i Contents Contents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Your Responsibility 1 Medical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 About the S8 Elite II 2 Contraindications 2 Warni...
Page 6 - Frequently Asked Questions; Accessories; Limited Warranty
ii Contents Replacing the Air Filter 24 Servicing 24 Frequently Asked Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 System Specifications . . . . . . . . . . . . . . . . ...
Page 7 - Introduction; is a; Your Responsibility; Please read this manual carefully before use.; Warning alerts you to possible injury.
1 Introduction Introduction Thank you for choosing the S8 Elite™ II flow generator. The S8 Elite II is a compact, all-in-one package with an inbuilt power supply and the option of integrated data management and humidification. This user manual contains the information you need for the correct use of...
Page 8 - Medical Information; About the S8 Elite II
2 Medical Information Medical Information About the S8 Elite II The S8 Elite II continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnoea (OSA) in adult patients. The S8 Elite II CPAP system is intended for home and hospital use. Contraindications T...
Page 9 - Cautions
3 Medical Information through the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than several minutes can, in some circumstances, lead to suffocation. This applies to most models of CPAP devices. • In the event of power failure or machine malfunction, remove the m...
Page 10 - Adverse Effects
4 Medical Information Note: The above are general warnings and cautions. Specific warnings, cautions and notes appear with the relevant instructions in the manual. Adverse Effects Patients should report unusual chest pain, severe headache or increased breathlessness to their prescribing physician. A...
Page 11 - The S8 Elite II System; S8 Elite II; Air tubing; Masks; ResMed mask system; Nasal Masks; Not all masks are available in all regions.
5 The S8 Elite II System The S8 Elite II System The S8 Elite II device is part of a system that consists of five elements: • S8 Elite II flow generator • Mask • Humidifier • Data management products • Accessories. S8 Elite II Please refer to the illustrations in section A of the illustration sheet. ...
Page 12 - WARNING; Humidifier
6 The S8 Elite II System ! WARNING For optimal product performance and data recording, ResMed mask systems should be used with the S8 Elite II. Please refer to “Warnings” on page 2. Humidifier Please refer to the illustration in section B of the illustration sheet. A humidifier may be required if yo...
Page 13 - How to Use the S8 Elite II; Set Up the S8 Elite II; CAUTION; Assemble the Mask
7 How to Use the S8 Elite II How to Use the S8 Elite II Please refer to the illustrations in section D of the illustration sheet. Set Up the S8 Elite II 1 Place the S8 Elite II on a table near the head of your bed. ! CAUTION Be careful not to place the device where it can be bumped or where someone ...
Page 14 - Start Treatment; SmartStart; Do not use SmartStart when using a mask with an anti-asphyxia valve,; Stopping Treatment
8 How to Use the S8 Elite II Start Treatment 1 Make sure the power is on. A welcome message is displayed briefly on the LCD screen, then the standby (Ramp) screen appears. The key and LCD backlights also turn on. 2 Lie down and arrange the air tubing so that it is free to move if you turn in your sl...
Page 15 - Using the Mask-Fitting Feature; Make sure that the mask setting on the MASK screen is the same as the; excelnt
9 How to Use the S8 Elite II Using the Mask-Fitting Feature You can use the S8 Elite II mask-fitting feature to help you fit your mask properly. This feature delivers air pressure for a three-minute period, prior to starting treatment, during which you can check and adjust your mask fit to minimise ...
Page 17 - How to Attach a Humidifier; Attach a HumidAire 3i Humidifier; Please refer to the illustration in section; Storing the HumidAire 3; Store the HumidAire 3i
11 How to Attach a Humidifier How to Attach a Humidifier Attach a HumidAire 3i Humidifier Please refer to the illustration in section E of the illustration sheet, and your HumidAire 3i manual. The HumidAire 3i humidifier attaches to the front of the S8 Elite II to provide heated humidification. No o...
Page 18 - How to use the Keypad; LCD Screen; The LCD screen displays the menus and treatment screens.; Keypad Keys; The S8 Elite II keypad has the following keys:; menu; cancel
12 How to use the Keypad How to use the Keypad The control panel of the S8 Elite II includes an LCD screen and keypad for navigating through the menus and delivering treatment. LCD Screen The LCD screen displays the menus and treatment screens. The LCD backlight turns off after two minutes of inacti...
Page 19 - How to Change the Settings; How to enter the menu screens; How to Change the Ramp Time
13 How to use the Keypad When the S8 device is in standby mode all keys remain lit. During therapy, the keypad backlight turns off after two minutes of inactivity and the Start/Stop key dims. How to Change the Settings You can change settings on the S8 Elite II by using the keypad and LCD. When the ...
Page 20 - How to Use the Standard Menu; How to Change the Expiratory Pressure Relief Setting
14 How to Use the Standard Menu How to Use the Standard Menu Figure 1 shows the standard menu. Figure 1: Standard menu screen You enter the standard menu from the standby (RAMP) screen, by pressing the Left (menu) key. How to Change the Expiratory Pressure Relief Setting Your clinician may have set ...
Page 21 - Other viewable information
15 How to Use the Standard Menu The following table shows the setting that should be selected for each mask type. Other viewable information You can also view the total number of hours of treatment (USED HRS); the number of sessions for which the S8 Elite II was used in a usage period (USAGE); and t...
Page 22 - How to Use the Detailed Menu; Figure 2 shows the detailed menu.; Left
16 How to Use the Detailed Menu How to Use the Detailed Menu Figure 2 shows the detailed menu. Figure 2: Detailed menu screens You enter the detailed menu from the standby (RAMP) screen, by pressing the Left (menu) key. The SETTINGS menu screen is displayed. USED HRS: 08000 exit USAGE: 1000/1068 exi...
Page 23 - Settings Menu
17 How to Use the Detailed Menu Settings Menu Select Enter ( Left key) on the Settings menu to access the options in this menu. How to Change the Expiratory Pressure Relief Setting Your clinician may have set up your device to deliver expiratory pressure relief (EPR). This can be used to help ease a...
Page 24 - How to Change the Leak Alert Setting; Results Menu; Contact your clinician for further details about Smart Data.; Viewing Usage Information; Options Menu; How to change the language setting; Servicing Menu
18 How to Use the Detailed Menu How to Change the Leak Alert Setting Leak alert indicates when a high mask leak is present. Scroll to LEAK ALERT and select Change ( Left key). Press the Up or Down keys to change the option to ON or OFF. Results Menu Select Enter ( Left key) on the Results menu to ac...
Page 26 - S8 Modules; Using a Blank Module; Attaching the Module
20 S8 Modules S8 Modules Note: Do not remove modules while power is connected to the S8 Elite II. Your S8 Elite II will have one of two modules fitted to the back: a blank module or a ResScan Data Card module. Using a Blank Module Please refer to the illustrations in section F of the illustration sh...
Page 27 - Insert the Data Card; II; Remove the Data Card; Using the Data Card to Update Settings; have updated your settings, this message is not displayed.
21 S8 Modules 1 Insert the Data Card • Switch on the S8 Elite II and wait until you see the standby (RAMP) screen. • Hold the Data Card with the arrow facing up, and insert it into the slot in the Data Card module ( H-1 ). • Push the card in smoothly until it stops moving ( H-2 ) and messages start ...
Page 29 - Cleaning and Maintenance; Daily Cleaning; Weekly Cleaning; Remove the air tubing from the S8 Elite II; Periodic Cleaning; Clean the exterior of the S8 Elite II
23 Cleaning and Maintenance Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask manual for detailed instructions. Daily Cleaning M ASK Clean the mask according to the instructions supplied with the mask. A IR TUBING Disco...
Page 30 - Replacing the Air Filter; To replace the air filter:; Servicing
24 Cleaning and Maintenance Replacing the Air Filter To replace the air filter: 1. Remove the air filter cover at the back of the S8 Elite II. 2. Remove and discard the old air filter. 3. Insert a new filter with the blue tinted side facing out from the device. 4. Replace the air filter cover. Inspe...
Page 31 - Do not use moisturiser when wearing a mask.
25 Frequently Asked Questions Frequently Asked Questions How will it feel when I am breathing with the S8 Elite II? When you first use your S8 Elite II, you may feel uncomfortable breathing against the airflow. This is normal. You will continue to breathe normally while you sleep. Take time to adjus...
Page 32 - Your S8 Elite II; Do not use your S8 Elite II; Can I use my S8 Elite II if mains (AC) power is not available?; You must use a ResMed DC-12 converter to connect the
26 Frequently Asked Questions A LTITUDE CHANGES Your S8 Elite II automatically adjusts to changes in altitude up to 2591 m. O N AN A IRCRAFT Please consult the medical services department of your carrier if you intend to use your S8 Elite II on an aircraft. The S8 Elite II complies with US FCC Part ...
Page 33 - Troubleshooting
27 Troubleshooting Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device. Problem Possible Cause Solution No display. Power not connected. Ensure the appropriate power cord i...
Page 37 - System Specifications; Operating pressure range: 4 to 20 cm H; Power Supply; Flexible plastic, 2 m
31 System Specifications System Specifications S8 Elite II (no Accessories) Performance: Operating pressure range: 4 to 20 cm H 2 O Dimensions (H x W x D): 112 mm x 164 mm x 145 mm Weight: 1.4 kg Power Supply • Input range for S8 Elite II with HumidAire 3i: 100–240V, 50/60Hz; 110V, 400Hz; 2.5A <1...
Page 38 - Noise Level; Data Card; Housing Construction:; Flame retardant engineering thermoplastic; Environmental Conditions; The sum of a measured noise emission value and its associated
32 System Specifications Noise Level Accessories Data Card Dimensions (L x W x D): 65.04 mm x 22.15 mm x 0.85 mm Data Card Module Dimensions (H x W x D): 49.7 mm x 42.2 mm x 12 mm All Devices (where applicable) Housing Construction: Flame retardant engineering thermoplastic Environmental Conditions ...
Page 39 - Symbols which appear on the device
33 System Specifications Symbols which appear on the device Attention, consult accompanying documents Type CF equipment Class II equipment Drip proof Dangerous voltage Start/Stop and mask-fit Environmental information WEEE 2002/96/EC is a European Directive that requires the proper disposal of elect...
Page 44 - Some models are not available in all regions.
38 Limited Warranty Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Note: Some models are not available in all regions...
Page 45 - Index
39 Index Index A accessories 6air filter 5, 27 replacing 24 air pressure 27air tubing 5 cleaning 23connect to air outlet 7kinked 27positioning 8 aircraft 26airflow breathing with 25 B blank module 20 C cleaning daily 23periodic 23weekly 23 contraindications 2 D daily cleaning 23Data Card 20, 29 copy...