Page 2 - USER; ROW ENG
Global leaders in sleep and respiratory medicine www.resmed.com 248139/3 05 11 VPAP III ST-A USER ROW ENG Front Cover.qxd 9/11/2005 10:31 AM Page 2
Page 3 - VPAP; Illustrations
HUMIDIFIERS A B 1 2 4 Removable front cover SETTING UP THE VPAP III ST-A D 1 2 4 LCD screen Handle Control panel Air outlet Air filter cover Serialcommunicationsport (9-pin) Power switch Power sockets © 20 05 R esMed Ltd. © 20 05 R esMed Ltd. © 20 05 R esMed Ltd. © 20 05 R esMed Ltd. OTHER ACCESSORI...
Page 5 - User Manual
Page 7 - Contents
iii Contents Contents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 User/Owner Responsibility 1 Medical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 What the VPAP III ST-A is Intended for 1 ...
Page 8 - System Specifications
iv System Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...
Page 9 - Introduction; User/Owner Responsibility; Please read this manual carefully before use.; Medical Information; What the VPAP III ST-A is Intended for
1 Introduction Introduction The VPAP™ III ST-A is a bilevel pressure support ventilator specifically designed for non-invasive mask ventilation.This user manual contains the information you need for the correct use of your VPAP III ST-A. User/Owner Responsibility The user or owner of this system sha...
Page 11 - Adverse Effects
3 Medical Information Adverse Effects You should report unusual chest pain, severe headache or increased breathlessness to your physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.The following side effects may arise during the course of therapy w...
Page 13 - The VPAP III ST-A System; WARNING; Masks; ResMed mask system; Nasal Masks
5 The VPAP III ST-A System The VPAP III ST-A System Please refer to the illustrations in section A of the illustration sheet. Please identify and familiarize yourself with the following components of the VPAP III ST-A: • VPAP III ST-A front view ( A-1 ) • VPAP III ST-A rear view ( A-2 ) • Power cord...
Page 14 - Humidifiers; Accessories
6 Notes: • ResMed VPAP III ST-A has been designed and manufactured to provide optimum performance using ResMed vented mask systems. While other vented mask systems may be used, performance and data outputs may be affected. • Not all masks are available in all regions. Humidifiers Please refer to the...
Page 15 - Setting up the VPAP III ST-A System; Setting up the VPAP III ST-A; CAUTION; CAUTION
7 Setting up the VPAP III ST-A System Setting up the VPAP III ST-A System Please refer to the illustrations in section D of the illustration sheet. Setting up the VPAP III ST-A 1 Place the VPAP III ST-A on a flat surface near the head of your bed. If the device is placed on the floor, ensure that th...
Page 16 - Attaching a Humidifier; Attaching a HumidAire 2i Humidifier
8 ! WARNING • Make sure the power cord and plug are in good condition and the equipment is not damaged. • The air filter cover protects the device in the event of accidental liquid spillage onto the device. Ensure that the air filter and air filter cover are fitted at all times. 3 Connect one end of...
Page 17 - Attaching a Passover Humidifier
9 Setting up the VPAP III ST-A System To set up the VPAP III ST-A with the HumidAire:1. Fill the HumidAire with water as described in the humidifier manual.2. Place the filled water chamber inside the HumidAire. Connect the medium (52 cm) air tubing to the right connector port, and the long air tubi...
Page 18 - Using the LCD Screen and Keypad; LCD Screen; Green; Yellow; Red
10 ! WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged. 7. Navigate to the humidifier setting (if available) in the VPAP III ST-A menu and select "PASSOVER". See “How to Use the Detailed Menu (if enabled by your clinician)” on page 20. The ...
Page 19 - Keypad Keys; Keypad and LCD Backlight
11 Setting up the VPAP III ST-A System Keypad Keys The VPAP III ST-A keypad has the following keys: Keypad and LCD Backlight To assist you in adjusting the VPAP III ST-A, the keypad and LCD are equipped with a backlight. The LCD backlight comes on when the unit is turned on or when you press a key, ...
Page 21 - Starting Treatment; Ramp Time
13 Starting Treatment Starting Treatment The VPAP III ST-A should be assembled beside your bed with the air tubing and mask system connected. See “Setting up the VPAP III ST-A” on page 7. 1 Turn the main power switch at the back of the device to on (I) . When the VPAP III ST-A is turned on, a welcom...
Page 22 - Stopping Treatment; Using the HumidAire 2i Warm-up Feature; Using the Mask-Fitting Feature
14 Stopping Treatment To stop treatment at any time, remove your mask and press the Start/Stop key or if your clinician has enabled the SmartStart function, simply remove your mask and treatment will end (SmartStop is not applicable with the "Mir Full" mask setting). Using the HumidAire 2i W...
Page 25 - Cleaning and Maintenance; Daily Cleaning; Mask; Humidifier; Weekly Cleaning; Periodic Cleaning
17 Cleaning and Maintenance Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this section. Daily Cleaning Mask Clean the mask according to the instructions supplied with the mask. Air tubing Disconnect the air tubing from the VPAP III ST-A (and humidi...
Page 26 - Testing the Alarm; Replacing the Air Filter; Servicing
18 ! CAUTION Do not attempt to open the VPAP III ST-A. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorized service agent. Testing the Alarm Test the alarm weekly. See “Testing the Alarm” on page 29. Replacing the Air Filter Please refe...
Page 27 - How to Use the VPAP III ST-A Menus; Ramp Screen
19 How to Use the VPAP III ST-A Menus How to Use the VPAP III ST-A Menus The VPAP III ST-A provides a set of functions arranged in menus and submenus. Via the LCD screen, the menus and submenus allow you to view and change the settings for a particular function. You can access the menus regardless o...
Page 28 - Left; Right; Settings Summary Screens
20 Figure 1: VPAP III ST-A Standard Menu Series • To access the VPAP III ST-A menus:Press the Left key (menu) while the VPAP (or Ramp) screen is displayed. • To scroll through items within the menu:Press the Up/Down key • To exit out of the menu:Press the Right key (exit) Settings Summary Screens Th...
Page 30 - Settings Menu
22 • To access the VPAP III ST-A menus:Press the Left key (menu) while the VPAP (or Ramp) screen is displayed. • To scroll through items within a menu or submenu:Press the Up/Down key. • To enter a submenu:Press the Left key (enter). • To change a setting option for a function:1. Press the Left key ...
Page 31 - SmartStartTM; Settings for Mask Types
23 How to Use the VPAP III ST-A Menus SmartStart™ The VPAP III ST-A has a function called SmartStart which can be enabled by your clinician. If SmartStart is enabled, VPAP III ST-A will start automatically when you breathe into the mask and will stop automatically when you take your mask off. This m...
Page 32 - Results Menu; Options Menu
24 Results Menu Note: This menu appears only if at least one Smart Data™ option has been enabled by the clinician. See the VPAP III Smart Data Diary for further details. Table 4: Results Menu Options Menu Table 5: Options Menu Function Function Description Setting Options Mask Fit(Smart Data) Displa...
Page 33 - Servicing Menu; Treatment Screens
25 How to Use the VPAP III ST-A Menus Servicing Menu Table 6: Servicing Menu * These items also appear in the Standard Menu. Treatment Screens After starting treatment, you can display one of the treatment screens below. Press the Up/Down key to switch between views. Figure 3: Treatment screen 1 Fig...
Page 35 - Non-vented Mask Alarm Screen
27 How to Use the VPAP III ST-A Menus Non-vented Mask Alarm Screen If the non-vented mask alarm has been disabled by your clinician, the following screen is displayed after starting therapy. To exit this screen, press any key to display the treatment screens, or stop treatment. NON–VENTED MASK DISAB...
Page 37 - The Alarms; Alarm Mute Key; Alarms Troubleshooting
29 The Alarms The Alarms The VPAP III ST-A unit is fitted with alarms to alert you to changes that will affect your treatment. Alarm Mute Key You can mute an alarm by pressing the Alarm Mute key once. If the problem is still present, the alarm will sound again after two minutes. An alarm LED will re...
Page 41 - Helpful Hints; Starting out
33 Helpful Hints Helpful Hints Starting out Mouth Leaks If using a nasal mask, try to keep your mouth closed during treatment. Air leaks from your mouth can decrease the effectiveness of your treatment. If mouth leaks are a problem, a full face mask or chin strap may help. Contact your clinician or ...
Page 42 - Inverter or UPS
34 Please refer to the VPAP III Series Battery Power Guide for the recommended battery type and capacity, or contact ResMed's Technical Services division for your specific application. ResMed DC-24/30 Converter ResMed supplies the DC-24/30 Converter (product code: 22015) which is compatible with the...
Page 43 - Troubleshooting
35 Troubleshooting Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the VPAP III ST-A. Problem Possible Cause Solution No display. Power not connected or switch at back is not on. ...
Page 47 - Pressure Variation
39 System Specifications Pressure Variation Pressure Volume Curve Note: The manufacturer reserves the right to change these specifications without notice. 0.00 5.00 10.00 15.00 20.00 25.00 0 1 2 3 Breaths Pressure (cm H 2 O) 10 BPM15 BPM20 BPM Max Pressure 2/3 Max Pressure 1/3 Max Pressure Max Press...
Page 48 - Symbols which may Appear on the Product; Environmental information
40 Symbols which may Appear on the Product Attention, consult accompanying documents Class II equipment Type CF equipment Drip proof Start/Stop Mask-Fit Alarm LEDs Alarm Mute QuickView AC switch only Environmental information WEEE 2002/96/EC is a European Directive that requires the proper disposal ...
Page 49 - Guidance and Manufacturer’s Declaration - Electromagnetic
41 System Specifications Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity Guidance and manufacturer’s declaration – electromagnetic emissions The VPAP III ST-A is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP...
Page 53 - Limited Warranty
45 Limited Warranty Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Note: Some models are not available in all regions...
Page 55 - Index
47 Index Index A AC locking clip 7Accessories 6Air filter 18, 37air tubing 37 cleaning 17 B Back-light, LCD and Keypad 10, 11battery power 33 C Cautions 2cleaning periodic 17weekly 17 Cleaning and Maintenance 17cleaning mask 17Cleaning Periodically 17converter 34 D Daily Cleaning 17DC converter 34lo...