Page 2 - User Guide
VPAP™ IVVPAP™ IV ST POSITIVE AIRWAY PRESSURE DEVICE User Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorised Representative) 96 Milton Park A...
Page 3 - Contents
Contents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications 1 Adverse Effects 1 The VPAP System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Masks 2 Humidifier 2 ResScan Data Card 2 S...
Page 4 - gli; Introduction; Contraindications; WARNING; The V
1 Introduction En gli sh Introduction The VPAP TM IV and the VPAP TM IV ST are intended to provide non-invasive ventilation for patients with respiratory insufficiency or obstructive sleep apnoea (OSA), in the hospital or home. Contraindications Positive airway pressure therapy may be contraindicate...
Page 5 - ResScan Data Card
2 Masks The following ResMed mask systems are recommended for use with these devices: For information on using masks, see your mask manual. For the latest available masks, see www.resmed.com. Humidifier If you are experiencing dryness of the nose, throat or mouth, the H4i heated humidifier is recomm...
Page 6 - Using the VPAP IV and VPAP IV ST; Using the VPAP; Setting up your VPAP
3 Using the VPAP IV and VPAP IV ST En gli sh Using the VPAP IV and VPAP IV ST Setting up your VPAP 1 Connect the power cord to the socket at the rear of your VPAP. 2 Plug the other end of the power cord into the power outlet. 3 Connect one end of the air tubing firmly onto the air outlet. 4 Connect ...
Page 7 - Attaching a H4i Humidifier; How to Use the Control Panel; device
4 Attaching a H4i Humidifier The H4i humidifier attaches to the front of a VPAP IV or VPAP IV ST device to provide heated humidification. These devices automatically detect the presence of the H4i and no other accessories are required for its use. For more information on using your H4i, please refer...
Page 8 - Using the Menus
5 Using the VPAP IV and VPAP IV ST En gli sh Using the Menus The VPAP IV and VPAP IV ST provide a set of functions which are arranged in menus and submenus. Via the LCD screen, the menus and submenus allow you to view and change the settings for a particular function. To navigate and make selections...
Page 10 - Stopping Treatment; Using the Mask-Fitting feature; and VPAP IV ST
7 Using the VPAP IV and VPAP IV ST En gli sh 6 After starting therapy, an introductory treatment screen will display: Stopping Treatment To stop treatment at any time, remove your mask and press . If your clinician has enabled SmartStart/Stop, simply remove your mask and treatment will end. Note: Sm...
Page 12 - Using the Data Card; Copying Data onto a Data Card; Updating Settings on your VPAP
9 Using the VPAP IV and VPAP IV ST En gli sh Using the Data Card If your clinician needs to review your treatment, they will ask you to use the Data Card to copy data from your VPAP device and return the card to them. Copying Data onto a Data Card 1 Switch on your VPAP and wait until you see the sta...
Page 13 - Traveling with the VPAP IV and VPAP IV ST; International Use; Use with DC Power
10 Traveling with the VPAP IV and VPAP IV ST International Use Your VPAP flow generator has an internal power adapter that enables it to operate in other countries. It will operate on power supplies of 100–240V and 50–60Hz. No special adjustment is necessary, but you will require an approved power c...
Page 14 - Cleaning and Maintenance; Daily; Monthly; Replacing the Air Filter
11 Cleaning and Maintenance En gli sh Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask and humidifier manuals for detailed instructions regarding the care of those devices. Daily Disconnect the air tubing and hang it i...
Page 15 - Servicing; CAUTION
12 Servicing This device is intended to provide safe and reliable operation when operated and maintained in accordance with the instructions provided by ResMed. To ensure that your device continues to provide optimum performance it is recommended that this product be inspected by an authorised ResMe...
Page 16 - Troubleshooting
13 Troubleshooting En gli sh Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device. No display Insufficient air delivered from the VPAP device Device does not start when you ...
Page 18 - Technical Specifications; Symbols Which Appear On The Device
15 Technical Specifications En gli sh Technical Specifications Note: • The manufacturer reserves the right to change these specifications without notice. • Pressure may be displayed in cm H 2 0 or hPa. Symbols Which Appear On The Device Caution; Follow instructions for use; Drip proof; Type CF equip...
Page 19 - General Warnings and Cautions; Warnings
16 General Warnings and Cautions ! Warnings • Advice contained in this manual should not supersede instructions given by the prescribing physician. • A patient should not connect a device to the data communication port unless instructed to do so by their health care provider or physician. Only ResMe...
Page 22 - Limited Warranty
19 Limited Warranty En gli sh Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Note: Some models are not available in a...