Page 2 - TELEDYNE BROWN EGINEERING; Warranty; IMPORTANT NOTICE
TELEDYNE BROWN EGINEERING Analytical Instruments Copyright © 1993 Teledyne Brown Engineering Analytical InstrumentsAll Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whol...
Page 3 - Table of Contents; Appendix
TELEDYNE BROWN EGINEERING Analytical Instruments Table of Contents 1.0 Introduction Main Features ............................................................................................... 1Micro-Fuel Cell ............................................................................................
Page 5 - TELEDYNE BROWN ENGINEERING; Introduction; Main Features
1 319R Introduction 1.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a widerange of gases and gas mixtures. Modular and compact design provideoutstanding perform...
Page 6 - warranted for 6 months, with a life expectancy of 6 months.; Warranty Conditions
1.0 Introduction 319R 2 TELEDYNE BROWN ENGINEERING Analytical Instruments 4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months. 5. Analog Voltage Output. The Model 319R provides customer connections to external devices, such as com...
Page 7 - Installation & Operation; Precautions; Upon receipt of the instrument, inspect the entire unit for; must; be free of entrained solids and water. A high
3 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage. Notify the shipper if any damage ...
Page 8 - Analysis Unit Connection
4 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level ofelectromagnetic noise in the operating environment. Analysis Unit Connect...
Page 9 - Alarm Terminal Connection; CAUTION: Do not scratch, puncture, or damage the sensing membrane of
5 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuitconnection. The explanation for the relay te...
Page 10 - Setting Alarms
6 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale. Thus, an alarm set to indicate 192 on the displaywhile on the 200 PPM O 2 range will activate at...
Page 11 - Calibration; To calibrate the 319R:
7 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Calibration To calibrate the 319R: 1. Place the power switch in the ON position. 2. Allow the LCD reading to stabilize. 3. Expose the sensor to ambient air (20.9%) or instrument grade air. If the analyzer goes ...
Page 13 - Maintenance; Replacing the Fuse
9 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using thefollowing procedure: 1. Disconnect the AC power and plac...
Page 14 - Replace the fuse holder into its receptacle, pushing in firmly; Replacing the Sensor; In addition, if the LCD displays “00.0” when the unit is plugged in,; NOTE: Do not stockpile cells, as the cell warranty begins on date of; Only one cell per instrument should be in reserve. The spare cell
3.0 Maintenance 319R 10 TELEDYNE BROWN ENGINEERING Analytical Instruments 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing theknob ...
Page 15 - The Class Z-2 cell used in the 319R is warranted for six (6) months of
11 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Warranty Conditions The Class Z-2 cell used in the 319R is warranted for six (6) months of service from the date of shipment. Customers having warranty claims mustreturn the cell in question to the factory for evaluation. If i...
Page 17 - Specifications; 5% calibration range
13 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Appendix Specifications Ranges : 0–200 PPM, 0–1999 PPM O 2 , and 0– 25% calibration range Sensitivity: 1% of full scale Signal Output : Standard: 0–10 VDC, negative ground 4-20 mA, negative groundplus contact closures for rangeidenti...
Page 18 - Spare Parts List; DESCRIPTION
14 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Spare Parts List QTY. P/N DESCRIPTION 1 C57304 PC board, main 1 A33748 Thermistor assembly 1 C6689-Z2 Micro-Fuel Cell, Z-2 4 F1189 Fuse, 1A (AC) 4 F51 Fuse, ½A (DC) 1 A59264 Probe to analyzer cable, 6 feet A minimum charge of US $20....
Page 19 - Drawing List; JP2 - connects pins 1 and 2
15 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Drawing List C-57490 Outline diagramA-58154 Probe interconnection diagramD-57302 Electronic PCB schematic diagramC-57305 Power supply PCB schematic diagramC-57307 Power supply PCB assembly diagramB-58123 Analysis unit (probe) outline...
Page 20 - Material Safety Data Sheet; Section I – Product Identification; Quantity
16 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Material Safety Data Sheet Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19.Mini-Micro-Fuel Cells, all classes. Manufacturer: Teledyne Brow...
Page 21 - Section V – Reactivity Data; Hazardous Polymerization:; Section IV – Fire and Explosion Hazard Data
17 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fireconditions. No specific agents recommended. Wear NIOSH/OSHA approved self-contained breathingapparatus a...
Page 22 - Emergency First Aid Procedures:; Section VII – Precautions for Safe Handling and Use
18 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Contact of electrolyte with skin or eyes will cause aburning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss ofsleep, loss of appetite, metallic taste and fatigue. Carcinogenicity: Le...
Page 23 - Section VIII – Control Measures; Other Protective Clothing:; Section IX – Disposal; Section X – References
19 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: Apron, face shield Ventilation: na Section IX – Disposal Both lead and potassium hydroxide are consider...