Milwaukee MW14 - User Manual
Milwaukee MW14 – User Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
USER MANUAL
IRON HIGH RANGE
MODEL:
MW14
ISTMW14 08/10
WARRANTY:
This instrument is warranted from all defects in materials and manufacturing for
a period of two years from the date of purchase.
If during this period, the repair or replacement of parts is required, where the
damage is not due to negligence or erroneous operation by the user, please return
the parts to either dealer or our office and the repair will be effected free of charge.
Note:
We reserve the right to modify the design, construction and appearance
of our products without advance notice.
SPECIFICATIONS:
MW 14
RANGE
0.00 to 5.00 ppm
RESOLUTION
0.01 ppm
ACCURACY
±0.04 ppm ±2% of reading @ 25 °C
TYPICAL EMC DEV. ±0.01 ppm
LIGHT SOURCE
Light Emitting Diode @ 525 nm
LIGHT DETECTOR
Silicon Photocell
METHOD Adaptation of the EPA Phenantroline method 315B, for
natural and treated waters. The reaction between iron
and reagent causes an orange tint in the sample.
ENVIRONMENT
0 to 50°C (32 to 122 °F)
max. 95% RH non-condensing
BATTERY TYPE
1 x 1.5V AAA
AUTO-SHUT OFF
After 3 minutes of non-use and 2 minutes
after reading
DIMENSIONS 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
WEIGHT
64 g (2.25 oz.)
OPTIONAL ACCESSORIES:
Reagent Sets
2720416
Iron powder reagent, 25 pcs.
Other Accessories
3000100
Glass cuvettes with caps (2 pcs)
3000200
Tissue fow wiping cuvettes (2 pcs)
3000300
1.5V AAA batteries (4 pcs)