Page 3 - Important note; Leica Microsystems Nussloch GmbH
3 Leica CM 3050 S – Cryostat Issued by: Leica Microsystems Nussloch GmbHHeidelberger Str. 17 - 19D-69226 NusslochGermanyPhone: +49 6224 143-0 Fax: +49 6224 143-200 Internet: http://www.histo-solutions.com Serial No.: .............. ...................................................... Year of Manuf...
Page 4 - Table of Contents
4 Instruction Manual V1.2 - 11/2006 Table of Contents 1. Important information ................................................................................................................ 7 1.1 Symbols used in this manual and .........................................................................
Page 7 - Leica CM 3050 S – Cryostat; Important information; Symbols used in this manual and
7 Leica CM 3050 S – Cryostat The chapters of the Leica CM3050 S cryostatinstruction manual: Chapter 1 Structure of the manual : • Table of contents • Important information on thismanual Chapter 2 Safety • Read this chapter beforeoperating the instrument! Chapter 3 Installation • Unpacking and instal...
Page 8 - General; Designated use
8 Instruction Manual V1.2 - 11/2006 General This instruction manual includes importantinstructions and information related to theo p e r a t i n g s a f e t y a n d m a i n t e n a n c e o f t h einstrument. The instruction manual is an important part of theproduct. It must be read carefully before ...
Page 9 - Integrated safety devices; Handwheel lock; Sicherheit
9 Leica CM 3050 S – Cryostat 2.1 General information on instrumentdesign and safe handling This instrument has been built and tested in ac-cordance with the following safety regulationson electrical measuring, control, regulating andlaboratory devices: • DIN EN 292 • DIN EN 61010-1 • EN 50082-1 • EN...
Page 10 - Safety; L O C K E D
10 Instruction Manual V1.2 - 11/2006 Instruments with sectioning motor: The sectioning motor is now blocked. All instruments: The message ‘LOCKED’ in the display of controlpanel 1 indicates that the handwheel has beenlocked: Unlocking: • To unlock, push locking pin ( 2 ) to the left into position ( ...
Page 11 - Control panel 2
11 Leica CM 3050 S – Cryostat 2.2.3 Emergency stop function (instruments with sectioningmotor only) The emergency stop is activated via the redemergency stop button in control panel 2 or viathe footswitch Activating the emergency stop function • Press emergency stop button or step onfootswitch force...
Page 12 - Safety instructions: handling and operation of the instrument; position
12 Instruction Manual V1.2 - 11/2006 2. Safety 2.3 Safety instructions: handling and operation of the instrument 2.3.1 Transport • To avoid severe damage to the instrument byrunning it while the compressor oil is dis-placed from its regular position:• Do not tilt the instrument,only transport in an ...
Page 13 - frozen accessories; Wear protective gloves.
13 Leica CM 3050 S – Cryostat 2.3.4 Handling microtome knives/blades • Danger!--> Microtome knives and disposable blades have extremely sharp cutting edges andcan cause serious injuries. Therefore: • Handle knives / blades with utmost care. • Never leave any knives / blades in unpro-tected places...
Page 14 - When working with radioactive specimens:
14 Instruction Manual V1.2 - 11/2006 2. Safety 2.3.9 Infectious/radioactive material • Danger!Use caution when working with potentiallyinfectious specimens:-->Risk of infection! • When working with potentially infectious /radioactive specimens: • Wear protective clothes (gloves, protec-tive boots...
Page 15 - microtome
15 Leica CM 3050 S – Cryostat 2. Safety 2.3.11 Removing/reinstalling the microtome • Before removing the microtome: • Switch instrument off. • Unplug from mains. • Place handwheel grip in lowest position andlock. --> When removing the microtome, the specimen head must always be lok-ked in the low...
Page 16 - Handwheel easily accessible.; Installation
16 Instruction Manual V1.2 - 11/2006 3.2 Unpacking and installation Unpacking instructions are always located in atransparent protective envelope on the outsideof the instrument shipping crate. Make sure to read and follow allsafety instructions provided inchapter 2.3.1 ‘Transport’ and on theunpacki...
Page 17 - Standard delivery
17 Leica CM 3050 S – Cryostat 3. Installation 3.3 Standard delivery 1 Cryostat with microtome ............................................................................. 0460 32037 1 Handwheel with marking, antibacterial ................................................... 0471 42558 1 Heat extract...
Page 18 - In addition to the above:
18 Instruction Manual V1.2 - 11/2006 Further accessories: Further accessories which you ordered will be included in the box containingthe standard delivery items. Knife holders are delivered with anti-roll guide, knife guard, and aseparate instruction manual.In case of non-configured instruments, th...
Page 19 - Installing/inserting accessories
19 Leica CM 3050 S – Cryostat 3.4 Installing/inserting accessories 3.4.1 Installing the handwheel • Insert pin ( 1 ) of the handwheel shaft into hole ( 2 ). • Place spring washer ( 3 ) onto screw ( 4 ) as shown. • Tighten screw ( 4 ) with an Allen key. • Cover with selfadhesive disc ( 5 ). For purpo...
Page 20 - Function; The footswitch performs the same task as the; Models with footswitch:; All instruments with sectioning motor.; Prior to switching on the instrument; Insert mains plug into wall outlet.; Connecting the footswitch; Insert footswitch into port (
20 Instruction Manual V1.2 - 11/2006 3.4.3 The footswitch Function The footswitch performs the same task as the RUN/STOP and RUN/ENABLE keys (activating/ deactivating motorized sectioning / trimming).In addition, the footswitch can be used to acti-vate the emergency stop function. Models with footsw...
Page 21 - Instrument Properties; Overview
21 Leica CM 3050 S – Cryostat 4. Instrument Properties 4.1 Overview 1 Cryostat CM3050 S 2 Control panel 1 3 Control panel 2 4 Storage shelf, left 5 Storage shelf, right 6 Rubber mat 7 Mains switch /Automatic cutoutfor instrument,Automatic cutoutfor sectioning motor,Footswitch port 8 Quick-freeze she...
Page 23 - Instrument properties; All specifications related to tempe-; Technical Data
23 Leica CM 3050 S – Cryostat 4. Instrument properties Operating temperature range (ambient temperature): + 18°C to + 40°C. All specifications related to tempe- rature are valid only up to an ambient temperature of + 22°C and relative air humidity of less than 60 %! Refrigeration system 50 Hz 60 Hz ...
Page 24 - Microtome
24 Instruction Manual V1.2 - 11/2006 4. Instrument properties Microtome Rotary microtomeSection thickness setting 0.5 - 300 μm Overall specimen feed 25 mm Vertical stroke 59 mm Specimen retraction 50 μm Maximum specimen size 40 x 55 mm Specimen orientation 8° (x-, y-, z-axis) Trimming (6 thickness s...
Page 25 - Operation; Operating the instrument for the first time
25 Leica CM 3050 S – Cryostat 5. Operation 5.1 Operating the instrument for the first time Prior to setting up the instrument, make sure to read the instruction manualcarefully and in full. When setting up the instrument, please follow the individual chapters in thefollowing order: • First, study ch...
Page 26 - Switching on; To switch on the instrument, place switch (; Automatic cutout for sectioning motor:; Push automatic cutout (
26 Instruction Manual V1.2 - 11/2006 5.2 Switching on/fuses Switching on • To switch on the instrument, place switch ( 1 ) in upper lock-in position. Mains switch / automatic cutout for theinstrument: In addition to its function as ON/OFF switch, switch ( 1 ) is also the automatic cutout for the in-...
Page 27 - Control panel 1
27 Leica CM 3050 S – Cryostat 5.3.1 Key functions in control panel 1 5.3 Control panel 1 ON/OFF key for cryochamber illumination. To select the menu items for setting /changing instrument parameters. In the status indication, pressing the ar-row keys displays the individual currentparameters.In the ...
Page 28 - Meaning of indication:; Display indications when switching on
28 Instruction Manual V1.2 - 11/2006 5. Operation Meaning of indication: Operator activity: yes no no Acknowledge indication bypressing once briefly. Initialization phase 1 Initialization phase 2 Software version Selected display language Sectioning motor speedlevel ‘slow’ / (fast) --- --- --- Selec...
Page 29 - Note on the arrow key function:; S E T
29 Leica CM 3050 S – Cryostat 5. Operation Once the instrument has gone through the initialsequence of indications, you can go on to themenu functions to select and/or change instru-ment parameters. If display is locked:Press key until display backgroundlights up. Next, press the menu key. With each...
Page 30 - Preset counter function:; P R E S E T
30 Instruction Manual V1.2 - 11/2006 5. Operation 2. Cryochamber temperature: Press key until menu item ‘Set chambertemperature’ is displayed: --> Continue as described for specimen temperature. 3. Preset counter (instruments with sectioning motor only): Press key until menu item ‘Presetcounter’ ...
Page 31 - Starting time of automatic defrost cycle:; or
31 Leica CM 3050 S – Cryostat The automatic defrost cycle takes place onceevery 24 hours at the selected defrost time. 7. Duration of defrost cycle: Press key until menu item ‘Defrost period’is displayed: • Use defrost keys ( or ) to adjust de- frost time. (Only if necessary - see below). Important ...
Page 32 - H O M E
32 Instruction Manual V1.2 - 11/2006 5. Operation 5.3.5 Status display Each individual parameter is displayed in a certain area of the display screen. Digits 1 and 2 indicate sectioning stroke position: • Sectioning stroke (outside sectioning window) • : Retraction stroke • [ ] : Sectioning stroke (...
Page 33 - Refrigeration state
33 Leica CM 3050 S – Cryostat 5. Operation These parameters can be displayed via the ar-row keys.When pressing , the parameters are dis- played in the sequence listed on this page;when pressing , they are displayed in re- verse order. The display can be scrolled contin-uously, i.e. after the last pa...
Page 34 - Control panel 2
34 Instruction Manual V1.2 - 11/2006 5. Operation 5.4 Control panel 2 Functional area A:• Sectioning modes• Sectioning window• Sectioning speed• Emergency stop function Functional area B:• Coarse feed• Trimming• Section thickness selection Functional area C:• Start / Stop sectioning Instruments with...
Page 35 - Start / stop motorized sectioning; Activating a sectioning mode
35 Leica CM 3050 S – Cryostat Start / stop motorized sectioning Sectioning mode 1 - Intermittent stroke: • Simultaneously press and hold RUN/STOP andRUN/ENABLE or depress and hold footswitch. • To stop the sectioning stroke, release thekeys or the footswitch:The specimen head stops immediately. Sect...
Page 36 - Functional areas A and C - Instruments with sectioning motor only
36 Instruction Manual V1.2 - 11/2006 The value has been stored when the LED in thesectioning window key lights up briefly. If the LED flashes, the value has not yet beenstored, because the user has tried to set a win-dow margin outside the possible window range(the area in which the sectioning windo...
Page 37 - Caution when working in continuous mode:
37 Leica CM 3050 S – Cryostat 5.4.3 Selecting sectioning speed • Sectioning speed is selected via the slidingpotentiometer. Speed can be changed whilesectioning is in motion. Caution when working in continuous mode: If the slider is set to ‘0’, the sectioning motionstops, however, the actual section...
Page 38 - Coarse feed ‘fast backward’; Functional area B - all models
38 Instruction Manual V1.2 - 11/2006 5. Operation 5.4.5 Coarse feed keys The four coarse feed keys in functional area Bfunction as follows: Coarse feed ‘fast backward’ - Travelling speed 1,000 μm/s • Press key to activate. Specimen head travels backward(= away from knife): a. until it reaches the re...
Page 39 - Selecting trimming thickness; Selectable section thickness settings:
39 Leica CM 3050 S – Cryostat 5. Operation Press this key to alternate between thetwo functions. illuminated:Trimming mode is active off:Sectioning mode is active. Selecting trimming thickness • Activate trimming mode • Select trimming thickness (= sectioningthickness at which trimming is carried ou...
Page 40 - Daily operation of the instrument
40 Instruction Manual V1.2 - 11/2006 5.5 Daily operation of the instrument • Select chamber temperature and, in instru-ments with specimen cooling, specimentemperature according to the type of speci-men to be sectioned (see also chapter 6.3‘Reference chart for temperature settings’). • Do not contin...
Page 41 - Installing the 90 ° prism; discs with 90 ° prism
41 Leica CM 3050 S – Cryostat 5.5.2 Freezing specimens onto specimen In instruments with specimen cooling, the 90 °prism, which is inserted into the specimen head,is an additional option to freeze specimens atmaximum temperature . Installing the 90 ° prism • Loosen screw (3) at specimen head. • Inse...
Page 42 - Lock handwheel in upper position.; Thermoblock
42 Instruction Manual V1.2 - 11/2006 5. Operation 5.5.4 Inserting specimen discs into the • Lock handwheel in upper position. If a knife has already been insertedinto the knife holder, cover knifeedge with knife guard! • Loosen clamping screw (3). • Insert specimen disc in opening (2) (rearsurface o...
Page 43 - Inserting the knife; When
43 Leica CM 3050 S – Cryostat 5. Operation Inserting the knife • Insert the knife / disposable blade into theknife holder. • Select appropriate clearance angle at knifeholder (settings between 4° and 6° are suit-able for the majority of applications). For details on how to insert and ad-just the kni...
Page 44 - Finishing work; S e e; longer period of time
44 Instruction Manual V1.2 - 11/2006 5. Operation 5.6 Finishing work 5.6.1 Daily routine steps Remove all specimens from the cry-ochamber and store safely! S e e also corresponding safety instruc-tions in chapter 2.3.7 ‘Defrosting /handling frozen tissue’! • Lock handwheel. • Remove knife from knife...
Page 45 - Defrosting
45 Leica CM 3050 S – Cryostat H O M E 5.7 Defrosting The Leica CM3050 S has 3 (instruments withoutspecimen cooling: 2) different defrosting options. - Automatic chamber defrost cycle - Manual chamber defrost cycle - Specimen head defrosting (manual) - instru-ments with specimen cooling only. Read th...
Page 46 - Instruments with specimen cooling:
46 Instruction Manual V1.2 - 11/2006 S E T T E M P O T - 3 5 ° C S E T T E M P C T - 3 0 ° C - 5. Operation 5.7.2 Manual chamber defrost cycle In addition to the automatic defrost cycle, amanual chamber defrost cycle can be activat-ed when necessary. • Press until menu item ‘Set chamber temperature’...
Page 47 - Troubleshooting, applications tips; Displayed error messages; To acknowledge the message, press; Temperature indications; P O W E R
47 Leica CM 3050 S – Cryostat 6. Troubleshooting, applications tips 6.1 Displayed error messages The autodiagnosis function of the CM3050 Sdisplays error messages in case of certain mal-functions. 1. Power failure This error message is displayed after the instru-ment had been disconnected from mains...
Page 48 - Potential problems - causes and remedies
48 Instruction Manual V1.1 02/2000 6. Troubleshooting, applications tips 6.2 Potential problems - causes and remedies Remedies - Eliminate draft sources or change place of installation ofthe cryostat. - Close sliding window. - Consider wearing a mask. - Select lower temperature.- Wait until knife an...
Page 49 - Problem
49 Leica CM 3050 S – Cryostat 6. Troubleshooting, applications tips Problem Sections curl on anti-roll plate. Scraping noise during section-ing stroke and specimen returnstroke. Ondulated sections. Chatter marks form duringsectioning. Condensation forms on anti-rollplate and knife during cleaning. C...
Page 50 - Causes
50 Instruction Manual V1.1 02/2000 6. Troubleshooting, applications tips Causes - Anti-roll plate protrudes too far beyond the knife edge. Adjust-ment was done in the directionof the knife. - Wrong temperature for the tissue to be sectioned. - Knife profile inappropriate for the specimen to be secti...
Page 52 - Reference chart for temperature settings (in minus °C); Tissue
52 Instruction Manual V1.1 02/2000 6. Troubleshooting, applications tips 6.3 Reference chart for temperature settings (in minus °C) Tissue 10 - 15 15 - 25 25 - 50 Adrenal Bone marrow ❄ Brain Bladder Breast - fatty Breast - little fat Cartilage Cervical Fatty Heart and vascular Intestinal Kidney Lary...
Page 53 - Disinfection, cleaning and maintenance; Safety instructions; • ‘Disinfection and Cleaning’; Spray disinfection with Leica Cryofect; Set chamber temperature -20 °C or warmer.; on disinfection and cleaning
53 Leica CM 3050 S – Cryostat 7. Disinfection, cleaning and maintenance 7.1 Safety instructions When disinfecting the instrument,make sure to observe the followingsafety instructions: • ‘Disinfection and Cleaning’ (Chapter 2.3.10) • ‘Handling microtome knives / blades’(Chapter 2.3.4) • ‘Defrosting /...
Page 54 - Disinfection with conventional; Switch the instrument off.; disinfectants; Allow cryochamber to dry thoroughly.; Cleaning
54 Instruction Manual V1.2 - 11/2006 7 Disinfection, cleaning and maintenance 7.3 Disinfection with conventional • Switch the instrument off. • Carefully remove knife / blade as well as spec-imens and accessories from the chamber. • Remove section waste from the chamberand dispose of it. • Remove sl...
Page 55 - Removing / reinstalling the microtome; • ‘Removing / reinstalling
55 Leica CM 3050 S – Cryostat 7.5 Removing / reinstalling the microtome 7.5.1 Removing the microtome When removing the microtomemake sure to observe the followingsafety instructions: • ‘Removing / reinstalling the microtome’(Chapter 2.3.11) • ‘Defrosting / handling frozen tissue’(Chapter 2.3.7) • ‘I...
Page 56 - Remove the microtome from the chamber.
56 Instruction Manual V1.2 - 11/2006 7.5.3 Reinstalling the microtome When reinstalling the microtomemake sure to observe the followingsafety instructions: • ‘Removing / reinstalling the microtome’(Chapter 2.3.11) • Display message ‘Dry Microtome’ (Chapter 2.3.12) • Lock handwheel in lower position....
Page 57 - Removal; Put the sliding window back in place.; Exchanging the fluorescent light lamp; • ‘Maintenance’; Removing the sliding window
57 Leica CM 3050 S – Cryostat Removal • The lamp (3) is located behind the glareshield (4). • Carefully hold lamp with both of your hands.Slightly pull the lamp downward at the leftside to remove it from fastening clip (5). • Pull the lamp towards the left to remove itfrom its socket (6). Installing...
Page 58 - Warranty and service; Warranty; Please provide the following information:; Decommissioning and disposal
58 Instruction Manual V1.2 - 11/2006 8. Warranty and service Warranty Leica Microsystems Nussloch GmbH guarantees that the contractual prod-uct delivered has been subjected to a comprehensive quality controlprocedure based on the Leica in-house testing standards, and that the prod-uct is faultless a...
Page 59 - Standard delivery; Appendix
59 Leica CM 3050 S – Cryostat An electrohydraulic lifting device for individualworking height adjustment is available as anoptional accessory for the CM3050 S cryostatseries. 1.1 Designated use The electrohydraulic height adjustment devicehas been designed exclusively for adjustmentof the instrument...
Page 60 - Take the height adjustment device out of the box.
60 Instruction Manual V1.2 - 11/2006 9. Appendix 1.4 Installation • Open the shipping box and remove all packing material including the Sty-rofoam pieces. • Take the height adjustment device out of the box. • At the place of installation, place the U-frame on the floor with the openside facing yours...
Page 61 - Maintenance and care
61 Leica CM 3050 S – Cryostat The control keys are located on top of the con-trol unit (see photograph). • To lift the instrument, press . • To lower the instrument, press . As long as a button is held, the instrument willgo up or down until reaching the upper or lowerlimit. 1.6 Maintenance and care...
Page 62 - Ordering information - accessories
62 Instruction Manual V1.2 - 11/2006 2.1 Knife holders Knife-holder CN for reusable microtome knives´w/anti-roll guide, adjustable knife guard and knife support bars for slim and wide knives ............................................. 0419 18649Knife support bar for knife holder CN, for slim knive...
Page 63 - We herewith declare, in exclusive responsibility, that the; EC Declaration of Conformity
. . . . . . . . . . . . . . . . . . . . . . . . . . . . Dag Graupner Managing Director We herewith declare, in exclusive responsibility, that the Leica CM3050 S – Cryostat was developed, designed and manufactured to conform with the• Council Directive 73/23/EEC (Low Voltage), • Council Directive 89/...
Page 64 - Decontamination Certificate; NB; Other biohazards; This equipment has been cleaned and decontaminated:
64 Instruction Manual V1.2 - 11/2006 11. Decontamination Certificate (master) Dear Customer,any product which is be returned to Leica Microsystems or serviced on site, must be cleaned anddecontaminated in the appropriate manner. This certificate, duly completed, must be placed in the unitprior to pa...
Page 65 - Important - to avoid refusal of shipment:
65 Leica CM 3050 S – Cryostat 11. Decontamination Certificate (master) Important - to avoid refusal of shipment: Place one copy in the unit prior to packaging or hand it over to the service engineer. Customer as-sumes all responsibility for the immediate return shipment of articles sent to Leica wit...