Page 3 - CONTENTS
© Adam Equipment Company 2009 1 CONTENTS 1.0 INTRODUCTION ............................................................................................ 3 2.0 SPECIFICATIONS ............................................................................................ 4 3.0 SAFETY PRECAUTIONS .............
Page 5 - INTRODUCTION
© Adam Equipment Company 2009 3 1.0 INTRODUCTION • The LBH models are supplied with a power supply adapter, to charge the internal rechargeable battery operation. • Four models: 3kg X 0.1g, 6kg X0.2g, 15kg X 0.5g, 30kg X 1g • Large LCD with backlight • 5 weighing units: g, kg, lb, oz, and a user set...
Page 6 - model
© Adam Equipment Company 2009 4 2.0 SPECIFICATIONS model LBH-3 LBH- 6 LBH-15 LBH-30 capacity 3Kg 6Kg 15Kg 30Kg Readability 0.1g 0.2g 0.5g 1g Repeatability 0.1g 0.2g 0.5g 1g Linearity ± 0.2g 0.4g 1g 2g Units kg / g / lb / oz / custom unit Stabilization time 2s 2s 2s 2s Operating temperature 0ºC to 40...
Page 7 - PRECAUTIONS
© Adam Equipment Company 2009 5 3.0 SAFETY PRECAUTIONS Caution: Please use only the original DC adapter supplied with the scale. Other cords or adapters may damage the scale. Notice: • Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents. • Battery should be ...
Page 8 - UP; Put the scale on a stable and even table.
© Adam Equipment Company 2009 6 4.0 SET UP 4.1 ASSEMBLING OF SCALE • Put the scale on a stable and even table. • Adjust the height of the feet to level the scale; the spirit level bubble should be in the center of the spirit level. • Put the pan support in place, inserting the pan support pins into ...
Page 9 - BEFORE
© Adam Equipment Company 2009 7 5.0 WEIGHING 5.1 BEFORE WEIGHING Check the setting of the auto shut off function, see section 6.2. If the auto shut off function is on, the scale will shut off after a period of time if the scale is not used and it has a stable reading. 5.2 WEIGHING • Press [On/Off] t...
Page 10 - USER
© Adam Equipment Company 2009 8 5.3 USER FUNCTIONS The parts counting, percent weighing and calibration functions can be accessed using the [Func/ ] key, if the functions have been enabled as described in section 6. If the function is turned off the function will not be available. Only functions tha...
Page 11 - “CHECK”
© Adam Equipment Company 2009 9 • The display will show “100.0 % ” . • When another item with a different weight value is put on the pan, the relative % value would be shown against the initial reference sample value. • Press the [Zero/ ESC ] key to exit and go back to weighing mode. 5.3.3 CHECK WEI...
Page 13 - SETTING
© Adam Equipment Company 2009 11 6.0 FUNCTION SETTING 6.1 ENTERING THE FUNCTION SETTING MODE • Press [On/Off] to shut off the power. • Hold [Func/ ] and press [On/Off] to turn the scale on. • Continue to press the [Func/ ] key until the auto shut off mode function is shown. • The display will show “...
Page 15 - On g
© Adam Equipment Company 2009 13 6.8 USER CALIBRATION FUNCTION • Press [Hold/ ] to switch on or off the function. • "OFF CAL" : the user calibration function is off. • "ON CAL" : the user calibration function is enabled. • Press the [Tare/ ↵ ↵ ↵ ↵ ] key to confirm the setting and ent...
Page 16 - AUTOMATIC; “CAL 0”
© Adam Equipment Company 2009 14 7.0 CALIBRATION • Calibration is recommended for best accuracy. It is suggested that the scale be calibrated when first installed and any time the scale is moved to a different environment. • To ensure best accuracy calibrate before use. • Ensure weights are ready be...
Page 17 - “Load”
© Adam Equipment Company 2009 15 Calibration selectable values for all model: • Press the [Tare/ ↵ ↵ ↵ ↵ ] key , the display will show “Load” then add the selected calibration mass value to the pan. • After the calibration mass is added, wait for the stable symbol to come on and then press the [Tare...
Page 18 - MESSAGE
© Adam Equipment Company 2009 16 8.0 ERROR MESSAGE ---ADO A/D counts are too high ---ADL A/D counts are too low --ERR- Wrong operation ------ Overload 9d Unit-E All the units are disabled Low battery indication
Page 19 - Manufacturer’s Declaration of Conformity
© Adam Equipment Company 2009 17 Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatibility Directive 2004/108/EC Low Voltage Directive 2006/...
Page 20 - ADAM EQUIPMENT
© Adam Equipment Company 2009 18 ADAM EQUIPMENT is an ISO 9001:2000 certified global company with more than 35 years experience in the production and sale of electronic weighing equipment. Adam products are predominantly designed for the Laboratory, Educational, Medical, retail and Industrial Segmen...