A user manual for the Fluke 80TK thermocouple module serves as a comprehensive guide to its role as a precision converter between K-type thermocouples and digital multimeters. The document covers essential operational aspects such as electrical specifications, accuracy ranges, environmental constraints, and safety considerations–including protection against electrical hazards. It also addresses practical usage with industrial surface probes, calibration methods for temperature measurement functions, and maintenance procedures like battery checks and replacements. While the schematic and part replacement details are technical, they primarily support understanding its hardware design rather than user interaction.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
Page: / 4

Table of Contents:

  • Page 1 – 80TK Thermocouple Module Specifications and Accessories
  • Page 2 – Operation and Safety Procedures for Industrial Surface Probe
  • Page 3 – Device Calibration and Support Information
  • Page 4 – 80TK Schematic, Parts List, and Warranty
Loading the manual
Page 1

PN 735985 February 1985 Rev.6, 11/02

©1985-2002 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their respective companies.

Introduction

The Fluke 80TK is a thermocouple converter for use with any
K-type thermocouple and a digital multimeter. The 80TK
accepts the output of any K-type thermocouple and converts it
to 1 millivolt per degree (Celsius or Fahrenheit).
A three-position switch acts as a power switch as well as
selecting Fahrenheit or Celsius scaling for the output. In
addition, the OFF position of the power switch allows you to
read the battery condition via the external multimeter. The
thermocouple input accepts wire leads or standard mini-
thermocouple connectors.
Temperature is measured by exposing or attaching the
thermocouple to the surface to be measured. The multimeter
displays the temperature directly in degrees Fahrenheit or
Celsius.

Specifications

Electrical

Measurement Range: -50 to 1000 degrees Celsius

-58 to 1832 degrees Fahrenheit

Resolution:
2000 count display:

0.1° C/F in 200 mV range

1.0° C/F in 2V range

3200 count display:

0.1 C/F in 300 mV range
1.0 C/F in 3V range

4000 count display:

0.1° C/F in 400mV range
1.0° C/F in 4V range

Note: Ignore decimal point on 2V, 3V, or 4V range.
Maximum Voltage at Thermocouple Input: 60V DC, 24V AC
RF Field: Exposure to an RF field interferes significantly with
this module’s capacity to accurately measure temperature with
thermocouples. The interference is temporary and the module
suffers no harm when exposed to an RF field of 3 V/m.
Accurate temperature measurement capability is completely
restored when the module is removed from the RF field.
Common sources of RF fields are handheld 2-way radios and
cellular telephones. If such a source is suspected of interfering
with this module, either turn off the transmitter or increase the
separation between the transmitter and the module.

Environmental

Ambient Operating Range: 0 to 50°C
Storage Temperature: -40 to 60°C

-40 to 140°F

Humidity:

0% to 90% (0 to 35°C)
0% to 70% (35 to 50°C)

Basic Accuracy: (@ 23

±

5°C Calibration)*

(Applies for one year after purchase or recalibration)
RANGE

ACCURACY

±

(% of reading +

degrees), 1 year

-20 to 350°C

0.5%

±

2°C

-4 to 662°F

0.5%

±

3.6°F

351 to 500°C†

1.75%

±

2°C

664 to 932°F†*

1.75%

±

3.6°F

501 to 1000°C

2.0%

±

2°C

933 to 1832°F

2.0%

±

3.6°F

-50 to -21°C

2.5%

±

2°C typical

-58 to -5.8°F

2.5%

±

3.6°F typical

Temperature Coefficient:
*

0°C to 18°C and 28°C to 50°C ambient, multiply the basic
accuracy specification by 0.1 for each degree above 28 or
below 18.

† For accuracy enhancement between 351°C to 500°C,

subtract 3 degrees from the reading. The accuracy is now
0.75%

±

2°C.

†* For accuracy enhancement between 663°F and 932°F,

subtract 5.4 degrees from the reading. The accuracy is now
0.75%

±

3.6°F.

Note

The basic accuracy specification does not include the
error of the probe or the DMM. Please refer to the
probe accuracy specification for additional details.

General

Weight: 4.5 ounces, 126 grams
Overall length: 4 5/8 inches, 11.75 mm
Battery: Standard 9V Battery (NEDA #1604, 6F22, 006P)
Battery Life: 1600+ hours, 6.5V minimum
Output: 3/4 inch spaced banana plugs, 10 megohm minimum
load resistance

Accessories

80PK-1 (included): Bead Probe 4 foot K-type thermocouple
bead probe, with Teflon tape insulation. Maximum insulation
temperature: 260°C. Accuracy:

±

2.2°C or

±

0.75% (whichever

is greater) from 0 to 800°C.

80PK-2A: Immersion Probe (8" or 20.32 cm from tip to probe
handle)

80PK-3A: Surface Probe (for flat or curved surfaces)

80PK-4A: Air Probe (protected by stainless steel baffle)

80PK-5A: Piercing Probe (suitable for food use, liquids and
gels)

80PK-6A: Exposed Junction Probe (exposed bead attached to
6" sheath)

®

80TK Thermocouple

Module

Instruction Sheet

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Was this manual helpful?

About this manual

Brand
Fluke
Model
80TK
Document type
Manual
Language(s)
English
Pages
4
File size
410.1 KB
Format
PDF

Summary

Page 1 - 80TK Thermocouple Module Specifications and Accessories

Function of the module as a converter for K-type thermocouples and digital multimeters, electrical specifications, accuracy ranges, environmental limits, and included probe types.

Page 2 - Operation and Safety Procedures for Industrial Surface Probe

Instructions for using the industrial surface probe, including measurement techniques, temperature procedures, open thermocouple detection, battery testing, replacement, self-test, and electrical safety warnings.

Page 3 - Device Calibration and Support Information

Calibration procedures for Celsius and Fahrenheit functions, required and recommended equipment lists, and contact information for Fluke Corporation.