Fluke 21 - Manual

Fluke 21

Fluke 21 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
Page: / 12

Table of Contents:

  • Page 2 – Amp; See Specifications for overload protection.; Autorange; The meter defaults to autorange when you turn on the meter.; Manual Range; Momentary; To return to autorange, press
  • Page 3 – Warning; To avoid electric shock, do not use the Touch Hold; Bar Graph
  • Page 4 – Volts AC
  • Page 5 – Page
  • Page 6 – meter beeps twice if it is in Touch Hold
  • Page 7 – Caution; Internal Fuse Test; OK
  • Page 8 – Battery Replacement; Note; Fuse Replacement
  • Page 9 – Service and Parts; To contact Fluke, call one of the following telephone numbers:; Item; Specifications; Accuracy specifications are given as:; Rated voltage
  • Page 12 – Overload protection for all functions and ranges: Rated voltage.; Function
Loading the manual

77/75/23/21 English Instruction Sheet

Page

1

®

77/75/23/21

Series III Multimeter

Instruction Sheet

W

Read First: Safety Information

Never use the meter if the meter or test leads look damaged.

Be sure the test leads and switch are in the correct position for

the desired measurement.

Never measure resistance in a circuit when power is applied.

Never touch the probes to a voltage source when the test

leads are plugged into the 10 A or 300 mA input jack.

Never apply more than the rated voltage between any input

jack and earth ground.

Be careful when working with voltages above 60 V dc or 30 V

ac rms. Such voltages pose a shock hazard.

Keep your fingers behind the finger guards on the test probes

when making measurements.

W

Warning

To avoid false readings, which could lead to
possible electric shock or personal injury, replace
the battery as soon as the battery indicator (

M

)

appears.

Symbols

W

Read First: Safety Information

Y

Dangerous Voltage May Be Present

T

Double Insulation

Overvoltage Installation Category per IEC 1010:

CAT II

CAT III

Typical locations include; main wall outlets,
local appliances, and portable equipment.

Typical locations include switches in the fixed
installation and equipment for industrial use
permanently connected to the fixed installation.

PN 650512 August 1997 Rev.2, 7/98

1997, 1998 Fluke Corporation. All rights reserved. Printed in U.S.A.

All product names are trademarks of their respective companies

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Amp; See Specifications for overload protection.; Autorange; The meter defaults to autorange when you turn on the meter.; Manual Range; Momentary; To return to autorange, press

77/75/23/21 English Instruction Sheet Page 2 Input Jacks V 600V 1000V FUSED CAT CAT Volts, Ohms, Diode Test Common Terminal Amp Milliamp hc1f.eps See Specifications for overload protection. Autorange The meter defaults to autorange when you turn on the meter. Manual Range Manual ranging is available...

Page 3 - Warning; To avoid electric shock, do not use the Touch Hold; Bar Graph

77/75/23/21 English Instruction Sheet Page 3 Automatic Touch Hold  Mode W Warning To avoid electric shock, do not use the Touch Hold  mode to determine if a circuit with high voltage is dead. The Touch Hold  mode will not capture unstable or noisy readings. The Touch Hold  mode automatically cap...

Page 4 - Volts AC

77/75/23/21 English Instruction Sheet Page 4 Standby If the meter is on but is inactive for an hour (20 minutes in diodetest), the display goes blank and displays four bar graphsegments. To resume operation, turn the rotary switch or press abutton. AC and DC Voltage ( K L m L ) RANGE CAT CAT + _ _ R...

Other Fluke Models

All Fluke Other