Page 2 - Teledyne API Model 201A NH; SAFETY MESSAGES; CAUTION: Hot Surface Warning
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that...
Page 3 - TABLE OF CONTENTS
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 TABLE OF CONTENTS SAFETY MESSAGES ....................................................................................................... 1 TABLE OF CONTENTS ........................................................................
Page 4 - LIST OF FIGURES; ECO
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 LIST OF FIGURES F IGURE 1 - M201A A MMONIA A NALYZER – B LOCK D IAGRAM ................................................................. 21 F IGURE 2 - M201A E XTERNAL P NEUMATIC C ONNECTION D IAGRAM WITH ECO (E XTERNAL C ALIBR...
Page 5 - LIST OF TABLES; CALS
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 LIST OF TABLES T ABLE 2.1 - Z ERO C ALIBRATION P ROCEDURE - Z ERO G AS T HROUGH THE S AMPLE P ORT ......................... 11 T ABLE 2.2 - E NTER THE E XPECTED S PAN G AS C ONCENTRATIONS P ROCEDURE ...............................
Page 7 - Principle of Operation
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 1.0 M201A AMMONIA ANALYZER This manual is to be used in conjunction with the M200A NOx Analyzer manual p/n 02246 supplied with this instrument. The M201A Ammonia Analyzer consists of a modified M200A Nitrogen Oxides Analyzer an...
Page 8 - Special Considerations for Ammonia Measurement
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 converts some of the ammonia to NO, thus the TN channel data contains a small ammonia signal. Secondly, the ammonia converter does not convert all of the ammonia or ambient NO2 to NO. Finally, the high temperature in the ammoni...
Page 10 - M201A Analyzer Specifications
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 1.2 M201A Analyzer Specifications Ranges 0-50 to 0-2000 ppb in 1 ppb increments Range Modes Single, Independent, AutoRange Noise at Zero 1 0.5 ppb RMS Noise at Span 1 < 1.0% of reading above 50 ppb Lower Detectable Limit 1 p...
Page 11 - CALIBRATION PROCEDURE; Zero/Span Calibration – Basic Instrument
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 2.0 CALIBRATION PROCEDURE Calibration of the M201A is done by first calibrating the TN and TNx channels with NO span gas diluted with zero air. In the second phase of calibration, ammonia span gas is introduced to measure the o...
Page 12 - Action Comment
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 Table 2.1 - Zero Calibration Procedure - Zero Gas Through the Sample Port Step No. Action Comment 1. Press CAL The M201A enters the calibrate mode from sample mode. Input zero gas through the sample port. 2. Wait 20 min Wait fo...
Page 13 - Press CAL
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 Table 2.3 - Span Calibration Procedure - Span Gas thru Sample Port Step No. Action Comment 1. Press CAL The M201A enters the calibrate mode. Input NO span gas through the sample port. 2. Press CAL-CONV-SET Set the converter eff...
Page 14 - CALZ
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 2.1.1 Zero/Span Calibration with Z/S Valves or ECO Option The zero gas enters through the zero gas inlet port on the rear panel. Make sure that the umbilical cord that makes the control connections between the analyzer and the ...
Page 18 - The instrument is now fully calibrated.
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 The instrument is now fully calibrated. 2.1.4 Channel Calibration Quality After the ammonia channel is calibrated via the converter efficiency parameter, it is important to check the calibration by sampling the NH3 calibration ...
Page 19 - MAINTENANCE SCHEDULE; Maintaining the M501NH Converter
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 3.0 MAINTENANCE SCHEDULE 3.1 M201A Maintenance Schedule The maintenance requirements of the M201A are essentially like a standard M200A NOx analyzer. Please refer to Table 8-1 in the M200A Operator Manual for the M200A maintena...
Page 20 - Converter Rebuild Kit
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 Table 3.2 – M501NH Converter Rebuild Parts List Part No. Description KIT000191 Converter Rebuild Kit KIT000139 Ceramic Bushings HE0000007 Ceramic Heater 220W 60 VAC Procedure: 1. Turn off power to the converter and allow to coo...
Page 21 - AutoZero Flow Check; This procedure should be performed with the sample pump running.; from the left. It will have a
Teledyne API Model 201A NH 3 Analyzer Instruction Manual, 04689, Rev. B1 3.3 AutoZero Flow Check Since the M201A is an ammonia analyzer, the flow through the AutoZero orifice is especially important. Check the AutoZero flow as follows: 1. This procedure should be performed with the sample pump runni...