2 Contents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 VPAP™ Adapt SV 3 User/owner responsibility 3 Definitions 3 Medical Information 3 Quick Setup Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....
Page 3 - English; Introduction; VPAPTM Adapt SV; CAUTION; User/Owner Responsibility; Please read this manual carefully before use.; Definitions; WARNING; Medical Information; Intended Use
3 Introduction English Introduction VPAP™ Adapt SV The VPAP Adapt SV flow generator has been designed to be reliable, comfortable and easy to use. You can observe your treatment and make adjustments to settings using a keypad and LCD screen. ! CAUTION USA federal law restricts this device to sale by...
Page 5 - bloating; Quick Setup Guide; Switch on the VPAP Adapt SV. Make sure the mask is unblocked.
5 Quick Setup Guide English The following side effects may arise during a course of therapy with the VPAP Adapt SV: • drying of the nose, mouth, or throat • bloating • ear or sinus discomfort • eye irritation • skin rashes • chest discomfort. Quick Setup Guide 1 Attach the air tubing and pressure se...
Page 6 - VPAP Adapt SV System; VPAP Adapt SV Components
6 7 Remove your mask to stop treatment. If SmartStop is enabled, treatment will stop automatically. If the SmartStop function has been disabled, press the Start/Stop key once. VPAP Adapt SV System VPAP Adapt SV Components The VPAP Adapt SV flow generator is supplied with air tubing and a power cord....
Page 7 - tube clips (to hold the pressure sensor tube onto the air tubing); Masks
7 VPAP Adapt SV System English Figure 2: Rear of VPAP Adapt SV See page 14 to connect a ResLink™. See page 30 to replace the air filter. Figure 3: Air tubing fully assembled All necessary air tubing components are packed with the VPAP Adapt SV: • a 6’6" (2 m) length of air tubing • a pressure se...
Page 8 - Assembling the VPAP Adapt SV System; To add a humidifier see page 10.; Connect the power cord; Plug the free end of the power cord into a power outlet.; CAUTIONS
8 Assembling the VPAP Adapt SV System Complete the following steps to assemble the VPAP Adapt SV system. Make sure the area around the unit is clean (dust free) and clear of bedding, clothes and any other potential blockages. To add a humidifier see page 10. ! WARNING Use only the air tubing system ...
Page 9 - To attach the pressure sensor tube to the unit,; screw the Luer connector onto; Attach the mask to the air tubing.; Connect the mask system to the free end of the air tubing.
9 VPAP Adapt SV System English 2 Connect air tubing Connect the air tube firmly onto the air outlet at the front of the unit . Figure 4: Front view of VPAP Adapt SV To attach the pressure sensor tube to the unit, screw the Luer connector onto the socket on the side of the unit. Figure 5: Twisting th...
Page 10 - Humidifiers; When using a humidifier, stop the VPAP Adapt SV by using the; Be very careful not to allow water to enter the VPAP Adapt SV; as this could affect therapy and/or damage the device.; Connecting a Humidifier; HumidAire 2i; HumidAire
10 Humidifiers A humidifier may be required if you are experiencing dryness of the nose, throat, or mouth. The following ResMed humidifiers are compatible with the VPAP Adapt SV: Figure 7: Humidifiers compatible with the VPAP Adapt SV When using a humidifier, stop the VPAP Adapt SV by using the Star...
Page 11 - HumidAire 2iTM; HumidAire 2i User’s Manual
11 VPAP Adapt SV System English HumidAire 2i™ The HumidAire 2i provides heated humidification for the VPAP Adapt SV. To attach the HumidAire 2i, remove the VPAP Adapt SV front cover and attach the humidifier docking station and water chamber. No other accessories are required for use. Please refer t...
Page 12 - HumidAireTM; unit
12 HumidAire™ You will need a medium size 21" (52 cm) air tube to connect the VPAP Adapt SV unit to the HumidAire™ (see Figure 7). 1 Make sure both the HumidAire and the VPAP Adapt SV are turned off. Fill the HumidAire with water as described in the humidifier manual. Place the filled water cham...
Page 13 - Switch on the HumidAire and the VPAP Adapt SV.; HumidAire User’s Manual; Set up the HumidAire and the VPAP Adapt SV
13 VPAP Adapt SV System English 5 Connect the mask system to the free end of the long air tubing. The final assembly should look like this: 6 Switch on the HumidAire and the VPAP Adapt SV. ! WARNING Make sure that the power cords and plugs are in good condition and the equipment is not damaged. 7 Pe...
Page 14 - ResLinkTM; ResLink User’s Manual; Anti-bacterial Filter; Only hydrophobic anti-bacterial filters should be used with a
14 ResLink™ The ResLink is a device that records data onto a SmartMedia™ card during your therapy. The card can be returned to your clinician so that they can observe your progress and adjust flow generator parameters if necessary. The ResLink connects to the back of VPAP Adapt SV as shown. For full...
Page 15 - Using DC Power to Run the VPAP Adapt SV; Supplemental Oxygen
15 VPAP Adapt SV System English Using DC Power to Run the VPAP Adapt SV The VPAP Adapt SV can be run on DC in case of power failure. ResMed supplies separately a converter that allows a 12V or 24V battery to be connected to the DC input of the VPAP Adapt SV. When using a DC power supply, the power i...
Page 16 - Using the VPAP Adapt SV; Starting Treatment; key once to commence; Stopping Treatment; key; Left; If the SMART DATA - AUTO APPEAR option is set to ON the screens
16 Using the VPAP Adapt SV ! WARNING Before starting therapy with a new mask, select the correct mask type in the Settings menu (page 21). When adding or removing a new component, such as a mask, humidifier, or anti-bacterial filter, perform the Air Circuit Learn function (page 21). Starting Treatme...
Page 17 - Control Panel Functions; start or stop treatment; LCD screen; Press the key once to start or stop treatment.
17 Using the VPAP Adapt SV English Control Panel Functions The control panel of the VPAP Adapt SV has a keypad that allows you to: • start or stop treatment • adjust settings on the flow generator • mute or stop alarms • view treatment data. Figure 13: The VPAP Adapt SV control panel LCD screen Disp...
Page 18 - Adjusting Settings
18 Left key (green) • Enters a menu. • Confirms and applies settings. • Performs the function indicated by the text above it in the LCD (eg, ’enter’, ’change’). Right key (red) • Performs the function indicated by the text above it in the LCD (eg, ’exit’). • Cancels operations. • Extended hold: exit...
Page 19 - Example; You wish to change the setting for the mask you are using.
19 Using the VPAP Adapt SV English Example You wish to change the setting for the mask you are using. Figure 14: Using the keys to adjust settings MASK: FULL FACEchange exit LEARN CIRCUIT?yes exit SMARTSTOP: OFFchange exit SOUND LEVEL: LOWchange exit (ASV MODE) VPAP Adapt SVmenu SETTINGSenter exit R...
Page 20 - Menus
20 Menus The options available in the Settings menu will depend on the therapy you are receiving (either ASV or CPAP), as set by your clinician. Figure 15: Navigating the Patient menu. SETTINGSenter exit VPAP ADAPT SVmenu RESULTSenter exit SERVICINGenter exit RAMP: 20minchange exit MASK: FULL FACEch...
Page 21 - Settings Menu; Mask; Learn Circuit; To run this function:; Select ’start’ with the
21 Using the VPAP Adapt SV English Settings Menu You will only see the options relevant to your therapy. In CPAP (Continuous Positive Airway Pressure) mode, the clinician sets a pressure that will be constant throughout treatment. In ASV mode, the VPAP Adapt SV continually measures your breathing an...
Page 22 - Results Menu; Average Pressure
22 7. When the function is complete, a screen will appear saying that the circuit has been learned successfully. The next menu item is then displayed. Commence treatment as described in “Starting Treatment” on page 16. If there is a problem with the circuit a screen saying ’Invalid circuit’ will be ...
Page 23 - Options Menu; Smart Data; Backlight
23 Using the VPAP Adapt SV English Mask-fit Gives a rating of how well the mask was fitting during your last session on the VPAP Adapt SV. The stars indicate how good the fit was (more stars = better fit). Average Pressure Provides information on the average air pressure during your most recent sess...
Page 24 - Servicing Menu; Treatment Screens; QuickView; TARG is not a relevant parameter in CPAP mode.
24 Servicing Menu This menu displays the following information: • SN—the flow generator’s serial number • PCB—the flow generator’s PCB number (PCB—printed circuit board) • SW—the version of software currently installed on the flow generator. Treatment Screens There are three Treatment screens displa...
Page 25 - Alarms; Alarm Mute; Low Mask Pressure Alarm; If a low pressure alarm sounds:
25 Alarms English Alarms The VPAP Adapt SV unit is fitted with alarms to alert you to changes that will affect your treatment. The most common reason for an alarm to sound is because the system has not been properly assembled. Check that the air tubing and pressure sensor tube have been properly att...
Page 26 - Power Fail Alarm; If a power fail alarm sounds:; Remove the mask from your face if the power fails.; High Pressure / High Average Pressure Alarms; High Average Pressure Alarm:; If a high pressure alarm sounds:
26 Will clear when: • the low pressure condition is fixed • treatment is stopped by pressing the Start/Stop key. Power Fail Alarm Triggered by: • power failure • machine is disconnected or switched off while delivering treatment. If a power fail alarm sounds: • the flow generator stops delivering ai...
Page 27 - High Leak Alert; If a high leak alert sounds:; The High Leak alert can be turned off in the Leak Alert option in the; Insufficient Pressure Support Alarm; If an insufficient pressure support alarm sounds:; Flow Blocked Alarm; If a flow blocked alarm sounds:
27 Alarms English High Leak Alert Triggered by: • high mask leak (greater than 30 L/min) for more than 20 seconds • the mask being removed while SmartStop is disabled. If a high leak alert sounds: • adjust the mask to minimize leak • check that the air tube and sensor tube are connected properly. Wi...
Page 28 - Cleaning and Maintenance; Daily; Clean the mask according to the mask user instructions.; Weekly
28 Will stop when: • blockage is removed. Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this manual. ! CAUTION Do not wash the pressure sensor tube. If fluid enters the pressure sensor tube allow it to dry completely by hanging it in a clean place ...
Page 29 - Do not wash or dry the mask frame at a temperature above 176; Periodically; regularly for damage.
29 Cleaning and Maintenance English Figure 19: Air tubing assembly removed for cleaning ! CAUTION • Do not use bleach, chlorine-, alcohol-, or aromatic-based solutions (including all scented oils), moisturising or antibacterial soaps to clean the cushion, mask, air tubing, or the VPAP Adapt SV. Thes...
Page 30 - Air Filter; Fitting the filter; Remove the filter cover at the back of the flow generator.; Hypoallergenic Air Filter; Fit as described above in “Air Filter”.; Servicing
30 Air Filter Inspect the air filter every month and check if it is blocked by dirt or contains holes. The filter should be replaced every six months, or more often if the flow generator is operating in a dusty environment. ! WARNING Do not wash the air filter once it has become soiled. The air filt...
Page 31 - Troubleshooting
31 Troubleshooting English Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorized ResMed service center...
Page 32 - Glossary; breaths per minute
32 Glossary Alarms When more than one alarm is activated at the one time, the highest priority alarm will be displayed. Backlight The backlight turns off if no button has been pressed for two minutes. It can also be kept permanently on by selecting On in the Backlight screen of the Options menu. bpm...
Page 33 - Technical Specifications
33 Technical Specifications English SmartStop A feature that stops the VPAP Adapt SV automatically when you remove the mask. You can turn SmartStop on or off in the Settings menu. Tidal volume The volume of air, in liters, breathed in or out in one breath. Technical Specifications Dimensions (H x W ...
Page 34 - Alarm specifications
34 Alarm specifications • High maximum mask pressure alarm Stops air delivery at 30 cm H 2 O >700 msec. • High mean mask pressure alarm Stops air delivery when mean mask pressure exceeds 22 cm H 2 O (averaged over approx 1 minute). • Low mask pressure alarm Mask pressure is less than 3 cm H 2 O f...
Page 35 - Glossary of symbols; Read the relevant operating manual before using
35 Technical Specifications English Displayed values Maximum low impedance flow at stated pressures Glossary of symbols Read the relevant operating manual before using Type CF equipment AC switch Class II Drip proof Value Range Accuracy Display Resolution Proximal cuff Pressure -3 to 40 cm H 2 O 0.5...
Page 40 - Limited Warranty; Some models are not available in all regions.
40 Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Note: Some models are not available in all regions. If the product ...
Page 41 - Index
41 Index English Index A air circuit learn function 21 air filter 30 hypoallergenic 30 air pressure monitoring 23 air tubing attaching to mask 9 cleaning 28 connecting 9 air tubing system 6 , 7 , 9 alarm flow blocked 27 high leak 27 high pressure 26 low mask pressure 25 power fail 26 alarm LED 18 an...