 ni.com Adjusting the Analog Output ....................................................35Adjusting the Oscillator Frequency...........................................52Adjusting the Calibration ADC .................................................55Closing the Ext...

Page 3 - External Calibration; Caution

© National Instruments Corporation 3 NI PXI-5422 Calibration Procedure What Is Calibration? Calibration consists of verifying the output accuracy of a device and correcting for any output error. Verification is measuring the performance of a device and comparing the results to the specifications of ...

Page 4 - External Calibration Options; Software and Documentation Requirements; Software; Self-Calibration Procedures

NI PXI-5422 Calibration Procedure 4 ni.com The output accuracy requirements of your application determine how often you should externally calibrate the NI PXI-5422. NI recommends that you perform a complete external calibration at least once every two years. You can shorten this interval based on th...

Other National Instruments Models

All National Instruments Other