Page 5 - Safety Messages; CAUTION
05386 Rev. D i Safety Messages Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol, and are placed throughout this manual and...
Page 6 - ii
Page 7 - iii; Table of Contents
05386 Rev. D iii Table of Contents Safety Messages ...................................................................................................................................... i Table of Contents..................................................................................................
Page 8 - iv; List of Figures
iv 05386 Rev. D 8.0 M200EU-NO Y SPARE PARTS LIST ...................................................................... 37 List of Figures FIGURE 1 REAR PANEL PNEUMATIC CONNECTIONS ................................................................................ 8 FIGURE 2 REPLACING THE PARTICULATE F...
Page 9 - NOy; . The converter is mounted on the analyzer at
05386 Rev. D 5 1.0 INTRODUCTION The M200EU-NOy converter is designed to support a NOx analyzer by converting multiple, unstable compounds grouped under the name NOy . The converter is mounted on the analyzer at the sample inlet to minimize flow time between sample in and converter, thereby optimizin...
Page 10 - This page intentionally left blank.
Model 200EU-NOy Option Manual Addendum 6 05386 Rev. D This page intentionally left blank.
Page 11 - Bypass Pump Chassis; Electrical and Pneumatic Connections; Mount the converter on a suitable mast outside of the enclosure.; Figure 1; to match up the correct tube with the rear panel bulkhead
05386 Rev. D 7 2.0 GETTING STARTED The NOy converter has been designed to operate with the M200EU low level NOx analyzer. 2.1 Unpacking The M200EU with NOy Option comes in three boxes: M200EU analyzer Bypass Pump Chassis External Converter housed in a stainless enclosure, and umbilical cable a...
Page 12 - Figure 1 Rear Panel Pneumatic Connections
Model 200EU-NOy Option Manual Addendum 8 05386 Rev. D Figure 1 Rear Panel Pneumatic Connections
Page 13 - Initial Operation; Note; After 30 min, the display on the Bypass Pump Chassis should read 315; Table 1 Final Test and Calibration Values
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 9 2.3 Initial Operation 1. After confirming proper supply voltage, turn on the instrument power. If you are unfamiliar with the M200EU with NOy Option, we recommend that you read the overview in Section 3 before proceeding. The power indicator li...
Page 15 - Table 2. Specifications for the converter and the Bypass Pump
05386 Rev. D 11 3.0 SPECIFICATIONS, WARRANTY 3.1 Specifications Table 2. Specifications for the converter and the Bypass Pump Converter Remotely mounted molybdenum converter with temperature controller in a NEMA – 4 enclosure. Converter Temperature 315 o C 7 o C with read-out on front panel of byp...
Page 16 - TELEDYNE ADVANCED POLLUTION INSTRUMENTATION, INC.; COVERAGE
Model 200EU-NOy Option Manual Addendum 12 05386 Rev. D 3.2 Warranty TELEDYNE ADVANCED POLLUTION INSTRUMENTATION, INC. 02024c Prior to shipment, API equipment is thoroughly inspected and tested. Should equipment failure occur, API assures its customers that prompt service and support will be availabl...
Page 17 - CONVERTER; Principle of Operation
05386 Rev. D 13 4.0 THE M200EU-NO Y CONVERTER 4.1 Principle of Operation The T-API Model M200EU with NOy Option is designed to measure the concentration of NO, NO2, and other compounds that are too unstable to be measured when taken in through the normal ambient air sample inlet system. Please refer...
Page 19 - SAMPLE; C or
05386 Rev. D 15 5.0 CALIBRATION AND ZERO/SPAN CHECKS Unlike most NOx analyzers, the M200EU with the NOy option does not have a sample inlet port on the rear panel of the M200EU. The sample port is located on the External Converter; therefore Zero/Span calibration is different than a normal NOx instr...
Page 20 - Calibration or Cal Check Procedure; Table 3
Model 200EU-NOy Option Manual Addendum 16 05386 Rev. D 5.1 Calibration or Cal Check Procedure The calibration of the instrument can be checked or adjusted using gas introduced through the calibration port “CAL IN” on the Bypass Pump Module. The calibration gas is routed to a TEE near the sample inle...
Page 21 - Table 5 Span Calibration Procedure
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 17 Table 4 Enter Expected Span Gas Concentrations Procedure For NO & NOy Step Action Comment 1. Press CAL-CONC-NOy This key sequence causes the M200EU to prompt for the expected NOy concentration. Enter the NOy span concentration value by press...
Page 23 - Maintenance Schedule; Table 6
05386 Rev. D 19 6.0 MAINTENANCE 6.1 Maintenance Schedule The schedule for preventative maintenance is presented in Table 6 below. Table 6 Preventative Maintenance Calendar Item Maintenance Interval Zero/Span Calibration Annually or after maintenance or repairs Zero/Span Checks Daily Particulate Filt...
Page 24 - Replacing the Sample Particulate Filters; Fold down the Bypass Pump Module front panel.; Figure; for an exploded view of the filter assembly.
Model 200EU-NOy Option Manual Addendum 20 05386 Rev. D 6.2 Replacing the Sample Particulate Filters The particulate filter should be inspected often for signs of plugging or contamination. It is also common for dirt particles to cause instrument drift, and affect accuracy. To check and change the fi...
Page 25 - Figure 2 Replacing the Particulate Filter
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 21 Figure 2 Replacing the Particulate Filter
Page 26 - Checking Analyzer Flow Rate; Replacing the Converter; Figure 3
Model 200EU-NOy Option Manual Addendum 22 05386 Rev. D 6.3 Checking Analyzer Flow Rate The external sample pump is capable of maintaining the reaction cell pressure at less than 5.0 In-Hg-A. If a higher pressure is noted, the pump may need servicing. Check the pump and pneumatic system for leaks or ...
Page 27 - Figure 3 NOy Converter Assembly; NOy Out
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 23 Figure 3 NOy Converter Assembly NOy Out NO Out Sample In
Page 28 - Inspecting Pneumatic Lines
Model 200EU-NOy Option Manual Addendum 24 05386 Rev. D 6.5 Inspecting Pneumatic Lines Particulate matter in the pneumatic lines will affect both flow rate and response time. It is important that the pneumatic system be periodically inspected and thoroughly cleaned if necessary. Clean by disassemblin...
Page 30 - Figure 5 Pneumatic diagram (OPT 86 option – Sample Driers)
Model 200EU-NOy Option Manual Addendum 26 05386 Rev. D Figure 5 Pneumatic diagram (OPT 86 option – Sample Driers)
Page 31 - Checking for Leaks; Turn off instrument power and pump power.
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 27 6.6 Checking for Leaks If a leak checker is not available, it is possible to leak check the instrument using the sample pump plus a shut-off valve. 1. Turn off instrument power and pump power. 2. Cap the sample inlet port, bypass out port, ozone...
Page 33 - temperature readout.
05386 Rev. D 29 7.0 TROUBLESHOOTING, ADJUSTMENTS CAUTION! The operations outlined in this section must be performed by qualified maintenance personnel only! This section of the manual contains information on diagnosing and repairing instrument performance problems. It provides troubleshooting proced...
Page 34 - Operation Verification – Diagnostic Techniques
Model 200EU-NOy Option Manual Addendum 30 05386 Rev. D 7.1 Operation Verification – Diagnostic Techniques 7.1.1 Pneumatic System
Page 35 - The pneumatic system is diagramed in
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 31 The pneumatic system is diagramed in
Page 36 - Figure 4; Setting the Converter Temperature; PV
Model 200EU-NOy Option Manual Addendum 32 05386 Rev. D Figure 4 . 7.1.2 Leak Check Refer to Section 5.6 for the leak check procedure. 7.1.3 Bypass Pump Diagnostic Procedures The bypass pump is capable of maintaining a total bypass flowrate of ~2.2 lpm. If flow checks indicate a fall off in bypass fl...
Page 38 - PVOF; hold until
Model 200EU-NOy Option Manual Addendum 34 05386 Rev. D SECONDARY MENU (The fixed characteristics of the system). PRESS DISPLAY ACTION SEL hold until p-n1 SET TO 0 – REVERSE ACTION SEL hold until p-n2 SET TO 2 = TYPE J THERMOCOUPLE SEL p-dF DIGITAL FILTER SET TO 5 SEL P-SL LOWER LIMIT SET TO -40 (-40...
Page 39 - Figure 6 Electrical Diagram
Model 200EU-NOy Option Manual Addendum 05386 Rev. D 35 Figure 6 Electrical Diagram
Page 41 - SPARE PARTS LIST; DESCRIPTION
05386 Rev. D 37 8.0 M200EU-NO Y SPARE PARTS LIST The following spare parts list supplements the spare parts list for the M200EU analzyer. Note: Use of replacement parts other than those supplied by API may result in non-compliance with European Standard EN 61010-1. PART NO. DESCRIPTION 000940400 ORI...