Page 2 - User Guide; VPAP; POSITIVE AIRWAY PRESSURE DEVICES
Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia. Distributed by: ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA. ResMed (UK) Ltd 96 Milton Park Abingdon Oxfordshire OX14 4RY UK. See www.resmed.com for other ResMed locations worldwide. For patent...
Page 3 - English
English 1 S9 VPAP ST indications for use The S9 VPAP ST is intended to provide non-invasive ventilation for patients weighing more than 13 kg with respiratory insufficiency or obstructive sleep apnoea (OSA). The S9 VPAP ST is intended for home and hospital use. S9 VPAP S indications for use The S9 V...
Page 4 - Masks and humidifiers
English 2 • ear or sinus discomfort • eye irritation • skin rashes. Masks and humidifiers Recommended masks and humidifiers are available on www.resmed.com on the Products page under Service & Support. For information on using your mask or humidifier, refer to the manual supplied with your mask ...
Page 5 - Getting started; Do not wash the air tubing in a washing machine or dishwasher.
English 3 Getting started 1. Make sure the power is connected.2. Adjust the ramp time if required.3. Fit your mask as described in your mask user guide.4. To start therapy, simply breathe into the mask and/or press . 5. Lie down and arrange the air tubing so that it is free to move if you turn in yo...
Page 6 - Troubleshooting; Problem/Possible cause; No display
English 4 Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open these devices. Problem/Possible cause Solution No display Power is not connected. Ensure the power cable is connected and...
Page 7 - Device does not stop when you remove your mask; High temperature fault, refer to user manual; Air filter is blocked.
English 5 Device does not stop when you remove your mask SmartStart/Stop is disabled. Note: SmartStart/Stop is not available if Leak Alert is enabled. Enable SmartStart/Stop. SmartStart/Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system bein...
Page 8 - Displays error message:; This message only appears once. If you re-insert the
English 6 Displays error message: Check ResMed 30/90W Power Supply Unit and fully insert the connector The DC plug is partially inserted into the back of the device or inserted too slowly. Fully insert the DC plug. A non-ResMed power supply unit is connected to the device. Remove the power supply un...
Page 9 - SD card
English 7 SD card An SD card has been supplied to gather therapy data from your device and provide settings updates from your clinician. When instructed to do so, disconnect the device from the power outlet, remove your SD card, insert it in the protective folder and send it to your clinician. For m...
Page 10 - The air flow for breathing produced by this therapy
English 8 Operating temperature +5ºC to +35ºC Note: The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (40ºC) the device remains safe. Operating humidity 10–95% non-condensing Operating altitude Sea ...
Page 11 - Pneumatic flow path; Blower
English 9 Notes: • The manufacturer reserves the right to change these specifications without notice. • The temperature and relative humidity settings displayed for Climate Control are not measured values. • Check with your clinician/service provider before using the SlimLine or ClimateLine air tubi...
Page 12 - Environmental information; Servicing; Product
English 10 Environmental information WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and electronic equipment. These devices should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use appropriate collection, re...
Page 14 - WARNINGS; Read the entire manual before using the device.
English 12 WARNINGS • Read the entire manual before using the device. • Use the device only as directed by your physician or healthcare provider. • Use the device only for the intended use as described in this manual. Advice contained in this manual should not supersede instructions given by the pre...
Page 15 - Do not use electrically conductive or antistatic air tubings.; CAUTIONS
English 13 • Do not use electrically conductive or antistatic air tubings. • Do not use the air tubing if there are any visible signs of damage. • Only ResMed air tubing and accessories should be used with the device. A different type of air tubing or accessory may alter the pressure you actually re...