Page 2 - Please read and understand this manual before using the system.; Record of information for future reference
Please read and understand this manual before using the system. Record of information for future reference Complete the following details when you receive your SULLIVAN ® Comfort. Unit prescribed by (physician) _______________________________ Sleep clinic _______________________________ Date prescri...
Page 3 - Contents; Components of SULLIVAN
Contents 3 Contents 1 Introduction ...................................................................................................... 4 How bi-level treatment works 4 Features of SULLIVAN ® Comfort 5 Delay timer 5 2 Components...
Page 4 - Introduction; How bi-level treatment works; inspiratory or IPAP pressure
SULLIVAN Comfort Operating Manual 4 1 Introduction How bi-level treatment works Obstructive sleep apnea or OSA occurs when the tongue and soft palate relaxduring sleep, blocking the airway and cutting off the brain’s supply of oxygen. Totreat OSA, nasal...
Page 5 - Features of SULLIVAN; Delay timer; Daily operation; Use in different countries; Use in; Indications for use; This is not a life support device.; It may stop operating with a
Section 1: Introduction 5 Features of SULLIVAN ® Comfort Delay timer If you find the initial sensation of pressure uncomfortable while trying to fallasleep you may use the delay timer. This allows you to select the time it will take(5, 10 or 20 minutes) for the air pressure in your mask to build up...
Page 6 - SULLIVAN Comfort bi-level system; a flow generator unit
SULLIVAN Comfort Operating Manual 6 SULLIVAN Comfort bi-level system 2 Components of SULLIVAN ® Comfort You should take time to become familiar with the components of your SULLIVAN ® Comfort bi-level system (see figure below). You should have: ◆ a flow...
Page 7 - Preparing for use; Assembling the mask system; Do not use bleach or chlorine-based solutions to; Modular Mask Frame
Section 3: Preparing for use 7 Preparing for use Assembling the mask system The mask frame has two accessports which should be plugged andare not used except for initialpressure setting by a technican. Ithas a swivelling mask elbow whichallows the air tubing to bepositioned without kinks. The airout...
Page 8 - Fit the headgear to the mask frame as shown below.; Fit headgear to mask frame; Setting up the flow generator; other end of the power cord into a power outlet.; Plug in power cord at rear of flow generator
SULLIVAN Comfort Operating Manual 8 4 Fit the headgear to the mask frame as shown below. • Thread the headgear strapsthrough the slots in the maskframe – the lower strapsshould also pass through theslots on the cushion. Note : If using the ResCap™, posi...
Page 9 - Switch the main switch at the rear of the flow generator on (; START; Allow the unit to run for a few seconds, then press the; STOP; Positioning the flow generator; Do not position a water humidifier on top of the flow generator.; If water is spilt inside the flow generator, turn off the main
Section 3: Preparing for use 9 2 Switch the main switch at the rear of the flow generator on ( I ). Press the START button on the front panel (see figure below). Allow the unit to run for a few seconds, then press the STOP button. STOP button Front panel 3 Positioning the flow generator • Place the ...
Page 10 - Connect the air tubing to the flow generator and mask; Connecting the air tubing; Note
SULLIVAN Comfort Operating Manual 10 Connect the air tubing to the flow generator and mask IMPORTANT: Make sure the power cord and plug are in good condition and the equipment is not damaged. Only use the same type of mask and air tubing as original...
Page 11 - Fitting the mask; To remove the mask
Section 4: Fitting the mask 11 Fitting the mask The mask frame and headgear should be assembled as described in section 3, Preparing for use . To fit the mask to your face: 1 Loosen the headgear straps and pull theheadgear over your head. Make sure thestraps are sufficiently loose so that thenasal c...
Page 12 - Adjust the air tubing – try both ways and; Starting; button to put the unit in standby.
SULLIVAN Comfort Operating Manual 12 5 Adjust the air tubing – try both ways and select whichever is more comfortable Using the START button Start the flow generator by pressing the START button. The pressure will risequickly to your set pressures – the...
Page 13 - Using the delay timer; If you want the pressure to rise very slowly, press the
Section 5: Daily operation 13 * The pressure strip light display gives an approximate indication of thepressure you are receiving. You will notice that the indicator lightsfluctuate as the unit switches between your IPAP and EPAP pressures. Using the delay timer If you prefer to have the air pressur...
Page 14 - Stopping; Press the; button to stop the flow generator. To remove the mask,; or a delay timer button. Adjust the; After daily use; Cleaning and Maintenance
SULLIVAN Comfort Operating Manual 14 Stopping 1 Press the STOP button to stop the flow generator. To remove the mask, stretch the headgear slightly and pull it over your head. Or release one of thebottom headgear straps. 2 When you are ready to go back ...
Page 15 - Helpful hints; When you use your SULLIVAN
Section 6: Helpful hints 15 Helpful hints • When you use your SULLIVAN ® Comfort™ bi-level system for the first time you may notice an unusual sense of pressure when breathing. Eventhough a reduced pressure is delivered when you breathe out, you may stillneed to consciously push out when breathing o...
Page 16 - Your SULLIVAN
SULLIVAN Comfort Operating Manual 16 Use in different countries • Your SULLIVAN ® Comfort™ has a switch mode power supply which means it will operate on the different power supplies in other countries. Theunit accepts 110-240 V, 50-60 Hz. • No special a...
Page 17 - Warnings related to treatment
Section 8: Warnings related to treatment 17 Warnings related to treatment The SULLIVAN ® Comfort™ bi-level system is designed to help you have a good night’s sleep. However, you need to be aware of possible problems that may ariseduring the course of your treatment. If you have concerns about any of...
Page 18 - Cleaning and maintenance; Daily; Do not hang the air tubing in direct sunlight as; Weekly; Remove the air tubing from the flow generator and from the mask.
SULLIVAN Comfort Operating Manual 18 9 Cleaning and maintenance You should regularly carry out the cleaning and maintenance described in thissection. These procedures do not require a specialist service organization. Daily 1 After daily use, wipe the in...
Page 19 - Periodically; Replacing the air filter; To remove the air filter box:
Section 9: Cleaning and maintenance 19 Remove air filter box atrear of flow generator Periodically All items of the mask and air tubing system are subject to normal wear and tearand may eventually need to be replaced. Check periodically for excessive wear inthese items. 1 When dirty, remove the head...
Page 21 - Servicing
Section 10: Servicing 21 10 Servicing Your SULLIVAN ® Comfort™ is designed to give years of trouble-free operation. It is not necessary to send your unit away for regular servicing provided youfollow the recommendations in section 9, Cleaning and maintenance . Should you feel that your unit is not p...
Page 22 - Troubleshooting; If you have any problems with your SULLIVAN; Do not attempt to open the flow generator.
SULLIVAN Comfort Operating Manual 22 Troubleshooting If you have any problems with your SULLIVAN ® Comfort™, try the following suggestions. If it is still not working properly, consult your equipment supplier ortake it to an authorized service agent. Do...
Page 23 - Accessories and spare parts; Accessories; A range of accessories is available to adapt your SULLIVAN; Modular Mask Frame Kit
Section 12: Accessories and spare parts 23 Accessories and spare parts Accessories A range of accessories is available to adapt your SULLIVAN ® Comfort™ to your needs. Contact your sleep clinic or equipment supplier for further information onthe following items. Modular Mask Frame Kit Adapts to a wi...
Page 24 - Humidifiers; Spare parts
SULLIVAN Comfort Operating Manual 24 SULLIVAN ® HCH Disk – a foam insert which fitsinside your mask frame HC100 Heated Humidifier Passover Humidifier Humidifiers Bi-level treatment sometimes results in a dry or runny nose. ResMed produces arange of humi...
Page 25 - System specifications; SULLIVAN; Flow generator unit; Glossary; Type CF equipment/Appareil de classe CF
Section 13: System specifications 25 13 System specifications SULLIVAN ® Comfort bi-level system Flow generator unit PERFORMANCEOperating pressure range: 4-18 cm H 2 O Maximum IPAP time range: 0.2-10.8 seconds DIMENSIONS (H x W x D)145 x 240 x 350 mm(5.7 x 9.5 x 13.8 in) WEIGHT4.1 kg (9.1 lb) POWER...