Page 2 - THIS PAGE IS INTENTIONALLY LEFT BLANK
Page 3 - PRINTED DOCUMENTS ARE UNCONTROLLED.; made to operate the instrument.; DOCUMENTS
PRINTED DOCUMENTS ARE UNCONTROLLED. We recommend that this document be read in its entirety before any attempt is made to operate the instrument. DOCUMENTS Document P/N Revision DCN Nomenclature Dated 05743 B 5359 M703, Manual, Instruction - Title 03/20/09 05744 B 5359 M703, Manual, Instruction - Te...
Page 5 - SAFETY MESSAGES
TELEDYNE API M703E Calibrator Operator’s Manual Safety Messages SAFETY MESSAGES Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could h...
Page 6 - M703E Dynamic Operator’s Manual; ii
TELEDYNE API Safety Messages M703E Dynamic Operator’s Manual ii 05744 Rev B USER NOTES:
Page 7 - iii
TELEDYNE API M703E Calibrator Operator’s Manual Table of Contents TABLE OF CONTENTS GENERAL INFORMATION ....................................................................................................... 1 1. INTRODUCTION .............................................................................
Page 8 - M703E Calibrator Operator’s Manual; iv
TELEDYNE API Table of Contents M703E Calibrator Operator’s Manual 6.2. Overview of Operating modes .....................................................................................................................37 6.3. STANDBY Mode ..................................................................
Page 10 - vi
TELEDYNE API Table of Contents M703E Calibrator Operator’s Manual 9.2.4. Motherboard........................................................................................................................................ 143 9.2.4.1. A to D Conversion ...................................................
Page 11 - vii
TELEDYNE API M703E Calibrator Operator’s Manual Table of Contents 11.4.4. I 2 C Bus .............................................................................................................................................. 188 11.4.5. Keyboard/Display Interface......................................
Page 12 - LIST OF FIGURES; Set up for Using the M703E to Measure an External O; viii
TELEDYNE API Table of Contents M703E Calibrator Operator’s Manual LIST OF FIGURES Figure 3-1: M703E Front Panel Layout ................................................................................................................10 Figure 3-2: M703E Rear Panel Layout..................................
Page 13 - LIST OF TABLES; M703E Specifications for O; ix
TELEDYNE API M703E Calibrator Operator’s Manual Table of Contents Figure 10-4: O 3 Generator Temperature Thermistor and DC Heater Locations ................................................. 174 Figure 10-5: Location of O 3 Generator Reference Detector Adjustment Pot ....................................
Page 15 - LIST OF APPENDICES; APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION
TELEDYNE API M703E Calibrator Operator’s Manual Table of Contents LIST OF APPENDICES APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX A-1: Model 703E Software Menu Trees, Revision C.0 APPENDIX A-2: Model 703E Setup Variables Available Via Serial I/O, Revision C.0 APPENDIX A-3: Model 703...
Page 16 - xii
TELEDYNE API Table of Contents M703E Calibrator Operator’s Manual xii 05744 Rev B USER NOTES:
Page 17 - SECTION I
TELEDYNE API M703E Calibrator Operator’s Manual GENERAL INFORMATION SECTION I – GENERAL INFORMATION 05744 Rev B 1
Page 18 - USER NOTES
TELEDYNE API GENERAL INFORMATION M703E Calibrator Operator’s Manual 2 05744 Rev B USER NOTES
Page 19 - M703E CALIBRATOR OVERVIEW; NOTE; SECTION I – GENERAL INFORMATION; INTRODUCTION
TELEDYNE API M703E Calibrator Operator’s Manual Introduction 1. INTRODUCTION 1.1. M703E CALIBRATOR OVERVIEW The Model 703E is a microprocessor-based ozone calibrator for calibration of ambient ozone analyzers, such as the T-API M400E. The M703E features an internal ozone photometer that provides ver...
Page 20 - SECTION II – OPERATING INSTRUCTIONS
TELEDYNE API Introduction M703E Calibrator Operator’s Manual 4 05744 Rev B GETTING STARTED Instructions for setting up, installing and running your calibrator for the first time. GLOSSARY: Answers to the most frequently asked questions about operating the calibrator and a glossary of acronyms and te...
Page 21 - SPECIFICATIONS, APPROVALS AND; SPECIFICATIONS; Table 2-1: M703E Analytical Specifications
TELEDYNE API M703E Calibrator Operator’s Manual Specifications, Approvals and Warranty 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. SPECIFICATIONS Table 2-1: M703E Analytical Specifications Linearity +/- 1.0% of full scale Precision 1.0 ppb Stability +/- 2.0 ppb (photometer feedback mode) Response...
Page 22 - Table 2-3: M703E Specifications for Ozone Generator; CE MARK COMPLIANCE; EMISSIONS COMPLIANCE
TELEDYNE API Specifications, Approvals and Warranty M703E Calibrator Operator’s Manual Table 2-3: M703E Specifications for Ozone Generator Maximum Output 6 ppm LPM Minimum Output 100 ppb LPM Response Time: 180 Sec. (98%) Optical Feedback Standard Table 2-4: M703E Specifications for O 3 Photometer Fu...
Page 23 - WARRANTY; COVERAGE
TELEDYNE API M703E Calibrator Operator’s Manual Specifications, Approvals and Warranty 2.3. WARRANTY WARRANTY POLICY (02024D) Prior to shipment, T-API equipment is thoroughly inspected and tested. Should equipment failure occur, T-API assures its customers that prompt service and support will be ava...
Page 25 - GETTING STARTED; UNPACKING AND INITIAL SETUP; Remove the set-screw located in the top, center of the Front panel.
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 3. GETTING STARTED 3.1. UNPACKING AND INITIAL SETUP CAUTION THE M703E WEIGHS ABOUT 16.1 KG (35.5 POUNDS) WITHOUT OPTIONS INSTALLED. TO AVOID PERSONAL INJURY, WE RECOMMEND USING TWO PERSONS TO LIFT AND CARRY THE CALIBRATOR. 1. Inspect th...
Page 26 - VENTILATION CLEARANCE; be sure to leave sufficient ventilation clearance.; AREA; M703E Front Panel Layout
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual VENTILATION CLEARANCE : Whether the calibrator is set up on a bench or installed into an instrument rack, be sure to leave sufficient ventilation clearance. AREA MINIMUM REQUIRED CLEARANCE Back of the instrument 10 cm / 4 inches Sides o...
Page 27 - PHOTO; M703E Internal Layout – Top View
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 1 LPM Flow Control Assy. AC Power Connector Gas Inlets & Outlets Motherboard Relay PCA 5 LPM Flow Control Assy. CPU PCA ON / OFFSwitch Front Panel DC Power supplies Back Panel Ethernet PCA installed here (Optional) PHOTO M E T ER Ph...
Page 28 - M703E Pneumatic Diagram
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual O 3 FLOW SENS O R O 3 Generator A ssembly O 3 G E NERA TOR Flow Control (100 cm 3 /min) Fl o w Cont ro l (5.0 lpm ) Fl o w Cont ro l (1.0 to 2 .0 LP M) CHA R COA L SCR U B B E R Figure 3-4: M703E Pneumatic Diagram 12 05744 Rev B
Page 29 - ELECTRICAL CONNECTIONS; HIGH VOLTAGES ARE PRESENT INSIDE THE CALIBRATORS CASE; CAUTION; DO NOT LOOK AT THE PHOTOMETER UV LAMP.; ANALOG OUT; M703E the TEST CHANNEL Connector
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 3.2. ELECTRICAL CONNECTIONS 3.2.1. POWER CONNECTION Verify the correct line voltage and frequency configuration on the serial number tag on the rear panel of the M703E. Attach the power cord to the calibrator and plug it into a power ou...
Page 30 - Status Output Connector; The pin assignments for the Status Outputs are:; Table 3-1: Status Output Pin Assignments
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual 3.2.3. CONNECTING THE STATUS OUTPUTS The status outputs report calibrator conditions via optically isolated NPN transistors, which sink up to 50 mA of DC current. These outputs can be used interface with devices that accept logic-level ...
Page 31 - CONNECTOR INPUT
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 3.2.4. CONNECTING THE CONTROL INPUTS The calibrator is equipped with 12 digital control inputs that can be used to Initiate various user programmable calibration sequences (see Section 6.5.1.5 for instructions on assigning the control i...
Page 32 - M703E Digital Control Input Connectors
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is the most convenient method. However, if full isolation is required, an external 5 VDC power supply should be used. Example o...
Page 33 - CONTROL OUTPUTS; M703E Digital Control Output Connector; The pin assignments for the control outputs are:; Table 3-3: M703E Control Input Pin Assignments; STATUS DEFINITION
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 3.2.5. CONNECTING THE CONTROL OUTPUTS The calibrator is equipped with 12 opto-isolated, digital control outputs. These outputs are activated by the M703E’s user-programmable, calibration sequences (see Section 6.5.1.6 for instructions o...
Page 34 - The M703E firmware supports dynamic IP addressing or DHCP.
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual 3.2.6. CONNECTING THE SERIAL PORTS If you wish to utilize either of the calibrator’s two serial interface COMM ports, refer to Section 7.1 of this manual for instructions on their configuration and usage. 3.2.7. CONNECTING TO A LAN OR T...
Page 35 - PNENUMATIC CONNECTIONS; Basic Pneumatic Setup of M703E
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 3.3. PNENUMATIC CONNECTIONS Figure 3-9: Basic Pneumatic Setup of M703E 3.3.1. DRY AIR IN When using the internal zero air pump, a source of dry air should be connected to the port labeled ‘Dry Air In’ on the rear panel. This air should ...
Page 37 - INITIAL OPERATION; The calibrator should automatically switch to; STANDBY; mode after completing the boot-up sequence.
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 3.4. INITIAL OPERATION If you are unfamiliar with the M703E theory of operation, we recommend that you read Chapter 9. For information on navigating the calibrator’s software menus, see the menu trees described in Appendix A.1. 3.4.1. S...
Page 38 - Name Color Behavior; Instrument is in; Unit is operating in; To view and clear warning messages, press:
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual 3.4.2. WARM UP The M703E Photometric calibrator requires a minimum of 30 minutes for all of its internal components to reach a stable operating temperature. During that time, various portions of the instrument’s front panel will behave ...
Page 39 - MESSAGE MEANING
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started Table 3-5: Possible Warning Messages at Start-Up MESSAGE MEANING ANALOG CAL WARNING The calibrator’s A/D converter or at least one analog input channel has not been calibrated. CONFIG INITIALIZED Stored Configuration information has bee...
Page 40 - Only appears when warning
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual 3.4.4. FUNCTIONAL CHECK 7. After the calibrator’s components have warmed up for at least 30 minutes, verify that the software properly supports any hardware options that are installed. 8. Check to make sure that the calibrator is functi...
Page 41 - GENERATOR
TELEDYNE API M703E Calibrator Operator’s Manual Getting Started 05744 Rev B 25 3.4.5. OPERATING MODES FOR THE O 3 GENERATOR The O 3 generator can be set to operate in three different modes: 3.4.5.1. CNST (CONSTANT) In this mode, the O 3 output of the generator is based on a single, constant, drive v...
Page 42 - Push the regulator knob back in to lock.; Location of Pressure Regulator Adjustment Knob
TELEDYNE API Getting Started M703E Calibrator Operator’s Manual 3.4.7. SETTING THE M703E’S OUTPUT FLOW RATE The output flow rate of the M703E should be adjusted to match the gas flow requirements of the analyzers connected to the output manifold. NOTE The minimum total flow should equal the sum of...
Page 43 - FREQUENTLY ASKED QUESTIONS AND; FAQ’S; ENTR; GLOSSARY; cubic centimeter
TELEDYNE API M703E Calibrator Operator’s Manual Frequently Asked Questions and Glossary 4. FREQUENTLY ASKED QUESTIONS AND GLOSSARY 4.1. FAQ’S The following list is a list from the T-API Customer Service Department of the 10 most commonly asked questions relating to the Model 703E Photometric Calibra...
Page 45 - OPTIONAL HARDWARE AND SOFTWARE
TELEDYNE API M703E Dynamic Dilution Calibrator Operator’s Manual Optional Hardware and Software 5. OPTIONAL HARDWARE AND SOFTWARE This includes a brief description of the hardware and software options available for the M703E Photometric Calibrator. For assistance with ordering these options, please ...
Page 46 - M703E Dynamic Dilution Calibrator Operator’s Manual; COMMUNICATION OPTIONS; M703E Multidrop Card
TELEDYNE API Optional Hardware and Software M703E Dynamic Dilution Calibrator Operator’s Manual 5.2. COMMUNICATION OPTIONS 5.2.1. RS232 MODEM CABLES (OPTS 60 AND 60A) The analyzer can have come standard with a shielded, straight-through DB-9F to DB-9F cable of about 1.8 m length, which should fit mo...
Page 47 - M703E Ethernet Card
TELEDYNE API M703E Dynamic Dilution Calibrator Operator’s Manual Optional Hardware and Software 5.2.4. ETHERNET (OPT 63) The ETHERNET option allows the calibrator to be connected to any Ethernet local area network (LAN) running TCP/IP. The local area network must have routers capable of operating at...
Page 48 - NIST TRACEABLE, PRIMARY STANDARD CERTIFICATION
TELEDYNE API Optional Hardware and Software M703E Dynamic Dilution Calibrator Operator’s Manual 32 05744 Rev B 5.2.5. ETHERNET + MULTIDROP (OPT 64) This option allows the instrument to communicate on both RS-232 and ETHERNET networks simultaneously. It includes the following: RS232 MODEM CABLES (O...
Page 49 - SECTION II
TELEDYNE API M703E Dynamic Dilution Calibrator Operator’s Manual OPERATING INSTRUCTIONS SECTION II – OPERATING INSTRUCTIONS 05744 Rev B 33
Page 51 - OPERATING THE M703E CALIBRATOR
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6. OPERATING THE M703E CALIBRATOR The M703E calibrator is a computer-controlled calibrator with a dynamic menu interface for easy and yet powerful and flexible operation. All major operations are controlled from the front...
Page 52 - A variety of; TEST FUNCTIONS; To view these; or; Test Functions; will scroll at; TEST MEASUREMENT; ACT=GENERATE XX PPB O3
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.1. TEST FUNCTIONS A variety of TEST FUNCTIONS are available for viewing at the front panel whenever the calibrator is at the MAIN MENU . These measurements provide information about the present operating status of the c...
Page 53 - OVERVIEW OF OPERATING MODES; Mode Field; STANDBY ACT =STANDBY; STANDBY Key; MODE MEANING
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.2. OVERVIEW OF OPERATING MODES The M703E calibrator software has a variety of operating modes. Most commonly, the calibrator will be operating in STANDBY mode. In this mode, the calibrator and all of its subsystems are ...
Page 54 - STANDBY MODE; Some functions under the; STBY
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.3. STANDBY MODE When the M703E Photometric Calibrator is in standby mode, it is at rest. Some functions under the SETUP MORE DIAG submenu, those which conflict with accurate creation of calibration gas mixtures (e...
Page 55 - GENERAL INFORMATION ABOUT THE GENERATE MODE; mode the Zero Air Pump (if enabled) and; AUTO: Basic Generation of Calibration Gas; The M703E will now enter; GENERATE
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 05744 Rev B 39 6.4. GENERAL INFORMATION ABOUT THE GENERATE MODE The GENERATE mode is the mode of operation where the M703E is actively producing calibration gas, either zero or some specified concentration of ozone. In th...
Page 56 - AUTOMATIC CALIBRATION SEQUENCES; SEQ: PROGRAMMING CALIBRATION; NAME; For each sequence, seven attributes must be programmed. They are:; Table 6-3: Automatic Calibration SEQUENCE Set Up Attributes; ATTRIBUTE NAME; Table 6-4: Calibration SEQUENCE Step Instruction; INSTRUCTION NAME
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.5. AUTOMATIC CALIBRATION SEQUENCES The M703E calibrator can be set up to perform automatic calibration sequences of multiple steps. These sequences can perform all of the calibration operations available for manual oper...
Page 59 - REPEAT COUNTER; To set th; e REPEAT COUNTER
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.5.1.3. Setting the Repeat Count for a Sequence The sequence can be set to repeat a certain number of times, from 1 to 100. It can also be set to repeat indefinitely by inputting a zero (0) into the REPEAT COUNTER . To s...
Page 63 - CC INPUT
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.5.1.5. Setting Up Control Inputs for a Sequence The M703E calibrator’ control inputs allow the entire sequence to be triggered from an external source. This feature allows the calibrator to operate in a slave mode so th...
Page 64 - CC OUTPUT
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.5.1.6. Setting Up Control Outputs for a Sequence The M703E calibrator’s control outputs allow the entire sequence to be triggered from an external source. This feature allows the calibrator to control devices that accep...
Page 65 - Table 6-5: Sequence Progress Reporting Mode; MODE; DESCRIPTION; STEP
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.5.1.7. Setting the PROGRESS Reporting Mode for the Sequences As sequences run, the M703E calibrator reports progress by displaying a message in the MODE field of the front panel display (See Figure 3-1). There are sever...
Page 66 - INSERT STEP; submenu by pressing:
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.5.2. ADDING SEQUENCE STEPS To insert an instruction step into a sequence, navigate to the INSERT STEP submenu by pressing: 50 05744 Rev B
Page 67 - step that; EXIT; ZERO AIR
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.5.2.1. The Generate Step This step operates and is programmed similarly to the GENERATE AUTO. At the end of the programming sequence, the M703E firmware will automatically insert a DURATION step that needs to be defin...
Page 69 - To insert an; EXECSEQ
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.5.2.4. The EXECSEQ Step The EXECSEQ step allows the sequence to call another, already programmed sequence. This is a very powerful tool in that it allows the user to create a “toolbox” of often-used operations that can ...
Page 73 - SETUP
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.6. SETUP CFG Pressing the CFG key displays the instrument’s configuration information. This display lists the calibrator model, serial number, firmware revision, software library revision, CPU type and other informati...
Page 74 - The M703E has a time of day clock that supports the; DURATION; step of the calibration sequence feature, time; returns to
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 58 05744 Rev B 6.7. SETUP CLK 6.7.1. SETTING THE INTERNAL CLOCK’S TIME AND DAY The M703E has a time of day clock that supports the DURATION step of the calibration sequence feature, time of day TEST function, and time s...
Page 76 - PASS; menu item, the system will prompt the user
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.8. SETUP PASS The M703E provides password protection of the calibration and setup functions to prevent unauthorized adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt the u...
Page 77 - password enables the; mode
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator Example: If all passwords are enabled, the following keypad sequence would be required to enter the VARS or DIAG submenus: SYSTEM ENTER SETUP PASS:0 8 1 8 ENTR EXIT SYSTEM ENTER SETUP PASS:0 0 0 0 ENTR EXIT SETUP X.X PRIM...
Page 78 - To configure the analyzer’s; , set the electronic signal type of each channel and calibrate the; TEST CHANNEL; function to be output on the channel.
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.9. SETUP DIAG TEST CHAN OUTPUT: USING THE TEST CHANNEL ANALOG OUTPUT The M703E calibrator comes equipped with one analog output. It can be set by the user to carry the current signal level of any one of the paramete...
Page 79 - Not used in the
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator To access the analog I/O configuration sub menu, press: AIO Configuration Submenu STANDBY ACT =STANDBY <TST TST> GEN STBY SEQ SETUP SETUP X.X PRIMARY SETUP MENU O3 SEQ CFG CLK PASS MORE EXIT SETUP X.X SECONDARY SETU...
Page 80 - to the list of Test Functions viewable via the Front Panel Display.
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.9.1.2. Selecting a Test Channel Function to Output The Test Functions available to be reported are: Table 6-8: Test Channels Functions Available on the M703E’s Analog Output TEST CHANNEL DESCRIPTION ZERO FULL SCALE NONE...
Page 81 - Channel and select a function press:; discards the new; accepts the
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 05744 Rev B 65 To activate the TEST Channel and select a function press: DIAG TEST CHANNEL:NONE PREV NEXT ENTR EXIT STANDBY ACT =STANDBY <TST TST> GEN STBY SEQ SETUP Make sure that the M703E is in standby mode. SETU...
Page 82 - RANGE SPAN
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.9.1.3. TEST CHANNEL VOLTAGE RANGE Configuration In its standard configuration the analog outputs is set to output a 0 – 5 VDC signals. Several other output ranges are available (see Table 7-5). Each range has is usable ...
Page 85 - AUTO CAL
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.9.2. TEST CHANNEL CALIBRATION TEST CHANNEL calibration needs to be carried out on first startup of the analyzer (performed in the factory as part of the configuration process) or whenever re-calibration is required. The...
Page 88 - Setup for Calibrating the TEST CHANNEL
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.9.2.3. Manual Calibration of the TEST CHANNEL configured for Voltage Ranges For highest accuracy, the voltages of the analog outputs can be manually calibrated. NOTE: The menu for manually adjusting the analog output si...
Page 89 - OFF
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 05744 Rev B 73 To adjust the signal levels of an analog output channel manually, press: From the AIO CONFIGURATION SUBMENU (See Section 6.9.1.1.) DIAG ANALOG I/O CONFIGURATION PREV NEXT ENTR EXIT Continue pressing SET>...
Page 90 - AIN CALIBRATION
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.9.3. AIN CALIBRATION This is the sub-menu calibrates the analyzer’s A-to-D conversion circuitry. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. ...
Page 91 - STD PRESSURE; VARS
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.10. SETUP MORE VARS: INTERNAL VARIABLES (VARS) The M703E has several-user adjustable software variables, which define certain operational parameters. Usually, these variables are automatically set by the instrument’...
Page 93 - OPERATING THE M703E CALIBRATOR AS AN O; PHOTOMETER; To convert the M703E from an O; Enc; Photometric; Capped; Minimum input gas flow for
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 6.11. OPERATING THE M703E CALIBRATOR AS AN O 3 PHOTOMETER The M703E can easily be configured to measure an external source of ozone. 6.11.1. SET UP FOR OPERATING THE M703E AS AN O 3 PHOTOMETER To convert the M703E from an...
Page 96 - is a combination of several parameters:; LEVEL
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.12.3. LEVELS A LEVEL is a combination of several parameters: An ID number for the LEVEL An action, (e.g. GENERATE, STANDBY) A target concentration value An output flow rate (if applicable) Configuration for on...
Page 97 - find the; LVL
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 05744 Rev B 81 6.12.5. PROGRAMMING NEW LEVELS To begin programming a new LEVEL find the LVL submenu by pressing: CHOOSE ACTION Submenu SETUP X.X ACTION TO PERFORM:GENERATE PREV NEXT ENTR EXIT STANDBY ACT =STANDBY <TST ...
Page 99 - MANUAL
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator 05744 Rev B 83 6.12.5.2. Creating a MANUAL LEVEL To create a level using the M703E’s MANUAL generation function, press: SETUP X.X INSERT STEP: MANUAL PREV ENTR EXIT CHOOSE ACTION Submenu SETUP X.X ACTION TO PERFORM:GENERA...
Page 100 - Levels are displayed according to the following Format:
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.12.5.3. Editing or Deleting a LEVEL To edit or delete an existing LEVEL , press: Levels are displayed according to the following Format: LEVEL ID: Any number between 0 and 99. This will be the number used to select th...
Page 101 - LEADS Level Display Format
TELEDYNE API M703E Calibrator Operator’s Manual Operating the M703E Calibrator Figure 6-5: LEADS Level Display Format 05744 Rev B 85
Page 102 - associated with LEADS; communicate status over the serial data port; STATUS BLOCKS
TELEDYNE API Operating the M703E Calibrator M703E Calibrator Operator’s Manual 6.12.6. CONFIGURING LEVEL STATUS BLOCKS There are two STATUS BLOCKS associated with LEADS LEVELS . BLOCK 1: This block corresponds to the physical CONTROL OUTPUT connections located on the back panel of the M703E (see F...
Page 103 - OPERATING THE M703E OVER THE SERIAL I/O; USING THE ANALYSER’S COMMUNICATION PORTS; DTE devices receive data on pin 2 and transmit data on pin 3.
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7. OPERATING THE M703E OVER THE SERIAL I/O PORTS 7.1. USING THE ANALYSER’S COMMUNICATION PORTS The M703E is equipped with two serial communication ports located on the rear panel accessible via 2 DB-9 conne...
Page 104 - ASSIGNMENTS; TXD
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual 7.1.2. COMM PORT DEFAULT SETTINGS AND CONNECTOR PIN ASSIGNMENTS Received from the factory, the calibrator is set up to emulate an RS-232 DCE device. RS-232 (COM1) : RS-232 (fixed), DB-9 male connector. o ...
Page 109 - COM; menu. This test sends a string of
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.1.5. COMM PORT TESTING The serial ports can be tested for correct connection and output in the COM menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running,...
Page 111 - Table 7-2: Terminal Mode Software Commands; COMMAND Function
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.1.7. TERMINAL OPERATING MODES The M703E can be remotely configured, calibrated or queried for stored data through the serial ports. As terminals and computers use different communication schemes, the cali...
Page 112 - COMMAND COMMAND; are all valid integers.
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual 7.1.7.2. Command Syntax Commands are not case-sensitive and all arguments within one command (i.e. ID numbers, keywords, data values, etc.) must be separated with a space character. All Commands follow the ...
Page 114 - SECURITY MODE
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual 7.1.7.5. COM Port Password Security In order to provide security for remote access of the M703E, a LOGON feature can be enabled to require a password before the instrument will accept commands. This is done...
Page 115 - REMOTE ACCESS BY MODEM; Once the cable has been connected, check to make sure:; MODEM ENABLE; communication mode is turned; ON; To change this setting press:
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.2. REMOTE ACCESS BY MODEM The M703E can be connected to a modem for remote access. This requires a cable between the calibrator’s COM port and the modem, typically a DB-9F to DB-25M cable (available from ...
Page 116 - If there is a problem initializing the; “MODEM NOT INITIALIZED”; Continue pressing
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual To Initialize the modem press: If there is a problem initializing the modem the message, “MODEM NOT INITIALIZED” will appear. STANDBY ACT =STANDBY <TST TST> GEN STBY SEQ SETUP SETUP X.X PRIMARY SETUP ...
Page 117 - MULTIDROP RS-232 SET UP; One Teledyne Instruments Option 62 installed.
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.3. MULTIDROP RS-232 SET UP The RS-232 multidrop consists of a printed circuit assembly that plugs onto the CN3, CN4 and CN5 connectors of the CPU card and the cabling to connect it to the calibrator’s mot...
Page 118 - and testing it before setting up the rest of the chain.; CALIBRATOR; Make Sure; RS232-Multidrop PCA Host
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual 10. Using straight-through, DB9 male DB9 Female cables, interconnect the host and the calibrators as shown in Figure 6-14. NOTE: Teledyne Instruments recommends setting up the first link, between the Host...
Page 119 - RS-485 CONFIGURATION OF COM2; The RS-485 port can be configured with or without a 150; CPU card Locations of RS-232/485 Switches, Connectors and Jumpers
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.4. RS-485 CONFIGURATION OF COM2 As delivered from the factory, COM2 is configured for RS-232 communications. This port can be re-configured for operation as a non-isolated, half-duplex RS-485 port capable...
Page 120 - GND
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual When COM2 is configured for RS-485 operation the port uses the same female DB-9 connector on the back of the instrument as when Com2 is configured for RS-232 operation, however, the pin assignments are diff...
Page 121 - REMOTE ACCESS VIA THE ETHERNET; port. Once the Ethernet option is installed and activated, the; INET; LED FUNCTION; BAUD RATE
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.5. REMOTE ACCESS VIA THE ETHERNET When equipped with the optional Ethernet interface, the calibrator can be connected to any standard 10BaseT Ethernet network via low-cost network hubs, switches or router...
Page 124 - DHCP
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual 7.5.2.1. Manually Configuring the Network IP Addresses There are several circumstances when you may need to configure the interface settings of the calibrator’s Ethernet card manually. The INET sub-menu may...
Page 125 - INSTRUMENT IP; and; SUBNET MASK; addresses by pressing:
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports STEP 2: Configure the INSTRUMENT IP , GATEWAY IP and SUBNET MASK addresses by pressing: Internet Configuration Keypad Functions KEY FUNCTION [0] Press this key to cycle through the range of numerals and ava...
Page 126 - HOSTNAME; Teledyne Instruments M703E calibrators is
TELEDYNE API Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual 110 05744 Rev B 7.5.3. CHANGING THE CALIBRATOR’S HOSTNAME The HOSTNAME is the name by which the calibrator appears on your network. The default name for all Teledyne Instruments M703E calibrators is M703E ....
Page 127 - APICOM REMOTE CONTROL PROGRAM; standing in front of the instrument.; APICOM Remote Control Program Interface
TELEDYNE API M703E Calibrator Operator’s Manual Operating The M703E over the Serial I/O Ports 7.6. APICOM REMOTE CONTROL PROGRAM APICOM is an easy-to-use, yet powerful interface program that allows the user to access and control any of Teledyne Instruments’ main line of ambient and stack-gas instrum...
Page 129 - M703E CALIBRATION AND VERIFICATION; VERIFYING AND CALIBRATING THE M703E’S O; external source of O; Note
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8. M703E CALIBRATION AND VERIFICATION Basic electronic calibration of the M703E Calibrator is performed at the factory. Normally there is no need to perform this factory calibration in the field however, the performan...
Page 131 - PHOTOMETER PERFORMANCE
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.1.2. VERIFYING O 3 PHOTOMETER PERFORMANCE To verify the performance of the M703E’s optional internal photometer perform the following steps: Wait A MINIMUM OF 10 MINUTES or until the ACT reading settles down STANDBY...
Page 132 - OFFSET
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.1.3. CALIBRATING THE O 3 PHOTOMETER The following procedure sets values held in the calibrator’s memory of for zero point OFFSET and SLOPE. 8.1.3.1. Photometer Zero Calibration To set the zero point offset for the M...
Page 134 - PHOTOMETER DARK CALIBRATION; Yes
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.1.4. O 3 PHOTOMETER DARK CALIBRATION The Dark Calibration Test turns off the Photometer UV Lamp and records any offset signal level of the UV Detector-Preamp-Voltage to Frequency Converter circuitry. This allows the...
Page 135 - CALIBRATING THE O
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.2. CALIBRATING THE O 3 GENERATOR The M703E calibrator’s software includes a routine for automatically calibrating the O 3 generator. A calibration table of drive voltages stored in the M703E’s memory is the basis fo...
Page 136 - GENERATOR CALIBRATION POINTS; To view these calibration points, press:
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.2.2. VIEWING O 3 GENERATOR CALIBRATION POINTS To view these calibration points, press: SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG EXIT Make sure that the M703E is in standby mode. SETUP X.X ENTER PASSWORD 8 1 8 E...
Page 137 - GENERATOR CALIBRATION
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.2.3. ADDING OR EDITING O 3 GENERATOR CALIBRATION POINTS To add a calibration point to the table or edit an existing point, press: 05744 Rev B 121
Page 138 - To delete an existing calibration point, press:
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.2.4. DELETING O 3 GENERATOR CALIBRATION POINTS To delete an existing calibration point, press: 122 05744 Rev B
Page 139 - GENERATOR CALIBRATION POINTS ON / OFF; To enable or disable an existing calibration point, press:
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.2.5. TURNING O 3 GENERATOR CALIBRATION POINTS ON / OFF To enable or disable an existing calibration point, press: STANDBY ACT =STANDBY <TST TST> GEN STBY SEQ SETUP Make sure that the M703E is in standby mode. ...
Page 140 - To run the automatic O
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.2.6. PERFORMING AN AUTOMATIC CALIBRATION OF THE O 3 GENERATOR To run the automatic O 3 generator calibration program, press: 124 05744 Rev B
Page 141 - M703E GAS PRESSURE SENSOR CALIBRATION; Pressure Calibration Monitor Points
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.3. M703E GAS PRESSURE SENSOR CALIBRATION The M703E Calibrator has two sensors that monitor the pressure of the gases flowing through the instrument. The data collected by these sensors is used to compensate the fina...
Page 142 - Generator Pressure Monitor Point Physical Location– M703E
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual Pressure Regulator Attach Pressure Monitor Here Outlets to O 3 Generator Figure 8-4: O 3 Generator Pressure Monitor Point Physical Location– M703E 126 05744 Rev B
Page 143 - PRESSURE SENSORS; Turn off the calibrator and open the top cover.
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.3.2. CALIBRATING THE REGULATOR AND PHOTOMETER PRESSURE SENSORS 1. Turn off the calibrator and open the top cover. 2. Connect a pressure meter to the Regulator Pressure measurement fitting. This fitting is located on...
Page 144 - M703E GAS FLOW CALIBRATION; M703E REGULATOR PRESSURE TO OUTPUT FLOW
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.4. M703E GAS FLOW CALIBRATION The M703E has two gas flow characteristics that affect its performance: the flow of gas thought the sample chamber of the instrument’s photometer and the total gas flow being output. Wh...
Page 145 - During the; the photometer; To perform a; PHOTO FLOW
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.4.1. CALIBRATING THE PHOTOMETER’S SAMPLE GAS FLOW NOTE The procedure described in this section requires an independent, calibrated gas flow meter/monitor be connected to the EXHAUST fitting on the back of the M703E....
Page 146 - Output Flow Calibration Monitor Point
TELEDYNE API M703E Calibration and Verification M703E Calibrator Operator’s Manual 8.4.2. CALIBRATING THE OUTPUT GAS FLOW 8.4.2.1. Output Gas Flow Set Up The procedure described in this section requires an independent, calibrated flow meter/monitor and the following set up: O 3 FLOW SENSOR O 3 G e n...
Page 147 - and the AC dry air pump; An external flow
TELEDYNE API M703E Calibrator Operator’s Manual M703E Calibration and Verification 8.4.2.2. Performing an Output Gas Flow Calibration During the PHOTO FLOW calibration, the M703E software automatically turns the DC pump downstream from the photometer OFF and the AC dry air pump ON . To perform a PHO...
Page 149 - SECTION III
TELEDYNE API M703E Calibrator Operator’s Manual TECHNICAL INFORMATION SECTION III – TECHNICAL INFORMATION 05744 Rev B 133
Page 151 - THEORY OF OPERATION; PNEUMATIC OPERATION; Location of Gas Flow Control Assemblies
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9. THEORY OF OPERATION 9.1. PNEUMATIC OPERATION 9.1.1. GAS FLOW CONTROL Gas flow rates are set by various flow control assemblies located in the gas stream(s). 9.1.1.1. Flow Control Assemblies O 3 FLOW SENS O R O 3 Generator A ssemb...
Page 152 - REG PRESSURE; called; PHOTO SPRESS; displayed on the front panel.
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual stays at least 2:1 the gas flow rate is unaffected by any fluctuations, surges, or changes in downstream pressure because such variations only travel at the speed of sound themselves and are therefore cancelled out by the sonic shoc...
Page 153 - ELECTRONIC OPERATION; M703E Electronic Block Diagram
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.2. ELECTRONIC OPERATION 9.2.1. OVERVIEW Photometer Lamp Power Supply Photometer Detector Preamp Absorption tube Photometer Detector O 3 Generator ‘ Optional Ethernet Card Analog Outputs Aout 1 Aout 4 Analog Outputs (D/A) External ...
Page 154 - M703E CPU Board Annotated
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual A variety of sensors report the physical and operational status of the calibrator’s major components, again through the signal processing capabilities of the motherboard. These status reports are used as data for the concentration c...
Page 156 - Relay Board PCA with AC Relay Retainer Removed
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.2.3. RELAY PCA The relay board is one of the central switching and power distribution units of the calibrator. It contains power relays, valve drivers and status LEDs for all heated zones and valves, as well as thermocouple amplif...
Page 157 - Heater Control Loop Block Diagram.; Generator Lamp Heater; Status LED Locations – Relay PCA
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.2.3.1. Valve Control The relay board also hosts two valve driver IC's, each of which can drive up four valves. In the M703E, the relay PCA controls only those valves associated with the O 3 generator and photometer options. 9.2.3....
Page 158 - LED COLOR; The most important of the status LEDs on the relay board is the red I
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual Table 9-1: Relay Board Status LEDs LED COLOR DESCRIPTION FUNCTION D1 Red Watchdog Circuit; I 2 C bus operation. Blinks when I 2 C bus is operating properly D2 Yellow Dry Air Pump Status When lit the zero air AC pump is running. D3-6...
Page 160 - STATUS; CONTROL; CONTROL; C Data Bus; C data to parallel inputs and
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.2.4.5. External Digital I/O The external digital I/O performs two functions. The STATUS outputs carry logic-level (5V) signals through an optically isolated 8-pin connector on the rear panel of the calibrator. These outputs convey...
Page 161 - RELAY PCA; M703E Power Distribution Block diagram
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.2.5. POWER SUPPLY AND CIRCUIT BREAKER The M703E calibrator operates in two main AC power ranges: 100-120 VAC and 220-240 VAC (both ± 10%) between 47 and 63 Hz. A 5 ampere circuit breaker is built into the ON/OFF switch. In case of...
Page 162 - Location of the AC Configuration Jumper for the Dry Air Pump
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.2.6. AC POWER CONFIGURATION The E-Series digital electronic systems will operate with any of the specified power regimes. As long as instrument is connected to 100-120 VAC or 220-240 VAC at either 50 or 60 Hz it will turn on and a...
Page 163 - Present on 50 Hz version of jumper set,; FRONT PANEL INTERFACE
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 110 VAC /115 VAC 220 VAC /240 VAC 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 Present on 50 Hz version of jumper set, and only functional for M300E Figure 9-9: Pump AC Power Jumpers (JP7) 9.3. FRONT PANEL INTERFACE MODE FIELD KEY DEFI...
Page 164 - data and response messages during interactive tasks.
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.3.1.1. Calibrator Status LEDs Three LEDS are used to inform the user of the instruments basic operating status Table 9-3: Front Panel Status LEDs Name Color Behavior Significance Main Message Field N/A Displays Warning messages an...
Page 165 - Keyboard and Display Interface Block Diagram
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.3.1.4. Keyboard/Display Interface Electronics FRONT PANEL Keypad Decoder Key Press Detect KEYBOARD Beeper Sample LED (Green) Cal LED (Yellow) Fault LED (Red) Display Data Decoder Display Power Watchdog From 5 VDC Power Supply I 2 ...
Page 166 - DISPLAY DATA DECODER; SOFTWARE OPERATION; Schematic of Basic Software Operation
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual DISPLAY DATA DECODER This decoder translates the serial data sent by the CPU (in TTY format) into a bitmapped image that is sent over a parallel data bus to the display. DISPLAY CONTROLLER This circuit manages the interactions betwe...
Page 167 - GENERATOR OPERATION; GENERATION; which also may be undesirable in a calibration application; Generator Internal Pneumatics
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.5. O 3 GENERATOR OPERATION 9.5.1. PRINCIPLE OF PHOTOLYTIC O 3 GENERATION Ozone is a naturally occurring substance that is sometimes called "activated oxygen". It contains three atoms of oxygen (O3) instead of the usual two...
Page 168 - Generator Valve and Gas Fixture Locations; GENERATOR ELECTRONIC OPERATION; Electronically the O
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.5.2. GENERATOR PNEUMATIC OPERATION The rate of flow through the O 3 generator is controlled by a flow control assembly located on the Regulator Sub- Assembly in the front of the M703E. Measure / Reference Valve for Photometer Benc...
Page 169 - Generator Electronic Block Diagram
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation O 3 G ene ra tor Sensor Inpu ts O 3 Generato r Lamp He at er Thermi st or Inte rface A/D Conve rter Figure 9-15: O 3 Generator Electronic Block Diagram UV Lamp O 3 Generator Reference Detector O 3 Generator Reference Detector PCA UV...
Page 170 - Generator Temperature Control; In order to operate at peak efficiency the UV lamp of the M703E’s O; UV Lamp; Generator Temperature Thermistor and DC Heater Locations
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.5.3.1. O 3 Generator Temperature Control In order to operate at peak efficiency the UV lamp of the M703E’s O 3 generator is maintained at a constant 48ºC. if the lamp temperature falls below 43ºC or rises above 53ºC a warning is i...
Page 171 - PHOTOMETER OPERATION; Concentration; LC; ln; at STP
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.6. PHOTOMETER OPERATION The Model M703E calibrator’s optional photometer determines the concentration of Ozone (O 3 ) in a sample gas drawn through it. Sample and calibration gasses must be supplied at ambient atmospheric pressure...
Page 172 - inHg; TIME INDEX; CYCLE REPEAT EVERY 6 SECONDS
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual In order to account for this effect the following addition is made to the equation: Equation 9-7 × × × = Ρ inHg 92 . 29 Κ 273 Τ L α 1 I I ln C o o Where: T = sample ambient temperature in degrees Kelvin P = ambient pressure in inche...
Page 173 - Photometer Gas Flow – Measure Cycle; Photometer Gas Flow – Reference Cycle
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation PHOTOMETER PRESSURE SENSOR O 3 GEN / PHOTOMETER PRESSURE / FLOW SENSOR PCA O 3 GAS INPUT PRESSURE SENSOR O 3 FLO W SENSO R O 3 Gen erato r As sembly O 3 G E NERAT OR Fl ow C ontrol (1 .0 LPM) REF/MEAS Valve On Back Panel M703E Chass...
Page 174 - Photometer Absorption Path
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.6.1.3. The Absorption Path In the most basic terms, the M703E photometer uses a high energy, mercury vapor lamp to generate a beam of UV light. This beam passes through a window of material specifically chosen to be both non-react...
Page 175 - Photometer Layout – Top Cover Removed
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.6.2. PHOTOMETER LAYOUT The Photometer is where the absorption of UV light by ozone is measured and converted into a voltage. It consists of several sub-assemblies: A mercury-vapor UV lamp. This lamp is coated in a material that ...
Page 176 - MOTHER BOARD; CPU Card; Photometer Electronic Block Diagram; Photometer Temperature Control; PHOTO LAMP TEMP
TELEDYNE API Theory of Operation M703E Calibrator Operator’s Manual 9.6.4. PHOTOMETER ELECTRONIC OPERATION Photometer Photometer Lamp Heater MOTHER BOARD Photometer Lamp Power Supply PC 104 Bus I 2 C Bus Photometer Pump Thermistor Interface Photometer Detector Preamp A/D Converter RELAY PCA I 2 C St...
Page 177 - generator assembly measure; - The flow rate of gas through the photometer measured in LPM.; the same as
TELEDYNE API M703E Calibrator Operator’s Manual Theory of Operation 9.6.4.2. Pneumatic Sensors for the O 3 Photometer The several sensors located on the pneumatic sensor just to the left rear of the O 3 generator assembly measure the absolute pressure and the flow rate of gas inside the photometer’s...
Page 179 - MAINTENANCE SCHEDULE & PROCEDURES; MAINTENANCE SCHEDULE
TELEDYNE API M703E Calibrator Operator’s Manual Maintenance Schedule & Procedures 10. MAINTENANCE SCHEDULE & PROCEDURES Predictive diagnostic functions including failure warnings and alarms built into the calibrator’s firmware allow the user to determine when repairs are necessary without pe...
Page 180 - M703E Photometric O3 Calibrator Operators Manual
TELEDYNE API Maintenance Schedule & Procedures M703E Photometric O3 Calibrator Operators Manual 164 05744 Rev B USER NOTES:
Page 183 - PERFORMING LEAK CHECKS
TELEDYNE API M703E Calibrator Operator’s Manual Maintenance Schedule & Procedures 10.2. PERFORMING LEAK CHECKS Leaks are the most common cause of analyzer malfunction; Section Error! Reference source not found. presents a simple leak check procedure. Section 10.2.1 describes a more thorough proc...
Page 184 - Pneumatic setup for performing Pressure Leak Checks
TELEDYNE API Maintenance Schedule & Procedures M703E Calibrator Operators Manual Once the fittings have been wetted with soap solution. Do not re-apply vacuum as it will draw soap solution into the instrument and contaminate it. Do not exceed 15 psi pressure. 11. Once the leak has been loc...
Page 186 - M703E Calibrator Operators Manual; Remove the instrument cover
TELEDYNE API Maintenance Schedule & Procedures M703E Calibrator Operators Manual NOTE The M703E calibrator cannot be leak checked with the pump in line due to internal leakage that normally occurs in the pump. 1. Remove the instrument cover 2. Locate the photometer pump. 3. Disconnect the two fi...
Page 187 - CLEANING OR REPLACING THE ABSORPTION TUBE; when cleaning or replacing the absorption tube may be required.; REBUILDING THE DRY AIR PUMP
TELEDYNE API M703E Calibrator Operator’s Manual Maintenance Schedule & Procedures 10.3. CLEANING OR REPLACING THE ABSORPTION TUBE NOTE: Although this procedure should never be needed as long as the user is careful only to supply the photometer with clean, dry and particulate free zero air, it is...
Page 188 - PHOTOMETER UV SOURCE LAMP ADJUSTMENT; PHOTO REFERENCE; test function value drops below 3000 mV.; UV DETECTOR GAIN ADJUST POT; Replace the cover on the analyzer.
TELEDYNE API Maintenance Schedule & Procedures M703E Calibrator Operators Manual 10.5. PHOTOMETER UV SOURCE LAMP ADJUSTMENT This procedure details the steps for adjustment of the UV source lamp in the optical bench assembly. This procedure should be done whenever the PHOTO REFERENCE test functio...
Page 189 - PHOTOMETER UV SOURCE LAMP REPLACEMENT; MAXIMUM
TELEDYNE API M703E Calibrator Operator’s Manual Maintenance Schedule & Procedures Figure 10-3: Photometer – Location of UV Detector Gain Adjustment & UV Lamp Set Screw 10.6. PHOTOMETER UV SOURCE LAMP REPLACEMENT This procedure details the steps for replacement of the UV source lamp in the op...
Page 190 - ADJUSTMENT OR REPLACEMENT OF OZONE
TELEDYNE API Maintenance Schedule & Procedures M703E Calibrator Operators Manual 10.7. ADJUSTMENT OR REPLACEMENT OF OZONE GENERATOR UV LAMP This procedure details the steps for replacement and initial adjustment of the ozone generator lamp. If you are adjusting an existing lamp, skip to Step 8. ...
Page 193 - GENERAL TROUBLESHOOTING & REPAIR OF; GENERAL TROUBLESHOOTING; deviations from the factory values and take corrective action.
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11. GENERAL TROUBLESHOOTING & REPAIR OF THE M703E CALIBRATOR This section contains a variety of methods for identifying and solving performance problems with the calibrator. NOTE The oper...
Page 194 - STANDBY LAMP DRIVER WARNING; To view or clear the various warning messages press:
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 18. Follow the procedures defined in Section 3.4.4 to confirm that the calibrator’s vital functions are working (power supplies, CPU, relay PCA, keyboard, PMT cooler, etc.). See Figure 3-3 f...
Page 195 - Table 11-1: Front Panel Warning Messages; WARNING FAULT
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator Table 11-1: Front Panel Warning Messages WARNING FAULT CONDITION POSSIBLE CAUSES CONFIG INITIALIZED Configuration and Calibration data reset to original Factory state. - Failed Disk on Chip -...
Page 196 - TEST FUNCTION
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual Table 11-1: Front Panel Warning Messages (cont.) WARNING FAULT CONDITION POSSIBLE CAUSES RELAY BOARD WARN The CPU cannot communicate with the Relay PCA. - I 2 C Bus failure - Failed relay PCA ...
Page 197 - Test Functions - Indicated Failures; PHOTO LAMP TEMP WARNING
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator Table 11-2: Test Functions - Indicated Failures TEST FUNCTION DIAGNOSTIC RELEVANCE AND CAUSES OF FAULT CONDITIONS. PHOTO MEASURE & PHOTO REFERENCE If the value displayed is too high the U...
Page 198 - Example of Signal I/O Function
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.1.3. USING THE DIAGNOSTIC SIGNAL I/O FUNCTION The Signal I/O parameters found under the DIAG Menu combined with a thorough understanding of the instruments Theory of Operation (found in Cha...
Page 199 - USING THE ANALOG OUTPUT TEST CHANNEL; Table 11-3: Test Channel Outputs as Diagnostic Tools
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.2. USING THE ANALOG OUTPUT TEST CHANNEL The signals available for output over the M703E’s analog output channel can also be used as diagnostic tools. See Section 6.9 for instruction on act...
Page 200 - USING THE INTERNAL ELECTRONIC STATUS LEDS; C bus and Relay; Mother Board; C Bus Watchdog Status LEDs; The most important is D1 (see, which indicates the health of the I; Menu Signal I/O to identify hardware
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.3. USING THE INTERNAL ELECTRONIC STATUS LEDS Several LEDs are located inside the instrument to assist in determining if the calibrators CPU, I 2 C bus and Relay PCA are functioning properly...
Page 201 - Status LEDs; Relay PCA Status LEDS Used for Troubleshooting
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.3.2.2. O 3 Status LEDs D15 (Yellow) - Photometer Lamp Heater D16 (Yellow) – O 3 Generator Lamp Heater D2 (Yellow) – Dry (zero) Air Pump Status D9 (Green) – External Zero Air Valve Status D...
Page 202 - SUBSYSTEM CHECKOUT; Test Channel D
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.4. SUBSYSTEM CHECKOUT The preceding sections of this manual discussed a variety of methods for identifying possible sources of failures or performance problems within the M703E calibrator. ...
Page 203 - Location of DC Power Test Points on Relay PCA
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.4.3. DC POWER SUPPLY If you have determined that the calibrator’s AC mains power is working, but the unit is still not operating properly, there may be a problem with one of the instrument...
Page 204 - Table 11-7: DC Power Supply Acceptable Levels; C BUS
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual A voltmeter should be used to verify that the DC voltages are correct per the values in the table below, and an oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate ...
Page 205 - SENSOR ASSEMBLY
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator If the calibrator starts operation with a normal display but pressing a key on the front panel does not change the display, then there are three possible problems. One or more of the keys...
Page 206 - PHOTOMETER PRESSURE SENSOR:
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual PHOTOMETER PRESSURE SENSOR: 1. Measure the pressure on the inlet side of S1 with an external pressure meter. 2. Measure the voltage across TP4 and TP1. The expected value for this signal sho...
Page 208 - FULL SCALE OUTPUT OF VOLTAGE RANGE; DIAG; PIN; SPARE
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual For each of the steps the output should be within 1% of the nominal value listed in the table below except for the 0% step, which should be within 0mV ±2 to 3 mV. Make sure you take into accou...
Page 209 - CONNECTOR
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.4.8.4. Control Inputs Table 11-11: M703E Control Input Pin Assignments and Corresponding Signal I/O Functions CONNECTOR INPUT CORRESPONDING I/O SIGNAL Top A CONTROL_IN_1 Top B CONTROL_IN_2...
Page 212 - BOX TEMP
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.4.11. TEMPERATURE PROBLEMS Individual control loops are used to maintain the set point of the UV lamp and ozone generator. If any of these temperatures are out of range or are poorly contro...
Page 213 - SIGNAL; TROUBLESHOOTING THE O; Readings at Zero
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.4.11.4. Ozone Generator Temperature There are three possible causes for the Ozone Generator temperature to have failed. The O 3 generator heater has failed. Check the resistance between ...
Page 214 - Readings
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.5.1.2. Noisy, Unstable, or Non-Linear Span O 3 Readings Check for leaks in the pneumatic systems as described in Section10.2. Check for proper operation of the meas/ref switching valve ...
Page 216 - between Pins 1 and 2 on the cable connector.
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.6. TROUBLESHOOTING THE O 3 GENERATOR 11.6.1. CHECKING THE UV LAMP POWER SUPPLY NOTE A schematic and physical diagram of the Lamp Power Supply can be found in Appendix D. WARNING Hazardous v...
Page 217 - TROUBLE SHOOTING THE OPTIONAL O
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.7. TROUBLE SHOOTING THE OPTIONAL O 3 GENERATOR The only significant components of the O 3 generator that might reasonable malfunction is the power supply assembly for the UV source lamp an...
Page 218 - REPAIR PROCEDURES; Disconnect the pneumatic connection from the flow assembly.; Critical Flow Restrictor Assembly Disassembly
TELEDYNE API General Troubleshooting & Repair of the M703E Calibrator M703E Calibrator Operators Manual 11.8. REPAIR PROCEDURES 11.8.1. REPAIRING SAMPLE FLOW CONTROL ASSEMBLY The critical flow orifice is housed in the flow control assembly (Teledyne Instruments part number: 001760400) located on...
Page 219 - the IC by gently prying it up from the socket.; TECHNICAL ASSISTANCE; USA
TELEDYNE API M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator 11.8.2. DISK-ON-CHIP REPLACEMENT PROCEDURE NOTE Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small to be felt by the human nervous system. Failure to use E...
Page 221 - A PRIMER ON ELECTRO-STATIC DISCHARGE; HOW STATIC CHARGES ARE CREATED; Triboelectric Charging; pellets during shipment can also build; Table 12-1: Static Generation Voltages for Typical Activities; MEANS OF GENERATION
TELEDYNE API M703E Calibrator Operator’s Manual A Primer on Eletro-Static Discharge 12. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Instruments considers the prevention of damage caused by the discharge of static electricity to be extremely important part of making sure that your analyzer continue...
Page 222 - HOW ELECTRO-STATIC CHARGES CAUSE DAMAGE; Semiconductor Reliability News; Table 12-2: Sensitivity of Electronic Devices to Damage by ESD; DAMAGE SUSCEPTIBILITY VOLTAGE; Potentially damaging electro-static discharges can occur:
TELEDYNE API A Primer on Eletro-Static Discharge M703E Calibrator Operators Manual 12.2. HOW ELECTRO-STATIC CHARGES CAUSE DAMAGE Damage to components occurs when these static charges come into contact with an electronic device. Current flows as the charge moves along the conductive circuitry of the ...
Page 223 - COMMON MYTHS ABOUT ESD DAMAGE; I didn’t feel a shock so there was no electro-static discharge; : The human nervous system is not able; I didn’t touch it so there was no electro-static discharge; : Electro Static charges are fields whose; It still works so there was no damage; : Sometimes the damaged caused by electro-static discharge can; Static Charges can’t build up on a conductive surface; : There are two errors in this statement.; BASIC PRINCIPLES OF STATIC CONTROL; Basic anti-ESD Work Station
TELEDYNE API M703E Calibrator Operator’s Manual A Primer on Eletro-Static Discharge 12.3. COMMON MYTHS ABOUT ESD DAMAGE I didn’t feel a shock so there was no electro-static discharge : The human nervous system is not able to feel a static discharge of less than 3500 volts. Most devices are damaged...
Page 224 - Simply touching a grounded piece of metal is insufficient; . While this may temporarily bleed off static
TELEDYNE API A Primer on Eletro-Static Discharge M703E Calibrator Operators Manual For technicians that work in the field, special lightweight and portable anti-ESD kits are available from most suppliers of ESD protection gear. These include everything needed to create a temporary anti-ESD work area...
Page 225 - MAINTENANCE
TELEDYNE API M703E Calibrator Operator’s Manual A Primer on Eletro-Static Discharge 12.4.2. BASIC ANTI-ESD PROCEDURES FOR ANALYZER REPAIR AND MAINTENANCE 12.4.2.1. Working at the Instrument Rack When working on the analyzer while it is in the instrument rack and plugged into a properly grounded powe...
Page 227 - DO NOT; component/assembly directly; In either case wait several seconds.
TELEDYNE API M703E Calibrator Operator’s Manual A Primer on Eletro-Static Discharge 12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service. Always pack electronic components and assemblies to be sent to Teledyne Instruments Customer Service in anti-ESD bins, tubes or bags. ...
Page 229 - APPENDIX A – Version Specific Software Documentation
TELEDYNE API M703E Calibrator Operator’s Manual APPENDIX A – Version Specific Software Documentation 05745 Rev C A-1 APPENDIX A – Version Specific Software Documentation APPENDIX A-1: Model 703E Software Menu Trees, Software Version C.0 APPENDIX A-2: Model 703E Setup Variables Available Via Serial I...
Page 231 - are
TELEDYNE API M 703E Calibrator O pera tor’s M a nual APPENDIX A- 1: So ftware Men u T rees, So ftware Ver s io n C.0 0574 5 Rev C A-3 APPENDIX A-1: Soft w are Menu Trees, Soft w are Version C.0 Figure A-1: Main Menu
Page 232 - TELEDYNE API; pera; PRIMARY SETUP MENU - Basics
TELEDYNE API APPENDIX A- 1: So ftware Men u T rees, So ftware Ver s io n C.0 M 703E Calibrator O pera tor’s M a nual A-4 0574 5 Rev C MAIN M ENU SETU P SE Q 1 O3 PASS CF G C LK MOR E See SECO NDAR Y SET U P Menu ON OFF TIME DATE MODEL TYPE AN D NUM BER PART NUMB ER SERIAL NUMB ER SOFTW ARE RE VISIO ...
Page 234 - SECONDARY SETUP Menu - Ba
TELEDYNE API APPENDIX A- 1: So ftware Men u T rees, So ftware Ver s io n C.0 M 703E Calibrator O pera tor’s M a nual A-6 0574 5 Rev C 1 Only appears if optional Ethernet PCA i s installed. When Ethernet PCA is present COM2 submenu disappears. 2 , & a re on ly ed itab le whe n is . 3 Alth oug h i...
Page 235 - SECONDARY SETUP Menu DIAG Submenu – Basics
TELEDYNE API M 703E Calibrator O pera tor’s M a nual APPENDIX A- 1: So ftware Men u T rees, So ftware Ver s io n C.0 0574 5 Rev C A-7 1 O nly occ urs if one of t he v olta ge r anges is se lect ed. 2 M anua l a djus tmen t m enu only app ears if ei ther the fea ture is 3 Menu is inacti ve whi le ins...
Page 236 - SECONDARY SETUP Menu DIAG Submenu – G
TELEDYNE API APPENDIX A- 1: So ftware Men u T rees, So ftware Ver s io n C.0 M 703E Calibrator O pera tor’s M a nual A-8 0574 5 Rev C Togg le these keys to se t the g as pressure. Togg le these keys to se t the fl ow rate of the O3 photometer sampl e gas Continue pressing to cycle throug h similar a...
Page 237 - PRIMARY SETUP Menu LVL Subm
TELEDYNE API M 703E Calibrator O pera tor’s M a nual APPENDIX A- 1: So ftware Men u T rees, So ftware Ver s io n C.0 0574 5 Rev C A-9 Figure A-7: PRIMARY SETUP Menu; LVL Subm enu – programming LEADS LEVELS (optional)
Page 239 - M703E SETUP VARIABLES FOR LATEST REVISION
TELEDYNE API M703E Calibrator Operator’s Manual APPENDIX A-2: Setup Variables For Serial I/O, Software Version C.0 05745 Rev C A-11 APPENDIX A-2: Setup Variables For Serial I/O, Software Version C.0 Table A-1: M703E Setup Variables, Software Version C.0 M703E SETUP VARIABLES FOR LATEST REVISION SETU...
Page 247 - C address 44 hex
TELEDYNE API M703E Calibrator Operator’s Manual APPENDIX A-4: Signal I/O Definitions, Software Version C.0 05745 Rev C A-19 APPENDIX A-4: Signal I/O Definitions, Software Version C.0 Table A-4: M703E Signal I/O Definitions, Software Version C.0 M703E I/O Signal List for Latest Revision Signal Name B...
Page 248 - Front panel I
TELEDYNE API APPENDIX A-4: Signal I/O Definitions, Software Version C.0 M703E Calibrator Operator’s Manual A-20 05745 Rev C M703E I/O Signal List for Latest Revision Signal Name Bit or Channel Number Description PHOTO_REF_VALVE 6 0 = photometer valve in reference position 1 = measure position ZA_SHU...
Page 249 - Rear board DAC MUX analog inputs
TELEDYNE API M703E Calibrator Operator’s Manual APPENDIX A-4: Signal I/O Definitions, Software Version C.0 05745 Rev C A-21 M703E I/O Signal List for Latest Revision Signal Name Bit or Channel Number Description PHOTO_LAMP_TEMP 2 Photometer lamp temperature O3_GEN_TEMP 3 O 3 generator lamp temperatu...
Page 250 - TERMINAL KEY ASSIGNMENTS
TELEDYNE API APPENDIX A-5: Terminal Command Designators, Software Version C.0 M703E Calibrator Operator’s Manual A-22 05745 Rev C APPENDIX A-5: Terminal Command Designators, Software Version C.0 Table A-5: Terminal Command Designators, Software Version C.0 COMMAND ADDITIONAL COMMAND SYNTAX DESCRIPTI...
Page 251 - APPENDIX B: Spare Parts List
TELEDYNE API Model 703E Calibrator Operator’s Manual APPENDIX B: Spare Parts List 05746 Rev B B-1 APPENDIX B: Spare Parts List NOTE Use of replacement parts other than those supplied by API may result in non-compliance with European standard EN 61010-1. 05834 - LIST, SPARE PARTS, M703E 05863 - L...
Page 252 - Model 703E Calibrator Operator’s Manual
TELEDYNE API APPENDIX B: Spare Parts List Model 703E Calibrator Operator’s Manual B-2 05746 Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK
Page 253 - Spare Parts List; Part Number
Spare Parts List M703E Part Number Description 000940100 CD, ORIFICE, .003 GREEN 006120100 ASSY, OZONE GEN LAMP (BIR) (OP5) 022710000 ABSORPTION TUBE, QUARTZ, M400A/E (KB) 040010000 ASSY, FAN REAR PANEL, E SERIES 040030700 PCA, PRESS SENSORS (2X), 700E PHOTO OPT 040300110 ASSY, CONFIG PLUG, 100-115V...
Page 255 - Recommended Spare Parts Stocking Levels; ABSORPTION TUBE, QUARTZ; Units; Recommended Spare Parts Stocking Level: Standard
Recommended Spare Parts Stocking Levels M703E 1 2-5 6-10 11-20 21-30 006120100 ASSY, OZONE GEN LAMP 1 2 4 022710000 ABSORPTION TUBE, QUARTZ 1 2 4 8 058330000 ASSY, INT PUMP, 115V/60Hz 1 058330100 ASSY, INT PUMP, 230V/50Hz 1 040010000 ASSY, FAN, REAR PANEL 1 1 2 4 8 040030700 PCA, PRESS SENSORS PHOTO...
Page 257 - Questionnaire; PARAMETER; What is measured O
Model M703E Calibrator Operator’s Manual Warranty/Repair Questionnaire Model 703E TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected] PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816 05747 Rev B C-1 CUSTOMER:_______________________________ PHONE: ________________...
Page 259 - APPENDIX D: Diagrams and Schematics; Table D-1: List of Included Diagrams and Schematics; Document Title; Interconnect Drawing M703E
TELEDYNE API Model 703E Calibrator Operator’s Manual APPENDIX D: Diagrams and Schematics 05748 Rev B D-1 APPENDIX D: Diagrams and Schematics Table D-1: List of Included Diagrams and Schematics Document # Document Title 05826 Interconnect Drawing M703E 05827 Interconnect List M703E 04420 SCH, PCA 041...
Page 266 - PCA, UV DETECTOR PREAMP; PCA VERSION TABLE
1 2 3 4 A B C D 4 3 2 1 D C B A APPROVALS DRAWN CHECKED APPROVED DATE SIZE DRAWING NO. REVISION SHEET The information herein is theproperty of API and issubmitted in strictest con- Unauthorized use by anyone fidence for reference only. for any other purposes isprohibited. This document orany informa...
Page 270 - CH
1 2 3 4 A B C D 4 3 2 1 D C B A A PPROV A LS D R A WN CHECKED A PPROV ED DATE S IZE D R A W I N G NO . R E V I S I ON S HEET The information herein is theproperty of API and issubmitted in strictest con- Unauthorized use by anyone fidence for reference only. for any other purposes isprohibited. This...
Page 281 - ONTROL INPUTS
1 2 3 4 5 6 A B C D 6 5 4 3 2 1 D C B A Title Number Revision Size Orcad B Date: 17-Jun-2008 Sheet of File: N:\Pcbmgr\05701dn.E-motherboard.gen4\Source\05701a.DDB Drawn By: Schematic for E Series Motherboard PCA 05702 6 8 05703 A D0D1D2D3D4D5D6D7 D [0..7] IOR D0D1D2D3D4D5D6D7 A1 2 A2 3 A3 4 A4 5 A5 ...