Milwaukee SM105 - Manual

Milwaukee SM105

Milwaukee SM105 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
Page: / 2
Loading the manual

GENERAL OPERATION:
• The meter is supplied

complete with a 9V bat-
tery. Remove the battery
compartment cover on
the back of the meter and
install the battery while
paying attention to its po-
larity.

• Connect the MA914BR/1

probe to the meter.

• Turn the instrument on

by pressing the ON/OFF
key. At start-up, the LCD
shows the percentage of
the remaining battery life
for a few seconds, and
then the current mea-
surement.

• Always remove the electrode protective

cap before taking any measurement. If
the electrode has been left dry, soak the
tip (bottom 2.5 cm) in M10000 rinse so-
lution for a few minutes to reactivate it.

• Immerse the tip (2.5 cm)

of the probe into the
sample and stir gently.

• Read the pH value when

the clock symbol stops
blinking.

• The temperature reading can be displayed

in °C or °F unit; to select the desired
scale, press and hold the “ON/OFF” key
until “TEMP” and the current temperature
unit are displayed on the secondary LCD.
Use the “SET” button to select the unit
and then press the “ON/OFF” key a couple
of times to return to normal mode.

• To activate the HOLD function, keep

pressed the “HOLD” key. The measured
value will be frozen on the display and
the “HOLD” message appear on the sec-
ondary LCD.

• After measurements, switch the meter

off by pressing the “ON/OFF” key. The
“OFF” message appears on the LCD;
release the button.

Notes:
• Before taking any measurement,

make sure that the meter has been
calibrated (the “CAL” tag is displayed
on the left lower corner of the LCD).

• Always replace the electrode protec-

tive cap after use with a few drops of
storage solution.

CALIBRATION PROCEDURE:

It is recommended to recalibrate the
meter at least once every three weeks.
A) Preparation:

Buffer solutions:

1. pH 7.01 / 6.86 (MA9007 / MA9006)
2. pH 4.01 (MA9004) for acidic calibra-

tion (pH < 7) or pH 10.01 / 9.18
(MA9010 / MA9009) for alkaline
range (pH > 7).

Use two beakers for each buffer solu-
tion: one beaker for rinsing the probe,
the other for calibration. In this way the
cross contamination between solutions
is minimized.

B) Procedure:
• Select the calibration buffer set: press

and hold the “ON/OFF” key until the
LCD shows “TEMP”. Press again this
key and the “BUFF” message will ap-
pear; then select the desired buffer
set with the “SET” key: “7.01 pH BUFF”
(for standard solutions: pH 4.01, 7.01,
10.01) or “6.86 pH BUFF” (for NIST
solutions: pH 4.01, 6.86, 9.18). Press
the “ON/OFF” key again to exit.

• Remove the protective cap, then im-

merse the probe in the first buffer so-
lution (7.01 or 6.86 pH).

• Press and hold the “ON/OFF” key until

“CAL” is displayed on the lower LCD.

USER MANUAL

PORTABLE pH METER

WITH EXTENDED RANGE

MODEL: SM105

pH & Temperature Meter