Page 3 - English; Table of Contents
2 R7 English Dear Customer,Thank you for purchasing this OMRON product. OMRON products are known for their reliabilityand advanced technology. By buying this OMRON R7 wrist blood pressure monitor withposition sensor you have chosen for a high quality and innovative instrument. We hope thatyou will b...
Page 4 - Before measuring your Blood Pressure; Notes on Safety Please read thoroughly before using this monitor.; Caution; Examples of signs
3 R7 English Before measuring your Blood Pressure Notes on Safety Please read thoroughly before using this monitor. • The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as to prevent the risk of damage to you and others. • The icons a...
Page 5 - Notes on safety; Important instructions for obtaining meaningful readings; Cautions; Before using your OMRON R7 for the first time
4 R7 English Notes on safety Important instructions for obtaining meaningful readings • Avoid eating and drinking (alcohol), smoking, sport and taking medicines before measuring your blood pressure as this could affect your blood pressure level. • Before you start sit comfortably on a chair and rela...
Page 6 - Names and functions of the parts; Main unit
5 R7 English Main unit Storage case Batteries for monitoring Instruction Manual Blood pressure pass (Two LR03 alkaline batteries) Names and functions of the parts Battery Cover Wrist Cuff Main unit Guarantee card HEM-637-E2 EN 7/8/10 9:17 AM Page 5
Page 7 - Pulse Display; Mark; How to insert the batteries; Use this monitor after inserting batteries in the main unit.
6 R7 English Systolic Blood Pressure Display Diastolic Blood Pressure Display Date/Time Display Pulse Display Button Push to move the displayed item forward. Push to move the displayed item backward. Mark Displayed when a storedmeasurement is shown. SETTING button Push this button to setvarious item...
Page 8 - How to set the Date and Time; How to measure Blood Pressure correctly; How to wrap the cuff
7 R7 English With new batteries (long life alkaline batteries) you are able to measure app. 300 times undernormal conditions. Please switch off the monitor after using it. If you forget to switch themonitor off it will switch off itself after app. 2 minutes. Please take the batteries out of themonit...
Page 10 - Incorrect Posture; How to obtain a correct posture; Measuring your blood pressure and pulse
9 R7 English Incorrect Posture How to obtain a correct posture Achieving a correct posture before measuring your blood pressure is important since anincorrect posture might give you incorrect readings. To achieve a correct posture please followthe instructions (see pictures):1) Sit down and stretch ...
Page 11 - How to measure Blood Pressure
10 R7 English How to measure Blood Pressure Inflation Start Screen Inflation in Progress Screen MarkIndicatingInflation inProgress PulseRateMark Sustain your elbow with your other hand or a table toavoid movement. • Adjust the height of your wrist by using your elbow as a fulcrum. When the ( ) will ...
Page 12 - How to enlarge the reading; The results will be shown on your display. By pushing the (; How to use the Memory function
11 R7 English How to enlarge the reading The results will be shown on your display. By pushing the ( ) buttons you can enlarge the readings. How to use the Memory function Date and Time of Measurement Your OMRON R7 blood pressure monitor is able to storemaximum 90 measurements including date and tim...
Page 13 - How to use the Graph function
12 R7 English Night Data Graph NightData Mark How to use the Graph function This monitor has the possibility to display the stored measurements as a graph. In order to geta clear history of your blood pressure the measurements are put in time zones:• Morning: Measurements made between 0:00 and 9:59•...
Page 14 - How to change the settings; Explanation of the graph display
13 R7 English Date ofMeasurement Month/Days of a Week PulseRate Blood Pressure Values(systolic/diastolic) (mmHg) Detail Screen of Data in Flashing Display Display screen where thesystolic blood pressurevalues higher than 170mmHg are measuredtwice Graph Display Screen ofHypertension Type The systolic...
Page 15 - Auto memory
14 R7 English Automatic Memory Setting (Refer to Page 14.) Position Sensor Setting (Refer to Page 15.) Measuring (Wrist) Setting (Refer to Page 15.) Alarm Setting (Refer to Page 15.) Time Setting (Refer to Page 16.) All Memory Erasure (Refer to Page 16.) Reset Settings (Refer to Page 16.) Automatic ...
Page 16 - Position sensor; Alarm
15 R7 English Position Sensor Screen Position Sensor OFF Screen Position Sensor ON and SIGNAL OFF Screen Position sensor The position sensor of your OMRON R7 can be switched ON or OFF. In the factory preset theposition sensor is enabled. If the position sensor is switched ON you can choose to have t...
Page 17 - Time
16 R7 English Time Screen Time You can change the time and date setting of your blood pressuremonitor. When reaching the time screen you will see the Gregoriancalendar. The year will be flashing. 1. By using the ( ) buttons you can set the current year. 2. By pushing the SET button you will lock the...
Page 18 - What is Blood Pressure?; Health and blood pressure; Lifestyle
17 R7 English What is Blood Pressure? Blood is pumped to artery by the beat(contraction and expansion) of a pumpcalled the heart. The pressure to pumpout blood to artery is called the "systolicblood pressure." The pressure when theheart expands as the blood havingcirculated the whole body re...
Page 19 - Hypertension guide; Blood pressure at the upper arm and wrist; Wrist Blood Pressure Monitor
18 R7 English (mmHg) According to the blood pressure classification by the WHO/ISH* (revised in 1999) * International Society of Hypertension (mmHg) 80 85 90 100 110 120 130 140 160 180 Systolic blood pressure Diastolicblood pressure Optimal blood pressure (target value) Normal blood pressure Normal...
Page 20 - Diseases influencing blood pressure values; Maintenance and storage; • Do not use petroleum spirits, thinners or similar solvents!
19 R7 English Diseases influencing blood pressure values Diabetes, hyperlipemia, or hypertension accelerates arterialsclerosis, which causes dangerous disease such ascerebral apoplexy or myocardial infarction or inducesarterial stenosis or peripheral circulatory disorder. In these cases, the wrist a...
Page 21 - Error indication; Cause
20 R7 English Error indication If the measurement is not made correctly, the following error can be displayed. Cause How to correct You moved your wrist during measurement. Measure your bloodpressure again withoutmoving your wrist untilthe measurement is com-pleted. (Refer to Page10.) Error After se...
Page 22 - Troubleshooting; Possible cause; Trouble
21 R7 English Troubleshooting Possible cause How to correct Batteries are worn out. Replace the batteries with newones. (See to Page 6.) Nothing is displayed whenyou push the on/offbutton. Measurement cannot bemade. Or the blood pressure values are displayedabnormally high (low). The wrist cuff is n...
Page 23 - Spare parts and accessories
22 R7 English Product Description : Automatic Blood Pressure Monitor Model : R7 (HEM-637-E2 / HEM-637-E7) Display : Dot matrix LCDdisplay Measurement : Oscillometric method Measurement Range : Pressure; 0 to 299 mmHg, Pulse rate; 40 to 180 /min. Accuracy : Pressure; within +/-3 mmHg, Pulse rate; wit...
Page 24 - Manufacturer; EC; Made in China
OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto,617-0002 JAPAN OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR HoofddorpTHE NETHERLANDSwww.omron-healthcare.com OMRON (DALIAN) CO., LTD. Dalian, CHINA OMRON HEALTHCARE UK LTD. Opal Drive, Fox Milne, Milton Keynes MK15 0DG, U.K. OMRO...