Toshiba LF410/LF602 - Manual
Toshiba LF410/LF602 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 4 – Avoid using the flowmeter in an; Converter LF602 dimensions and weights:
- Page 5 – ANSI class 150 and class 300 dimensions:; The dimension of L1 is changed when the material of
- Page 8 – Figure 9. Separate type converter LF602; External Connections; Use the appropriate switch rating as shown below:; flowmeters Wiring Diagram
- Page 9 – flowmeters wiring Diagram
- Page 10 – Wiring Precautions; Figure 12. Electrical Conductivity and
- Page 11 – Combined type; Separate type; Grounding of the Combined type and Separate type
- Page 12 – Make sure the full-scale flow rate used for the; Figure 13. Flow Rate and Flow Velocity; Calibration Range; Table 1. Standard Flow Range; PED matrix in each flange connection.; Table 2 PED matrix in each flange connection
- Page 13 – About establishment environment; Ordering Information; Consult Toshiba representative before ordering
- Page 14 – Table 3. Specification Code (Wafer type detector LF410 Series)
- Page 15 – Table 4. Specification Code for converters
No.
EJL-114B
Introduction
The electromagnetic flowmeter uses Faraday’s Law of
electromagnetic induction to measure the process flow.
The device consists of two units: a detector, through
which the fluid to be measured flows and in which
low-level signals proportional to flow rates are
obtained; and a converter, which supplies excitation
current to the detector, and amplifies the signals from
the detector and then processes and converts the
signals into the 4–20mAdc current signal
or
communication signal. Thanks to the unique patented
magnetic field distribution technology, the meter is
highly immune for upstream flow disturbances.
Combined with multi-functional converter LF600
(combined type) or LF602 (separate type) equipped
with its original noise-suppression circuit and
arithmetic operation capability, LF410 has high
tolerance to noise, giving stable output even for slurry
fluid measurement. IR (Infrared) switches enable
parameter setting of the converter without removing
the cover. Flow direction can be set in either way, and
its 128 x 128 dot matrix LCD display allows the LCD
rotated to 90, 180 and 270 degrees with a software.
The AF900 hand-held terminal (HART*
1
communicator) can be used to communicate with the
flowmeter from a remote place. PROFIBUS-PA*
2
interface is available as option.
*1: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial
sensors recommended by the HCF
(HART
Communication
Foundation).
*2: PROFIBUS is the communication protocol for the
factory automation and process automation that
PROFIBUS Organization recommends. Instead of
analog control with a conventional analog signal
(4-20mA), it is one kind of the fieldbus which digitized
all signals. Flowmeters support PROFIBUS-PA.
Power
supply
Converter
I/O
Detector
Combined type
LF410/LF600
LF414/LF600
Converter
Power
supply
Signal cable
Detector
Separate type
LF410/LF602
LF414/LF602
Terminal box
I/O
Figure1. Configuration
LF410/LF600
LF410
LF602
LF414/LF600 LF414
Figure2. LF410 series Flowmeters
Specifications
Overall Specifications
Measurement range in terms of flow velocity:
0
–
0.3
m/s to 0
–10
m/s (0
–1.0
ft/s to 0
–
32.8
ft/s).
0
–
0.1
m/s to 0
–
0.3
m/s (0
–
0.3
ft/s to 0
–
1.0
ft/s)
range is available optionally.
Accuracy: See the following graph.
Pulse output:
Vs > 0.5 m/s (1.64 ft/s): +/-0.5 % of rate.
Vs < 0.5 m/s (1.64 ft/s): +/-0.3% of rate
+/- 1 mm/s (0.039 inch/s).
Current output: plus +/-8 uA (0.05 % of
span)
Note:
Span = Range in the magmeters.
0.0
1.0
2.0
3.0
4.0
5.0
0
1
2
3
Flow Velosity (m/s)
Error
(% of rate)
Absolute figure
Accuracy
Note:
The accuracy above is measured under standard
operating conditions using the weighing method
at Toshiba's flow calibration facility.
*3: CE mark is under application.
Field Intelligent Device Series
Electromagnetic Flowmeter
LF410 /LF600
LF410 /LF602
15 to 200 mm (1/2" to 8")
0 1 2 10
*3
Certification number
Z01207
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
LF410/LF600 LF410/LF602 4 Cable connection ports: Cable glands — LF600 and LF602 without cFMus Approval: Provided as standard OD of cable φ 11~13mm Material Nylon 66 G (PF) 1/2 male screws. Note: When PROFIBUS option is specified , cable grands size is φ 6~8mm for signal cable, φ 11~13mm for power c...
LF410/LF600 LF410/LF602 5 Installation Dimensions Unit : mm L1 Φ D1 (L 2) 64 143 (35) 225 131 BS16, DIN PN16 and JIS 10K dimensions: Meter size (mm) L1 (mm) L2 (mm) D1 (mm) Weight (kg) 15 70 252 49 approx. 5 Note: 1 inch = 25.4mm ANSI class 150 and class 300 dimensions: Meter size (inch) L1 (inch) L...
LF410/LF600 LF410/LF602 8 矢視A 信号ケーブルグランド 電源ケーブルグランド 封止栓 励磁ケーブルグランド 電流出力ケーブルグランド 接地端子(M4) 取付金具 4-φ11 取付金具 82 (36 ) 15 6 13 15 1 17 8 20 0 151 74±0.3 25 LCD 表示器 赤外線スイッチ Excitation cable ground 15 1 151 74±0.3 4 – Φ 11 Attachment LCD display 17 8 20 0 IR Switch Cable ground 15 6 (36) 25 82 13 I/O cable...