Page 2 - Note
NI 9229/9239 2 ni.com This doc u ment describes how to u se the National Instr u ments 9229 and National Instr u ments 9239 and incl u des specifications and terminal assignments. In this doc u ment, the NI 9229 and NI 9239 are referred to incl u sively as the NI 9229/9239. Visit ni.com/info and ent...
Page 3 - Safety Guidelines; Hot Surface; Safety Guidelines for Hazardous Voltages; not; Caution
© National Instruments Corp. 3 NI 9229/9239 Safety Guidelines Operate the NI 9229/9239 only as described in these operating instr u ctions. Hot Surface This icon denotes that the component may be hot. To u ching this component may res u lt in bodily inj u ry. Safety Guidelines for Hazardous Voltages...
Page 5 - Safety Guidelines for Hazardous Locations
© National Instruments Corp. 5 NI 9229/9239 Safety Guidelines for Hazardous Locations The NI 9229/9239 is s u itable for u se in Class I, Division 2, Gro u ps A, B, C, D, T4 hazardo u s locations; Class I, Zone 2, AEx nA IIC T4, and Ex nA IIC T4 hazardo u s locations; and nonhazardo u s locations on...
Page 6 - Special Conditions for Hazardous Locations Use in Europe
NI 9229/9239 6 ni.com Caution For Zone 2 applications, install a protection device between the inp u t signal and the NI 9229/9239 inp u t terminal. The device m u st prevent the inp u t channel-to-COM voltage from exceeding 85 V if there is a transient overvoltage condition. Caution For Zone 2 appl...
Page 7 - Special Conditions for Marine Applications
© National Instruments Corp. 7 NI 9229/9239 Special Conditions for Marine Applications Some mod u les are Lloyd’s Register (LR) Type Approved for marine applications. To verify Lloyd’s Register certification, go to ni.com/certification and search for the LR certificate, or look for the Lloyd’s Regis...
Page 9 - Specifications
© National Instruments Corp. 9 NI 9229/9239 The NI 9229/9239 has fo u r 2-terminal detachable screw-terminal connectors. Note Yo u m u st u se 2-wire ferr u les to create a sec u re connection when connecting more than one wire to a single terminal on the NI 9229/9239. Yo u can connect gro u nd-refe...
Page 12 - Wiring for High-Vibration Applications
NI 9229/9239 12 ni.com Wiring for High-Vibration Applications If an application is s u bject to high vibration, National Instr u ments recommends that yo u either u se ferr u les to terminate wires to the detachable screw-terminal connector or u se the NI 9971 backshell kit to protect the connection...
Page 13 - Passband
© National Instruments Corp. 13 NI 9229/9239 Understanding NI 9229/9239 Filtering The NI 9229/9239 u ses a combination of analog and digital filtering to provide an acc u rate representation of in-band signals while rejecting o u t-of-band signals. The filters discriminate between signals based on t...
Page 15 - Stopband
© National Instruments Corp. 15 NI 9229/9239 Stopband The filter significantly atten u ates all signals above the stopband freq u ency. The primary goal of the filter is to prevent aliasing. Therefore, the stopband freq u ency scales precisely with the data rate. The stopband rejection is the minim ...
Page 17 - Sleep Mode
© National Instruments Corp. 17 NI 9229/9239 Sleep Mode This mod u le s u pports a low-power sleep mode. S u pport for sleep mode at the system level depends on the chassis that the mod u le is pl u gged into. Refer to the chassis man u al for information abo u t s u pport for sleep mode. If the cha...
Page 18 - Input Characteristics
NI 9229/9239 18 ni.com Specifications The following specifications are typical for the range – 40 to 70 °C u nless otherwise noted. All voltages are relative to the AI– signal on each channel u nless otherwise noted. The specifications are the same for the NI 9229 and the NI 9239 u nless otherwise n...
Page 19 - Module
© National Instruments Corp. 19 NI 9229/9239 Data rate range ( f s ) u sing external master timebase Minim u m .................................... 390.625 S/s Maxim u m ................................... 51.2 kS/s Data rates 1 ( f s ) ................................... , n = 1, 2, …, 31 Inp u t v...
Page 20 - Acc; Measurement Conditions
NI 9229/9239 20 ni.com Acc u racy, NI 9229 Measurement Conditions Percent of Reading (Gain Error) Percent of Range * (Offset Error) Calibrated max (– 40 to 70 °C) ±0.13% ±0.05% Calibrated typ (25 °C, ±5 °C) ±0.03% ±0.008% Uncalibrated max (– 40 to 70 °C) ±1.2% ±0.55% Uncalibrated typ (25 °C, ±5 °C) ...
Page 21 - Stability
© National Instruments Corp. 21 NI 9229/9239 Acc u racy, NI 9239 Inp u t noise NI 9229....................................... 320 μ V rms NI 9239....................................... 70 μ V rms Stability Gain drift .................................... ±5 ppm/°COffset drift NI 9229 ...................
Page 23 - Bellcore Iss; Contact NI for Bellcore MTBF specifications
© National Instruments Corp. 23 NI 9229/9239 Alias-free bandwidth ........................ 0.453 · f s –3 dB prefilter bandwidth( f s = 50 kS/s)...................................... 24.56 kHz CMRR ( f in = 60 Hz) NI 9229....................................... 116 dBNI 9239.............................
Page 25 - Safety; Safety Voltages
© National Instruments Corp. 25 NI 9229/9239 Safety Safety Voltages Connect only voltages that are within the following limits. Channel-to-earth gro u nd isolation Contin u o u s ................................. 250 V rms , Meas u rement Category II Withstand.................................... 2,3...
Page 27 - Safety Standards; Hazardous Locations
© National Instruments Corp. 27 NI 9229/9239 Safety Standards This prod u ct is designed to meet the req u irements of the following standards of safety for electrical eq u ipment for meas u rement, control, and laboratory u se: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and oth...
Page 28 - Environmental
NI 9229/9239 28 ni.com Environmental National Instr u ments C Series mod u les are intended for indoor u se only b u t may be u sed o u tdoors if installed in a s u itable enclos u re. Refer to the man u al for the chassis yo u are u sing for more information abo u t meeting these specifications. Op...
Page 29 - Shock and Vibration; Electromagnetic Compatibility
© National Instruments Corp. 29 NI 9229/9239 Shock and Vibration To meet these specifications, yo u m u st panel mo u nt the system and either affix ferr u les to the ends of the terminal wires or u se the NI 9971 backshell kit to protect the connections. Operating vibration Random (IEC 60068-2-64)....
Page 30 - CE Compliance; Environmental Management
NI 9229/9239 30 ni.com Note For EMC compliance, operate this device with shielded cabling. CE Compliance This prod u ct meets the essential req u irements of applicable E u ropean directives, as amended for CE marking, as follows: • 2006/95/EC; Low-Voltage Directive (safety) • 2004/108/EC; Electroma...
Page 31 - NI and the; Waste Electrical and Electronic Equipment (WEEE); EU Customers; must; RoHS; National Instruments; about China RoHS compliance, go to
© National Instruments Corp. 31 NI 9229/9239 For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment . This page contains the environmental reg u lations and directives with which NI complies, as well as other environmental information not incl u ...
Page 32 - Calibration; Where to Go for Support
NI 9229/9239 32 ni.com Calibration Yo u can obtain the calibration certificate and information abo u t calibration services for the NI 9229/9239 at ni.com/ calibration . Calibration interval ........................... 1 year Where to Go for Support The National Instr u ments Web site is yo u r comp...
Page 33 - Belgi
© National Instruments Corp. 33 NI 9229/9239 A u stralia 1800 300 800, A u stria 43 662 457990-0, Belgi u m 32 (0) 2 757 0020, Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech Rep u blic 420 224 235 774, Denmark 45 45 76 26 00, Finland 358 (0) 9 725 72511, France 01 57 66 24...