Milwaukee SM102 - User Manual

Milwaukee SM102

Milwaukee SM102 – User Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
Page: / 2
Loading the manual

M ilw a u k e e

M ilw a u k e e

WARRANTY:

This instrument is warranted from all defects in
materials and manufacturing for a period of two
years from the date of purchase.

The probe is warranted for a period of six months.

If during this period the repair or replacement of
parts is required, where the damage is not due to
negligence or erroneous operation by the user, please
return the parts to either distributor or our office and
the repair will be effected free of change.
Note: We reserve the right to modify the design,
construction and appearance of our products without
advance notice.

USER MANUAL

PORTABLE uP-BASED

pH METER

MODEL: SM102

Smart pH Meter

OPERATION:

• The meter is supplied complete with a

9V battery. Slide off the battery compart-
ment cover on the back of the meter
and install the battery while paying at-
tention to its polarity.

• Always remove the electrode protective

cap before taking any measurement. If
the electrode has been left dry, soak
the tip (bottom 2.5 cm) in rinse solution
(M10000B) for a few minutes to reacti-
vate it.

• C o n n e c t t h e p H

electrode and the
temperature probe
to the sockets on
t h e m e t e r ’ s t o p .
The temperature
probe can be used
independently to
take temperature
measurements, or
it can be used in
c o n j u n c t i o n w i t h
the pH electrode to utilize the meter’s
ATC capability.

• Turn the instrument on by pressing the

ON/OFF key.

• Before taking any mea-

surements, make sure
that the meter has al-
ready been calibrated.

• Immerse the tip (2.5 cm)

of the pH electrode and
the temperature probe
into the sample and stir
gently.

• Take the pH reading when the hour-

glass symbol stops blinking.

• To display the mea-

sured temperature,
press and hold the
°C key.
When the key is released, the display
will return to the pH reading.

• The HOLD function is activated by

keeping pressed the HOLD key.
The measured value is frozen on
the display and the “HOLD” tag
lights up.

• After measurements, switch the

meter off and store the electrode
with a few drops of storage solu-
tion in the protective cap.

• Error message on display:

CALIBRATION PROCEDURE:

A) Preparation:

Reference solutions:
1. pH 7.01 (MA 9007)
2. pH 4.01 (MA 9004) for measuring

in the acid range (pH lower than 7)
or pH 10.01 (MA 9010) for measur-
ing in the alkaline range (pH higher
than 7).

Use two beakers for each reference so-
l u t i o n . O n e b e a k e r f o r r i n s i n g t h e
electrode, the other for calibration. This
way contamination between solutions
during calibration is minimized.

B) Procedure:

Remove the electrode
protective cap and rinse
t h e t i p w i t h s o m e
pH7.01 solution, then
immerse the electrode
and the temperature
probe into a pH7.01
buffer solution.

Press the CAL key
to enter the Cali-
bration mode. The
d i s p l a y f l a s h e s
“pH 7.01” and the
hourglass lights up.

Blinking full scale value: reading is
over range;

“Eb”: battery low voltage indication;

“Ec”: wrong buffer solution;

“Clr”: the calibration data have been
lost and recalibration is needed.